AEMC INSTRUMENTS MR193-BK (01) Image

Free English AEMC INSTRUMENTS MR193-BK (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS MR193-BK (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

What do the symbols on the AEMC INSTRUMENTS MR193-BK mean?

A (in triangle) WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
Lightning bolt (in triangle) CAUTION! Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.
Clamp with lightning bolt Application or withdrawal authorized on conductors carrying dangerous voltages. Type A current sensor as per IEC 61010-2-032.
Strikethrough clamp with lightning bolt Must not be applied to or removed from conductors at dangerous voltages. Type B current sensor as per IEC 61010-2-032.
Square within a square Equipment is protected by double insulation.
Battery icon Battery
Exclamation mark (in circle) Important instructions to read and to fully understand.
“i” (in circle) Useful information or tip to read.
CE The CE marking guarantees conformity with European directives and with regulations covering EMC.
Crossed-out trash can The trash can with a line through it means that in the European Union, the product must undergo selective disposal for the recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.

What are the definitions of Measurement Categories (CAT) relevant to the AEMC INSTRUMENTS MR193-BK?

CAT IV: Measurement category IV corresponds to measurements taken at the source of low-voltage installations. Example: power feeders, counters and protection devices.

CAT III: Measurement category III corresponds to measurements on building installations. Example: distribution panel, circuit-breakers, machines or fixed industrial devices.

CAT II: Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations. Example: power supply to domestic electrical appliances and portable tools.


What precautions should be taken before using the AEMC INSTRUMENTS MR193-BK?

The protection assured by the current probe can be compromised if it is used in a way that is not recommended by the manufacturer.

• Comply with the rated maximum voltage and current, and the measurement category. Do not use the current probe on networks where the voltage or category exceeds those specified.

• Comply with the conditions of use (e.g. temperature, humidity, altitude, degree of pollution, location).

• Do not use the current probe if its housing is open, deteriorated, or incorrectly reassembled. Before each use, check the integrity of the insulation of the unit, jaws, clamps, housing, and leads.

• Do not subject the current probe to water or other liquids.

• Keep the jaw contacts of the clamp absolutely clean.

• Use suitable personal protective equipment when hazardous voltages may be accessible in the installation where the measurement is made.

• Any repairs must be carried out by accredited skilled personnel.


What should I do upon receiving my AEMC INSTRUMENTS MR193-BK?

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim.


What are the control features of the AEMC INSTRUMENTS MR193-BK AC Current Probe?

1. Jaws

2. Protective guard

3. Jaw opening lever

4. Two-position range switch

5. Shielded lead

6. Zero adjustment

7. Custom 4-pin input connector


How do I operate the AEMC INSTRUMENTS MR193-BK current probe?

The current probes and flexible sensors are used to measure the current flowing in a conductor or bus bar without opening the circuit. They also insulate the user from dangerous voltages in the circuit.

For three-phase measurements, use the color-coded ID markers to associate a color for each current input to match the phase identifiers on the measured system.

IMPORTANT NOTE: Always connect the probes or sensors with the indicating arrows pointing toward the load.

• Connect the current probe or sensor to the current terminals of the instrument.

• For the MR193 probe: Set the switch to 1mV/A; the ON indicator will light up.

• Connect the probe to the instrument.

• Adjust the zero by turning the potentiometer with no conductor in the jaws of the clamp.

• When the measurement is finished, turn the probe switch to OFF.


How do I clamp the AEMC INSTRUMENTS MR193-BK probe onto a conductor?

• Press the jaw opening lever on the probe to open the jaws.

• Clamp the probe around the conductor to be tested. For best results, center the conductor in the jaws of the clamp.


What are the electrical and battery specifications for the AEMC INSTRUMENTS MR193-BK?

NOTE: The measurement ranges specified are for the probes and sensors. In some cases, they may differ from the ranges that can be measured by the instrument with which they are used.

Model Measurement Range Compatibility
MR193 10 to 1000AAC; 10 to 1300A PEAK AC+DC All PowerPads and PEL

Battery: 9V Alkaline NEDA 1604A, 6LR61

Battery Life: MR193 – 100H typical

NOTE: Battery life is twice the typical value when using Lithium batteries.


What are the environmental operating conditions for the AEMC INSTRUMENTS MR193-BK?

• Indoor use.

• Operating Temperature: 14° to 131°F (-10° to 55°C); 10% to 85% RH

• Storage Temperature: -40° to 158°F (-40° to 70°C); 10% to 90% RH

• Degree of pollution: 2

• Altitude: < 2000 m


What are the mechanical specifications for the AEMC INSTRUMENTS MR193-BK?

Model Lead Length (nominal) Clamping Diameter Dimensions Weight
MR193 10 ft (3m) One 1.6″ (42mm) or two 0.98″ (25mm) or two bus bars 1.96 x 0.19″ (50 x 5mm) 8.8 x 3.82 x 1.73″ (224 x 97 x 44mm) 19 oz (540g)

What are the safety ratings for the AEMC INSTRUMENTS MR193-BK?

• Protection index IP 40 for the probes and IP 30 jaws open, according to IEC 60 529

• Drop test: According to IEC 61010-1

• Electrical safety according to IEC 61010-2-032.

• Maximum applicable voltage for MR193: 600V CAT III; 300V CAT IV


How do I clean the AEMC INSTRUMENTS MR193-BK?

CAUTION: Risk of electric shock. Disconnect the instrument from any source of electricity.

• Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth.

• Do not use alcohol, solvents, or hydrocarbons.

• Do not splash water directly on the instrument.


How do I replace the battery in the AEMC INSTRUMENTS MR193-BK?

1. Disconnect the MR193 completely and turn the rotary switch to OFF.

2. Use a screwdriver to unscrew the screws and remove the battery compartment cover on the backside of the unit.

3. Withdraw the battery from its compartment.

4. Disconnect the old battery without pulling on the wires and replace with a new one, observing the polarity.

5. Put the battery into its compartment.

6. Put the cover back in place and screw the screws back in.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS MR193-BK (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.