What are the safety instructions for using the UNI-T UT501C?
To use the product safely and correctly, please observe the following safety instructions:
Danger:
1. Do not measure in circuits with voltage above AC 750V.
2. Do not test in flammable places. Sparks may cause an explosion.
3. Never use the Tester if its surface or the operator’s hands are wet.
4. Do not touch the live part of test probe during measurement.
5. Never open the battery cover during measurement.
6. Do not touch the circuit under test when measuring insulation resistance.
Warning:
1. Stop using the Tester if it behaves abnormally (such as bare test lead, damaged tester).
2. Be cautious when the test is conducted under voltage above 30Vrms, 42.4Vacrms or 60Vdc, such voltages may pose a shock hazard!
3. Do not replace the battery in damp environments.
4. Ensure secure connection between all test leads and test ports of the Tester.
5. Please switch off the Tester and remove the test leads before replacing the batteries.
Caution:
1. Before measuring insulation resistance, the circuit to be measured must be fully discharged and separated from other power circuits.
2. If the test leads appear damaged, replace them by test leads with same model and electric specification.
3. Stop using the Tester when the low battery symbol indicates low battery. If the Tester is not used for an extended period of time, please store the Tester after removing the battery.
4. Do not use or store the Tester in high temperature, high humidity, flammable, explosive or strong magnetic field environments.
What do the symbols on the UNI-T UT501C indicate?
| Symbol | Description |
|---|---|
| Danger Triangle | Identify conditions that could cause severe or fatal harm. |
| Warning Triangle | Avoid electric shock hazard. |
| Caution Triangle | Avoid damage to the Tester and make accurate measurement. |
| Lightning Bolt | Possible electric shock |
| Double Square | Double or reinforced insulation |
| DC | Direct current |
| AC | Alternating current |
| Ground Symbol | Earth (ground) terminal |
What are the technical specifications for the Insulation Resistance Test on the UNI-T UT501C?
UT501C Insulation Resistance Test Index
| Voltage rating | 100V | 250V | 500V | 1000V |
|---|---|---|---|---|
| Measurement range | 0.00MΩ~100MΩ | 0.00MΩ~500MΩ | 0.00MΩ~2.00GΩ | 0.00MΩ~5.00GΩ |
| Open-circuit voltage | DC 100V+10% | DC 250V+10% | DC 500V+10% | DC 1000V+10% |
| Test current rating | Loaded at 100kΩ, 1.0mA~1.1mA | Loaded at 250kΩ, 1.0mA~1.1mA | Loaded at 500kΩ, 1.0mA~1.1mA | Loaded at 1MΩ, 1.0mA~1.1mA |
| Short-circuit current | < 2.0mA approximately | |||
| Accuracy range | 0.00MΩ~99.9MΩ: ±(3%+5); 100MΩ~5.00GΩ: ±(5%+5) | |||
Error limit: ± (a% Reading+ Number of digit), one year warranty.
Ambient temperature: (23±5) °C
Ambient humidity: 45%~75%
What are the AC Voltage Test specifications for the UNI-T UT501C?
| Item | Specification |
|---|---|
| Test range | 30V~750V (50Hz/60Hz) |
| Resolution | 1V |
| Accuracy | ± (2%+3) |
What are the main functions and features of the UNI-T UT501C?
The UNI-T UT501C includes the following functions and features:
Auto-range function: The device automatically adjusts ranges.
Auto power off function: The Tester powers off automatically after 10 minutes of inactivity.
Data hold: The data hold function is enabled when measuring the insulation resistance function, it is disabled for AC voltage measurement.
Measurement Indication: The measured insulation resistance itself is charged (>30V), the symbol (High voltage warning) will flash when insulation resistance test function is turned off.
Display: After starting up, full display will appear on the LCD for about 1 second, display screen: 80.0*50.0mm visible area, maximum display is “9999”.
Low battery indication: When the battery power is ≤7±0.2V, the symbol lights up; when the battery power is ≤6.4V, the symbol flashes 10 times and the Tester shuts down automatically.
Powered by adapter (without charging function): The symbol is displayed in priority when charged by adapter, battery symbol is not displayed.
Overrange Indication: Overrange indication for insulation resistance measurement is “>110MΩ”, “>550MΩ”, “>2.20GΩ”, “>5.50GΩ”, or “>22.0GΩ”; overrange indication for AC voltage is “>750V AC”.
Auto backlight function: The backlight lights up for about 5 seconds when the Tester is turned on. The backlight automatically turns on when measurement starts for insulation resistance test, and it automatically turns off when the measurement stops. The backlight function is disabled for AC voltage measurement.
Red warning light: The light is on when high voltage is output.
What are the components of the external structure on the UNI-T UT501C?
