What do the different indicators on the LCD screen of the EUTECH INSTRUMENTS DO 110 mean?

The LCD screen of the EUTECH INSTRUMENTS DO 110 has a primary and secondary display, along with several indicators. The primary display shows the measured % Saturation or mg/L (or ppm) values, while the secondary display shows the temperature.

Indicator Description
SETUP Setup mode indicator
MEAS Measurement mode indicator
CAL Calibration indicator
MEM Memory recall mode indicator
% Percentage Saturation of Oxygen
ppm parts per million
mg/l milligram per liter
°C °F Temperature indicator
ATC Automatic Temperature Compensation indicator
ERR Error indicator
MEM Memory location indicator
Low battery indicator Indicates low battery
Electrode indicator Indicates electrode status
Calibration standard indicator Indicates calibration standard
ON READY/Auto HOLD set up enable indicator
OFF READY set up disable indicator
HOLD Hold indicator
READY Ready indicator

What are the functions of the keypad buttons on the EUTECH INSTRUMENTS DO 110?

KEY FUNCTION
ON/OFF Powers on and shuts off the meter. The meter will start in a measurement mode similar to the last measurement mode it was in before being switched off.
CAL / MEAS

CAL – Activates the DO % Saturation or mg/l calibration mode and when used with the MODE key, it activates the temperature calibration mode.

MEAS – Allows return to measurement mode when canceling or terminating any operation.

HOLD / ENTER

HOLD – Activates/Deactivates freezing of the measured reading while in measurement mode.

ENTER – Confirms the calibration values in Calibration mode and the selection in SETUP and Print Function mode.

Scroll through the memory location and the stored data during memory recall.

MI (Up Arrow) MI (Memory Input) – Captures the measured readings with its corresponding temperature values and stores them in the memory.
MR (Down Arrow)

MR (Memory Recall) – Retrieves the stored data from memory.

Sets the calibration values during mg/l or temperature calibration.

Scrolls through each SETUP and its sub group menu.

Scroll through Print Function menu and its parameters.

SETUP Activates the parameter setting menu to allow you to customise meter configuration, view calibration point and electrode data, select auto power off, reset meter, clear memory, set % Saturation Offset and set dissolved oxygen parameters.
MODE Selects the measurement parameter option between % Saturation with temperature and mg/L (or ppm) with temperature.
PRINT Allows print of current measurement or stored data to either the printer or the computer.

How do I insert the batteries into my EUTECH INSTRUMENTS DO 110?

Your EUTECH INSTRUMENTS DO 110 meter requires 4 “AAA” alkaline batteries for operation. Follow these steps to insert them:

1. To open the battery compartment, press down the catch of the battery cover.

2. Note the polarity markings and insert the batteries into the battery compartment correctly.

3. Replace the battery cover into its original position, ensuring the catch is locked in position.


How do I connect and disconnect the DO probe on the EUTECH INSTRUMENTS DO 110?

The EUTECH INSTRUMENTS DO 110 uses a special notched 6-pin connector. Keep the connector dry and clean. Do not touch it with soiled hands.

To Connect the Probe:

1. Line up the notch and 6 pins on the meter with the holes in the 6-pin connector.

2. Push down and screw the metal sleeve to lock the probe connector into place.

To Disconnect the Probe:

1. Unscrew the metal sleeve and slide it up the probe connector.

2. While holding onto the metal sleeve, pull the probe away from the meter. CAUTION: Do not pull on the probe cord, as the probe wires might disconnect.


How do I power the EUTECH INSTRUMENTS DO 110 using an AC/DC adapter?

To operate the EUTECH INSTRUMENTS DO 110 from a power main, follow these steps:

1. Switch off the meter and the power source of the adapter.

2. Select the correct AC/DC Adapter that matches your input mains voltage.

3. Ensure the adapter has the correct output voltage: 9 to 15 V DC, Current: >=50 mA.

4. Gently insert the power adapter D.C. jack into the meter’s power socket.

5. Switch on the power source of the adapter, followed by the meter.


How do I connect the RS232C cable to my EUTECH INSTRUMENTS DO 110?

