Free English AEMC INSTRUMENTS SL361 (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS SL361 (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

What are the meanings of the international electrical symbols on the AEMC INSTRUMENTS SL361?

Symbol

Meaning

Square within a square

Signifies that the instrument is protected by double or reinforced insulation.

Exclamation mark in a triangle

CAUTION – Risk of Danger! Indicates a WARNING. Whenever this symbol is present, the operator must refer to the user manual before operation.

I with an arrow pointing right

To identify the phase, or the direction, of the primary current.

Lightning bolt in a square

Application or withdrawal authorized on conductors carrying dangerous voltages. Type A current sensor as per IEC 61010-2-032.

Plus and minus signs

This symbol signifies a voltage limiting circuit.

Battery symbol

Battery

USB symbol

USB socket

Letter ‘i’ in a circle

This symbol indicates Important information to acknowledge.

CE

This product complies with the Low Voltage & Electromagnetic Compatibility European directives.

Crossed-out wheeled bin

In the European Union, this product is subject to a separate collection system for recycling electrical and electronic components in accordance with directive WEEE 2012/19/EU.

Eco Conception logo

Chauvin Arnoux® and AEMC® Instruments have adopted an Eco-Design approach in order to design this instrument. Analysis of the complete lifecycle has enabled us to control and optimize the effects of the product on the environment. In particular this instrument exceeds regulation requirements with respect to recycling and reuse.


What are the safety precautions for using the AEMC INSTRUMENTS SL361?

Failure to observe the safety instructions may result in electric shock, fire, explosion, and destruction of the instrument and of the installations. The following precautions must be followed:

• The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.

• Use personal protection equipment (PPE) that meets or exceeds the environment in which the operator is working.

• If you use this instrument other than as specified, the protection it provides may be compromised, thereby endangering you.

• Do not use the instrument on networks of which the voltage or category exceeds those mentioned in this document.

• Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.

• Before each use, check the condition of the insulation on the leads and housing. Any item of which the insulation is deteriorated (even partially) must be set aside for repair or scrapping.

• When handling the instrument, keep your fingers behind the guard.

• Do not expose the current probe to sprays of water.

• All troubleshooting and metrological checks must be done by competent, accredited personnel.

• Use ONLY a wall plug adapter rated ≤1 A.


What are the measurement categories defined for the AEMC INSTRUMENTS SL361?

The device is compliant with safety standard IEC/EN 61010-2-032 or BS EN 61010-2-032, for voltages up to 600 V in CAT III or 300 V in CAT IV.

CAT IV: Corresponds to measurements performed at the primary electrical supply (< 1000 V). Example: primary overcurrent protection devices, ripple control units, and meters.

CAT III: Corresponds to measurements performed in the building installation at the distribution level. Example: hardwired equipment in fixed installation and circuit breakers.

CAT II: Corresponds to measurements performed on circuits directly connected to the electrical distribution system. Example: measurements on household appliances and portable tools.


What is included with the AEMC INSTRUMENTS SL361 shipment?

The AC/DC Current Probe Model SL361 (Cat. #2153.09) includes:

• (1) 9 V (6LR61) alkaline battery

• a user manual


What are the main features of the AEMC INSTRUMENTS SL361 and what instruments is it compatible with?

The AEMC INSTRUMENTS SL361 can be used to measure current from 100 mA to 100 Apeak without breaking the circuit. It measures the waveform and amplitude of the current in the form of a voltage, with a bandwidth from DC to 100 kHz. It can be powered by a battery or 5 V via the micro-USB connector.

Instrument Features:

• a reset push button,

• a range overload indicator,

• a power supply indicator,

• automatic standby to prolong battery life.

Compatibility:

The Model SL361 is compatible with any analog or digital oscilloscope or other voltage-measuring instrument which has the following features:

• BNC input connector

• Range capable of displaying (0.2 to 0.5) V per division

• Minimum input impedance of >1 MΩ

• Jaw opening can accommodate up to a single 4/0 AWG (11.684 mm diameter) conductor.


How do I turn on the AEMC INSTRUMENTS SL361?

Switch the instrument ON by pushing the slide switch to the 10 mV/A or the 100 mV/A setting.

• The 10 mV/A setting corresponds to the 100 A range.

• The 100 mV/A setting corresponds to the 10 A range.

The ON indicator lights green. If it blinks, there is less than 4 hours of use. If it fails to light, you must replace the battery or plug into USB power. It takes 10 seconds to set up the instrument for use.


How do I perform a zero adjustment on the AEMC INSTRUMENTS SL361?

1. Switch the probe on.

2. Connect the probe to the measuring instrument. The phase is on the core of the BNC connector.

3. Make sure that the probe is not on a conductor and that its jaws are correctly closed.

4. Place the probe in the position it will be in during the measurement.

5. Press the zero adjustment button.

6. The OL indicator lights for approximately three seconds to indicate that the zero adjustment is in progress in the two ranges.

7. If the zero has been correctly adjusted, the OL indicator goes off. If it remains on, the zero could not be adjusted.

In case of adjustment failure, check that the probe is not on any conductor and its jaws are correctly closed, then press the zero adjustment button again. Alternatively, switch the probe off and back on; the last stored adjustment will be used.


How do I take a measurement with the AEMC INSTRUMENTS SL361?

NOTE: The zero must be adjusted before each measurement. WARNING: Always connect the probe to the instrument before clamping onto the circuit under test.