The external structure consists of the following:
1. EARTH: Terminal for insulation resistance sampling test
2. GUARD: Negative input terminal for voltage test lead
3. ACV: Positive input terminal for AC voltage test lead
4. LINE: Terminal for high voltage output test lead
5. Liquid Crystal Display
6. High voltage output warning indicator
7. Insulation resistance test buttons
8. Functional rotary switch
9. Alligator clip
10. Test probe
11. Alligator clip
12. Test lead
13. Power adapter port
What do the LCD items represent on the UNI-T UT501C?
| Item No. | Description |
|---|---|
| 1 | Less than |
| 2 | Negative value |
| 3 | Greater than |
| 4 | Battery symbol |
| 5 | Powered by adapter |
| 6 | Audible and visual alarm symbol |
| 7 | Data hold |
| 8 | Alternating current |
| 9 | High voltage alarm |
| 10 | Voltage unit |
| 11 | Insulation resistance unit (GΩ) |
| 12 | Insulation resistance unit (MΩ) |
| 13 | Displayed value |
How do I perform an AC Voltage Test with the UNI-T UT501C?
Caution: Do not input AC voltage over 750Vrms. It’s possible to display higher voltage, but it may damage the Tester or pose a shock hazard. Disconnect the test lead from measured circuit and remove them from the input terminal after all measurement operations are completed.
Steps:
1. Turn the rotary switch to ACV position for AC voltage measurement.
2. Connect red test lead to “ACV” input terminal, and connect black test lead to “GUARD” input terminal.
Note: Considering external interference during AC voltage test, ACV and GUARD test probes can be used to short-circuit for zero calibration.
How do I perform an Insulation Resistance Test with the UNI-T UT501C?
Caution:
1. Two test leads must be separated during measurement.
2. Clamp the insulation resistor before high voltage output.
3. Do not short-circuit two test leads at high voltage output.
4. Do not conduct test with battery cover opened, otherwise it may pose an electric shock.
Steps:
1. Select a test voltage: Before measuring insulation resistance, the circuit to be measured must be fully discharged and separated from power circuit. Connect red test lead to “LINE” input terminal, and connect black test lead to “EARTH” input terminal. Connect red and black alligator clips to the measured circuit, the voltage is output from LINE terminal.
2. Continuous measurement operation: After selecting a test voltage, press and rotate TEST to LOCK position for continuous measurement, output insulation resistance test voltage, the indicator lights up red. After the test is completed, press and rotate TEST button to unlock and stop the test.
What are the common connection methods for cable insulation resistance tests using the UNI-T UT501C?
Common two-wire measurement connection method:
Test whether there is leakage current in the two conductors of the same cable. Connect the two ports of the insulation resistance test to the metal body of the two conductors in the same cable. Measure the insulation resistance between the two conductors.
Common single-wire measurement connection method:
Measure the resistance value between the conductor wire and the bare metal wire to determine whether there is leakage current in the conductor.
How do I test transformer insulation resistance with the UNI-T UT501C?
Test of Insulation Resistance between Primary Winding and Secondary Winding Grounding: Connect the Tester to test insulation between the Primary Winding and Secondary Winding Grounding.
Test of Insulation Resistance between Primary Winding Grounding and Secondary Winding: Connect the Tester to test insulation between Primary Winding Grounding and Secondary Winding.
Test of Insulation Resistance between Secondary Windings: Connect the Tester to test insulation between Secondary Windings.
Test of Winding and Magnetic Core: Connect the Tester to test insulation between the Winding and the Magnetic Core.
How do I replace the batteries in the UNI-T UT501C?
Caution: Do not mix new and old batteries. The same brand of battery is recommended. Pay attention to the battery polarity when installing batteries. Do not measure if the battery compartment is open.
Low voltage symbol: ≤7V
Steps:
1. To avoid electric shock, turn the functional rotary switch to OFF position to power off the Tester and remove test leads before replacing the batteries.
2. Unscrew and remove the battery cover to replace with 6pcs 1.5V AA alkaline batteries.
3. Tighten the screws after the replacement.
Can I use a power adapter with the UNI-T UT501C?
Yes, the UNI-T UT501C can be powered by an adapter.
1. Power off the Tester and remove all test leads before switching batteries and power adapter.
2. Use dedicated power adapter only (Model: XZF48-15600D).
3. Connect the power adapter to the power port of the Tester before measurement.
How do I clean the UNI-T UT501C housing?
1. Wipe the Tester housing with a soft cloth or sponge dampened with clear water.
2. Do not use abrasives or solvents.
3. Do not immerse the Tester in water to avoid damage. If water gets into the Tester accidentally, please have it repair by a professional.
4. If the Tester is wet, allow time for drying before storing it.