Free English HANTEK HRDO4104D (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.
CLICK HERE TO DOWNLOAD HANTEK HRDO4104D (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
What are the general safety precautions for using the HANTEK HRDO4104D oscilloscope?
Read the following safety precautions carefully to avoid injury and to prevent damage to this product or any product connected. To avoid possible dangers, please use this product in accordance with the regulations.
Only professionally authorized personnel can perform repairs.
Use the right power cable.
Use the power cable approved by the country in which the product is used only.
Connect and disconnect correctly.
Before connecting the probe to the circuit being measured, please connect the probe to the oscilloscope first. Before disconnecting the probe from the oscilloscope, please disconnect the probe and the circuit under test first.
Ground the product.
To avoid electric shocks, the product is grounded through a grounding conductor of the power cable. The grounding conductor must be connected to the ground before connecting the input or output terminals of the product. Ensure that the product is properly grounded.
Connect the probe properly.
The ground wire of the probe is the same as the ground potential. Do not connect the ground wire to high voltage.
View all terminal rating values.
To avoid fire or excessive current, please check all rating values and signs on the product.
Do not operate with the cover open.
Do not run the product with the cover or panel open.
Avoid circuit exposure.
Do not touch exposed connectors and components after power is switched on.
Do not operate if the product is suspected to be faulty.
If you suspect that the product has been damaged, please ask qualified maintenance personnel to check it.
Maintain proper ventilation.
Do not operate in a humid environment.
Do not operate in inflammable or explosive environment.
Please keep the product surface clean and dry.
What should I do upon receiving my HANTEK HRDO4104D oscilloscope?
After receiving the oscilloscope, please follow these steps to check the instrument:
Check the shipping package:
Check whether there is any damage caused by transportation. If the packaging cartons or protective foam pads are seriously damaged, please keep them until the whole machine and accessories pass the electrical and mechanical testing.
Check the accessories:
The details of the accessories are provided in Appendix A: Accessories at the end of the user manual. If you find any accessory missing or damaged, please contact the dealer responsible for this business.
Check the machine:
If you find the instrument is damaged, not working properly, or unable to pass the performance test, please contact the dealer responsible for this business.
How do I prepare the HANTEK HRDO4104D oscilloscope for first use?
1. To Adjust the Supporting Legs:
There is a bracket design at the bottom of the machine. Adjust it to tilt the oscilloscope slightly upwards for better stability and observation. Close the bracket when not in use.
2. To Connect to AC Power:
The AC power supply specifications are 47-440Hz, 90-264VAC/127-320VDC, 80Watts MAX. Connect the oscilloscope to the power supply using the provided power cord. Ensure the instrument is correctly grounded to avoid electric shock.
3. Turn-on Checkout:
When the machine is properly plugged in, press the power button on the front panel. During startup, all button lights will light up for a few seconds as the oscilloscope performs self-checks. The startup screen will appear after completion.
4. Function Inspection:
p>Click on [Default Setup] to restore factory settings.
p>Connect the grounding crocodile clamp of the probe to the grounding terminal.
p>Connect the probe to the input terminal of channel 1 and the compensation signal output terminal.
p>Set the probe attenuation ratio to X10 and click [Auto Scale].
p>Observe the waveform to see if the square wave signal is displayed normally. Check other channels (CH2-CH4) using the same method.
5. Probe Compensation:
When connecting a probe to any input channel for the first time, you must match the probe with the channel.
p>Perform the first four steps of the “Function Inspection”.
p>Compare the waveform with the compensation diagrams (Perfectly compensated, Over compensated, Under compensated).
p>If necessary, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform is “Perfectly compensated”.
