What are the meanings of the icons on the HERAEUS M 110?

Icon Explanation
! in a triangle This icon marks chapters and sections of the instruction manual dealing with safety topics.
SSS symbol Caution hot surface! Danger of injury!
Hand with pointing finger symbol Marks information on most effective use of the unit.
O Unit OFF
I Unit ON
Thermometer with up arrow symbol Over-temperature protection, upper limit cut-out
Crossed out thermometer with up arrow symbol Deactivation of over-temperature protection
Fan symbol Exhaust fan
off / -on button symbol Push-button out = Exhaust fan OFF
Push-button in = Exhaust fan ON

What are the general safety instructions for operating the HERAEUS M 110?

To ensure the safety of personnel, the environment, and samples, it is of paramount importance that the operator uses the HERAEUS M 110 correctly. To avoid errors and damage, especially personal injuries, read the manual before use and follow all instructions.

Prohibited Uses:

• The unit is not to be used to heat foods.

• The unit is not to be used for drying or heat treatment of substances which release gases or vapours that are flammable or dangerously explosive when mixed with air.

• This furnace is unsuitable for the heat treatment of combustible dusts and fibrous materials.

Operational Warnings:

• Metal surfaces around the mouth of the furnace become hot at high operating temperatures. There is a DANGER OF INJURY. Avoid contact with these surfaces.

• Special measures must be taken during external temperature measurement to ensure protection from electric shocks.

• Exhaust gases from heat treatments must be safely conducted out of the room. Observe all applicable environmental protection regulations.

• Export versions of the unit without over-temperature protection systems may only be operated under constant observation.

Equipment Checks and Maintenance:

• Test the mains supply cable and connecting plug for defects before use. Do not connect the unit if defects are detected.

• The voltage on the nameplate must match the voltage of the mains power supply.

• The door-operated safety switch must be tested at regular intervals (at least once per year).

• The over-temperature protection system must be functionally tested at appropriate intervals (at least every 3 months), and its electrical functions tested at least once each year.

• Work on electrical components may only be carried out by authorized service personnel with the unit disconnected from the mains power supply.

• Use only authorized replacement parts. Using other parts can present unforeseeable risks.


What are the intended applications for the HERAEUS M 110?

The HERAEUS M 110 muffle/chamber furnace is designed for thermal technology applications in the laboratory.

• Volume of furnace: 9 l

• Heat treatment of samples and objects from 300°C to 1100°C

• Loads of up to 5 kg

It is generally designed for installation and operation in laboratories, such as those found in manufacturing trades and industry, schools, universities, hospitals, and biology.

Specific uses include:

• aging

• analysis

• tempering

• decomposing

• baking

• annealing

• hardening

• soldering

• oxidizing

• reducing

• incineration

• preheating

The unit is designed for continuous operation. The optional over-temperature protection “upper limit cut-out” enables operation without constant supervision.


How do I test the over-temperature protection feature on the HERAEUS M 110?

The upper limit cut-out (TWB) should be periodically tested (at least every 3 months) to ensure it is functioning properly.

Functional Test Procedure:

1. Precondition: The temperature set on the temperature controller has been reached and remains constant.

2. Using a tool like a coin or screwdriver, adjust the TWB setting to approx. 50°C below the temperature currently displayed on the temperature controller.

3. The TWB should activate, and the red “malfunction” display light will illuminate. This confirms the TWB is functioning properly.

4. After the test, reset the TWB to the desired temperature limit for your application.

Note: The unit must be reset manually using the key (deactivation of over-temperature protection symbol) after it has cooled down.

If the “malfunction” light is indicated during normal operation, check the settings of the TWB and the temperature controller. If the malfunction persists, contact the service department.


How do I operate the Digicon® temperature controller on the HERAEUS M 110?

The Digicon® temperature controller displays and adjusts temperature in increments of 10°K.

Command Action Display Condition
Turn on unit (Use main power switch) Furnace temperature is displayed (e.g., 20)
Select preset temperature Press and hold down the dial Display changes from actual temperature to the set nominal temperature (e.g., 920)
Raise preset value While holding the dial down, turn it in the clockwise direction Nominal value rises (e.g., 980)
Lower preset value While holding the dial down, turn it in the counterclockwise direction Nominal value decreases (e.g., 850)
Reset unit to new nominal value Release the dial The furnace temperature is displayed again (e.g., 750), and the unit will begin adjusting to the new setpoint. The “HEATING” indicator will pulse as the controller maintains the temperature.

How do I operate the HERAEUS M 110 with the Thermicon® P controller as a simple temperature regulator (without programming)?

