Free English HANTEK HRDO4054E (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD HANTEK HRDO4054E (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

What are the general safety precautions for using the HANTEK HRDO4054E digital oscilloscope?

To ensure safe operation of your HANTEK HRDO4054E, please follow these safety precautions:

• Only professionally authorized personnel can perform repairs.

• Use the right power cable: Use only the power cable approved for the country in which the product is used.

• Connect and disconnect correctly: Before connecting a probe to a circuit, connect the probe to the oscilloscope first. Before disconnecting the probe from the oscilloscope, disconnect the probe from the circuit under test first.

• Ground the product: The product is grounded through the grounding conductor of the power cable. To avoid electric shocks, ensure the product is properly grounded before connecting input or output terminals.

• Connect the probe properly: The ground wire of the probe is at the same potential as the ground. Do not connect the ground wire to high voltage.

• View all terminal rating values: To avoid fire or excessive current, check all rating values and signs on the product before connecting.

• Do not operate with the cover open: Do not run the product with the cover or panel open.

• Avoid circuit exposure: Do not touch exposed connectors and components when the power is on.

• Do not operate if the product is suspected to be faulty: If you suspect damage, have it inspected by qualified maintenance personnel.

• Maintain proper ventilation.

• Do not operate in a humid, inflammable, or explosive environment.

• Keep the product surface clean and dry.


What are the environmental and working conditions required for the HANTEK HRDO4054E?

The required working environment for the HANTEK HRDO4054E is as follows:

Temperature:

• Operating: 0°C to 50°C

• Non-operating: -30°C to 70°C

Humidity:

• Operating: Below +30°C: ≤95%RH; +30°C to +40°C: ≤75%RH; +40°C to +50°C: ≤45%RH (all without condensation).

• Non-operating: Below 65°C: ≤95% RH (without condensation).

Altitude:

• Operating: below 3 km

• Non-operating: below 15 km

Ventilation:

The oscilloscope uses a fan for cooling. Ensure that air intake and exhaust areas are free from obstructions. Provide at least 10 cm of clearance beside, above, and behind the instrument for adequate ventilation.

Installation (Overvoltage) Category:

This product is powered by mains conforming to installation (overvoltage) category II.

Pollution Degree:

The product is rated for Pollution Degree 2, which means normally only nonconductive pollution occurs, with temporary conductivity caused by condensation to be expected (e.g., an indoor environment).


How do I clean my HANTEK HRDO4054E digital oscilloscope?

To clean the outer surface of your HANTEK HRDO4054E and its probes, follow these steps:

1. Use a lint-free cloth to remove dust from the outside of the oscilloscope and probe. Be careful to avoid scratching the display filter.

2. For a more thorough cleaning, clean the oscilloscope with a soft cloth soaked in water or an aqueous solution of 75% isopropyl alcohol.

Note: To avoid damaging the surface, do not use any corrosive or chemical cleaning reagents.

Warning: Before powering on the device again, ensure it is completely dry to avoid electrical short circuits or personal injury from moisture.


What are the key product features of the HANTEK HRDO4054E?

The HANTEK HRDO4054E offers the following key features:

• 12bit vertical resolution, up to 10GSa/s sampling rate, and bandwidths up to 3GHz.

• 8 independent instruments in one: 4-Channel oscilloscope, 16-channel logic analyzer, spectrum analyzer, 2-Channel arbitrary wave generator, digital voltmeter, frequency meter and accumulator, protocol analyzer, and protocol generator.

• Built-in dual-channel 200MHz signal source, ±7V voltage output, 1GSa/s sampling rate.

• 10.1 inch multi-touch capacitive screen with 256-level waveform gray scale and color temperature display.

• Rich serial protocol triggering and decoding functions.

• Waveform capture rate higher than 1.1 million waveforms per second.

• Up to 51 kinds of waveform parameters are automatically measured.

• Multiple data analysis functions: independent search, navigation buttons, event lists, histograms, Bode plots, power analysis, and frequency meters.


What models are available in the HANTEK HRDO4000 series oscilloscope lineup?