What are the functions of the controls on the front panel of the HANTEK HRDO4104D?
| No. | Description |
| 1 | Touch Screen |
| 2 | Function Menu Operation Keys |
| 3 | Multifunction Knob |
| 4 | Default Setting Key |
| 5 | Touch Lock Key |
| 6 | Common Operation Keys |
| 7 | Horizontal Control System |
| 8 | Vertical Control System |
| 9 | Run/Stop Key |
| 10 | Single Trigger Control Key |
| 11 | Auto Waveform Display Key |
| 12 | Arbitrary Waveform Generator Setting Key |
| 13 | Trigger Control System |
| 14 | Probe Compensation Signal Output Terminal/Ground Terminal |
| 15 | Analog Channel Input Terminals |
| 16 | Digital Channel Input Interface |
| 17 | USB HOST Interface |
| 18 | Power Key |
How do I use the Horizontal control system on the HANTEK HRDO4104D?
Search: Press this key to enter the search settings menu to find events in collected data based on set criteria.
Zoom: Press this key to turn the delayed scanning function on or off.
Horizontal Time Base Knob: Rotate this knob to modify the horizontal time base. Clockwise decreases the time base (zooms in), and counterclockwise increases it (zooms out). Press the knob to switch between “Coarse” and “Fine” adjustment modes.
Horizontal Position Knob: Rotate this knob to modify the horizontal position (trigger position). Pressing the knob resets the horizontal position to zero.
Navigation Control Key: This combination key performs recording and playback navigation, time navigation, and event navigation.
How do I use the Vertical control system on the HANTEK HRDO4104D?
CH1, CH2, CH3, and CH4 analog input channel switches: Press these buttons to turn the respective channels on or off. The four channels and their waveforms are identified with different colors.
Channel vertical offset knob: Rotate this knob to modify the vertical offset (up/down position) of the current channel’s waveform. Press the knob to reset the vertical offset to zero.
Channel vertical gear knob: Rotate this knob to modify the vertical gear (V/div) of the current channel. Clockwise decreases the gear (increases amplitude), and counterclockwise increases the gear (decreases amplitude). Press this knob to switch between “coarse” and “fine” adjustment modes.
What are the functions of the Waveform and Trigger control systems on the HANTEK HRDO4104D?
Waveform Control System:
p>REF: Press to open the reference waveform setting menu.
p>Math: Press to open the mathematical operation function menu (A+B, A-B, A×B, A/B, FFT).
p>LA: Press to open the Logic Analyzer control menu.
p>Decode: Press to open the decoding settings menu.
Wave Gen (Arbitrary Waveform Generator):
p>G1, G2: Press to open the signal generator channel G1 or G2.
p>Burst: Burst button for the signal generator.
p>Signal source knob: Modify the frequency and amplitude of the signal source. Press to select between adjusting frequency or amplitude.
Trigger Control System:
p>Trigger knob: Modify the trigger level. Rotate clockwise to increase and counterclockwise to decrease. Press to set the trigger level to 50% of the waveform peak value.
p>Trig Menu: Press to open the trigger operation menu.
p>Force Trig: Press to force the oscilloscope to generate a trigger signal.
p>Mode: Press to switch the trigger mode between Auto, Normal, or Single.
What do the other function keys on the front panel of the HANTEK HRDO4104D do?
AUTO: Press for automatic waveform setting. The oscilloscope adjusts vertical, horizontal, and trigger settings for an optimal display.
Run/Stop: Toggles the oscilloscope’s operating status. Green backlight indicates “Running”; red indicates “Stopping”.
Single: Sets the triggering method to Single, where it acquires one waveform and then stops.
Multifunction Knob: Rotate to modify various setting parameters.
Default: Press to restore the oscilloscope to its factory default settings.
Touch Lock Key: Press to disable the touch screen function.
Function Menu Keys:
p>Quick Action: Performs quick operations like screenshots, saving waveforms/setups, all measurements, and statistical resetting.
p>Analyze: Enters the analysis menu.
p>Cursors: Enters the cursor measurement menu.
p>Save/Recall: Performs save and recall functions.
p>Utility: Enters the system function settings menu (interfaces, sound, language, etc.).
p>Help: Accesses the built-in help system.
p>Meas: Enters the measurement settings menu.
p>Acquire: Enters the sampling settings menu.
p>Display: Enters the display settings menu.
