Free English HANTEK HRDO4104C (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.
CLICK HERE TO DOWNLOAD HANTEK HRDO4104C (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
What are the general safety precautions for using the HANTEK HRDO4104C?
To ensure safe operation of your HANTEK HRDO4104C, please adhere to the following safety precautions:
Only professionally authorized personnel can perform repairs.
Use the right power cable: Use only the power cable approved for the country in which the product is used.
Connect and disconnect correctly: Before connecting a probe to a circuit, connect the probe to the oscilloscope first. Before disconnecting the probe from the oscilloscope, disconnect the probe from the circuit under test first.
Ground the product: To avoid electric shocks, the product is grounded through the power cable’s grounding conductor. This must be connected to earth ground before connecting any input or output terminals.
Connect the probe properly: The ground wire of the probe is at ground potential. Do not connect the ground wire to a high voltage.
View all terminal rating values: To avoid fire or excessive current, check all rating values and signs on the product before connecting.
Do not operate with the cover open: Do not run the product with its cover or panel open.
Avoid circuit exposure: Do not touch exposed connectors and components when the power is on.
Do not operate if the product is suspected to be faulty: If you suspect damage, have it inspected by qualified maintenance personnel.
Maintain proper ventilation: Ensure the product has adequate ventilation.
Operating environment: Do not operate in a humid, inflammable, or explosive environment. Keep the product surface clean and dry.
How should I care for and clean my HANTEK HRDO4104C oscilloscope?
Care:
Do not expose the LCD to direct sunlight for a long time when storing or placing the oscilloscope.
Cleaning:
Clean the outer surface of the instrument frequently as required by operating conditions by following these steps:
1. Use a lint-free cloth to remove dust from the outside of the oscilloscope and probe. Be careful not to scratch the display filter.
2. Clean the oscilloscope with a soft cloth soaked in water. For a more thorough cleaning, you can use an aqueous solution of 75% isopropyl alcohol.
Note: To avoid damaging the surface of the oscilloscope or probe, do not use any corrosive or chemical cleaning reagents.
Warning: Before powering on the device again, ensure it is completely dry to avoid electrical short circuits or personal injury from moisture.
What are the main product features of the HANTEK HRDO4104C series?
The HANTEK HRDO4104C series digital oscilloscope boasts the following features:
• 12-bit vertical resolution, up to 10GSa/s sampling rate, and maximum 3GHz bandwidth.
• 8 independent instruments in one: 4-Channel oscilloscope, 16-channel logic analyzer, spectrum analyzer, 2-Channel arbitrary wave generator, digital voltmeter, frequency meter and accumulator, and protocol analyzer/generator.
• Built-in dual-channel 200MHz signal source with ±7V voltage output and 1GSa/s sampling rate.
• A 10.1-inch multi-touch capacitive screen with 256-level waveform grayscale and color temperature display.
• Rich serial protocol triggering and decoding functions.
• Waveform capture rate higher than 1.1 million waveforms per second.
• Up to 51 kinds of waveform parameters are automatically measured.
• Multiple data analysis and processing functions: independent search, navigation buttons and event lists, histograms, Bode plots, power analysis, and frequency meters.
How do I perform an initial examination of my HANTEK HRDO4104C after receiving it?
After receiving your HANTEK HRDO4104C, follow these steps to check the instrument:
1. Check the shipping package:
Inspect for any damage caused by transportation. If the packaging cartons or protective foam pads are seriously damaged, keep them until the entire machine and accessories pass electrical and mechanical testing.
2. Check the accessories:
Verify that all accessories listed in Appendix A are present and undamaged. If anything is missing or damaged, contact your dealer.
3. Check the machine:
If you find the instrument is damaged, not working properly, or fails its performance test, contact your dealer.
How do I set up and perform a function inspection on the HANTEK HRDO4104C?
Follow these steps for initial setup and function inspection:
1. Adjust the Supporting Legs: Adjust the bracket at the bottom of the machine to tilt the oscilloscope slightly upwards for better observation and stability.
2. Connect to AC Power: Use the provided power cord to connect the oscilloscope to a power supply. The input specifications are 47-440Hz, 90-264VAC/127-320VDC, 80Watts MAX. Ensure the instrument is correctly grounded.
3. Turn-on Checkout: Press the power button on the lower left of the front panel. During startup, all button lights will illuminate, and the oscilloscope will perform a self-check. The startup screen will appear upon completion.
4. Function Inspection:
• Click [Default Setup] in the button area to restore the oscilloscope to factory settings.
• Connect the grounding crocodile clamp of the probe to the grounding terminal.
• Connect the probe to the input terminal of channel 1 (CH1) and the compensation signal output terminal.
