FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF DOCUMENT.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF USER GUIDE.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF USER MANUAL.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF OWNER GUIDE.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF OWNER MANUAL.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF REFERENCE GUIDE.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF INSTRUCTION GUIDE.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF REFERENCE MANUAL.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF INSTRUCTION MANUAL.
FREE ENGLISH KINETIK WELLBEING WW-200 (01) PDF OPERATING INSTRUCTIONS.
CLICK HERE TO DOWNLOAD KINETIK WELLBEING WW-200 (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
What are the main components of the KINETIK WELLBEING WW-200?
The main components of the KINETIK WELLBEING WW-200 are:
1. Indicator Light
2. Intensity – button
3. Mode / On-Off button
4. Intensity + button
What are the key features of the KINETIK WELLBEING WW-200?
The key features of the KINETIK WELLBEING WW-200 include:
• Targeted pain relief
• 6 TENS programmes
• Rechargeable
• Drug free pain relief
• Clinically validated
What is included in the box with the KINETIK WELLBEING WW-200?
The KINETIK WELLBEING WW-200 box includes:
• Control Unit
• Belt Clip
• Wires for Precision Pads
• 1x Type-C Charging Cable
• Pair of Precision Pads x2
• Storage Pouch
• User Manual
What is the intended use of the KINETIK WELLBEING WW-200 TENS device?
The KINETIK WELLBEING WW-200 is a Transcutaneous Electrical Nerve Stimulation (TENS) unit intended to provide temporary relief from muscle soreness caused by exercise, household tasks, or work activities. It is also effective in alleviating chronic pain, intractable pain, and pain associated with arthritis. It is important to apply the electrode pads only on intact skin. The TENS device is suitable for adult users, including laypersons and professionals.
Are there any contraindications for using the KINETIK WELLBEING WW-200?
Yes. Patients with implanted electronic devices such as pacemakers, life-sustaining medical devices (e.g., artificial hearts or lungs), or electrocardiographs should not use the KINETIK WELLBEING WW-200 device.
How do I prepare the electrode pads for the KINETIK WELLBEING WW-200?
1. Attach the end of the electrode cord with the two metal buttons to the two electrode pads.
2. Connect the other end of the electrode cord with one plug into the type-C socket on the bottom of the main unit.
3. Each electrode pad is protected by a layer of transparent film. Remove the layer of film before applying the pads to the skin.
Note:
• Hold the plug when pulling it out. Do not pull on the wire.
• Never stick two adhesive electrode pads together.
• The electrode pads must fit precisely inside the conductive surface.
• The electrode pad has a service life and it is recommended to use the electrode pads less than 20 times in total. The specific number of times depends on usage and storage conditions.
How do I use the KINETIK WELLBEING WW-200 TENS device for treatment?
1. Clean the areas of skin you intend on placing the electrode pads.
2. Place the electrode pads around the treatment areas.
3. Hold down the power button for 2 seconds to turn it on.
4. Press the power button to choose your desired treatment mode.
5. After selecting your treatment mode, the default intensity is set to zero. Use the + or – button to adjust the intensity and start the treatment.
6. Once the 15-minute treatment cycle is completed, the intensity will change to 0 and the main unit will automatically shut off.
7. Press and hold the power button for 2 seconds to power off the main unit during the treatment.
Note: Precision pads must be used in pairs. A single pad only will not work.
How do I set the volume on the KINETIK WELLBEING WW-200?
The KINETIK WELLBEING WW-200 has 4 volume settings, from 0 (mute) to 3, with a default of 1.
To change the volume:
1. Continuously hold the – button. After 3 seconds, press the + button. The indicator light will flash slowly to show it is ready to change the volume setting.
2. Press the + button until you have chosen your desired volume.
3. Release the – button to save the settings and power off the device.
To activate mute:
Press the – and + buttons simultaneously for two seconds to set the volume to 0 (mute), then the device will turn off.