The HANTEK HRDO4000 series includes the following models:

Model Bandwidth Channel Signal source Digital Channels
HRDO4304E 3GHz 4 1 16
HRDO4254E 2.5GHz 4 1 16
HRDO4204E 2GHz 4 1 16
HRDO4104E 1GHz 4 1 16
HRDO4084E 800 MHz 4 1 16
HRDO4054E 500 MHz 4 1 16
HRDO4304C 3GHz 4
HRDO4254C 2.5GHz 4
HRDO4204C 2GHz 4
HRDO4104C 1GHz 4
HRDO4084C 800 MHz 4
HRDO4054C 500 MHz 4
HRDO4304D 3GHz 4 1
HRDO4254D 2.5GHz 4 1
HRDO4204D 2GHz 4 1
HRDO4104D 1GHz 4 1
HRDO4084D 800 MHz 4 1
HRDO4054D 500 MHz 4 1

How do I perform an initial check and setup for the HANTEK HRDO4054E?

Follow these steps for initial setup and inspection:

1. General Examination:

• Check the shipping package for any damage. If the packaging is seriously damaged, keep it until the machine and accessories pass all tests.

• Check the accessories against the list provided in the user manual to ensure nothing is missing or damaged.

• Check the machine for any physical damage.

2. Preparation Before Use:

• To Adjust the Supporting Legs: Adjust the bracket at the bottom of the machine to tilt the oscilloscope slightly upwards for better viewing and operation.

• To Connect to AC Power: Use the provided power cord to connect the oscilloscope to a power supply matching the specifications: 47-440Hz, 90-264VAC / 127-320VDC, 80 Watts MAX. Ensure the instrument is correctly grounded to avoid electric shock.

3. Turn-on Checkout:

• Press the power button on the lower left corner of the front panel. During startup, all button lights will illuminate briefly, and the oscilloscope will perform a series of self-checks. The startup screen will appear upon completion.

4. Function Inspection:

• Click [Default Setup] to restore factory settings.

• Connect the grounding crocodile clamp of the probe to the grounding terminal.

• Connect the probe to the channel 1 (CH1) input and the compensation signal output terminal.

• Set the probe attenuation ratio to X10 and click [Auto Scale].

• Observe the waveform to see if a square wave signal is displayed normally. If the waveform is not a perfect square wave, proceed to probe compensation.

5. Probe Compensation:

• When connecting a probe for the first time, perform compensation to match the probe to the input channel.

• Follow the first four steps of “Function Inspection”.

• Use a non-metallic screwdriver to adjust the variable capacitance on the probe until the on-screen waveform is “Perfectly compensated” (a flat-topped square wave).


What are the functions of the controls on the front panel of the HANTEK HRDO4054E?

No. Description
1 Touch Screen
2 Function Menu Operation Keys
3 Multifunction Knob
4 Default Setting Key
5 Touch Lock Key
6 Common Operation Keys
7 Horizontal Control System
8 Vertical Control System
9 Run/Stop Key
10 Single Trigger Control Key
11 Auto Waveform Display Key
12 Arbitrary Waveform Generator Setting Key
13 Trigger Control System
14 Probe Compensation Signal Output Terminal/Ground Terminal
15 Analog Channel Input Terminals
16 Digital Channel Input Interface
17 USB HOST Interface
18 Power Key

What are the functions of the ports on the rear panel of the HANTEK HRDO4054E?

No. Description Function
1 Handle Allows for easy carrying of the oscilloscope.
2 LAN Interface Used to connect the oscilloscope to a network.
3 USB DEVICE Interface Connects the oscilloscope to a computer for remote control via SCPI commands.
4 Reference clock Provides a more accurate sampling clock signal or synchronizes clocks with other oscilloscopes.
5 Aux output terminal Outputs a trigger signal or a Pass/Fail test pulse.
6 EXT input terminal Used to input external trigger signals to the oscilloscope.
7 G1,G2 signal source output terminal Outputs the waveform set by the signal generator.
8 HDMI (Option) Connects to an external display like a monitor or projector.
9 RS232/485 (Option) Serial communication interface.
10 AC Power Cord Connector Power input terminal for the oscilloscope.