What are the functions of the ports on the rear panel of the HANTEK HRDO4104D?
| No. | Description | Function |
| 1 | Handle | Vertically pull up for easy access; press down when not in use. |
| 2 | LAN Interface | Used to connect the oscilloscope to a network. |
| 3 | USB DEVICE Interface | Connects to a computer for remote control via SCPI commands. |
| 4 | Reference clock | Input for an external reference clock for more accurate sampling or synchronization. |
| 5 | Aux output terminal | Outputs a trigger signal or a Pass/Fail test pulse. |
| 6 | EXT input terminal | Used to input external trigger signals to the oscilloscope. |
| 7 | G1,G2 signal source output terminal | Outputs the waveform set by the signal generator. |
| 8 | HDMI | (Option) Connects to an external display like a monitor or projector. |
| 9 | RS232/485 | (Option) Serial communication interface. |
| 10 | AC Power Cord Connector | Power input terminal. |
How do I use the touch screen controls on the HANTEK HRDO4104D oscilloscope?
The HANTEK HRDO4104D supports multi-touch and gesture operations:
1. Tap:
Tap a menu or function on the screen with one finger to operate menus, use the pop-up numeric or virtual keyboard, close pop-up boxes, or operate other windows.
2. Drag:
After holding down a target, drag it to the desired position. This can be used to drag a waveform to change its displacement/offset or drag a window to change its position.
3. Pinch & Stretch:
Bring two fingers together (pinch) or apart (stretch) on the screen. This can amplify or reduce the waveform. Vertical pinching/stretching adjusts the vertical gear (V/div). Horizontal pinching/stretching adjusts the horizontal gear (s/div).
4. Rectangle Drawing:
Click on the [Hantek] icon, then click on AreaSel to switch to rectangular drawing mode. Drag your finger on the screen to draw a rectangle. A menu will appear allowing you to select functions like “Histogram”, “Enlarge Level” (horizontal zoom), “Enlarge Vertical” (vertical zoom), or “Enlarge Waveform” (both horizontal and vertical zoom).
How do I use the built-in Help System on the HANTEK HRDO4104D?
The help system provides instructions for various function buttons and menu keys.
1. Click the [Help] button in the button area to enter the help module.
2. Press any other key to obtain help information for that specific key.
3. Press the [Help] button again to exit the help module.
For example, after clicking [Help], clicking the [QuickAction] button will pop up the explanation menu for [QuickAction].
How do I enable or disable an analog channel on the HANTEK HRDO4104D?
To Enable an Analog Channel (e.g., CH1):
p>Press the [CH1] button on the front panel. The button and the corresponding channel knob will light up.
p>Alternatively, click the channel status label at the bottom of the screen. The label and the corresponding [CH1] button will light up.
To Disable an Analog Channel:
p>If the channel is already open and selected, press its corresponding button (e.g., [CH1]) once to close it.
p>If the channel is open but not selected, press its button once to select it, and then press it again to close it.
p>Use the touch screen to close the channel by clicking on the number on the channel menu at the bottom of the screen.
How do I configure the vertical system settings for a channel on the HANTEK HRDO4104D?
Press the desired channel button (e.g., [CH1]) to open its settings menu. You can configure the following:
p>Vertical Scale (V/div): Adjust by rotating the vertical gear knob for that channel or by using the touch screen menu.
p>Vertical Offset: Adjust by rotating the vertical offset knob for that channel or by dragging the waveform up/down on the touch screen.
p>Channel Coupling: In the channel menu, select Coupling. Options are DC (passes both AC and DC components), AC (blocks DC component), and GND (disconnects the input signal).
p>Bandwidth Limit: Select BW Limit in the menu to limit the bandwidth (e.g., to 20MHz) to reduce noise. Options are Close, 20M, 100M, 200M, 350M.