What are the different treatment modes and sensations for the KINETIK WELLBEING WW-200?
| Mode | Sensation |
|---|---|
| Composite | Less intense, numbing electrical stimulation |
| Knock | Knocking feeling that alternates from fast to slow and back to fast |
| Rub | A slight sense of kneading |
| Tapping | Tapping sensations that vary in duration, intensity, and interval |
| Muscle Relaxation | A special waveform designed to relieve muscle fatigue |
| Press | Simulates manual pressing, providing a repeated pressing sensation |
How do I charge the KINETIK WELLBEING WW-200?
Ensure the main unit is fully charged before first use. The KINETIK WELLBEING WW-200 will indicate it needs to be charged when the indicator light continuously blinks.
1. Ensure the device is off before charging.
2. Insert the TYPE-C port of the charging cable into the main unit and insert the USB port of the charging cable into an USB compatible plug. Plug it into a socket and switch it on.
3. The TENS Device will indicate that it is charging when there is an initial beep and the indicator light flashes. This will change to a steady light once it is fully charged.
Avoid positioning the device in a manner that obstructs access to the disconnection device.
During normal use, when the indicator light continuously blinks and continuous buzzes appear, it indicates that the power is insufficient and needs to be charged.
Warning:
• Do not plug or unplug the power cord into the electrical outlet with wet hands.
• Do not overload power outlets, ensure to plug the device into the appropriate voltage outlet.
• If the AC adapter is abnormal, please change the adapter.
• Do not use any other type of charging cable as it may harm the main unit.
• Don’t use this main unit when charging.
How should I dispose of the KINETIK WELLBEING WW-200 and its accessories?
The main unit, the lithium batteries, the electrode pads, and the electrode cords must be disposed of according to local regulations at the end of their usage.
How should I care for and maintain my KINETIK WELLBEING WW-200?
• Avoid dropping or subjecting the device to strong impacts.
• Avoid high temperatures. Do not immerse the device in water, as this will result in damage.
• Do not disassemble or attempt to repair the device or components.
• Lithium battery replacement should only be performed by a qualified service centre technician.
• Clean the main unit with a lightly moistened cloth (or a cloth soaked in a neutral cleaning solution) and wipe gently. Do not use chemicals (like thinner, benzene).
• Wash the electrode pads when the adhesive surface becomes dirty, or they are difficult to attach.
• To wash the pads, first, remove the electrode cord from the electrode pads. Then wash the electrode pads softly with your fingertips under slow-running cold water for several seconds (do not use a sponge/cloth/sharp object like a nail on the adhesive side, and do not use detergents, chemicals, or soap.) Dry the pads and let the adhesive surface air dry completely. Do not wipe with tissue paper or cloth.
• Please wait for approximately 2 hours for the device to warm up or cool down before use when the device is transferred from the temperature of storage and transportation to the normal operating temperature environment.
• Your device was tested before being sold and does not require calibration or regular maintenance.
What should I do if the KINETIK WELLBEING WW-200 will not turn on properly?
If the main unit cannot be turned on properly, the battery may be exhausted. Please charge the battery.
Why am I feeling no stimulation from the KINETIK WELLBEING WW-200?
If you are feeling no stimulation, please check the following:
• Is the electrode cord properly connected? Ensure the main unit and the pads are connected.
• Has the protective film been removed? Remove the protective film from the pads before using.
• Is the intensity greater than 0? Press the intensity + button to increase intensity.
• Are the pads firmly adhered? Ensure the pads are firmly adhered to the skin for optimal performance.
Why is the stimulation from my KINETIK WELLBEING WW-200 weak?
If the stimulation feels weak, please check the following:
• Are the electrode pads overlapped? Separate the pads and ensure the pads are individually attached to the skin.
• Is there dirt on the electrode pads? Gently clean the pads to maintain their effectiveness.
• Is the intensity too low? Increase the intensity to a level that suits your personal comfort.
• Are the pads placed correctly? Please correctly connect the electrode pads and attach them to the appropriate body parts.
What should I do if my skin becomes red or uncomfortable when using the KINETIK WELLBEING WW-200?
If your skin becomes red or uncomfortable, check the following:
• Are the electrode pads too dry? Rinse with a small amount of running water, scrub gently with your fingers for a few seconds, and dry naturally before using.