To use the Thermicon® P controller without programming functions, ensure you enter “Pr = 0” to switch off the programming operation.

Command Key(s)/Action Display Remarks / Condition
Turn unit on All display elements illuminate for approx. 20 sec for a self-test, then the furnace temperature is displayed.
Display preset temperature Press the temperature key (thermometer icon). The nominal (set) value is displayed for approx. 15 sec. (e.g., 0850).
Select preset temperature Hold down the temperature key (thermometer icon). The display shows the nominal value (e.g., 0850).
Change preset temperature While holding the temperature key, use the “digit” key to select a position and the “up” key to change the value. The new value is displayed (e.g., 0900).
Reset to new preset temperature Release the temperature key. The display is temporarily dimmed while the value is stored. After approx. 15 sec., the display shows the new stored value, followed by the actual temperature in the furnace chamber.

How do I set the timer on the Thermicon® P controller to turn the HERAEUS M 110 ON after a set time?

This example shows how to switch the unit on after 6 hours to heat the chamber to 800 °C.

Command Key(s) / Action Display Remarks / Condition
Turn unit on Controller performs self-test, then displays furnace temperature.
Select preset temperature Set the desired temperature (e.g., 800 °C) as per standard operation. Display shows “SP 0800”.
Select operating mode of timer Press the “up” key and the “&” key simultaneously. Select “tf 0” (heating ON after time). Display shows “tF 0000”.
Switch to time setting function Press the time key (clock icon) and release. The “tf 0” indicator flashes “0”.
Enter time Hold down the time key. Press the “digit” key to select position and the “up” key to change the value. For this example, select “06.00”. Display shows “06.00”.
Confirm time setting Release the time key. The heating will be turned on after the 6-hour period has elapsed. After approx. 15 seconds, the display returns to showing “tf 0” and the current furnace temperature.

Note: The unit must remain turned on for the timer to function.


How do I set the timer on the Thermicon® P controller to turn the HERAEUS M 110 OFF after a set time?

This example shows how to switch the unit off after 6 hours of operation at 800 °C.

Command Key(s) / Action Display Remarks / Condition
Turn unit on Controller performs self-test, then displays furnace temperature.
Select preset temperature Set the desired temperature (e.g., 800 °C) as per standard operation. Display shows “SP 0800”.
Select operating mode of timer Press the “up” key and the “&” key simultaneously. Select “tf 1” (heating OFF after time). Display shows “tF 0001”.
Switch to time setting function Press the time key (clock icon) and release. The “tf 1” indicator flashes “1”.
Enter time Hold down the time key. Press the “digit” key to select position and the “up” key to change the value. For this example, select “06.00”. Display shows “06.00”.
Confirm time setting Release the time key. The heating will be turned off after the 6-hour period has elapsed. After approx. 15 seconds, the display returns to showing “tf 1” and the current furnace temperature.

Note: If no timer function is required, select “tf 0” and enter the time “00.00”.


How do I program a time/temperature profile on the Thermicon® P controller for the HERAEUS M 110?

The controller can execute programs of up to 9 steps. This example shows how to set up a 5-step program.

Step 1: Enter Programming Mode

Command Key(s) / Action Display Remarks / Condition
Select programming operation Hold down the program key (wavy line icon). Display shows “P- 0000” and flashes.
Select program function Press the “up” key to select “prog. on I”. Display shows “P- 0001”.
Select number of programming steps Release the program key quickly and hold it down again. Then, press the “up” key to select the number of steps (e.g., n=5). Display shows “n 0005” and flashes.
Select type of return command Release the program key quickly and hold it down again. Then, press the “up” key. (r=0: no return; r=1: repeat program). Display shows “r 0000” and flashes.

Step 2: Define Program Steps

Command Key(s) / Action Display Remarks / Condition
Define starting point (P0) Release the program key quickly and hold it down again. Use “digit” and “up” keys to enter the starting temperature (e.g., 30°C). Display shows “P0 0030”.
Enter time for step 1 (t1) Release and hold the program key. Use “digit” and “up” keys to enter the time (e.g., 2.00 hours). Display shows “t1 02.00”.
Enter temperature for step 1 (P1) Release and hold the program key. Use “digit” and “up” keys to enter the temperature (e.g., 700°C). Display shows “P1 0700”.
Enter time for step 2 (t2) Release and hold the program key. Use “digit” and “up” keys to enter the time (e.g., 1.30 hours). Display shows “t2 01.30”.
Enter temperature for step 2 (P2) Release and hold the program key. Use “digit” and “up” keys to enter the temperature (e.g., 700°C). Display shows “P2 0700”.
Repeat the same procedure for all remaining program steps (n3 to n5 in this example).
End entry Release the program key and wait for “tP” to appear on the display. Display shows “tP 30”.
Run programmed sequence Press the “digit” and “up” keys simultaneously. The controller executes the program. The display shows the current program step and value.

