How do I unpack and assemble the MEIJI TECHNO 5550D microscope?
Follow these steps for unpacking and initial assembly of your MEIJI TECHNO 5550D microscope:
Unpacking:
1. The microscope stand and accessories are packed in separate styrofoam containers. Remove the microscope stand from the container and place it on a solid bench.
2. Do not throw away any boxes or packing materials until the contents have been checked against the packing list.
Bulb Installation:
1. Mount the bulb in the bulbholder by pushing and twisting it 1/4 turn clockwise.
Transmitted Illuminator Installation:
1. Mount the Transmitted illuminator on the microscope base by inserting it into the mounting ring at the rear of the base with the screw (a) and the filter holders (b) uppermost.
2. Clamp in place with the clamp screw.
Oblique Illuminator Installation:
1. Insert the bulb into the Oblique illuminator as described in the bulb installation step.
2. Remove the Clamp Ring from the microscope body by turning it counterclockwise.
3. Put the Ring arm of the Oblique illuminator over the fine focusing knob and clamp it in place using the clamp ring.
4. Plug in and adjust the position of the illuminator to light up the specimen correctly. The oblique illuminator can be mounted on either side of the microscope.
Vertical Illuminator Installation:
1. Insert the bulb as previously described.
2. Remove the dust cap from the top of the microscope body and loosen the knurled Lock ring by turning it counterclockwise.
3. Plug in and insert the vertical illuminator into the microscope body, and clamp with the knurled Lock ring.
Main Tube, Eyepiece, and Objectives:
1. Mount the Main tube on the microscope by screwing it into the body.
2. Insert the Eyepiece into the Main tube. Adjust the direction of the reticle so that the dotted crosslines are correctly horizontal and vertical, then lock it with the Lock ring.
3. Focus the reticle sharply by turning the upper part of the eyepiece.
4. Mount the objective/objectives on the objective nosepiece. For Model MTM253, mount the objectives starting with the lowest magnification, and increasing magnification as the nosepiece is turned counterclockwise.
Stage Preparation:
1. Remove the metal transit rods, which prevent movement of the stage during shipping, by loosening the clamp screws. Be careful as the stage is spring-loaded.
2. Install the Digital micrometer head by inserting the stem into the mounting bracket and clamping it in place. Note: Do not over-tighten the clamp screw as this will affect the operation of the micrometer head.
3. Place the Stage glass in a recess of the stage.
4. Mount the Stage clips on the stage by inserting pins into the holes on the rotatable stage. Two different sizes of stage clips are supplied.
Final Check:
Caution: Ensure that the voltage selector on the rear of the microscope stand is set to your supply voltage before plugging the mains cord into the outlet.
How do I operate the illumination controls on the MEIJI TECHNO 5550D?
The light controls are located on the front panel of the microscope base. Here are the functions:
1. Power switch: Main power ON/OFF.
2. Switch for Oblique illuminator, left: Controls the left oblique illuminator.
3. Switch for Vertical illuminator: Controls the vertical illuminator.
4. Intensity control for oblique illuminator: Adjusts the brightness of the oblique illuminators.
5. Intensity control for Vertical/Transmitted illuminator: Adjusts the brightness of the vertical and transmitted illuminators.
6. Switch for Oblique illuminator, right: Controls the right oblique illuminator.
7. Switch for Transmitted illuminator: Controls the transmitted illuminator.
Bulb Types:
• MA357 8V, 12W frosted bulb is used for the Transmitted illuminator.
• MA358 8V, 12W clear bulb is used for Vertical and Oblique illuminators.
Operational Notes:
a. All switches are 3-step, with a neutral point in the middle position. It is suggested to control light intensity within 8V.
b. Centering adjustment of the lamp filament can be done with two centering screws.
c. When observing a specimen of high reflectivity, set the Vertical illuminator by turning the beam-splitter knob to align the Ph mark with the white dot on the body.
d. Use the transmitted illuminator to examine a specimen of low light reflection or to measure inclined surfaces.
e. Use the transmitted illuminator to measure the diameter or pitch of holes, rod diameters, or when observing transparent specimens.
How do I properly focus the MEIJI TECHNO 5550D microscope?
a. Place a specimen on the stage under the stage clips. If the stage clips are not required, remove them by pulling upward.
b. Focus the objective on the specimen by rotating the large coarse focusing knob until the image of the specimen is bright and clear. Then turn the Locking-lever to lock the body.
c. Then, bring the specimen into sharper focus by using the fine adjustment knob.
(The working distance for a 10X objective is 6.6mm)
How do I measure the X and Y axes of a specimen with the MEIJI TECHNO 5550D?
a. Rotate the stage so that the movement of the specimen coincides with the horizontal line of the eyepiece reticle.
b. Using the micrometer head, move the stage to bring the starting point of the measurement under the center of the crossline reticule. Now, read the graduation on the Micrometer.
c. Move the stage to bring the end point of the measurement under the center of the crossline reticle. Read the graduation. The difference between the readings in b. and c. is the dimension of the object being measured.
d. For operation of the Digimatic Head (Digital micrometer head), please refer to the manual supplied with that instruction.