• Do the electrode pads stick closely to the skin? Stick the electrode pads closely to the skin.
• Are the electrode pads dirty? Clean the electrode pads.
• Is the surface of the electrode pads scratched? Replace the electrode pads with new ones.
What should I do if the KINETIK WELLBEING WW-200 cuts power during use?
If the power cuts off during use, please check the following:
• Are the electrode pads detached from the skin? Turn off the power and stick the electrode pads firmly to the skin.
• Are the electrode cords disconnected? Turn off the power and connect the electrode cords.
• Has the battery been exhausted? Charge the battery.
How does the KINETIK WELLBEING WW-200 work to relieve pain?
The electrical impulses {Output pulse frequency: 0-100 Hertz (Hz); Output voltage: max. 110 peak-to-peak voltage (Vpp) (500 ohms); Output pulse width: 0-250 µs} generated by the TENS device are transmitted to the nerves under the skin to block or shut out the pain message from the source of pain to the brain. Additionally, these electrical impulses also increase the production of the body’s natural painkillers, such as endorphins. Furthermore, low-frequency vibrations can promote blood circulation and relieve pain.
What are the technical specifications of the KINETIK WELLBEING WW-200?
Product name: Precision Period Pain Reliever Transcutaneous Electrical Nerve Stimulators (TENS)
Model: AD-2300S
Classification: Internally powered; Class II, Type BF applied part; IP22; No AP or APG; Continuous operation; not intended to be used in an oxygen-rich environment
Size: Approx. 49.8mm X 49.8mm X 12mm (1.96″ X 1.96″ X 0.5″)
Weight: Approx. 18.5g (0.65 oz.) (exclude pads)
Electrode pads: a. Circular Electrode Pads: Approx. Ø50mm (Φ2.0″); Model: EP5036ΝΟ1
Electrode cords: Approx. 600 mm(23.6″); Electrode Cords model: TC6-3.8-0600
Output channel: 1
Number of treatments: 6 treatments
Intensity levels: 20
Output pulse frequency: 0-100 Hz
Output voltage: max. 110 Vpp (500 ohms)
Output pulse width: 0-250 µs
Timer: 15-minute countdown timer
Power source: Medical AC adapter: DC 5V 1A, Lithium battery: 1X3.7V Li-ion 100mAh
Environmental temperature for operation/storage: 5°C to 40°C (41°F to 104°F)
Environmental humidity and pressure for operation/storage: ≤80%RH, 80-105kPa
Environmental temperature for transport: -20°C to 55°C (-4°F to 131°F)
Environmental humidity for transport: ≤90%RH
Environmental pressure: 80 kPa to 105 kPa
Main unit life: Three (3) years
Battery life: Approx. 1 month (when used for one time a day)
What do the symbols on the KINETIK WELLBEING WW-200 and its packaging mean?
Symbol for “THE INSTRUCTION MANUAL MUST BE READ”: The sign background-color: blue. The sign graphical symbol: white.
Symbol for “TYPE BF APPLIED PARTS”: The electrode pads are type BF applied part.
Symbol for “ENVIRONMENT PROTECTION”: Electrical products should not be disposed of as household waste. Please recycle where facilities exist. Check with your local Authority or retailer for recycling advice.
Symbol for “MANUFACTURER”
Symbol for “COMPILES WITH REGULATION (EU) 2017/745”
Symbol for “DATE OF MANUFACTURE” and “MADE IN CHINA”
SN: Symbol for “SERIAL NUMBER”
IP22: The first characteristic numeral symbol for “Degrees of protection against access to hazardous parts and against solid foreign objects.” The second characteristic numeral symbol for “Degrees of protection against ingress of water.”
MR Unsafe Symbol: Symbol for “MR Unsafe”
EC REP: Symbol for “EUROPEAN REPRESENTATION”
Recycle Symbol: Symbol for “RECYCLABLE”
MD: Symbol for “MEDICAL DEVICE”
Class II Equipment Symbol: Class II Equipment
LOT: Batch Code
REF: Catalogue Number / Model Number
UDI: Unique Device Identifier
Importer’s Information Symbol: Importer’s Information
What are the important warnings I should know before using the KINETIK WELLBEING WW-200?