p>Probe Ratio (Attenuation): Select Attenuation to match the oscilloscope’s setting to the probe you are using (e.g., 1X, 10X). This ensures correct amplitude readings.
p>Input Impedance: Select Impedance to choose between 1M Ω (high impedance) or 50 Ω (for matching with 50 Ω sources).
p>Waveform Invert: Toggles the inversion of the waveform display.
p>Amplitude Unit: Change the vertical unit to W, A, V, or U.
p>Fine tuning: Switch the vertical scale adjustment between coarse (standard 1-2-5 steps) and fine for more precise resolution.
p>Channel Delay: Set a delay time from -100ns to 100ns to correct for probe cable transmission delay (skew).
p>Offset Cal: Corrects for small DC offset deviations in the channel’s zero point.
p>Vertical Expansion: Choose whether changing the vertical gear expands/compresses the waveform around the screen center or around the GND (zero) level.
p>Channel Label: Set a custom label for the channel, which can be selected from a library or typed in manually.
How do I configure the horizontal system settings on the HANTEK HRDO4104D?
To Adjust the Horizontal Position:
This adjusts the trigger point’s position relative to the center of the screen.
p>Rotate the horizontal position knob.
p>Click the horizontal displacement label at the top of the screen and use the pop-up menu.
p>Drag the waveform left or right using a touch screen gesture.
To Adjust the Horizontal Time Base (s/div):
This changes the time value represented by each horizontal grid division.
p>Rotate the horizontal time base knob.
p>Click the horizontal time base label at the top of the screen and use the pop-up menu.
p>Use a horizontal pinch/stretch gesture on the touch screen.
To Pan or Zoom a Stopped Acquisition:
After stopping acquisition (pressing Run/Stop), you can use the horizontal position knob to pan through the captured data and the horizontal time base knob to zoom in or out.
To Use Delayed Sweep (Zoom):
Press the [Zoom] button. The screen will split into two areas. The top area shows the original waveform, and the bottom shows a zoomed-in portion. Use the horizontal time base knob to change the zoom factor and the horizontal position knob to move the zoomed window along the original waveform.
How do I configure the sample and acquisition system on the HANTEK HRDO4104D?
Click the [Acquire] button to enter the sampling settings menu.
1. Timebase Mode:
p>YT Mode: Standard mode where the Y-axis is voltage and the X-axis is time.
p>XY Mode: Displays a Lissajous pattern, where the X-axis is the voltage of one channel and the Y-axis is the voltage of another. Useful for measuring phase differences.
p>ROLL Mode: For slow signals (50ms/div or slower), the waveform scrolls across the screen from right to left.
2. Acquisition Mode:
p>Normal: Standard sampling mode, best for most waveforms.
p>Average: Averages multiple waveform acquisitions to reduce random noise and improve vertical resolution.
p>Peak: Captures the maximum and minimum values in each sample interval to display the waveform envelope and avoid aliasing on narrow pulses.
p>High Resolution: Averages adjacent points within a single acquisition to reduce noise and create a smoother waveform.
3. Memory Depth:
Select the number of points to store in a single acquisition. A higher memory depth allows for a higher sample rate at longer time base settings. Options vary depending on whether one, two, or all four channels are active.
4. Horizontal Expansion:
Choose the reference point (Center, Left, Right, Trigger, User) around which the waveform expands or compresses when adjusting the time base.
How do I set up a basic trigger on the HANTEK HRDO4104D?
Click the [Trig Menu] button or tap the trigger status area at the top of the screen to open the trigger menu.
1. Trigger Type: Select the type of trigger. The most common is Edge.
2. Trigger Source: Select the input source for the trigger (e.g., CH1, CH2, EXT, etc.). Choose the channel with the signal you want to stabilize.
3. Slope Type (for Edge Trigger): Choose whether to trigger on the Rising edge, Falling edge, or Either edge of the signal.
