Free English PCE INSTRUMENTS PCE-GA 10 (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.
CLICK HERE TO DOWNLOAD PCE INSTRUMENTS PCE-GA 10 (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
What are the safety precautions for using the PCE INSTRUMENTS PCE-GA 10?
Please read the manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel. Damage or injuries caused by non-observance of the manual are excluded from liability.
• The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
• The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under any circumstances.
• Non-observance of the safety notes can cause damage to the device and injuries to the user.
• The meter must not be used as a gas detector.
What are the main features of the PCE INSTRUMENTS PCE-GA 10?
The PCE INSTRUMENTS PCE-GA 10 has the following features:
• Detection of flammable gases
• Vibration alarm when gases are detected
• Automatic calibration
• Measurement range up to 10000 ppm
• Audible alarm, 85 dB
• Sensor exchangeable
• Shelf life of the sensor approx. 5 years
• Rechargeable battery for mobile use
What gases can the PCE INSTRUMENTS PCE-GA 10 detect?
The PCE INSTRUMENTS PCE-GA 10 can detect the following gases and compounds in which these gases occur:
• Ammonia
• Benzene
• Ethane
• Ethanol
• Ethylene
• Formaldehyde
• Hexane
• ISO-Butane
• Methane
• Propane
• P-Xylene
• Hydrogen sulfide
• Toluene
• Hydrogen
What are the technical specifications of the PCE INSTRUMENTS PCE-GA 10?
| Measurement range | Low concentration: 0 … 1000 ppm, High concentration: 0 … 10000 ppm (methane only) |
| Audible alarm | Volume: 85 dB |
| Sensitivity | <50 ppm (methane) |
| Measuring interval | <2 seconds |
| Display | 5 red LEDs |
| Calibration | Automatic |
| Warm-up time | 50 seconds |
| Battery | Polymer Li-Ion battery 18500 3.7 V 1400 mAh |
| Mains adaptor | Primary side: 100 … 240 V, 50/60 Hz, 0.2 A Secondary side: 5 V, 1 A |
| Auto Power Off | Turns off automatically when battery is low or after 10 minutes of inactivity |
| Sensor shelf life | approx. 5 years (sensor is exchangeable) |
| Probe | Semi-rigid, approx. 500 mm |
| Weight | Approx. 430 g |
What is included in the delivery contents of the PCE INSTRUMENTS PCE-GA 10?
The delivery contents include:
• 1 x gas leak detector PCE-GA 10
• 1 x charger
• 1 x user manual
Can you describe the components of the PCE INSTRUMENTS PCE-GA 10?
The components of the PCE INSTRUMENTS PCE-GA 10 are as follows:
1. LED warm-up
2. LED low sensitivity
3. Automatic zero calibration on/off
4. Vibration alarm on / off
5. Meter on / off
6. LED display
7. LED high sensitivity
8. Sensitivity high / low
9. Auto Power Off
10. Hold on / off
11. Audible alarm on / off
What steps should be taken before the first use of the PCE INSTRUMENTS PCE-GA 10?
• Fully charge the battery before the first use. The charging socket is located on the right side above the sensor arm bracket.
• Switch on the meter using the “On / Off” key and leave the device switched on for about one hour for the sensor to acclimatise. Deactivate the automatic power-off function with the APO key and set the sensitivity to low. The alarm LEDs will gradually go off. As soon as only LED 1 flashes, the acclimatization is complete.
• If you do not use the device for more than two months, repeat this procedure.
What checks should be performed before each use of the PCE INSTRUMENTS PCE-GA 10?
• Check the meter for mechanical damage.
• Only switch on the unit in an environment unpolluted by gas.
• Carry out a gas test with the appropriate test gas.
• Attention! When using a lighter, only move the sensor head very slowly towards the gas source. A 100 % concentration exposure can destroy the sensor.
• If the gas is not detected, the device is not ready for use.
• After the gas test, wait until all alarm LEDs have gone off when low sensitivity is set.
How do I operate the PCE INSTRUMENTS PCE-GA 10 for leak detection?
• Make sure the gas to be detected is listed in the manual.
• Only turn on the meter in an environment unpolluted by gas.
• Press and hold the “On / Off” key for about 1 second. The meter will vibrate and a signal tone will sound.
• Let the meter warm up for 50 seconds. The lower LED will flash.
• The end of the warm-up time is confirmed by the unit by all alarm LEDs flashing 4 times and the signal tone sounding 4 times.
• When the L and H LEDs are lit simultaneously, charge the device.
• Select the desired sensitivity with the “H/L” key (8).
• Clean the areas to be examined from oil and dust. Oil and dust can impair the detection.
• Slowly move the sensor head to the area to be examined.
• If there is a gas leak, the instrument will signal that gas escapes (see alarm chart), depending on the sensitivity setting. Keep in mind the reaction time of at least 2 seconds. Move the sensor head back a little to exactly locate the leak.
• Switch off the meter after the leak detection. To do this, press and hold the “On / Off” key for about 1 second.
Can you explain the alarm chart for methane (CH4) on the PCE INSTRUMENTS PCE-GA 10?
The alarm chart for methane (CH4) indicates the device’s response at different concentrations, depending on the sensitivity setting.
| Approx. reading in ppm | Alarm LEDs | Frequency | |||||
|---|---|---|---|---|---|---|---|
| high | low | 1 | 2 | 3 | 4 | 5 | audible alarm x/s |
| 0 | 0 | Off | Off | Off | Off | Off | no alarm |
| 0 … 100 | 0 … 1000 | On | Off | Off | Off | Off | 1 |
| 100 … 400 | 1000 … 4000 | On | On | Off | Off | Off | 1.2 |
| 400 … 700 | 4000 … 7000 | On | On | On | Off | Off | 1.65 |
| 700 … 1000 | 7000 … 10000 | On | On | On | On | Off | 3.25 |
| >1000 | >10000 | On | On | On | On | On | 6.25 |
How does the Auto Zero function work on the PCE INSTRUMENTS PCE-GA 10?
What is the Auto Power Off function on the PCE INSTRUMENTS PCE-GA 10?
How can I adjust the sensitivity on the PCE INSTRUMENTS PCE-GA 10?
How do I control the alarms on the PCE INSTRUMENTS PCE-GA 10?
How can I troubleshoot common issues with the PCE INSTRUMENTS PCE-GA 10?
| Error | Possible cause | Solution |
|---|---|---|
| Meter cannot be turned on | Battery flat | Charge battery |
| L and H LEDs are flashing simultaneously | Battery very low | Charge battery |
| All alarm LEDs and the L and H LEDs are flashing simultaneously | Sensor defective | Replace sensor ESS-PCE-GA 10 |
What are the disposal instructions for the PCE INSTRUMENTS PCE-GA 10 and its batteries?
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU, the device can be returned. It will either be re-used or given to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
CLICK HERE TO DOWNLOAD PCE INSTRUMENTS PCE-GA 10 (01) PDF DOCUMENT
