AEMC INSTRUMENTS MN193-BK (01) Image

Free English AEMC INSTRUMENTS MN193-BK (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS MN193-BK (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

What precautions should be taken before using an AEMC INSTRUMENTS MN193-BK current probe?

The protection assured by the current probe can be compromised if it is used in a way that is not recommended by the manufacturer.

■ Comply with the rated maximum voltage and current, and the measurement category. Do not use the current probe on networks where the voltage or category exceeds those specified.

■ Comply with the conditions of use (e.g. temperature, humidity, altitude, degree of pollution, location).

■ Do not use the current probe if its housing is open, deteriorated, or incorrectly reassembled. Before each use, check the integrity of the insulation of the unit, jaws, clamps, housing, and leads.

■ Do not subject the current probe to water or other liquids.

■ Keep the jaw contacts of the clamp absolutely clean.

■ Use suitable personal protective equipment when hazardous voltages may be accessible in the installation where the measurement is made.

■ Any repairs must be carried out by accredited skilled personnel.


What do the safety symbols found on AEMC INSTRUMENTS current probes mean?

Warning symbol WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
High voltage symbol CAUTION! Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.
Type A current sensor symbol Application or withdrawal authorized on conductors carrying dangerous voltages. Type A current sensor as per IEC 61010-2-032.
Type B current sensor symbol Must not be applied to or removed from conductors at dangerous voltages. Type B current sensor as per IEC 61010-2-032.
Double insulation symbol Equipment is protected by double insulation.
Battery symbol Battery
Exclamation mark symbol Important instructions to read and to fully understand.
Information symbol Useful information or tip to read.
CE The CE marking guarantees conformity with European directives and with regulations covering EMC.
WEEE symbol The trash can with a line through it means that in the European Union, the product must undergo selective disposal for the recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.

What are the definitions of Measurement Categories (CAT) for the AEMC INSTRUMENTS MN193-BK?

CAT IV: Measurement category IV corresponds to measurements taken at the source of low-voltage installations. Example: power feeders, counters and protection devices.

CAT III: Measurement category III corresponds to measurements on building installations. Example: distribution panel, circuit-breakers, machines or fixed industrial devices.

CAT II: Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations. Example: power supply to domestic electrical appliances and portable tools.


What should I do when I receive my AEMC INSTRUMENTS MN193-BK shipment?

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim.


What are the main control features of the AEMC INSTRUMENTS MN193-BK AC Current Probe?

1. Jaws

2. Protective guard

3. Two-position range switch (MN193 only)

4. Jaw opening lever

5. Shielded lead

6. Custom 4-pin input connector


How do I operate the AEMC INSTRUMENTS MN193-BK current probe?

The current probes are used to measure the current flowing in a conductor or bus bar without opening the circuit. They also insulate the user from dangerous voltages in the circuit. The choice of current probe depends on the amperage to be measured and the diameter of the cables or size of the bus bar.

■ For three-phase measurements, use the color-coded ID markers to associate a color for each current input to match the phase identifiers on the measured system.

IMPORTANT NOTE: Always connect the probes or sensors with the indicating arrows pointing toward the load.

■ Connect the current probe or sensor to the current terminals of the instrument.

For Probes:

■ Press the jaw opening lever on the probe to open the jaws.

■ Clamp the probe around the conductor to be tested. For best results, center the conductor in the jaws of the clamp.


What are the electrical specifications for the AEMC INSTRUMENTS MN193-BK and other compatible probes?

NOTE: The measurement ranges specified are for the probes and sensors. In some cases, they may differ from the ranges that can be measured by the instrument with which they are used with. For complete specifications: Refer to the product user manual that is supplied with each compatible instrument.

Model Measurement Range Compatibility
AmpFlex® 193-24-BK (1) 24″ (610mm) 200mA to 10kAAC (2) (12,000A) (3) 8333, 8336 & PEL Series
AmpFlex® 193-36-BK (1) 36″ (910mm) 200mA to 10kAAC (2) (12,000A) (3) 8333, 8336 & PEL Series
AmpFlex® 196A-24-BK (1) 24″ (610mm) 200mA to 10kAAC (2) (12,000A) (3) 8435
MiniFlex® MA193 (1) 10″ (250 mm) 200mA to 3000A (10,000A peak) 8333, 8336 & PEL Series
MiniFlex® MA193 (1) 14″ (355 mm) 200mA to 3000A (10,000A peak) 8333, 8336 & PEL Series
J93 50 to 3500; 50 to 5000 (DC only) 8333, 8336, 8435 and PEL
MN93 2 to 240AAC (I >200A not permanent) All PowerPads and PEL
MN193 5A: 0.005 to 6AAC
100A: 0.1 to 120AAC
All PowerPads and PEL
MR193 10 to 1000AAC;
10 to 1300A PEAK AC+DC
All PowerPads and PEL
SL261 100mV/A: 100mA to 10A peak
10mV/A: 1 to 100A peak
8333, 8336, 8435 and PEL
SR193 1 to 1200AAC (I > 1000A not continuously) All PowerPads and PEL