• Set the probe attenuation ratio to X10 and click the [Auto Scale] button.
• Observe the waveform to see if a square wave signal is displayed normally. Check CH2-CH4 using the same method.
How do I perform probe compensation on the HANTEK HRDO4104C?
When connecting a probe to any input channel for the first time, you must perform this adjustment to match the probe with the channel. Uncompensated probes can lead to measurement errors.
1. Perform the first four steps of the “Function Inspection” procedure.
2. Compare the displayed waveform with the compensation diagrams (Perfectly compensated, Over compensated, Under compensated).
3. If necessary, use a non-metallic screwdriver to adjust the variable capacitance on the probe until the waveform on the screen matches the “Perfectly compensated” diagram.
Tip: The probe compensation signal is only for probe compensation and cannot be used for calibration.
What are the functions of the main controls on the front panel of the HANTEK HRDO4104C?
The front panel of the HANTEK HRDO4104C is divided into several control sections:
Horizontal:
• Search: Enters the search settings menu to find events in collected data.
• Zoom: Toggles the delayed scanning (zoom) function on or off.
• Time Base Knob: Rotate to change the horizontal time base (s/div). Press to switch between “Coarse” and “Fine” adjustment.
• Position Knob: Rotate to change the horizontal position of the waveform. Press to reset the position to zero.
• Navigation Control Keys: Perform recording, playback, time, and event navigation.
Vertical:
• CH1, CH2, CH3, CH4 Buttons: Enable or disable the respective analog input channels.
• Vertical Offset Knob: Rotate to adjust the vertical offset of the current channel. Press to reset the offset to zero.
• Vertical Gear Knob: Rotate to change the vertical scale (V/div) of the current channel. Press to switch between “coarse” and “fine” adjustment.
Waveform Control System:
• REF: Opens the reference waveform setting menu.
• Math: Opens the mathematical operation function menu (A+B, A-B, A×B, A/B, FFT).
• LA: Opens the logic analyzer control menu.
• Decode: Opens the decoding settings menu.
Wave Gen (Waveform Generator):
• G1, G2: Opens the signal generator channel G1 or G2.
• Burst: Activates the burst function.
• Signal Source Knob: Modifies the frequency and amplitude of the signal source. Press to select which parameter to adjust.
Trigger:
• Trigger Knob: Adjusts the trigger level. Press to set the trigger level to 50% of the waveform peak.
• Trig Menu: Opens the trigger operation menu.
• Force Trig: Forces the oscilloscope to generate a trigger signal.
• Mode: Switches the trigger mode between Auto, Normal, or Single.
Other Keys:
• AUTO: Automatically adjusts settings for an optimal waveform display.
• Run/Stop: Starts or stops waveform acquisition.
• Single: Sets the trigger mode to acquire a single waveform then stop.
• Multifunction Knob (V0): Modifies setting parameters in various menus.
• Default: Restores the oscilloscope to its factory default settings.
• Touch Lock: Disables the touch screen function.
Function Menu:
• Quick Action: Performs quick operations like screenshots, saving waveforms, and saving settings.
• Analyze: Enters the analysis menu.
• Cursors: Enters the cursor measurement menu.
• Save/Recall: Performs save and recall functions.
• Utility: Enters the system function settings menu.
• Help: Accesses the built-in help system.
• Meas: Enters the measurement settings menu.
• Acquire: Enters the sampling settings menu.
• Display: Enters the display settings menu.
What are the functions of the ports on the rear panel of the HANTEK HRDO4104C?
The rear panel of the HANTEK HRDO4104C includes the following ports and features:
1. Handle: Can be pulled up for carrying the oscilloscope.
2. LAN interface: Used to connect the oscilloscope to a network.
3. USB VIEW interface: Connects the oscilloscope to a computer for remote control via SCPI commands or software.
4. Reference clock: Input for an external reference clock to provide a more accurate sampling clock or to synchronize multiple oscilloscopes.
5. Aux output terminal: Can output a trigger signal (Trigger output) reflecting the capture rate or a pulse for Pass/fail test results.
6. EXT input: Used to input an external trigger signal.
7. G1, G2 signal source output terminal: Outputs the waveform set by the signal generator.
8. HDMI interface (option): Connects to an external display like a monitor or projector.
9. RS-232/485 interface (option): Serial communication interface for the oscilloscope.
10. Power socket: Power input terminal for the AC power cord.
How do I use the touch screen controls on the HANTEK HRDO4104C?
The HANTEK HRDO4104C features a 10.1-inch capacitive touch screen supporting several gestures:
1. Tap:
Tap a menu or function on the screen with one finger to operate menus, use the pop-up numeric keyboard, set file names with the virtual keyboard, or close pop-up windows.