4. Trigger Level: Adjust the voltage level that the signal must cross to activate the trigger. Use the dedicated Trigger knob to move the trigger level line up and down on the display.
5. Trigger Mode:
p>Auto: The oscilloscope will sweep even without a valid trigger, allowing you to see a signal. It will lock to a stable waveform if a trigger is found.
p>Normal: The oscilloscope only updates the display when a valid trigger event occurs. If there is no trigger, the last acquired waveform remains on screen.
p>Single: The oscilloscope waits for one valid trigger event, captures the waveform, and then stops acquiring.
6. Trigger Holdoff: Set a time period during which the oscilloscope will ignore subsequent trigger events. This is useful for stabilizing complex waveforms like pulse trains where you want to trigger only on the first pulse.
What are the different trigger types available on the HANTEK HRDO4104D?
The HANTEK HRDO4104D offers a variety of advanced trigger types:
p>Edge: Triggers on a rising, falling, or either edge crossing a specific voltage level.
p>Pulse: Triggers on positive or negative pulses that meet a specific width condition (e.g., greater than, less than, or equal to a set time).
p>Video: Triggers on standard analog video signals (NTSC, PAL/SECAM), allowing triggering on specific lines or fields.
p>Slope: Triggers on the rise or fall time of a signal edge, based on the time it takes to cross between two voltage levels.
p>Overtime: Triggers if the time interval between a rising and falling edge (or vice versa) is greater than a set timeout.
p>Window: Triggers when the input signal passes through either a high or low trigger level.
p>Runt: Triggers on pulses that cross one voltage level but fail to cross a second level.
p>Logic/Pattern: Triggers on a logical combination (AND, OR, NOT) of states (High, Low, Don’t Care) across multiple channels.
p>Delay: Triggers when the time between an edge on source A and an edge on source B meets a preset time condition.
p>Setup/Hold: Triggers on violations of setup and hold time between a clock and data signal.
p>Nth Edge: Triggers on the Nth edge of a burst after a specified idle time.
p>Serial Bus Triggers: UART, LIN, CAN, SPI, I2C. Triggers on specific protocol events like start of frame, specific data, address, or errors.
How do I perform math operations on the HANTEK HRDO4104D?
Press the [Math] button to open the math operation menu.
1. Select Operator:
Choose the desired math function. Options include:
p>Arithmetic: A+B, A-B, A*B, A/B.
p>FFT: Fast Fourier Transform to view the signal in the frequency domain.
p>Logic: A&&B (AND), A||B (OR), A^B (XOR), !A (NOT).
p>Function: Integral (Intg), Differential (Diff), Square Root (Sqrt), Logarithms (Lg, Ln), Exponent (Exp), Absolute Value (Abs).
p>Digital Filtering: LowPass, HighPass, BandPass, BandStop.
p>Expression: Create complex mathematical operations by combining multiple operators and sources.
2. Set Sources:
For operations requiring one or two inputs (Source A, Source B), select the channels (CH1-CH4) or reference waveforms (REF) to be used in the calculation.
3. Adjust Display:
Use the Scale (vertical gear) and Offset (vertical position) settings within the Math menu to adjust the display of the resulting math waveform for optimal viewing. An AutoSetting button is also available to automatically adjust these parameters.
4. Configure Specific Settings:
Some operations have unique settings. For example, FFT has settings for window function, frequency range, and peak search. Logic operations have threshold settings for each channel.
How do I perform automatic measurements on the HANTEK HRDO4104D?
Press the [Meas] button to enter the measurement settings menu.
1. Press Add to open the measurement selection panel.
2. Choose a measurement Category: Horizontal (time-related), Vertical (voltage-related), or Other (delay, phase, area).
3. Select the specific measurement parameter you want to add (e.g., Freq, Period, VMax, Pk-Pk).
4. Select the Source A (and Source B for two-channel measurements) for the measurement.
The selected measurements will be displayed at the bottom of the screen with their real-time values. You can add multiple measurements.