To connect the RS232C cable for transmitting data to a printer or computer:

1. Open the printer port cover located at the bottom end of the meter. Do not use excessive force.

2. Noting the orientation of the RS232C connector, plug the male connector into the RS232C port of the meter.

3. Fasten the RS232C connector by tightening the two screws at the side of the male connector.


What is the RS232C configuration for the EUTECH INSTRUMENTS DO 110?

The EUTECH INSTRUMENTS DO 110 meter has a 9-pin female RS232C connector with the following pinout:

PIN NO. DESCRIPTION
1
2 Transmit Data
3
4
5 CTS (Clear to Send)
6
7 GND (Ground)
8
9

If connecting to a 25-pin RS232C connector, the following cable configuration may be used:

DO 110 (9-pin) 25 pin connector
2 (TxD) (RxD) 3
5 (CTS) (RTS) 4
7 (GND) (GND) 7

How do I perform a temperature calibration on the EUTECH INSTRUMENTS DO 110?

It is necessary to calibrate or verify the temperature reading before DO calibration. To perform a temperature calibration:

1. Switch the meter on. The ATC annunciator will appear on the LCD.

2. Press the CAL/MEAS key to enter calibration mode.

3. While in calibration mode, press the MODE key to enter temperature calibration mode.

4. Dip the DO probe into a solution of known temperature (e.g., a temperature bath) and allow the reading to stabilize.

5. Use the MI/▲ or MR/▼ keys to adjust the primary display to the correct temperature value. You can adjust the reading in increments of 0.1°C and offset it up to ±5°C from the original reading.

6. Once the correct temperature is selected, press the HOLD/ENTER key to confirm. The meter will automatically return to the measurement mode.

NOTE: To exit without confirming the calibration, press CAL/MEAS instead of HOLD/ENTER.


How do I calibrate the EUTECH INSTRUMENTS DO 110 to 100% Saturation?

Follow these steps to calibrate your EUTECH INSTRUMENTS DO 110 for 100% air saturation:

1. Press the MODE key to select % Saturation mode.

2. Rinse the probe well with deionized water and blot the end dry. Do not touch the membrane.

3. Hold the probe in the air with the sensor facing down and press the CAL key. The primary display will show the current measurement, and the secondary display will show “100.0” (or the value corrected for barometric pressure). Wait for the reading to stabilize.

4. Press the ENTER key to confirm the calibration. The meter automatically calibrates to 100% air saturation and returns to the measurement mode.

NOTE: To exit without confirming the calibration, press CAL/MEAS instead of ENTER.


How do I calibrate the EUTECH INSTRUMENTS DO 110 to 0% Saturation?

Follow these steps to perform a zero-point calibration on your EUTECH INSTRUMENTS DO 110:

1. Press the MODE key to select % Saturation mode.

2. Rinse the probe well with deionized water.

3. Immerse the DO probe in a 0% solution (zero oxygen solution). Stir gently and wait for the reading to stabilize.

4. Press the CAL key. The primary display will show the current measurement, and the secondary display will show “0.0”.

5. Press the ENTER key to confirm the calibration. The meter automatically calibrates to 0% saturation and returns to the measurement mode.

NOTE: To exit without confirming the calibration, press CAL/MEAS instead of ENTER.


How do I calibrate the EUTECH INSTRUMENTS DO 110 in mg/L (ppm) mode?

Before calibrating in mg/L or ppm, ensure the barometric pressure and salinity parameters are set correctly in the advanced setup menu. To calibrate:

1. From the measurement mode, press the MODE key to select mg/L (or ppm).

2. Rinse the probe well with deionized water and blot the end dry. Do not touch the membrane.

3. Dip the probe into a solution with a known DO value.

4. Wait for the reading to stabilize, then press the CAL key.

5. Adjust the reading to the known value of the solution using the MI/▲ or MR/▼ keys.

6. Press ENTER to confirm the calibration.

NOTE: To exit without confirming the calibration, press CAL/MEAS.


How do I set up and use Manual Temperature Compensation (MTC) on the EUTECH INSTRUMENTS DO 110?

First, you must set the ATC mode in the SETUP menu Program P 1.3 to “NO”. The [ATC] annunciator will disappear from the LCD. To set the manual temperature:

1. Press the CAL/MEAS key to enter DO calibration mode.

2. While in DO calibration mode, press the MODE key to enter temperature calibration mode. The primary display acts as the adjustable temperature setting.