2. Drag:
After touching and holding a target, drag it to the desired position. This can be used to drag a waveform to change its vertical offset or to move a window to a new position.
3. Pinch & Stretch:
Bring two fingers together (pinch) or move them apart (stretch) on the screen to zoom in or out on a waveform. Pinching in the vertical direction adjusts the vertical scale (V/div), and pinching in the horizontal direction adjusts the horizontal time base (s/div).
4. Rectangle Drawing:
Click on the [Hantek] logo, then click on AreaSel to enter rectangular drawing mode. Drag your finger on the screen to draw a rectangle. A menu will appear allowing you to select functions like “Histogram”, “Enlarge Level” (horizontal zoom), “Enlarge Vertical” (vertical zoom), or “Enlarge Waveform” (both horizontal and vertical zoom) for the selected area.
How do I enable, disable, and adjust the vertical settings for an analog channel on the HANTEK HRDO4104C?
To Enable an Analog Channel (e.g., CH1):
• Press the [CH1] button on the front panel. The button and the corresponding channel knob will light up.
• Alternatively, click the channel status label at the bottom of the screen. The label will light up.
To Disable an Analog Channel:
• If the channel is currently selected, press its corresponding button (e.g., [CH1]) once to close it.
• If the channel is open but not selected, press its button once to select it, then press it again to close it.
• Use the touch screen to click on the channel number on the channel menu at the bottom of the screen to close it.
To Adjust the Vertical Scale (V/div):
• Rotate the vertical gear knob corresponding to the channel. Rotate clockwise to decrease the V/div value (zoom in vertically) and counterclockwise to increase it (zoom out vertically).
• Use the touch screen: click the channel menu at the bottom of the screen, and in the pop-up menu, set the vertical gear.
To Adjust the Vertical Offset:
• Rotate the vertical offset knob for the channel to move the waveform up or down. Press the knob to reset the offset to zero.
• Use the touch screen: drag the waveform up or down to adjust the offset.
How do I configure Channel Coupling, Bandwidth Limit, and Input Impedance on the HANTEK HRDO4104C?
Channel Coupling:
Open the vertical menu for the desired channel and select the “Coupling” menu.
• DC: Both DC and AC components of the signal pass through.
• AC: The DC component of the signal is blocked.
• GND: Both DC and AC components are blocked, disconnecting the input signal.
Bandwidth Limit:
Open the channel menu and select the “BW Limit” menu. This reduces noise on the displayed waveform.
• Off: Full bandwidth for the channel.
• 20MHz/100M/200M/350M: Limits the bandwidth to the selected frequency, attenuating higher frequency components.
Input Impedance:
Open the channel menu and click the “Impedance” menu tab.
• 1M Ω: High input impedance, suitable for most general-purpose measurements.
• 50 Ω: Used to match the oscilloscope with a device that has a 50 Ω output impedance.
How do I adjust the Horizontal Position and Time Base on the HANTEK HRDO4104C?
To Adjust the Horizontal Position:
The horizontal position refers to the displacement of the trigger point relative to the center of the screen.
• Using the Knob: Rotate the horizontal position knob in the “Horizontal” control area. Rotating it moves the waveform and its trigger point left or right. Press the knob to reset the horizontal displacement to 0.00s.
• Using the Touch Screen: Click on the horizontal displacement label at the top of the screen to open a pop-up menu where you can enter a value directly or use increase/decrease buttons.
• Using Drag Gesture: Drag the waveform horizontally across the screen.
To Adjust the Horizontal Time Base (s/div):
The horizontal time base is the time value represented by each horizontal grid division.
• Using the Knob: Rotate the horizontal time base knob in the “Horizontal” control area. Rotating clockwise decreases the time per division (zooms in), and counterclockwise increases it (zooms out).
• Using the Touch Screen: Click on the horizontal time base label at the top of the screen to open a pop-up menu for direct entry or adjustment.
• Using Pinch Gesture: Use a horizontal pinch or stretch gesture on the waveform to adjust the time base.
How do I use the different Timebase and Acquisition Modes on the HANTEK HRDO4104C?
Open the [Acquire] menu to access these settings.
Timebase Modes:
• YT Mode: This is the standard mode where the Y-axis represents voltage and the X-axis represents time. At time base settings of 100 ms/div or slower, the oscilloscope enters a scanning mode.
• XY Mode: Displays a Lissajous pattern where the X-axis and Y-axis both represent voltage levels from two different channels. This is useful for measuring phase differences between two signals of the same frequency.
• ROLL Mode: The waveform scrolls from right to left across the screen, which is useful for observing low-frequency signals. The horizontal gear range is 50ms to 1 ks.