(1) 10 to 6500AAC for Model 8435

(2) 200mA to 10,000A measurement range for the PEL 100 Series.

(3) 12,000A is specified for the PEL 100 Series only.


What are the environmental operating conditions for the AEMC INSTRUMENTS MN193-BK probe?

Indoor use.

Operating Temperature: 14° to 131°F (-10° to 55°C); 10% to 85% RH

Storage Temperature: -40° to 158°F (-40° to 70°C); 10% to 90% RH

Degree of pollution: 2

Altitude: < 2000 m


What are the mechanical specifications for the AEMC INSTRUMENTS MN193-BK and other compatible probes?

Model Lead Length (nominal) Clamping Diameter Dimensions Weight
AmpFlex® 193-24-BK 24″ (610mm) 10 ft (3m) 7.64″ (190mm) 6.6 x 6.2 x .98″ (170 x 158 x 25mm) 7.7 oz (270g)
AmpFlex® 193-36-BK 36″ (910mm) 10 ft (3m) 11.46″ (290mm) 11 x 10.4 x .98″ (280 x 265 x 25mm) 9.5 oz (220g)
AmpFlex® 196A-24-BK 24″ (610mm) 10 ft (3m) 7.64″ (190mm) 6.6 x 6.2 x .98″ (170 x 158 x 25mm) 7.7 oz (270g)
MiniFlex® MA193 -10-BK 10″ (250mm) 10 ft (3m) 2.75″ (70mm) 4.0 x 2.5 x 1.1″ (103 x 64 x 28mm) 1.94 oz (55g)
MiniFlex® MA193-14-BK 14″ (350mm) 10 ft (3m) 3.94″ (100mm) 4.0 x 2.5 x 1.1″ (103 x 64 x 28mm) 2.11 oz (60g)
J93 10 ft (3m) 2.84″ (72mm) 13.23 x 5.00 x 1.65″ (336 x 127 x 42mm) 3.75 lbs (1.7kg)
MN93 10 ft (3m) 0.8″ (20mm) 5.47 x 2.00 x 1.18″ (135 x 51 x 30mm) 24 oz (690g)
MN193 10 ft (3m) 0.8″ (20mm) 5.47 x 2.00 x 1.18″ (135 x 51 x 30mm) 24 oz (690g)
MR193 10 ft (3m) One 1.6″ (42mm) or two 0.98″ (25mm) or two bus bars 1.96 x 0.19″ (50 x 5mm) 8.8 x 3.82 x 1.73″ (224 x 97 x 44mm) 19 oz (540g)
SL261 6.5 ft (1.9m) 0.46″ (11.8mm) 9.09 x 1.42 x 2.64″ (231 x 36 x 67mm) 11.6 oz (330g)
SR193 10 ft (3m) 2″ (52mm) 8.5 x 4.4 x 1.8″ (216 x 111 x 45mm) 24 oz (690g)

What are the safety specifications for the AEMC INSTRUMENTS MN193-BK current probe?

■ Protection index IP 40 for the probes and IP 30 jaws open, according to IEC 60 529

■ IP 65 for the AmpFlex® according to IEC 60 529

■ IK 04 according to IEC 50102

■ Drop test: According to IEC 61010-1

■ Electrical safety according to IEC 61010-2-032.

Maximum applicable voltage for MN93/MN193:

600V CAT III; 300V CAT IV


How do I clean my AEMC INSTRUMENTS MN193-BK current probe?

CAUTION: Risk of electric shock. Disconnect the instrument from any source of electricity.

■ Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth.

■ Do not use alcohol, solvents, or hydrocarbons.

■ Do not splash water directly on the instrument.


What is the recommended procedure for calibration of the AEMC INSTRUMENTS MN193-BK?

To ensure that your instrument meets factory specifications, we recommend that it be scheduled back to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS MN193-BK (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.