To view statistics for the measurements (Current, Average, Max, Min, Standard Deviation), click on the Statistic menu tab and turn it on.
To display all available measurements for a channel at once, press [Meas] -> All Measure and select the desired channel.
How do I use cursor measurements on the HANTEK HRDO4104D?
Press the [Cursors] button to enter the cursor settings menu.
1. Select Mode:
p>None: Cursors are off.
p>Manual: Allows you to freely place two cursors on the screen to take measurements.
p>Track: The cursors lock onto the waveform of the selected source. As you move one cursor, the other tracks along the waveform.
2. In Manual Mode:
p>Select Type: Choose X (vertical cursors for time/frequency) or Y (horizontal cursors for voltage).
p>Select Source: Choose the channel (CH1-CH4) or Math waveform to measure.
p>Adjust Cursors: Select AX or BX (for X cursors) or AY or BY (for Y cursors) and use the multifunction knob V0 to move the selected cursor. Select AXBX or AYBY to move both cursors together.
A measurement window will appear on the screen showing the cursor positions (AX, BX, AY, BY) and the differences between them (ΔX, ΔY, 1/ΔX).
How do I use the Pass/Fail test function on the HANTEK HRDO4104D?
1. Navigate to the Pass/Fail menu by pressing [Utility] -> Pass/Fail.
2. Enable: Turn the Pass/Fail test function on.
3. Source: Select the channel (CH1-CH4) to test.
4. Mask:
p>Click Create Rules to define the test mask.
p>Set the horizontal (X Mask) and vertical (Y Mask) tolerance range around the waveform. This creates a template that the source signal must fit within to pass.
5. Option: Configure the output actions when a test fails.
p>Aux Output: Enable to send a pulse out of the rear Aux output terminal on a pass or fail event.
p>Error Action: Choose to enable a Buzzer on a failure event.
p>Stop Output: Choose whether to stop the test operation upon failure.
6. Operate: Press the play button in the menu to start the test. The screen will display a running count of total tests, passes, and fails.
How do I perform serial bus decoding on the HANTEK HRDO4104D?
The HANTEK HRDO4104D supports UART, I2C, SPI, LIN, and CAN decoding.
1. Press the [Decode] button to enter the decoding settings menu.
2. Bus Type: Select the protocol you want to decode (e.g., UART).
3. Bus Status: Turn the bus decoding function ON.
4. Source: Enter the Source Settings menu. Assign the correct oscilloscope channels to the bus signals (e.g., for UART, assign a channel to Tx and Rx; for I2C, assign channels to SCL and SDA).
5. Threshold: For each assigned channel, set the voltage threshold that defines the logic high and low levels.
6. Setting: Enter the protocol-specific settings menu. Configure parameters like Baud Rate, Data Bits, Parity (for UART), or signal type (for CAN).
7. Display: Configure how the decoded data is displayed.
p>Format: Choose Hex, Dec, Bin, or ASCII.
p>Event Table: Turn on the event table to see a detailed, time-stamped log of all decoded data packets.
Once configured, the decoded data will be overlaid on the waveform display, and the event table will provide a detailed breakdown.
How do I save and load files on the HANTEK HRDO4104D?
Press the [Save/Recall] button to open the save/recall menu. You can save to internal memory or a connected external USB drive.
To Save Files:
1. Select the type of file to save:
p>Save Image: Saves a screenshot of the display. You can set the format (*.png, *.bmp, etc.), file name, and whether to invert colors.
p>Save Wave: Saves the waveform data points. Select the data source (Screen or Memory), format (*.bin, *.csv), and file name.
p>Save Setup: Saves the current oscilloscope configuration. Set the file type (*.stp, *.ptr) and file name.
2. Select the storage destination (internal Local Disk (C) or external Removable USB Disk (D)) and press the Save button in the menu.
