What are the important usage notices for the FLOMEC TM075-N-LP?
Use TM Series meters with water and other chemicals compatible with wetted components. Do not use to meter fuel or incompatible chemicals. This TM Series meter features a computer for local electronic display plus a pulse output cable delivering raw pulses from the meter electronics to provide a digital signal to customer interfacing equipment. The computer display measures in gallons or litres.
These meters are not legal for trade applications.
TM Series meters are very sensitive to electric noise if operated within 1 to 2 inches of some electric motors or other sources of electronic noise.
How should I install my FLOMEC TM075-N-LP water meter?
Install your meter in-line either horizontally or vertically or at the end of the hose adjacent to the nozzle. Installation to metal connections is not recommended. Install as follows:
1. Plan to install turbine with a minimum straight pipe length as follows:
• Upstream from the turbine, allow a minimum straight pipe length of 10 times the internal diameter of the turbine.
• Downstream from the turbine, allow a minimum straight pipe length of 5 times the internal diameter of the turbine.
2. For Spigot (Pipe) End use only primer and solvents approved for PVC gluing.
For NPT Fittings wrap all connections with 3 to 4 wraps of thread tape (optional to use pipe thread sealant). Make sure the tape does not intrude into the flow path.
3. Attach meter with arrow pointed in the direction of fluid flow.
4. For NPT Fittings – Hand tighten the meter at the housing ends. Do not use a wrench or similar tool to tighten. This can damage the housing.
What safety warnings and cautions should I be aware of when using the FLOMEC TM075-N-LP?
WARNINGS:
• Compatibility of this product’s material and the process fluid and/or environment should be considered prior to putting into service.
• Product should never be operated outside its published specifications for temperature or pressure. See specifications for your model.
• Make sure flow and pressure have been eliminated from process pipe prior to installing or removing product.
• Always use appropriate thread sealant when connecting product to process piping.
CAUTIONS:
• To protect against leakage, seal all pipe threads with an appropriate sealing compound. Make sure the sealing compound does not intrude into the flow path.
• NOTE: If connecting to new male pipe threads, burrs and curls can adversely affect accuracy. Correct the problem prior to turbine installation.
• Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.
How does the pulse output feature work on the FLOMEC TM075-N-LP?
The output pulse through the cable is a raw pulse. For each rotation of the turbine rotor, a specific number of pulses are transmitted via the attached cable. Pulse amounts can vary depending on pipe/meter size, flowrate, fluid type, etc.
The pulse cable White wire is signal and the Black wire is common (GND).
NOTE: Some interface devices may not have an internal pull-up resistor. Use a minimum 820 ohms resistor if necessary. The resistor should be connected between the 5-30 VDC supply (V+) and the White – PULSE wire going to the equipment’s input.
What are the typical pulses per gallon for the FLOMEC TM series meters?
A Calibration Report with specific calibration information for your TM Series meter is also included for your reference. The table below shows reference information.
| Meter Size | Typ. Pulses per Gallon |
|---|---|
| TM050 | 2,400 |
| TM075 | 1,100 |
| TM100 | 570 |
| TM150 | 213 |
| TM200 | 100 |
How can I verify the accuracy of my FLOMEC TM075-N-LP?
Before using, check the meter’s accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs steadily. Then, stop or divert the flow using a valve or nozzle.
2. Meter an exact known volume into an accurate container. For best results, meter with one continuous full stream.
3. Check the volume against the display or recording equipment. If the amount metered is accurate, further calibration is not necessary. If not, refer to the Calibration Section for further instructions.
How do the computer display’s Batch and Cumulative totals work on the FLOMEC TM075-N-LP?
The computer maintains two totals. The Cumulative Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled TOTAL 1, Batch Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display reading of 999,999 the computer will highlight an X10 icon. This indicates to the operator that a zero must be added to the 6 digits shown. When the next rollover occurs, the computer will highlight an X100 icon. This indicates to the operator that two zeros must be added to the 6 digits shown.