How do I use the touch screen controls on the HANTEK HRDO4054E?

The HANTEK HRDO4054E features a 10.1 inch capacitive touch screen that supports multi-touch and gestures. The menus and enabled label buttons on the screen are all touch-operable.

1. Tap:

Tap a menu or function on the screen with one finger to operate menus, use the pop-up numeric keyboard, set tag names with the virtual keyboard, or close pop-up windows.

2. Drag:

After touching and holding a target, drag it to the desired position. This can be used to change a waveform’s displacement/offset or to move a window.

3. Pinch & Stretch:

Bring two fingers together (pinch) or move them apart (stretch) on the screen to reduce or enlarge the waveform. Vertical pinching/stretching adjusts the vertical gear (V/div), and horizontal pinching/stretching adjusts the horizontal gear (s/div).

4. Rectangle Drawing:

Click on the Hantek logo, then click on [AreaSel] to switch to rectangular drawing mode. Drag your finger on the screen to draw a rectangle over a part of the waveform. A menu will appear with options to:

• Select “Histogram” to draw a histogram range and open the histogram menu.

• Select “Enlarge Level” to horizontally expand the waveform within the rectangle.

• Select “Enlarge Vertical” to vertically expand the waveform within the rectangle.

• Select “Enlarge Waveform” to expand the waveform in both directions simultaneously.


How do I set the vertical system parameters on the HANTEK HRDO4054E?

You can adjust various vertical system parameters for each channel (CH1-CH4) on your HANTEK HRDO4054E:

Enable/Disable Channel: Press the corresponding channel button (e.g., [CH1]) or tap the channel status label at the bottom of the screen to turn the channel on or off.

Adjust Vertical Scale (V/div): Rotate the vertical gear knob for the selected channel to increase or decrease the waveform amplitude display. You can also tap the channel menu at the bottom and set the value in the pop-up menu.

Adjust Vertical Offset: Rotate the vertical offset knob for the selected channel to move the waveform up or down. You can also adjust it by dragging the waveform on the touch screen.

Channel Coupling: Open the vertical menu for the channel and select Coupling. Options are DC (passes both AC and DC components), AC (blocks DC component), and GND (disconnects the input signal).

Bandwidth Limit: In the channel menu, select BW Limit to engage a 20MHz, 100MHz, 200MHz, or 350Mhz low-pass filter to reduce waveform noise.

Probe Ratio: In the channel menu, select Attenuation to set the probe attenuation ratio (e.g., 1X, 10X, 100X) to match the probe you are using.

Input Impedance: In the channel menu, select Impedance to switch between 1MΩ (high impedance) and 50Ω (for matching 50Ω sources).

Waveform Invert: In the channel menu, select Invert to turn the waveform inversion on or off.

Amplitude Unit: In the channel menu, select Unit to change the amplitude unit to W, A, V, or U.

Fine Tuning: In the channel menu, select Fine to switch between coarse and fine adjustment for the vertical gear, allowing for higher resolution changes.

Channel Delay: Set a delay time from -100ns to 100ns to correct for probe cable transmission delay (zero offset).


How do I configure the horizontal system on the HANTEK HRDO4054E?

The horizontal system controls the time axis of the waveforms. Here’s how to adjust its settings:

To Adjust the Horizontal Time Base (s/div):

The horizontal time base represents the time value per horizontal grid division.

• Use the horizontal time base knob (marked with a wave and arrows) to adjust the time base. Rotate clockwise to decrease the time base (zoom in) and counterclockwise to increase it (zoom out).

• Use a pinch/stretch gesture on the touch screen to adjust the time base.

• Tap the horizontal time base label at the top of the screen to open a pop-up menu for direct value entry.

To Adjust the Horizontal Position:

This adjusts the horizontal position of the trigger point relative to the screen’s center.

• Rotate the horizontal position knob to move all waveforms left or right. Press the knob to reset the position to zero.

• Drag the waveform horizontally on the touch screen.

Delayed Sweep (Zoom):

This function horizontally amplifies a section of the waveform for detailed viewing.

• Press the [Zoom] button in the horizontal control area. The screen will split into two display areas.