• Use alongside implanted electronic devices(e.g., pacemakers, intracardiac defibrillators, etc.) may result in electric shocks, burns, interference, or even be life-threatening.
• Simultaneous connection of a PATIENT to high-frequency surgical medical electrical (ME) equipment may result in burns at the site of the stimulator electrodes and possible damage to the device.
• For Hospitals and Clinics: In the presence of, or when attached to the body, electronic monitoring equipment (e.g., cardiac monitors, ECG alarms, etc.) may not operate properly when the electrical stimulation device is in use.
• Operation within proximity (e.g., 3 feet or 1 meter) to shortwave or microwave therapy equipment may produce instability in the stimulator output.
• Pregnant women should not use the device.
• This device is designed for adults and should never be used on infants or young children. Consult your physician or other health care professionals before use on older children.
• Never let children or persons who are incapable of expressing their own use the TENS device. Keep the device safely stored and inaccessible to children to prevent from swallowing the batteries or small parts.
• Never use the device while sleeping or driving. Never use the device in humid environments (e.g., bathroom), as it may cause intense stimulation.
• If you are allergic to the device’s material, please do not use it.
• See the section regarding ELECTROMAGNETIC COMPATIBILITY INFORMATION for information regarding potential electromagnetic interference (EMI). Please do not use the device within the environment of the following devices: Magnetic Resonance Imaging (MRI), computerized axial tomography (CT), diathermy, Radio Frequency Identification (RFID), active HF SURGICAL EQUIPMENT, and electromagnetic security systems such as metal detectors.
• Place the electrode pads on intact skin only. Do not place on cuts, wounds, or areas with skin conditions.
• Please do not share electrode pads to prevent the risk of infection and cross-contamination.
• Please do not use any other electrode pads and/or electrode cords other than those supplied by the original equipment manufacturer, otherwise injury, discomfort, or equipment damage could result.
• Medical AC adapter (input: AC 100-240V, 50/60Hz, 0.2A; output: DC 5V, 1A) is suitable for this device, please use the AC adapter that was supplied by the manufacturer.
• Keep all electrode cords away from babies and children to prevent suffocation and death from winding cords around the neck.
• If the electrode pads are damaged or the adhesive starts to weaken, please replace them immediately to prevent potential skin injuries.
• Do not use this device for anything other than its intended purpose.
• Limit therapy time to no more than 30 minutes per session if using the electrode pads on the same part of the body.
• Power off the main unit before shifting or moving the position of the electrode pads.
• During stimulation and therapy, please do not let any metal (e.g., a part of a leather belt, wristwatch, jewellery, etc.) touch the electrode pads.
• The device might not meet its performance specifications or may cause safety hazards if stored or used outside the specified temperature and humidity ranges listed in the specifications section.
• The output wave parameters are not to be influenced by load resistance, except output voltage.
Where should I avoid placing the electrode pads of the KINETIK WELLBEING WW-200?
Never apply the pads to these body areas:
a. Avoid placing the electrode pads near to the heart, or on both sides of the thorax simultaneously (lateral or front and back), or across your chest (especially on the two large pectoral muscles). Otherwise, it can increase the risk of ventricular fibrillation and may induce cardiac arrest.
b. Avoid placing the electrode pads on the following areas: head, eyes, mouth, any area of the face, neck (especially the carotid sinus), private parts, or affected skin. Once you feel unwell or your skin appears abnormal while using the device, please stop using it immediately and seek medical attention.
What is the electromagnetic compatibility (EMC) information for the KINETIK WELLBEING WW-200?
• The essential performance in normal working mode is output frequency OHz~100Hz, intensity max-110 Vpp (500 ohms) pulse waveform.
• When EMI affects the above performance, please stop using the device.
• Use of this equipment and accessories adjacent to, or stacked with, other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
• The use of accessories and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and improper operation.
• Equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
| Phenomenon | Compliance | Electromagnetic environment |
|---|---|---|
| Conducted and radiated RF emissions | CISPR 11 Group 1, Class B | The device is intended to be used in home healthcare environment |
| Harmonic distortion | IEC 61000-3-2 NA | The power of the device is below the 75-watt limit |
| Conducted and radiated RF emissions | IEC 61000-3-3 NA | The standard for low-power equipment, which does not produce voltage fluctuations and flickers, specifies that the EUT (Equipment Under Test) power should be less than 75W. |
| Phenomenon | Basic EMC Standard | Immunity Test Levels Home Healthcare Environment |
|---|---|---|
| Electrostatic Discharge | IEC 61000-4-2 | ±8 kV contact ±2kV, ±4kV, ±8kV, ±15kV air |
| Radiated RF EM field | IEC 61000-4-3 | 10V/m 80MHz-2.7GHz 80% AM at 1kHz |
| Proximity fields from RF wireless communications equipment | IEC 61000-4-3 | Refer to table 4 |
| Rated power frequency magnetic fields | IEC 61000-4-8 | 30A/m 50Hz or 60Hz |
| Proximity magnetic fields | IEC 61000-4-39 | Refer to table 3 |
| Test frequency | Modulation | Immunity Test Level (A/m) |
|---|---|---|
| 30 kHz | CW | 8 |
| 134.2 kHz | Pulse modulation 2, 1 kHz | 65 |
| 13.56 MHz | Pulse modulation 50 kHz | 7.5 |
| Test Frequency (MHz) | Band (MHz) | Immunity Test Levels Home Healthcare Environment |
|---|---|---|
| 385 | 380-390 | Pulse modulation 18Hz, 27V/m |
| 450 | 430-470 | FM, ±5 kHz deviation, 1k Hz sine, 28 V/m |
| 710 | 740-787 | Pulse modulation 217Hz, 9V/m |
| 745 | ||
| 780 | ||
| 810 | 800-960 | Pulse modulation 18Hz, 28V/m |
| 870 | ||
| 930 | ||
| 1720 | 1700-1990 | Pulse modulation 217Hz, 28V/m |
| 1845 | ||
| 1970 | ||
| 2450 | 2400-2570 | Pulse modulation 217Hz, 28V/m |
| 5240 | 5100-5800 | Pulse modulation 217Hz, 9V/m |
| 5500 | ||
| 5785 |
| Phenomenon | Basic EMC Standard | Immunity Test Levels Home Healthcare Environment |
|---|---|---|
| Electrostatic Discharge | IEC 61000-4-2 | ±8 kV contact ±2kV, ±4kV, ±8kV, ±15kV air |
| Conducted disturbances induced by RF fields a) | IEC 61000-4-6 | 3 V 0.15 MHz – 80 MHz 6 V in ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz |
| Phenomenon | Basic EMC Standard | Immunity Test Levels Home Healthcare Environment |
|---|---|---|
| Electrostatic Discharge | IEC 61000-4-2 | ±8 kV contact ±2kV, ±4kV, ±8kV, ±15kV air |
| Electrical fast transients / bursts | IEC 61000-4-4 NA | SIP/SOPS whose maximum cable length is less than 3 m in length are excluded. |
| Surges Line-to-ground | IEC 61000-4-5 NA | This test applies only to output lines intended to connect directly to outdoor cables. |
| Conducted disturbances induced by RF fields | IEC 61000-4-6 | 3 V 0.15 MHz – 80 MHz 6 V in ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz |
| Phenomenon | Basic EMC Standard | Immunity Test Levels Home Healthcare Environment |
|---|---|---|
| Electrical fast transients / bursts | IEC 61000-4-4 | ±2 kV 100kHz repetition frequency |
| Surges Line-to-ground | IEC 61000-4-5 | ±0.5 kV, ±1 kV |
| Conducted disturbances induced by RF fields | IEC 61000-4-6 | 3V, 0.15MHz-80MHz 6V in ISM and amateur radio bands between 0.15MHz and 80MHz 80% AM at 1kHz |
| Voltage dips | IEC 61000-4-11 | 0% UT; 0.5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0% UT; 1 cycle and 70% UT; 25/30 cycles Single phase: at 0° |
| Voltage interruptions | IEC 61000-4-11 | 0% UT: 250/300 cycles |