To Load Files:
1. In the [Save/Recall] menu, select the type of file to load:
p>Load Wave: Loads a previously saved waveform file (*.wfm).
p>Load Setup: Loads a previously saved settings file (*.stp).
2. Navigate to the file’s location using the disk management interface.
3. Select the file and press the Load button in the menu.
Disk Management:
In the [Save/Recall] menu, press Disk to open the file manager. Here you can browse files, create new folders (NewFolder), rename, copy, paste, and delete files on both internal and external storage.
How do I use the Arbitrary Waveform Generator (AWG) on the HANTEK HRDO4104D?
Press the [G1] or [G2] button to enter the signal generator settings for the respective channel.
To Output Basic Waveforms:
1. In the generator menu, click the Wave menu tab.
2. Select a basic waveform type from the pop-up menu: Sine, Square, Ramp, Pulse, DC, Noise, Sinc, Exp.Rise, Exp.Fall, ECG, Gauss, Lorentz, Haversine.
3. Configure the waveform parameters:
p>Frequency: Set the frequency of the waveform.
p>Amplitude: Set the peak-to-peak voltage.
p>Offset: Set the DC offset voltage.
p>Start Phase: Set the starting phase of the waveform.
p>Other parameters like Duty Cycle (for Square), Symmetry (for Ramp), or Pulse Width (for Pulse) may also be available.
4. Impedance: Select the output impedance (HighZ or 50 Ω).
To Output Arbitrary Waveforms:
1. In the generator menu, select Arb from the Wave menu.
2. You can then:
p>Load from CH: Capture a waveform from an oscilloscope channel and use it as the arbitrary waveform.
p>Load Stored: Load a previously saved arbitrary waveform file (*.arb).
p>Create/Edit: Manually create or edit a waveform point-by-point using the waveform editor.
What should I do if the waveform display is ladder-like on my HANTEK HRDO4104D?
1. The horizontal time base might be too low. Increase the horizontal time base to improve the display effects.
2. If the display type is “Vector”, the lines between the sample points may cause ladder-like display results.
What should I do if the button light does not light up after connecting the power cord to my HANTEK HRDO4104D?
1. Check if the power switch of the oscilloscope is turned on.
2. Check if the fuse is intact. If a fuse needs to be replaced, please use a fuse that meets the specifications of this product.
3. If the button board still does not light up, please contact technical support.
Why is there no waveform displayed on the screen of my HANTEK HRDO4104D?
1. Check if the probe is correctly connected to the BNC connector of the signal input channel.
2. Check if the channel is open (CH1~CH4 menu buttons).
3. Detect whether the signal channel to be tested has signal output.
4. If it is a DC signal and the amplitude is relatively large, please increase the amplitude range.
5. You can press the automatic measurement button ([Auto Scale]) to automatically detect the signal first.
What can I do if the touch functions on my HANTEK HRDO4104D cannot be used normally?
1. Check if the touch screen function is turned on. If it is not, click [Touch Lock] in the button area to turn on the touch screen function.
2. Check if there is any dirt on the screen or fingers. If there is, please clean it first.
3. Check if the oscilloscope is close to a strong magnetic field. If it is, please stay away first.
What should I do if the USB storage device cannot be recognized by my HANTEK HRDO4104D?
1. Check if the USB drive is functioning properly.
2. Confirm the capacity of the USB flash drive. It is recommended to use a USB flash drive that does not exceed 16G.
3. Confirm that the format of the USB drive is FAT32 format.
4. Reopen the oscilloscope and reinsert the USB drive for inspection.
What steps can I take if my HANTEK HRDO4104D instrument cannot start normally?
1. Check if the power switch of the machine is turned on.
2. Check if the power plug of the oscilloscope is properly connected.
3. Check if the fuse is intact. If a fuse needs to be replaced, please use a fuse that meets the specifications of this product.
4. Turn on the machine again and see if it can be turned on normally.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
CLICK HERE TO DOWNLOAD HANTEK HRDO4104D (01) PDF DOCUMENT