• The top half shows the original waveform (main time base).

• The bottom half shows the horizontally expanded waveform (delayed scanning time base).

• Use the horizontal time base knob to change the zoom factor and the horizontal position knob to move the zoomed window across the original waveform.


How can I set the sample and acquisition system on the HANTEK HRDO4054E?

Press the [Acquire] key to enter the sampling settings menu. Here you can configure the following:

1. Timebase Mode:

YT Mode: The standard mode where the Y-axis is voltage and the X-axis is time.

XY Mode: Displays a Lissajous pattern where the X-axis and Y-axis both represent voltage levels, useful for measuring phase differences.

ROLL Mode: For slow signals, the waveform scrolls from right to left across the screen. This mode is available for time base settings of 50ms/div and slower.

2. Acquisition Mode:

Normal: Samples the signal at equal time intervals. This provides the best display for most waveforms.

Average: Averages multiple sampled waveforms to reduce random noise and improve vertical resolution. You can set the number of averages from 2 to 1024.

Peak: Captures the maximum and minimum values within each sampling interval to display the signal envelope and avoid aliasing on narrow pulses.

High Resolution: Averages adjacent points from a single acquisition to reduce noise and produce a smoother waveform.

3. Memory Depth:

This is the number of waveform points stored in a single acquisition. A higher memory depth allows for a higher sample rate at longer time base settings. You can select from various options depending on whether you are in single, dual, or full channel mode (e.g., Auto, 25K, 250K, 2.5M, up to 1G).

4. Horizontal Expansion:

Determines the reference point for horizontal expansion/compression. Options include expanding from the center, left, right, or trigger point of the screen.

5. Segmented Acquisition:

Allows the capture and storage of multiple trigger events into segments, ignoring the dead time between events.


How do I set up a trigger on the HANTEK HRDO4054E?

Setting a trigger correctly is essential for a stable waveform display. You can enter the trigger menu by pressing the [Trig Menu] button or tapping the trigger status area at the top of the screen.

1. Set Trigger Type:

In the [Trig Menu], select Type. The HANTEK HRDO4054E offers numerous trigger types, including:

Edge: Triggers on a rising, falling, or either edge of the signal.

Pulse: Triggers on pulses of a specified width (positive or negative).

Video: Triggers on standard analog video signals (NTSC, PAL/SECAM).

Slope: Triggers on the slew rate (positive or negative slope time) of the signal.

• Other advanced types like Runt, Window, Logic, Delay, Setup/Hold, and serial bus triggers (UART, I2C, SPI, CAN, LIN).

2. Set Trigger Source:

Select the Source menu to choose the input for the trigger circuit. Options include any analog channel (CH1-CH4), an external trigger signal ([EXT In]), or digital channels (D0-D15).

3. Set Trigger Level:

Rotate the Trigger knob to move the trigger level line up or down on the display. The trigger occurs when the source signal crosses this voltage level. Press the knob to set the level to 50% of the waveform’s peak value.

4. Set Trigger Mode:

Auto: The oscilloscope acquires a waveform even without a valid trigger. If no trigger occurs, it will auto-trigger after a certain time, which may result in an unsynchronized display.

Normal: The oscilloscope only acquires and updates the display when a valid trigger event occurs. If no trigger occurs, the display remains unchanged.

Single: The oscilloscope waits for a single trigger event, captures one waveform, and then stops acquiring.

5. Set Trigger Holdoff:

In the [Trig Menu], adjust the Holdoff time. This sets a period after a trigger during which the oscilloscope will not accept another trigger, which is useful for stabilizing complex, repetitive waveforms like pulse trains.


What types of mathematical operations can the HANTEK HRDO4054E perform?

Press the [Math] button to access the mathematical operations menu. The HANTEK HRDO4054E supports several types of operations:

Arithmetic Operations: Perform point-by-point math between two sources (A and B). Includes Addition (A+B), Subtraction (A-B), Multiplication (A*B), and Division (A/B).

FFT (Fast Fourier Transform): Calculates and displays the frequency spectrum of a source waveform. You can configure the source, frequency range (span/center or start/end), vertical units (dBm/dBV or Vrms), and use different windowing functions.

