HANNA INSTRUMENTS HI98192 (01) Image

FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF USER GUIDE
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF USER MANUAL
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF OWNER GUIDE
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF OWNER MANUAL
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF REFERENCE GUIDE
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF INSTRUCTION GUIDE
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF REFERENCE MANUAL
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF INSTRUCTION MANUAL
FREE ENGLISH HANNA INSTRUMENTS HI98192 (01) PDF OPERATING INSTRUCTIONS


CLICK HERE TO DOWNLOAD HANNA INSTRUMENTS HI98192 (01) PDF MANUAL

If this is not the document you want for this product, click here to see if we have any other documents for this product.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

What are the specifications for the HI98192 meter?

SPECIFICATIONS

Measurement Range
EC 0.001 to 9.999 µS/cm*, 10.00 to 99.99 µS/cm, 100.0 to 999.9 µS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm
Actual conductivity 1000.0 mS/cm
0 to 400 mS/cm (Shows values up to 1000.0 mS/cm)
Resistivity 1.0 to 99.9 Ω·cm, 100 to 999 Ω·cm, 1.00 to 99.9 ΚΩ·cm, 10.0 99.9 ΚΩ·cm, 100 to 999 ΚΩ·cm, 1.00 to 9.99 ΜΩ·cm, 10.0 to 100.0 ΜΩ·cm*
TDS 0.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 g/L, 10.00 to 99.99 g/L, 100.0 to 400.0 g/L (autoranging)
Salinity % NaCl 0.0 to 400.0 % NaCl
SeaWater Scale 0.00 to 80.00 ppt
Practical Salinity 0.01 to 42.00 (PSU)
Temperature -20.0 to 120.0 °C (-4.0 to 248.0 °F)

ACCURACY

EC ±1% of reading (±0.01 µS/cm or 1 digit, whichever greater)
Resistivity ±1% of reading (10 Ω·cm or 1 digit, whichever greater)
TDS ±1% of reading (±0.05 ppm or 1 digit, whichever greater)
Salinity ±1% of reading
Temperature ±0.2 °C (±0.4 °F, excluding probe error)

* The 0.001 µS/cm EC range and 100 MΩ·cm Resistivity range are not available with the 4 m cable probe.


How do I switch the meter ON and OFF?

To switch ON and OFF: press ON/OFF (①).


How do I take measurements?

With the meter off, connect the electrode’s DIN connector to the appropriate socket, remove the protective cap and immerse the electrode in the solution with the sleeve’s holes completely submerged. Switch ON the meter. Stir briefly and wait for a stable reading. For accurate readings tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.


How do I select the range, conductivity, resistivity, TDS, or salinity?

To select the range, conductivity, resistivity, TDS or salinity: press RANGE.


How do I enter USP or Salinity range while in EC range?

To enter USP or Salinity range while in EC range: press MODE.


How do I enter Calibration mode?

To enter Calibration mode: press CAL while in EC or Salinity range.


How do I change the calibration solution?

To change calibration solution: press the ▲/▼ keys.


What is the EC (Resistivity/TDS) calibration procedure?

Up to five calibration points are available:

• Keep the probe in air for the first point.

• Wait for “CFM”. Press CFM to confirm offset calibration.

• Place the probe in the appropriate standard solution. The instrument will detect the standard.

• Wait for “CFM” blinking. Press CFM to accept the calibration. Repeat until all five calibration points are done.

• To perform a one, two, three or four point calibration press ESC after the appropriate point was confirmed.

Note: Resistivity and TDS readings are automatically derived from the EC reading and no specific calibration for TDS is needed.


What is the NaCl calibration procedure?

• Place the probe in 100.0% NaCl standard solution.

• Wait for “CFM” blinking.

• Press CFM to accept the calibration. The meter returns to measurement mode.


How do I use the SETUP mode?

• To enter/exit SETUP mode: press SETUP while in measurement mode.

• To toggle between the displayed parameters: press the ▲/▼ keys.

• To enter/exit SETUP changing mode: press Modify or simply select option.


How do I use the GLP (Good Laboratory Practice) mode features?

• To store the current reading into the memory: press LOG while in measurement mode.

• To view logged data: press RCL while in measurement mode.

• To view more details: press More.

• To delete records: press Delete key.

• To delete all records: press Delete All.


How do I use the AutoEnd mode?

• To freeze the first stable reading on the LCD: press AutoEnd while in measurement mode; the “WAIT” tag will blink until the reading will stabilize, then “Hold” will be displayed.

• To return to the measurement mode: press Continue.