3. Check the temperature of your sample using an accurate thermometer.

4. Press the MI/▲ or MR/▼ keys to set the temperature on the meter to the value measured in step 3.

5. Press HOLD/ENTER to confirm the selected temperature and return to the measurement mode. The meter will now compensate DO readings for the manually set temperature.

NOTE: To exit without confirming the manual temperature, press CAL/MEAS instead of HOLD/ENTER.


How do I take a measurement with the EUTECH INSTRUMENTS DO 110?

To take a Dissolved Oxygen measurement:

1. Rinse the probe with deionized or distilled water before use. If it is dehydrated, soak it for 30 minutes in tap water.

2. Switch on the meter. The MEAS and ATC indicators will appear.

3. Dip the probe into the sample, ensuring the tip is completely immersed. Stir the sample gently to create a homogenous sample and tap the probe gently to remove air bubbles.

4. Allow time for the reading to stabilize. When stable, the READY annunciator appears. Note the reading.

5. Press the HOLD key to freeze the reading for delayed observation. Press HOLD again to release the value.

6. To toggle between % Saturation and mg/l (ppm) modes, press the MODE key.

Using the READY indicator: If activated in the SETUP menu, the READY annunciator lights up when the reading is stable.

Using the AUTO HOLD feature: If activated in the SETUP menu, the meter will automatically “HOLD” the reading when it is stable for more than 5 seconds. The “HOLD” indicator appears. Press HOLD/ENTER to release the reading.


How do I store a reading into the memory of the EUTECH INSTRUMENTS DO 110?

The EUTECH INSTRUMENTS DO 110 can store up to 100 sets of data.

1. During any measurement function (MEAS), press the MI/▲ key to input the current reading into memory.

2. The display will flash “MEM”, “Sto”, and the memory number. The meter then returns to measurement mode.

3. If the memory is full (100 readings), the display will momentarily flash “MEM – FUL”. The first value stored will be erased to create space for the new value.


How do I recall stored readings from the memory of the EUTECH INSTRUMENTS DO 110?

Memory recall is in “Last In First Out” order. To recall readings:

1. Press the MR/▼ key once to retrieve the last reading stored. The memory location screen (MEM, “Loc”, and memory number) will flash on the display.

2. Press the HOLD/ENTER key to recall the reading stored under that memory number.

3. To display the next stored reading, press HOLD/ENTER again. The display automatically moves to the next lower memory location screen.

4. To view a stored reading in a specific memory location, use the MI/▲ or MR/▼ keys to scroll to the desired location number, then press HOLD/ENTER to display the reading.

NOTE: Readings are retained even if the unit is turned off. To erase all readings, use the Memory Clear function in the SETUP menu.


How do I set up my EUTECH INSTRUMENTS DO 110 to print data?

To use the print function with a printer or computer:

1. Ensure the printer has a 9-pin or 25-pin RS232C serial port.

2. The communication settings of the printer or computer’s data acquisition software must match the following:

– Baud Rate: 9600

– Parity Bit: None

– Stop Bit: One

– Data Bit: 8

– Flow Control: None

3. Use the appropriate 9-pin or 9-to-25-pin cable to connect the meter to the printer/computer.

4. Turn on the DO 110 meter.

5. For a printer, ensure it is powered on and paper is loaded. For a computer, ensure the data acquisition software is installed and activated.

6. To send data, press the PRINT key on the meter.


How can I manually print the current measurement from my EUTECH INSTRUMENTS DO 110?

To manually print a single data point that is currently being measured:

1. From measurement mode, press the HOLD key to freeze the reading.

2. Press the PRINT key.

3. The “Prn – CUr” annunciator will appear and flash for about 0.5 seconds as the data is transmitted.

4. The meter returns to measurement mode.

5. Repeat steps 1 to 3 for the next measurement.


How do I print data on a timed interval from my EUTECH INSTRUMENTS DO 110?

This feature automates printing of ‘live’ measurement values at a set interval from 5 to 30 seconds.