When using the Digital Micrometer Heads, please refer to the instruction manual for the DIGITAL MICROMETER HEAD as supplied by the manufacturer.
How do I measure the height or depth of a specimen with the MEIJI TECHNO 5550D?
The Digital indicator is used for measuring height/depth.
a. Attach the extension rod to the Stem before inserting it into the microscope.
b. Insert the Digital indicator into the gauge tube on the top of the microscope body and clamp with the knurled lock ring.
c. Focus the objective sharply on the highest point of the specimen to be measured and follow the instruction manual of the Digital indicator for measurement of height.
How do I set up the MEIJI TECHNO 5550D for television use with a camera?
For television, the MA580 TV Camera Adapter must be used.
a. Remove the Vertical illuminator and insert the MA580 TV Camera Adapter in its place.
b. Remove the C-mount from the top of the MA580 Adapter and thread it into the CK3800 TV Camera.
c. When using a TV Camera, vertical illumination is required. The MA581 vertical illuminator adapter must be used. The MA581 is mounted as follows:
c-1 Remove the objective and thread the reflector unit into the nosepiece.
c-2 Adjust the direction of the reflector unit to the correct position by loosening the four set screws.
c-3 Attach the illuminator housing to the reflector unit with the filter holders facing upward, and tighten the screws.
d. To direct all light to the TV camera, turn the beam-splitter knob so that the Op mark coincides with the white dot on the microscope body.
Please refer to the Instruction Manual for the CK3800 CCD Color TV Camera when using this camera.
How do I produce a crossline on the TV screen with the MEIJI TECHNO 5550D?
To produce a crossline on the TV screen, the MA151/30 1X Crossline Mount is used.
a. Remove the upper half of the MA580 TV Camera Adapter and mount the MA151/30 Crossline mount in its place.
b. If the crossline is not horizontal on the TV screen, it can be adjusted by rotating the tube of the MA151/30 mount.
How do I set up the MEIJI TECHNO 5550D for photography with a PMX100 Large Format Camera?
a. Remove the Vertical illuminator from the top of the microscope body.
b. Mount the MA582 Camera Adapter on the microscope body by inserting it in place of the vertical illuminator.
c. Mount the Camera on the MA582 Camera adapter and tighten the Clamp screw.
Please refer to the Instruction Manual for the PMX100 Large Format Photomicrographic Camera when using this camera.
How do I use the EX-1 Exposure meter with the MEIJI TECHNO 5550D and PMX100 camera?
The EX-1 Exposure meter is available to use with the PMX100.
a. Connect the EX-1 Exposure meter to the PMX100 by inserting the light sensor of EX-1 into the opening of the PMX100 and turn the switch of the EX-1 on.
b. Set the film speed number.
c. Read the speed on the Display of the EX-1 and set the camera shutter at the same speed and release the cable release to complete exposure.
Please refer to the separate Instruction Manual when using the EX-1 Exposure meter.
How do I set up the MEIJI TECHNO 5550D for photography with a 35mm SLR Camera Body?
For photomicrography with a 35mm SLR Camera body, the MA150/60 Camera attachment, Photo eyepiece, and T2 Adapter should be used.
a. Insert a photo eyepiece, either MA512 2.5x or MA508 5x, into the MA582 Camera Adapter.
b. Mount the MA150/60 Camera attachment on the MA582 Camera adapter and tighten the clamp screw.
c. Mount the T2 adapter that matches your 35mm SLR Camera body on the MA150/60 attachment and attach the 35mm SLR Camera body to it.
d. Focus the objective on the specimen looking through the Finder eyepiece of the MA150/60 Attachment.
Please refer to the separate Instruction Manual for MA150/60 Camera attachment.
What are the general care instructions for the MEIJI TECHNO 5550D microscope?
Always cover the instrument with the plastic dust cover provided when the microscope is not in use. Keep the eyepiece in the microscope body at all times to prevent dust from falling on the internal optics.
Store the microscope in a safe, clean, and dry place when not in use for an extended period of time.
How should I clean the MEIJI TECHNO 5550D microscope?
Clean exposed lens surfaces carefully with a pressurized air source, soft brush, or clean soft cloth. Too much finger pressure may damage lens coatings.
To remove oil, fingerprints, and grease smudges, moisten the cleaning cloth with a very small amount of alcohol.
Painted or plastic surfaces should be cleaned only with a cloth moistened with water and a small amount of detergent.
DO NOT ATTEMPT TO MAKE ADJUSTMENTS TO THE INTERNAL OPTICS OR MECHANISMS.
If the microscope does not seem to be functioning properly or you have questions about its operation, call your supplier (or an authorized repair service) for advice.
How do I change the operating voltage on the MEIJI TECHNO 5550D?
1. Unscrew the Fuse holder and remove the fuse.
2. Remove the larger diameter ring labeled “voltage selector” by pulling it away from the microscope. You will see contact pins and voltage markings (e.g., 110V, 120V, 220V, 240V).
3. The ring has a notch cut out for voltage selection. Reinsert the ring so the notch aligns with your desired voltage, making it visible.
4. Install a power cord with the correct pin pattern for your region.