1. Once the zero has been adjusted, press the mobile arm of the probe handle to open the jaws.

2. Clamp the cable carrying the current to be measured.

3. The arrow on the probe must point in the presumed direction of current flow (from load source to load).

4. Release the mobile arm gently and make sure that the jaws are correctly closed.

5. The value measured is displayed on the measuring instrument.

If the OL indicator lights, it means the current is too strong to be measured. If you are in the 100 mV/A range, switch to the 10 mV/A range.

NOTE: Remember to unclamp the probe from the conductor before disconnecting it from your meter or instrument.


How does the Auto Off feature work on the AEMC INSTRUMENTS SL361?

At the end of 10 minutes of operation with no sign of the user’s presence (a press on the zero adjustment button or an action on the switch), the probe switches to standby and the ON indicator goes off.

To wake up the probe, press the zero adjustment button or set the switch to a position other than OFF.


How can I disable the Auto Off feature on the AEMC INSTRUMENTS SL361?

To disable auto off (operation in permanent mode), press the zero adjustment button when starting up the instrument. The ON indicator blinks to indicate that the request has been taken into account, then lights steady orange when you release the zero adjustment button.

When the probe is switched off (switch set to OFF), the auto off function is re-activated.


What do the indicator lights on the AEMC INSTRUMENTS SL361 mean?

ON Indicator

State

Meaning

Off

Instrument Off

Lighted green

Instrument On

Blinking green

The battery will have to be replaced in less than 4 h.

Lighted orange

Permanent operation (auto off deactivated)

OL Indicator

State

Meaning

Off

The measurement is correct.

Lighted red

The measurement exceeds the range.

Lighted red for 3 seconds

The zero adjustment is in progress.


Can I power the AEMC INSTRUMENTS SL361 externally?

Yes, for long-term measurements, you can connect the clamp to an external power source using an optional line power adapter. You can use any micro-USB adapter that supplies 5 VDC with a minimum 50 mA output or more.

When the instrument is powered via the micro-USB connector, automatic standby is disabled.

WARNING: The type B micro-USB connector must not be in contact with conductors or uninsulated parts at dangerous voltage.


How do I clean the AEMC INSTRUMENTS SL361?

• Disconnect the instrument completely and turn the rotary switch to OFF. Also make sure that no cable is clamped.

• Use a damp soft cloth and dry rapidly with a dry cloth or forced air.

• Do not use alcohol, solvents or hydrocarbons.

• The air gaps of the clamp must always be kept clean.

• Do no leave the probe in very damp places, or exposed to splashes of any liquid substances.


How do I replace the battery in the AEMC INSTRUMENTS SL361?

WARNING: Risk of electric shock. Disconnect all input(s) from the unit or remove the probe from any conductor before opening the battery cover to change the battery.

If the green battery indication LED does not light up when the probe is turned on, the battery will need to be replaced. Do not replace the battery while the probe is in use. Disconnect the current probe from the circuit and turn the current probe OFF.

To replace the battery:

1. Use a screwdriver to remove the screw securing the battery compartment.

2. Slide the battery compartment cover away and off from the current probe.

3. Connect the battery to the snap-on connector; take care with the polarity.

NOTE: You can use a rechargeable Ni-MH battery, but the life will be shorter. The instrument does not recharge rechargeable batteries.

4. Place the battery in its compartment.

5. Slide the battery compartment cover back in place; make sure that it is completely and correctly closed.

6. Screw the battery compartment cover screw back in place.


What are the power sources for the AEMC INSTRUMENTS SL361?

The instrument is powered by a 9 V battery (type 6LR61 or NEDA 1604A). The typical battery life is 80 h with an alkaline battery.

The instrument can also be powered by an external supply (5 V, 50 mA) via the type B micro-USB connector.


What are the environmental operating conditions for the AEMC INSTRUMENTS SL361?

The device must be used in the following conditions:

Range of reference: (18 to 28)°C (64.4 to 82.4) °F from (20 to 75) % RH

Operating range: (-10 to 50) °C (14 to 122) °F from (10 to 45) % RH and (-10 to 35) °C (14 to 95) °F from (10 to 85) % RH

Storage range: (-30 to 80) °C (-22 to 176) °F from (0 to 30) % RH and (-30 to 50) °C (-22 to 122) °F from (0 to 85) % RH

Indoor use.

Pollution degree: 2

Altitude: < 6562 ft, < 2000 m

Transport altitude: ≤ 39,370 ft, ≤ 12,000 m


What are the mechanical specifications of the AEMC INSTRUMENTS SL361?

Dimensions (L x W x H): (9.09 x 1.42 x 2.64) in (231 x 36 x 67) mm

Weight: approximately 11.6 oz (330 g) with battery

Measurement Lead Length: 6.5 ft (2 m)

USB Cord Length: 5.9 in (15 cm)

Clamping capacity: .46 in (11.8 mm) in diameter

Protection by the housing: IP 20 per IEC 60529


What are the frequency characteristics of the AEMC INSTRUMENTS SL361?

10 mV/A 100 mV/A
Bandwidth to 3 dB down (DC to 100) kHz
Rise time (10 to 90) % and fall time (90 to 10) % 3 µs
Response time to 10 % 1.8 µs
Insertion impedance at 10 kHz 2 mΩ
Insertion impedance at 50 kHz 10 mΩ

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS SL361 (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.