Free English TEMPNOTE T32SU (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.
CLICK HERE TO DOWNLOAD TEMPNOTE T32SU (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
How do I start recording with the TEMPNOTE T32SU data logger?
To start recording, long press the START button for a minimum of 3 seconds. The display will show a recording symbol (▶) or “WAIT” to indicate a successful start.
How do I stop recording with the TEMPNOTE T32SU data logger?
To stop recording, long press the STOP button for a minimum of 3 seconds. The display will show a stop symbol (■) to indicate a successful stop.
How do I generate a report from the TEMPNOTE T32SU data logger?
To generate a report, remove the plastic cap and plug the logger into the USB port of a computer. A PDF and CSV report will be automatically generated and can be found in the device’s folder.
How can I check the status of my TEMPNOTE T32SU data logger?
To check the status, push the START button one time. The display confirmation will vary based on the current status:
| Status | Action | Display Confirmation |
|---|---|---|
| Not Started | Push Start Button 1 Time | Temperature + Battery Symbol |
| Recording | Push Start Button 1 Time | Temperature + Battery Symbol + Recording Symbol (▶) |
| Stopped | Push Start Button 1 Time | Temperature + Battery Symbol + Stop Symbol (■) |
What do the different symbols on the LCD display of the TEMPNOTE T32SU mean?
| 1 | ✓ | Normal | 6 | (Battery Icon) | Battery Power |
| X | Alarm | ||||
| 2 | ▶ | In Recording Status | 8 | (USB Symbol) | Interface Indication |
| ■ | Stop Recording | ||||
| 3 & 7 | ↑ | Alarm area: H1 H2 (High Temperature Alarm) | 9 | (e.g., 8888.8) | Temperature Value |
| ↓ | L1 L2 (Low Temperature Alarm) | ||||
| 4 | WAIT | Start Delay Status | 10 | (e.g., °C) | Temperature Unit |
| 5 | (Stop Symbol with line) | Stop button disable |
How do I interpret the different operational screens on the TEMPNOTE T32SU display?
1. When the logger is in idle condition (before starting), pressing the START button once will show the current temperature and a checkmark symbol (✓).
2. When the T32SU is in recording mode (▶) and within pre-set temperature limits, it will display a checkmark symbol (✓).
3. If the T32SU is recording and has crossed the upper or lower temperature limits, it will display an ‘X’ symbol and an alarm arrow (↑ or ↓).
4. When the T32SU has been stopped, a stop symbol (■) is shown on the display.
5. If the T32SU has been started and is in the start delay period, the “WAIT” symbol will be visible on the display. If the stop button is disabled, an additional symbol will appear.
What are the technical specifications of the TEMPNOTE T32SU?
| Recording Option | Single Use |
| Temperature Range | -30°C to +70°C |
| Temperature Accuracy | ± 0.5 °C (-30°C to +70°C) |
| Temperature Resolution | 0.1°C |
| Data Storage Capacity | 32,000 Readings |
| Shelf Life / Battery | 2 Years |
| Recording Interval | User Configurable (10 seconds to 18 hours) |
| Recording Duration | 300 + Days (Customizable) |
| Startup Mode | Button Start or Timed Start |
| Protection Level | IP65 |
| Product Size | 96 mm x 49 mm x 7.5 mm |
| Product Weight | 25g |
| Certifications | EN12830, CE, RoHS |
| Validation Certificate | Hardcopy in Package |
| Report Generation | Plug-N-Play |
| Report Format | Auto, PDF and CSV |
| Password Protection | Optional on request |
| Connection Interface | USB 2.0 Interface |
| Alarm Configuration | High and Low Alarm |
| Reprogrammable | Tempnote Analyzer Software |
| Accessories Incl. | Shipping Documents, Calibration Certificate |
What are the factory default parameters for the TEMPNOTE T32SU?
The TEMPNOTE T32SU comes with the following default settings. To change these parameters, you need to use the Tempnote Software.
| Project | Parameters |
|---|---|
| Temperature Unit | °C |
| Temperature Alarm Limit | -30°C to +70°C |
| Alarm Delay | 10 Minutes |
| Recording Interval | 15 Minutes |
| Start Delay | 0 Minutes |
| Equipment Time | UTC +5:30 |
| LCD Display Time | 1 Minute |
| Start Mode | Press Button To Start |