Note: The program is retained in memory during power failures but must be restarted manually once power is restored.


What do the error messages on the Thermicon® P controller of the HERAEUS M 110 mean?

Display / Condition Remarks Remedy
Display shows furnace temperature (e.g., 825) instead of timer info Power failure occurred during timer operation. Timer function is halted, and controller operation is off. Reset the timer.
Display shows “Er 0” Incorrect entry or controller is defective. Controller operation is off. Turn the unit off and on, then reprogram the controller. If the fault persists, contact Service.
Display shows “Er 99” Temperature sensor problem or defective controller. Controller is off. Contact Service.

What are the setup and installation requirements for the HERAEUS M 110?

Location and Placement:

• Place the furnace in a stable and perfectly upright position on a firm, fireproof surface (e.g., laboratory table or support frame).

• It is not possible to stack the furnaces.

• The room temperature can be up to 40°C.

• Do not cover or obstruct the ventilation or exhaust vents of the unit.

Distance from Other Surfaces and Objects:

• Left side: 300 mm (30 cm)

• Right side: 100 mm (10 cm)

• Top: 400 mm (40 cm)

• Rear (with spacer): 25 mm

• Min. rear spacing to wall: 10 cm

Room Ventilation:

• Proper ventilation must be installed where the unit is set up. Do not operate in non-ventilated, enclosed spaces.

• Special ventilation measures are required when several units are set up in the same room.


What are the electrical connection requirements for the HERAEUS M 110?

• The unit is powered via a permanent flexible mains power cable and a grounded plug.

• Before use, test the mains supply cable and connecting plug for any faults. Do not connect the unit if faults are detected.

• The voltage specified on the nameplate must match the voltage delivered by the mains power supply.

• Only plug the furnace into a properly installed outlet equipped with a PE conductor (Protection Class 1).

• The mains power circuit should additionally provide protection against electric shocks caused by leakage current.

Mains circuit protection required: T 16 A fuse or B 16 circuit break-switch.

• For any required power extensions, use a rubber-sheathed flexible cable of at least 1.5 mm diameter (H 07 RN-F).


How do I use an external thermocouple for temperature measurement with the HERAEUS M 110?

There is an aperture (5 mm dia.) in the middle of the swivel door that serves as an access for shielded thermocouples to provide external measurement of temperatures inside the furnace chamber.

IMPORTANT SAFETY INSTRUCTIONS:

• To avoid damage to temperature measurement devices, only connect them to power circuits that are entirely independent of the mains power supply (FRG: DIN VDE 0100 Section 410).

• To ensure protection against electric shock, only shielded thermocouples which are properly connected to the PE conductor are to be placed through this aperture.

• Due care must be taken to ensure that the thermocouple does not come into contact with the walls of the furnace.

• An additional leakage current flows over the thermocouple shield. Under unfavourable conditions, the sum of all leakage currents can activate the leakage current protection switch. It is recommended to connect the furnace to a separate leakage current protection switch (rated at ≤ 30 mA).


What is the startup procedure for the HERAEUS M 110?

After setting up and installing the unit, proceed as follows to start the unit:

Command Control / Remarks
Turn on unit Set Mains switch to “I”.
Select nominal temperature Adjust the temperature controller (Digicon or Thermicon® P).
Adjust over-temperature protection Adjust the TWB (position 3) to the upper limit required.
Turn on exhaust fan (if applicable) Select the function mode and time on the timer if necessary.
Load samples Load samples into the furnace, wait until the furnace chamber temperature is reached if necessary. Refer to operating guidelines for sample placement.

What are the operating guidelines for the HERAEUS M 110?

• Wear required personal safety clothing to protect hands, face, and body; remove jewellery.

• To avoid overheating heat-conducting samples, distribute them evenly throughout the furnace chamber and do not place them too close to the heating elements.

• Take care not to damage the heating elements when loading samples.

• Avoid any direct contact of aggressive chemicals with the ceramic and fibre material surfaces inside the chamber, as soiled surfaces can cause premature damage to insulation and heating.

• Limit operation of the furnace at its highest temperatures only as long as required to complete the task, as the life of the heating coils is highly dependent on the temperature load.

• If used often, it is best to run the unit in continuous operation at medium temperatures, as continually heating and cooling over a large range can adversely affect the unit’s life.

• Do not cover the ventilation and extraction vents on the outside of the unit, and keep them clean and clear of objects.