Press the DISPLAY button briefly to switch between the TOTAL 1, TOTAL 2 BATCH and FLOWRATE. Press DISPLAY briefly to display the TOTAL 2 BATCH. Hold the DISPLAY button for 3 seconds to reset the Batch Total to zero.
When fluid is flowing through the meter, a small propeller icon is highlighted.
NOTE: Totalization counts total units without differentiating between gallons, litres or field calibrated units.
How do I use the Flowrate feature on my FLOMEC TM075-N-LP?
To use this feature, press and release the DISPLAY button until the FLOWRATE icon appears. The factory set time base will be highlighted to the right of FLOWRATE (M = minutes, H = hours, D = days). When FLOWRATE is invoked, the display will be indicating the rate of flow.
How do I activate the FLOMEC TM075-N-LP meter?
The computer is on continuously and always ready to perform. The computer is powered by field replaceable batteries. When the display becomes dim, faded or the low battery message appears (LobAtt), the batteries need to be replaced.
How do I switch between factory calibration units on the FLOMEC TM075-N-LP?
All units are configured with a “factory” calibration. Both gallons and litres are available (“GL” or “LT” will be displayed). While holding the CALIBRATE button, briefly press DISPLAY to toggle between gallons and litres. This factory calibration (indicated with FAC) is permanently programmed into the computer and is not user adjustable.
NOTE: Your computer may have other units of measure programmed into it. If so, holding the CALIBRATE button and momentarily pressing the DISPLAY button will toggle through all factory set units. Other possible units are: IGL (imperial gallon), QT (quart), CF (cubic feet), CM (cubic meter), BL (42 gal. barrel), CC (cubic centimeter) or OZ (ounce).
Switching between different units will not corrupt the Total’s contents. For example, in GL mode, the computer totalizes 10.00 gallons, if the user switches to LT mode, the display will read 37.85 litres (the same volume, different unit).
What are the minimum flowrate requirements for calibrating the FLOMEC TM Series Meters?
| TM Series Meters | |
|---|---|
| 1/2 inch meter | 1 GPM (3.8 LPM) |
| 3/4 inch meter | 2 GPM (7.5 LPM) |
| 1 inch meter | 5 GPM (18.8 LPM) |
| 1 1/2 inch meter | 10 GPM (37.5 LPM) |
| 2 inch meter | 20 GPM (75 LPM) |
How do I perform a field calibration (Correction Factor Method) on the FLOMEC TM075-N-LP?
For the most accurate results, dispense at a flowrate which best simulates your actual operating conditions. Avoid “dribbling” more fluid or repeatedly starting and stopping the flow. This can result in less accurate calibrations. For best results, the meter should be installed and purged of air before field calibration.
1. To calibrate, press and hold the CALIBRATE and DISPLAY buttons for about 3 seconds until you see FLdCAL. Release both buttons and you will see CF – 00.0. You are now in the field calibration mode and values from -99.9% to +99.9% can be entered.
2. The +/- position appears either as an “underscore” character for plus, or as a “hyphen” character for minus. The DISPLAY button selects the position and the CALIBRATE button toggles this character.
3. The DISPLAY button can then be pushed to select the numeric positions. Press the CALIBRATE button to scroll from 0 to 9. Enter the percentage of change you want the display to correct. When satisfied with the value, press both CALIBRATE and DISPLAY buttons simultaneously. CALEnd will be displayed and unit will go back to normal operation, less the FAC (factory calibration) icon.
4. All enabled units-of-measure remain visible and selectable – the entered correction will be applied to all enabled units.
How do I return my FLOMEC TM075-N-LP to factory calibration?
To return to factory calibration (FAC), press and hold both CALIBRATE and DISPLAY buttons for about 3 seconds until FACCAL is displayed. Then release buttons. Unit should return to normal operation and FAC icon is visible.
How do I maintain the turbine rotor on my FLOMEC TM075-N-LP?
The meter is virtually maintenance-free. However, it is important the rotor moves freely. Keep the meter clean and free of contaminants.