1. From the measurement mode, press the PRINT key.

2. The “dAt/CUr” screen appears. Press HOLD/ENTER.

3. The “Sec 5” screen appears (default is 5 seconds). Use the MI/▲ or MR/▼ keys to set the desired time interval.

4. Press HOLD/ENTER to confirm.

5. The meter enters measurement mode with the MEM annunciator flashing.

6. At each pre-set time interval, the screen will flash “Prn/CUr” as it prints the data.

7. To stop printing, press the PRINT key. The MEM annunciator will disappear.


How do I print all data stored in the memory of my EUTECH INSTRUMENTS DO 110?

You can print all stored data from the meter’s memory.

1. From measurement mode, press the PRINT key.

2. Use the MI/▲ or MR/▼ key to select the “dAt – StO” option and press HOLD/ENTER.

3. At the “SEC – 0” option, you can use the MI/▲ or MR/▼ key to select a time interval (in seconds) between each printed data set. This helps prevent printer/computer buffer overflow.

4. Press HOLD/ENTER. The “Prn” display will flash until all data is printed.

5. To stop the printing process at any time, press the PRINT key. This will exit to measurement mode.

NOTE: All stored memory will remain in the meter after printing.


How do I print selected data from the memory of my EUTECH INSTRUMENTS DO 110?

To print data from a specific memory location:

1. From measurement mode, press the MR/▼ key. The “MEM” indicator appears, showing the last memory input location.

2. Use the MI/▲ or MR/▼ key to select a specific memory location, then press HOLD/ENTER to view the stored data.

3. When ready to print this data point, press the PRINT key.

4. The screen flashes “Prn/CUr” for about 0.5 seconds and then returns to the stored data display for that memory location.

5. Press HOLD/ENTER again to move to the next stored data location and repeat the process if desired.


How do I access and navigate the Advanced Setup Functions on the EUTECH INSTRUMENTS DO 110?

The advanced setup mode lets you customize meter parameters. To access it:

1. Press the SETUP key to enter Setup mode.

2. Use the MI/▲ and MR/▼ keys to scroll through the main program groups (P1.0, P2.0, etc.).

3. Press the HOLD/ENTER key to enter a selected program group and view its sub-parameters.

4. Use the MI/▲ and MR/▼ keys to change settings within a sub-parameter.

5. Press HOLD/ENTER to confirm a change and move to the next sub-parameter.

6. Press CAL/MEAS at any time to exit the setup menu and return to measurement mode.

The main setup groups are:

  • P1.0: Unit Configuration (COF)
  • P2.0: View Calibration Data (CAL)
  • P3.0: View Electrode Data (ELE)
  • P4.0: Auto Power Off (AtO)
  • P5.0: Reset to Factory Default (rSt)
  • P6.0: Memory Clear (CLr)
  • P7.0: Dissolved Oxygen Parameters (dPr)
  • P8.0: % Saturation Offset Adjustment (OFS) (Only in % Saturation mode)

How do I configure the READY indicator and Auto Hold function on the EUTECH INSTRUMENTS DO 110?

In setup program P1.1, you can choose how the meter indicates a stable reading.

1. Press SETUP, then use the MI/▲ or MR/▼ keys to navigate to P1.0 (COF). Press HOLD/ENTER to enter.

2. The display shows P1.1 (rdy). Press HOLD/ENTER again.

3. Use the MI/▲ or MR/▼ keys to cycle through the options:

ON (READY indicator only): The “READY” icon appears when the reading is stable.

OFF: The “READY” indicator is disabled for a faster response.

ON and HOLD (Auto Hold): The meter automatically freezes (“HOLD”) the reading when it is stable for more than 5 seconds. Press HOLD/ENTER to release.

4. Press HOLD/ENTER to confirm your selection and proceed to the next setup parameter (P1.2).


How do I change the temperature units between °C and °F on my EUTECH INSTRUMENTS DO 110?

You can select between Celsius (°C) and Fahrenheit (°F) in setup program P1.2.

1. Press SETUP and navigate to P1.0 (COF).

2. Press HOLD/ENTER twice to select parameter P1.2.

3. Use the MI/▲ or MR/▼ keys to toggle between “°C” and “°F”.

4. Press HOLD/ENTER to confirm your selection.


How do I switch between Automatic (ATC) and Manual (MTC) Temperature Compensation on the EUTECH INSTRUMENTS DO 110?

