NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

How do you switch the meter ON or OFF?

To switch ON or OFF: press ON/OFF (①).


How do you select the DO, BOD, OUR or SOUR range?

To select the DO, BOD, OUR or SOUR range: press RANGE.


How do you read contextual help?

To read contextual help: press HELP.


How do you return to the previous screen?

To return to the previous screen: press ESC.


What is the procedure for probe conditioning?

Remove the protective cap.

Moist the sensor for 5 minutes by soaking it in electrolyte solution (HI7041S).

Rinse the membrane cap with electrolyte solution while shaking it gently.

Refill with clean electrolyte solution.

Tap the probe repeatedly to remove any air bubbles that may be trapped inside.

With the sensor facing down, slowly screw the cap clockwise and wait 15 minutes before proceeding.


How do you take D.O. measurements?

To take D.O. measurements: with meter off, connect the electrode’s DIN connector to appropriate socket and immerse the electrode in the solution. Switch ON the meter. Stir briefly and wait for a stable reading. For accurate readings make sure you have a calibrated meter and a conditioned probe.


How do you select the D.O. unit?

To select the unit, D.O. (mg/L) or D.O. (%): press MODE.


How do you enter or exit Calibration mode?

To enter/exit Calibration mode: press CAL.


What is the temperature compensation feature of the D.O. probe?

The included D.O. probe has a built-in temperature sensor allowing automatic temperature compensation (0.0 °C to 50.0 °C).


What is the D.O. calibration procedure for zero oxygen solution?

Up to two point automatic calibration or 1 point manual calibration are available. From calibration screen select DO.

Place the probe in zero oxygen solution (HI7040).

Wait for “CFM”. Press CFM to accept the calibration.

For one-point calibration in 0.0 % press ESC after the appropriate accepted solution. The meter returns to measurement mode.


What is the D.O. calibration procedure for 100.0% calibration?

For 100.0% calibration dry the probe and take measurements in the air.

Wait for “CFM”. Press CFM to accept the calibration. The meter returns to measurement mode.

Note: For one-point calibration in 100.0 % start calibration with 100.0 % point. The meter returns to measurement mode.


What is the pressure calibration procedure?

From calibration screen select: Pressure.

Use ▲/▼ keys to enter the real local barometric pressure from a reference barometer.

Wait for “CFM”. Press CFM to accept calibration. The meter returns to measurement mode.


What is the temperature calibration procedure at 0 °C?

From calibration screen select T.

Place the DO probe into a vessel at 0 °C.

Use ▲/▼ keys and set the calibration point at the temperature read from the reference thermometer.

Wait for “CFM”. Press CFM to accept the calibration.


What is the temperature calibration procedure at 50 °C?

For one point calibration press ESC.

Place the DO probe into a vessel at 50 °C and proceed as in one point calibration.


How do you enter or exit SETUP view mode?

To enter/exit SETUP view mode, press SETUP while in measurement mode.


How do you toggle between displayed parameters in SETUP mode?

To toggle between the displayed parameters, press the ▲/▼ keys.


How do you enter or exit SETUP changing mode?

To enter/exit SETUP changing mode: press Modify or simply select option.


How do you store the current reading into memory?

To store the current reading into the memory: press LOG while in measurement mode.


How do you view logged data?

To view logged data: press RCL while in measurement mode.


How do you view more details about logged data?

To view more details: press More.


How do you delete records?

To delete records: press Delete key.


How do you delete all records?

To delete all records: press Delete All.


How do you freeze the first stable reading on the LCD using AutoEnd mode?

To freeze the first stable reading on the LCD: press AutoEnd while in measurement mode; the “WAIT” tag will blink until the reading will stabilize, then “Hold” will be displayed.


How do you return to measurement mode from AutoEnd mode?

To return to the measurement mode: press Continue.