If the rotor does not turn freely, apply a penetrating lubricant on the rotor, shaft and bearings. Remove any debris or deposits from the rotor using a soft brush or small probe. Be careful not to damage the turbine rotor or supports.
CAUTION: Blowing compressed air through the turbine assembly could damage the rotor.
CAUTION: Do not allow liquids to dry inside the turbine.
CAUTION: Handle the rotor carefully. Small scratches or nicks can affect accuracy.
How do I replace the batteries in my FLOMEC TM075-N-LP?
If the display becomes dim, blank or the low battery message appears (LobAtt), replace the batteries. The computer display is powered by two 3-volt lithium batteries which may be replaced while the meter is installed. When batteries are removed or lose power, the batch and cumulative totals and the field and factory calibrations are retained.
1. Remove the four Phillips-head screws from the face of the meter and lift the faceplate from the turbine.
2. Remove the old batteries and clean any corrosion from the terminals.
3. Install new batteries. Make sure the positive post is in the correct position.
4. When the batteries are replaced, the faceplate will power ON. Check the display to ensure normal functions have resumed before assembling again.
5. Reseat batteries, if necessary, and position the faceplate on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the faceplate.
WARNING: (Battery) – Avoid mechanical or electrical abuse. Batteries may explode or cause burns, if disassembled, crushed or exposed to fire or temperatures in excess of 212°F (100°C). Do not short circuit or install with incorrect polarity. DO NOT INCINERATE.
CAUTION: Batteries should ONLY be replaced with P/N 113520-1 Kit (Includes two each P/N 902004-2 Batteries). Do not mix old with new. Do not use other brands or technologies.
What are the specifications for the FLOMEC TM075-N-LP?
U.S. Measurement
• Unit of Measure: Gallon
• Inlet and Outlet: 3/4 inch NPT
• Flow Range: 2 – 20 GPM
• Accuracy with Computer: ± 3.0% of reading (Accuracy can be improved with field calibration)
• Operating Temperature: +32° to +140° F (Do not allow fluid to freeze inside meter.)
• Storage Temperature: -40° to +158° F
• Product Weight: .67 lbs.
• Dimensions – Inches (W x H x L): 2.1 x 2.9 x 6.1 (With NPT Fitting)
• Cable Length: 5 FT.
General Specifications
• Design Type: Turbine
• Wetted Components:
– Housing: PVC
– Journal Bearings: Ceramic
– Shaft: Tungsten Carbide
– Rotor and Supports: PVDF
– Retaining Washer: Stainless Steel
• Max. Working Pressure: 225 PSIG @ 73° F
What replacement parts are available for the FLOMEC TM Series meters?
The following replacement parts and accessories are available:
| Part No. | Description |
|---|---|
| 113520-1 | Battery Replacement Kit |
| 116000-1 | Calibration Container, Large (5 gallon) |
| 125508-03 | 1/2 inch, Turbine Assy Kit |
| 125508-04 | 1/2 inch NPT, PVC Turbine Assy Kit |
| 125510-03 | 3/4 inch, Turbine Assy Kit |
| 125510-04 | 3/4 inch NPT, PVC Turbine Assy Kit |
| 125512-03 | 1 inch, Turbine Assy Kit |
| 125512-04 | 1 inch NPT, PVC Turbine Assy Kit |
| 125514-03 | 1 1/2 inch, Turbine Assy Kit |
| 125514-04 | 1 1/2 inch NPT, PVC Turbine Assy Kit |
| 125516-03 | 2 inch, Turbine Assy Kit |
| 125516-04 | 2 inch NPT, PVC Turbine Assy Kit |
| 901002-52 | Seal |
Computer Kits:
| Part No. | Description |
|---|---|
| 125509-04 | 1/2 inch, Computer Assy Kit w/Pulse |
| 125511-04 | 3/4 inch, Computer Assy Kit w/Pulse |
| 125513-04 | 1 inch, Computer Assy Kit w/Pulse |
| 125515-04 | 1 1/2 inch, Computer Assy Kit w/Pulse |
| 125517-04 | 2 inch, Computer Assy Kit w/Pulse |