You can select the temperature compensation method in setup program P1.3. The meter default is ATC.

1. Press SETUP and navigate to P1.0 (COF).

2. Press HOLD/ENTER three times to reach parameter P1.3 (AEC).

3. Press HOLD/ENTER again. The upper display shows “ATC”.

4. Use the MI/▲ or MR/▼ keys to select between “YES” (activates ATC) and “NO” (activates MTC).

5. Press HOLD/ENTER to confirm your selection.


How do I switch the measurement mode between mg/L and ppm on my EUTECH INSTRUMENTS DO 110?

This option is available in setup program P1.4 when the meter is in mg/L (ppm) measurement mode.

1. From the mg/L (ppm) measurement mode, press SETUP.

2. Navigate to P1.0 (COF) and press HOLD/ENTER four times to enter the P1.4 mode selection menu (“dO”).

3. Use the MI/▲ and MR/▼ keys to select the desired mode (mg/l or ppm).

4. Press HOLD/ENTER to confirm the selection and return to the subgroup menu.

5. Press CAL/MEAS to return to the measurement mode.


How can I view previous calibration data on my EUTECH INSTRUMENTS DO 110?

This is a “view only” mode that recalls previous calibration data.

1. From measurement mode, press the SETUP key.

2. Use the MI/▲ or MR/▼ keys to scroll to parameter P2.0 (CAL).

3. Press the HOLD/ENTER key to view the previous calibration data (e.g., “100.0” for % Saturation).

4. Press HOLD/ENTER again to return to the subgroup menu.


How do I view the electrode data and diagnostics on my EUTECH INSTRUMENTS DO 110?

The Electrode Data menu (P3.0 ELE) provides diagnostic information about the probe’s performance. This is a “view only” menu.

1. From measurement mode, press the SETUP key.

2. Use the MI/▲ or MR/▼ keys to scroll to parameter P3.0 (ELE).

3. Press HOLD/ENTER repeatedly to cycle through the available data points. The data shown depends on whether you entered the setup menu from % Saturation mode or mg/L (ppm) mode.

In % Saturation mode, you can view:

– EL.1: Electrode Slope Factor (K)

– EL.2: % Saturation Offset

– EL.3: 100% Saturation mV value

– EL.4: 0% Saturation mV value

In mg/L (or ppm) mode, you can view:

– EL.1: Electrode Slope Factor

– EL.2: 100% Saturation mV value

– EL.3: 0% Saturation mV value


How do I enable or disable the Auto-Off feature on the EUTECH INSTRUMENTS DO 110?

The Auto-Off feature automatically powers off the instrument 20 minutes after the last key press to conserve battery power. It is activated by default.

1. From measurement mode, press the SETUP key.

2. Use the MI/▲ or MR/▼ keys to scroll to parameter P4.0 (AtO).

3. Press the HOLD/ENTER key.

4. Use the MI/▲ or MR/▼ keys to select “YES” (to enable Auto-Off) or “NO” (to disable Auto-Off).

5. Press HOLD/ENTER to confirm your selection and return to the P4.0 menu.


How do I reset the calibration data on my EUTECH INSTRUMENTS DO 110?

This function resets only the calibration values for % Saturation and mg/L (or ppm) back to factory defaults.

1. From measurement mode, press the SETUP key.

2. Scroll to parameter P5.0 (rSt) and press HOLD/ENTER. You are now in the Calibration Reset menu (P5.1 CAL).

3. Use the MI/▲ or MR/▼ keys to toggle between “NO” (retains current settings) and “YES” (erases calibration settings).

4. Press the HOLD/ENTER key to confirm your selection.

– If you select “NO”, the meter proceeds to the User Reset program (P5.2).

– If you select “YES”, all LCD segments will light up for one second, calibration values are erased, and the meter returns to measurement mode.


How do I perform a full user reset on my EUTECH INSTRUMENTS DO 110?

A user reset clears all data, including calibration, memory, and customized setup functions, returning the meter to factory default settings.