Logic Operations: Perform logical operations on waveforms based on a set threshold. Includes AND (A&&B), OR (A||B), XOR (A^B), and NOT (!A).

Function Operations: Apply a mathematical function to a single source waveform. Functions include Intg (integral), Diff (differential), Sqrt (square root), Lg (base 10 logarithm), Ln (natural logarithm), Exp (exponent), and Abs (absolute value).

Digital Filtering: Apply a digital filter to a source waveform. Filter types include LowPass, HighPass, BandPass, and BandStop.

Expression: Allows you to create and edit complex mathematical expressions by combining multiple operators and sources.


How do I perform automatic measurements on the HANTEK HRDO4054E?

The HANTEK HRDO4054E can automatically measure up to 51 different waveform parameters. To set up automatic measurements, press the [Meas] button.

Adding a Measurement:

1. In the measurement menu, go to Add.

2. Select the Category of measurement: Horizontal (time-related), Vertical (voltage-related), or Other (delay, phase, etc.).

3. Tap the desired parameter from the on-screen list (e.g., Freq, Period, VMax, Pk-Pk).

4. Select the Source A (and Source B for two-source measurements) for the measurement (e.g., CH1, CH2, Math).

The measurement results will be displayed at the bottom of the screen.

Measurement Statistics:

You can view statistics for your measurements. In the measurement menu, select Statistic. This will display the current (Cur), average (Avg), maximum (Max), minimum (Min), standard deviation (Dev), and count (Cnt) for each active measurement.

All Measure:

To display all available parameters for a specific channel simultaneously, go to [Meas] -> All Measure and select the desired channel. A table with all measurement results for that source will appear on the screen.

Cursor Measurement:

Press the [Cursors] button for manual measurements:

Mode: Choose between Manual or Track.

Select: Choose between X cursors (vertical lines for time/frequency) and Y cursors (horizontal lines for voltage).

Adjust: Use the multifunction knob V0 to position cursors A and B. The measurement window will display the values at each cursor (AX, AY, BX, BY) and the difference between them (ΔX, ΔY).


How do I configure and use the serial bus decoding function on the HANTEK HRDO4054E?

The HANTEK HRDO4054E can decode several common serial protocols. To begin, press the [Decode] button on the front panel.

General Setup Steps:

1. Select Bus Type: In the decode menu, press Bus Type and select the protocol you want to decode (e.g., UART, I2C, SPI, CAN, LIN).

2. Enable Bus Status: Press Bus Status to turn the decoder ON.

3. Configure Source(s): Press Source to enter the source settings menu. Here you will assign the physical channels (CH1-CH4, D0-D15) to the protocol’s signal lines (e.g., for UART: Tx and Rx; for I2C: SCL and SDA).

4. Set Thresholds: For each assigned channel, set the logic threshold voltage. When adjusting, a dashed line will appear on the screen to indicate the level.

5. Configure Protocol Parameters: Go to the Setting menu to define protocol-specific parameters. This is a critical step.

• For UART: Set Baud rate, Data bits, Parity, Stop bits, and Endian (bit order).

• For I2C: The R/W bit can be included or not.

• For SPI: Set Timeout, Clock Edge (Slope), Endian, and Bus Width.

• For LIN: Set Baud Rate, Parity, and protocol Version.

• For CAN: Define the Signal type (CAN_H, CAN_L, etc.), Baud rate, and Sample Position.

6. Configure Display: In the Display menu, you can set the format of the decoded data (Hex, Dec, Bin, ASCII), the on-screen position of the decode bus, and open an Event Table for a detailed, chronological view of the decoded data packets.

Once configured, the decoded data will be overlaid on the waveform, showing values, packets, and errors in real-time.


How do I use the Arbitrary Waveform Generator (Option) on the HANTEK HRDO4054E?

The built-in Arbitrary Waveform Generator (AWG) can be accessed by pressing the [G1] or [G2] button in the “Wave Gen” section of the front panel.