What is the shutdown procedure for the HERAEUS M 110?

Follow these steps to shut down the furnace:

1. Remove samples loaded into the furnace chamber.

2. Wait for the furnace and room temperatures to cool down.

3. Turn off the unit. If it is not used for a longer period, disconnect it from the mains and pull out the plug.

4. Clean out the furnace chamber.


How do I clean and maintain the HERAEUS M 110?

The unit requires very little maintenance.

Cleaning:

• The bearings of the swivel door can be lubricated with graphite oil.

• Clean the outer surfaces and control elements with a mild detergent and a soft cloth.

• Clean away any spillage or residue from the furnace chamber.

• If the unit is connected to an extraction system, the exhaust ducts should be part of the cleaning procedures.

Testing:

To ensure the unit remains in proper working condition, it should be tested at least once a year for the following:

• Mechanical functions

• Function according to technical specifications

• Electric circuits (FRG: UVV VBG 4)

• Safety equipment installed in the unit


What are the order numbers for accessories and replacement parts for the HERAEUS M 110?

Replacement / Accessory Order No.
Instruction manual 50 042 200
Circuit diagram 50 008 214
Tray 50 006 408
Exhaust duct 50 006 394
Exhaust fan (replacement part) 50 006 439
Exhaust fan (retrofit kit) 50 024 794
Base plate 50 005 547

What are the technical specifications of the HERAEUS M 110?

MECHANICAL SPECIFICATIONS

Dimensions (W x H x D) – Housing 576 x 752 x 720 mm
Dimensions (W x H x D) – Work chamber 200 x 150 x 300 mm
Dimensions (W x H x D) – Base plate 200 x 10 x 320 mm
Volume – Work chamber 9 l
Weight – Unit about 80 kg
Weight – Maximum load 7 kg

THERMAL SPECIFICATIONS

Operating temperatures 300 to 1100 °C
Temperature deviation temporal – with exhaust flue 1 / 1 K
Temperature deviation temporal – with exhaust fan 1 K
Spatial deviation (without flue/with flue/with fan) at 500°C ± 18 / ± 24 / – K
Spatial deviation (without flue/with flue/with fan) at 700°C ± 13 / ± 20 / – K
Spatial deviation (without flue/with flue/with fan) at 1100°C ± 7 / ± 9 / ± 9 K
Heat rise time (without load) up to 1100°C – with/without flue about 125 / 130 min
Heat rise time (without load) up to 1100°C – with exhaust fan about 180 min
Cool down time (to 100°C, without load) from 1100°C – with and without flue about 18 h
Cool down time (to 100°C, without load) from 1100°C – with exhaust fan 8.5 h
Heat radiation to surrounding areas at 1100°C 1.4 kW

VENTILATION / EXHAUST GASES SPECIFICATIONS

Air replacement in furnace chamber at 1100°C – without / with flue about 13 h⁻¹
Air replacement in furnace chamber at 1100°C – with exhaust fan about 192 h⁻¹
Exhaust gas flow (without load) at 1100°C – without / with flue 117 l / h
Exhaust gas flow (without load) – with exhaust fan 50 m³/h
Diameter of flue port / exhaust fan 30 / 60 mm
Exhaust gas temp. at port (max. op. temp) – without / with flue 270 (330) °C
Exhaust gas temp. at port (max. op. temp) – with exhaust fan 150 (200) °C

ELECTRICAL SPECIFICATIONS

Rated voltage (± 10 %) 1 / PE AC, 230 V
Rated frequency 50 / 60 Hz
Nominal power consumption 3 kW
Current load 13 A
Protection class I
Protection type IP 20
Noise level (DIN 45 635) < 58 dB (A)

What materials are used in the construction of the HERAEUS M 110?

PART MATERIALS USED
Housing Zinc-plated sheet steel, painted RAL 9002
Furnace chamber Silicon carbide panels, aluminium silicate fibre¹, fireclay brick (chamotte)
Door insulation Aluminium silicate fibre¹
Mouth of furnace Fireclay brick (chamotte)
Thermal insulation of furnace chamber Aluminium silicate fibre¹
Base plate Fireclay brick (chamotte)
Tray Heat-resistant steel, material no 1.4841
Heater coils Cr Fe Al
Control panel Mixture of ABS and PC
Electrical components Components coated with various plastics, partially mounted with epoxy resins on circuit boards strengthened with glass fibre.

¹ Fibre materials used for insulation are classified as hazardous to health in certain countries. No serious risks to the working environment arise from normal operation. Improper handling during repairs or dismantling the unit can lead to significant levels of contamination.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD HERAEUS M 110 (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.