1. From measurement mode, press the SETUP key.

2. Scroll to parameter P5.0 (rSt).

3. Press HOLD/ENTER twice to skip the Calibration Reset and go to the User Reset menu (P5.2 USr).

4. Use the MI/▲ or MR/▼ keys to select “YES” (resets to factory default) or “NO” (retains current settings).

5. Press ENTER to confirm.

– If you select “NO”, the meter retains settings and goes back to program P5.0.

– If you select “YES”, all LCD segments will light up for one second, all data is reset, and the meter returns to measurement mode.


How do I clear all stored memory on the EUTECH INSTRUMENTS DO 110?

Use this parameter to clear all 100 memory locations. NOTE: This action will wipe out all memory permanently.

1. From measurement mode, press the SETUP key.

2. Scroll to parameter P6.0 (CLr) and press HOLD/ENTER.

3. Use the MI/▲ or MR/▼ keys to toggle between “NO” (retains memory) and “YES” (clears all memory).

4. Press HOLD/ENTER to confirm your selection and return to the subgroup menu.


How do I set the barometric pressure on my EUTECH INSTRUMENTS DO 110?

The barometric pressure must be set for accurate DO measurements and calibration. This is done in setup program P7.0.

1. From the mg/L or ppm measurement mode, press SETUP.

2. Scroll to parameter P7.0 (dPr) and press HOLD/ENTER.

3. The meter enters the barometric pressure unit selection menu. Use the MI/▲ and MR/▼ keys to toggle between mmHg (“HG”) and kilo Pascal (“PA”). Press HOLD/ENTER to confirm.

4. The display now shows the pressure value. Use the MI/▲ and MR/▼ keys to set the current barometric pressure.

5. Press HOLD/ENTER to confirm the setting. The meter will then move to the Salinity setting display.

6. Press CAL/MEAS to return to measurement mode.


How do I set the salinity value on my EUTECH INSTRUMENTS DO 110?

Salinity compensation is required for accurate DO measurement in saline solutions. This setting is available in the mg/L (ppm) Set Up mode only.

1. Follow steps 1-5 for setting the Barometric Pressure to arrive at the Salinity setting menu (“SAL”).

2. Use the MI/▲ and MR/▼ keys to enter the salinity of your solution in ppt (parts per thousand).

3. Press HOLD/ENTER to confirm the value. The meter will return to the subgroup menu “dPr”.

4. Press CAL/MEAS to return to the measurement mode.


How do I adjust the % Saturation Offset on my EUTECH INSTRUMENTS DO 110?

This feature allows you to offset the meter’s value to match a reference meter. This is done in setup program P8.0 from the % Saturation measurement mode.

1. Dip the DO electrode in the sample solution and allow the reading to stabilize.

2. Check the reading of a reference DO meter in the same sample.

3. Press the SETUP key to enter the % Saturation Set Up menu.

4. Scroll to parameter P8.0 (OFS) and press HOLD/ENTER. The display will show the last offset adjusted value.

5. Use the MI/▲ and MR/▼ keys to enter the new value to match your reference meter. You can adjust within a +/- 10.0% offset.

6. Press HOLD/ENTER to confirm the offset adjustment. The display returns to the subgroup menu.

7. Press CAL/MEAS to exit to the measurement mode.

Note: When a new user calibration is performed, the offset will be reset to zero.


What are the general care instructions for the EUTECH INSTRUMENTS DO 110 probe?

Proper care will ensure maximum probe life and accurate readings.

– Deposits on the membrane surface act as a barrier to oxygen diffusion and must be cleaned off regularly.

– After use, rinse the probe in clean water and wipe it with a soft cloth or paper to prevent hardening of deposits.

– If growth develops on the probe, use a disinfecting chemical to clean it.

– NOTE: Although the membrane is strong, wipe it gently during cleaning. If the membrane is damaged or torn, the probe will no longer function and the membrane cap must be replaced.

– There are no special probe storage requirements.


How do I replace the membrane cap on the EUTECH INSTRUMENTS DO 110 probe?

Replacement is required when you cannot calibrate the probe or if the membrane is damaged (e.g., punctures, wrinkles).