To Output Basic Waveforms:

1. Press [G1] to open the signal generator settings for channel 1.

2. Click the Wave menu tab and select a basic waveform: Sine, Square, Ramp, Pulse, DC, Noise, Sinc, Exp.Rise, Exp.Fall, ECG, Gauss, Lorentz, or Haversine.

3. Adjust the waveform parameters:

Frequency: Set the frequency of the waveform.

Amplitude: Set the peak-to-peak voltage.

Offset: Set the DC offset voltage.

Start Phase: Set the starting phase of the waveform (0° to 360°).

Other parameters: Depending on the waveform, you can also set Duty Cycle (for Square), Symmetry (for Ramp), or Pulse Width/Rise Time/Fall Time (for Pulse).

4. Impedance: Select the output impedance (HighZ or 50Ω).

To Output an Arbitrary Waveform:

1. In the Wave menu, select Arb.

2. You can then choose to:

Load from CH: Load a waveform currently captured on one of the oscilloscope channels (CH1-CH4).

Load Stored: Load an arbitrary waveform file (*.arb) from internal or external memory.

Create/Edit: Use the built-in editor to create a new arbitrary waveform point by point.

Modulation and Burst:

In the generator settings, go to More -> Setting -> Type to select advanced functions:

Modulation: Apply AM, FM, or PM to a carrier wave. You can configure the modulating waveform, frequency, and depth/deviation.

Burst: Output a specified number of waveform cycles (N Cycle) or a continuous wave (Infinite) upon receiving a trigger. The trigger source can be internal (timed) or manual.


What should I do if the waveform display on my HANTEK HRDO4054E is ladder-like?

A ladder-like waveform display can be caused by two main factors:

1. The horizontal time base might be set too slow for the signal being viewed. Try increasing the horizontal time base (i.e., decreasing the s/div value) to see if it improves the display.

2. If the display type is set to “Vector”, the lines drawn between sample points can sometimes create a ladder-like effect, especially with a low sampling rate relative to the signal frequency. Check the display settings.


What should I do if my HANTEK HRDO4054E does not display a waveform on the screen?

If you have a signal connected but no waveform is displayed, follow these troubleshooting steps:

1. Check that the probe is correctly and securely connected to the BNC connector of the signal input channel.

2. Ensure the channel you are using is turned on. You can check this by looking at the channel menu buttons (CH1-CH4) or the on-screen channel indicators.

3. Verify that the circuit or device you are testing is actually outputting a signal.

4. If you are measuring a DC signal with a large amplitude, you may be off-screen. Try increasing the vertical scale (V/div) or adjusting the vertical offset.

5. Press the [Auto Scale] button. This will cause the oscilloscope to automatically adjust the vertical, horizontal, and trigger settings to find and display the signal.


Why can’t I use the touch functions on my HANTEK HRDO4054E?

If the touch screen is not responding, check the following:

1. Make sure the touch screen function is not locked. Press the [Touch Lock] key in the button area to enable or disable the touch function. The key will be lit when the touch screen is locked.

2. Check for any dirt, dust, or moisture on the screen or your fingers, and clean if necessary.

3. Check if the oscilloscope is near a strong magnetic field. If so, move the device away from the source of interference.


What do the default settings on the HANTEK HRDO4054E configure?

Pressing the [Default Setup] button restores the oscilloscope to a known starting state. The default settings are as follows:

System Setting Value
Horizontal Horizontal Time Base 1us
Horizontal Position 0s
Delayed Sweep Off
Timebase Mode YT
Horizontal Expansion Center
Vertical Vertical Scale 1V
Vertical Offset CH1: 3V, CH2: 1V, CH3: -1V, CH4: -3V
Channel Coupling DC
BW Limit 20M
Attenuation 1X
Input Impedance 1MΩ
Invert Off
Acquire Acquisition Mode Normal
Memory Depth 6.25K
Trigger Trigger Type Edge Trigger
Source Selection CH1
Edge Type Rising
Trigger Holdoff 8ns
Display Display Type Vector
Persistence Time Min
Intensity 100%
Grid Line
AWG AWG Off
Wave Sine
Cursor Mode Off
Utility Sound On
Math Operator A+B
LA state Off

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD HANTEK HRDO4054E (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.