What is included with the HANNA INSTRUMENTS HI702 meter?
Each HANNA INSTRUMENTS HI702 meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide
Please examine the product carefully upon receipt. If any damage occurred during shipment, contact your local Hanna Instruments Office.
What are the technical specifications of the HANNA INSTRUMENTS HI702?
| Range | 0.00 to 5.00 ppm |
| Resolution | 0.01 ppm |
| Accuracy | ±0.05 ppm ±5% of reading @25 °C/77 °F |
| Light Source | Light Emitting Diode @575 nm |
| Light Detector | Silicon Photocell |
| Method | Adaptation of the EPA approved method. The reaction between copper and the bicinchoninate reagent causes a purple tint in the sample. |
| Environment | 0 to 50 °C (32 to 122 °F); max 95% RH non-condensing |
| Battery Type | 1.5V AAA (1 pc.) |
| Auto-Shut off | After 10 minutes of non-use |
| Dimensions | 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5″) |
| Weight | 52 g (1.84 oz.) |
How do I perform a measurement with the HANNA INSTRUMENTS HI702?
1. Turn the meter on by pressing the button. All segments are displayed. When the display shows “Add”, “C.1” with “Press” blinking, the meter is ready.
2. Fill the cuvette with 10 mL of unreacted sample and replace the cap. Place the cuvette into the meter and close the meter’s cap.
3. Press the button. When the display shows “Add”, “C.2” with “Press” blinking the meter is zeroed.
4. Remove the cuvette from the meter and unscrew the cap. Add the content of one packet of HI702-0 reagent. Replace the cap and shake gently for about 15 seconds. Place the cuvette back into the meter.
5. Press and hold the button until the timer is displayed on the LCD (the display will show the countdown prior to the measurement) or, alternatively, wait for 45 seconds and press the button.
6. The instrument displays the concentration of copper in ppm. The meter automatically turns off after 10 minutes.
What are some tips for getting an accurate measurement with my HANNA INSTRUMENTS HI702?
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI731318 or a lint-free cloth prior to insertion.
• Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent is added, as accuracy will be affected.
• After the reading it is important to immediately discard the sample, otherwise the glass might become permanently stained.
How do I manage and replace the battery in my HANNA INSTRUMENTS HI702?
To save the battery, the instrument shuts down after 10 minutes of non-use. One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the instrument will display “bAd” then “bAt” for 1 second and then turns off. To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative end first.
• Insert the battery cover and replace the screw with a screwdriver.
What do the error and warning messages on the HANNA INSTRUMENTS HI702 mean?
| L.H. | Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette. |
| L.Lo | Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette. |
| Inu | Inverted Cuvettes: The sample and the zero cuvette are inverted. |
| Blinking “0.00” | Under Range: A blinking “0.00” indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement. |
| Flashing value | Over Range: A flashing value of the maximum concentration indicates the reading is over range. Dilute the sample and re-run the test. |
| bAt | Battery Low: The battery must be replaced soon. |
| bAd then bAt | Dead Battery: This indicates that the battery is dead and must be replaced. Once this indication is displayed, normal operation of the instrument will be interrupted. Change the battery and restart the meter. |
What accessories are available for the HANNA INSTRUMENTS HI702?
| Reagent Sets | |
|---|---|
| HI702-25 | Reagents for 25 Copper High Range tests |
| Other Accessories | |
|---|---|
| HI702-11 | Copper HR Certified Standard Kit |
| HI731225 | Cuvette Black Cap for Checker® HC Colorimeters (4 pcs.) |
| HI731318 | Cloth for wiping cuvettes (4 pcs.) |
| HI731321 | Glass cuvettes and Seal Cap for Checker® HC Colorimeters (4 pcs.) |
| HI731353 | Cuvette Seal Cap for Checker® HC Colorimeters (4 pcs.) |
| HI740028P | 1.5V AAA batteries (12 pcs.) |
| HI93703-50 | Cuvette cleaning solution (230 mL) |
What are the general recommendations for using the HANNA INSTRUMENTS HI702?
Before using Hanna Instruments products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipment, thus requiring the operator to take all necessary steps to correct such interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC performance. To avoid damages or burns, do not put the instrument in a microwave oven. For your and the instrument safety do not use or store the instrument in hazardous environments.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
CLICK HERE TO DOWNLOAD HANNA INSTRUMENTS HI702 (01) PDF DOCUMENT