1. Unscrew the old membrane cap housing.

2. Hold the probe under hot running water and use a stiff plastic brush to gently brush away the white oxide on the cylindrical anode. Do not use metal cleaning material.

3. If the cathode has deposits, remove them with a light scouring powder. Do not polish the cathode.

4. Take the new membrane cap housing and fill it with electrolyte solution. Inspect the bottom for leaks. If it leaks, use a different new cap.

5. If the assembly is leak-free, fill the membrane cap housing with electrolyte to the brim.

6. Tap the side of the housing gently to remove any air bubbles sticking to the membrane.

7. Screw the cap onto the probe. Excess electrolyte will drain out.

8. Replace the probe guard.

9. Calibrate the probe after the % saturation readings have stabilized.


What do the error messages on the EUTECH INSTRUMENTS DO 110 mean?

ERROR MESSAGE INDICATES POSSIBLE CAUSE CORRECTIVE ACTION
Err. Annunciator Wrong keypad input Wrong input in selected mode. Release key. Select valid operations depending on mode.
Battery icon lights up Low Battery Battery power is low Replace batteries with fresh ones as soon as possible.

How do I troubleshoot problems with my EUTECH INSTRUMENTS DO 110?

Problem Probable Causes Solution
Nothing is displayed when the ON/OFF key is selected. Batteries not in place. Insert batteries.
Batteries (+ & – poles) not in correct polarity. Re-insert batteries in the correct polarity.
Weak batteries. Replace batteries or attach AC adapter.
Fluctuating readings when probe is shaken or bumped lightly or when membrane is touched. Probe has lost electrolyte – a sloshing noise will be heard when the probe is shaken. Unscrew the membrane cap, discard the electrolyte, membrane and O-ring.
Torn or damaged membrane. Dry the internals of the probe, especially the cathode with a soft cloth. Switch the meter ON and observe the display. If the display reads zero, the probe and cable circuitry are alright; service probe and change membrane. If the display does not read zero, but some other value, there is likely moisture somewhere else (see next point).
Wet connections in the wiring or within probe.
With membrane cap removed and probe internals thoroughly dry, the reading from the probe is not zero and/or is erratic. Moisture has entered the system – either into the probe itself or at junctions or points in the cable. This moisture creates a secondary galvanic action. Locate the source of moisture by process of elimination. If moisture has entered a junction box or a cable joint, thoroughly dry out the area and take measures to prevent reoccurrence.
It is not possible to calibrate the probe in air – the display will not read high enough after fully adjusting the offset. Probe has dried out – no electrolyte inside. Service probe and change membrane.
Probe is overdue for servicing – excessive build-up of anode oxide. Use a stiff nylon brush to remove the loose oxide built-up from the anode. Do not use a wire brush. If the anode is badly corroded, replace with a new DO probe.
A deposit has built-up on the silver cathode, which is inhibiting the reduction of oxygen at its surface. If a deposit is suspected on the silver cathode, clean it with 400 grit wet/dry emery paper or with some scouring powder. The deposit is sometimes visible as a brownish stain. NOTE: The cathode must not be polished – the surface must remain dull.
Display values are erratic when membrane is lightly touched. Membrane has bulged outwards. The membrane has been damaged. If the membrane has been damaged, change it and service the probe.

How do I troubleshoot printing problems between my computer and the EUTECH INSTRUMENTS DO 110?

POSSIBLE CAUSES SOLUTIONS
You have not clicked “ENABLE CONNECTION” in the Data Acquisition program. Click on “ENABLE CONNECTION” in the Data Acquisition program.
The “Communication Settings” in the Data Acquisition program is different from meter’s setup. Match the COM port number, baud rate, parity and stop bits information between the Data Acquisition program and the meter.
The COM port number in the Data Acquisition program is wrong. Change the COM port number (1 or 2) in the Data Acquisition program.
Your computer’s COM port setting may be wrong. Check your computer’s hardware settings (through Windows OS, BIOS, or any other OS) and refer to the computer’s manual or consult with the computer’s manufacturer.
You may have used the wrong communication cable. Make sure you use the RS232C cable supplied with the meter. Check the RS232C configuration as described in the meter’s instruction manual.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD EUTECH INSTRUMENTS DO 110 (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.