Acquisition Modes:
• Normal: The oscilloscope samples the signal at equal time intervals to reconstruct the waveform. This is the best mode for most applications.
• Average: The oscilloscope averages multiple sampled waveforms to reduce random noise and improve vertical resolution. You can set the number of averages from 2 to 1024.
• Peak: The oscilloscope captures the maximum and minimum values of the signal within each sampling interval. This mode is useful for capturing glitches and avoiding aliasing.
• High Resolution: This mode averages adjacent points of a single sampled waveform to reduce random noise and produce a smoother display. It is used when the sampling rate is higher than the storage rate of the acquisition memory.
How do I set up the trigger system on the HANTEK HRDO4104C?
You can enter the trigger menu by pressing the [Trig Menu] button or clicking the trigger tab at the top of the screen. The main trigger settings are:
1. Trigger Type:
Press the [Trig Menu] button, then select the “Type” menu. The HANTEK HRDO4104C supports numerous trigger types, including Edge, Pulse, Video, Slope, Overtime, Window, Runt, Logic, Pattern, and various serial bus triggers (UART, I2C, SPI, CAN, LIN).
2. Trigger Source:
Select the signal source for the trigger. This can be any of the analog channels (CH1-CH4), the external trigger input (EXT), or any of the digital channels (if a logic probe is connected).
3. Trigger Mode:
This determines how the oscilloscope acquires data when a trigger condition occurs.
• Auto: The oscilloscope will sweep even without a trigger event, allowing you to see a signal. If a trigger event occurs, the display will become stable.
• Normal: The oscilloscope only acquires and updates the waveform when a valid trigger event is detected.
• Single: The oscilloscope waits for a single trigger event, captures one waveform, and then stops acquiring.
4. Trigger LEVEL:
Rotate the “Trigger” knob to adjust the voltage level that the trigger source signal must cross to cause a trigger. Pressing the knob sets the level to 50% of the waveform’s amplitude.
5. Trigger Holdoff:
Set a specific time period after a trigger during which the oscilloscope will not accept another trigger. This is useful for stabilizing the display of complex or periodic waveforms that have multiple valid trigger points within one cycle.
How do I use the Math functions on the HANTEK HRDO4104C?
Press the [Math] button to enter the mathematical operation menu. The HANTEK HRDO4104C supports several types of math operations:
Arithmetic Operations: Perform point-by-point calculations on waveforms.
• Operations: A+B (Addition), A-B (Subtraction), A*B (Multiplication), A/B (Division).
• Sources: Select Source A and Source B from any open channel (CH1-CH4) or a reference waveform (REF).
FFT (Fast Fourier Transform):
Calculates and displays the frequency spectrum of a source signal.
• Source: Select the channel to analyze.
• X frequency range: Set the frequency range using Span/Center or Start/End frequencies.
• Window: Select a windowing function (Rectangular, Hanning, Hamming, Blackman-Harris, etc.) to reduce spectral leakage.
• Peak Search: Automatically find and label peaks in the spectrum.
Logic Operation:
Performs logical operations on waveforms based on a set threshold.
• Operations: A&&B (AND), A||B (OR), A^B (XOR), !A (NOT).
• Sources & Thresholds: Select source channels and set the corresponding voltage thresholds for defining logical high and low.
Function Operation:
Applies a mathematical function to a single source waveform.
• Functions: Intg (Integral), Diff (Differential), Sqrt (Square Root), Lg (Base-10 Log), Ln (Natural Log), Exp (Exponent), Abs (Absolute Value).
Digital Filtering:
Apply a digital filter to a source waveform.
• Filters: LowPass, HighPass, BandPass, and BandStop.
How do I perform automatic measurements on the HANTEK HRDO4104C?
The oscilloscope can automatically measure up to 51 different waveform parameters. To access the measurement menu, press the [Meas] button.
To Add a Measurement:
1. In the [Meas] menu, select “Add”.
2. Choose a “Measurement Category”: Horizontal (time-related), Vertical (voltage-related), or Other (delay and phase).
3. Select the desired “Source” channel (CH1-CH4, Math, etc.).
4. Tap on the specific measurement parameter you want to add (e.g., Freq, Period, VMax, Vpp, etc.). The result will be displayed at the bottom of the screen.
To View All Measurements:
Press [Meas] -> All Measure, and select a channel (e.g., CH1). A table will appear on the screen displaying all available measurement parameters for that channel simultaneously.
To Use Measurement Statistics:
1. In the [Meas] menu, select “Statistic”.
2. A table will appear showing the current (Cur), average (Avg), maximum (Max), minimum (Min), and standard deviation (Dev) for all active measurements.
To Remove Measurements:
In the [Meas] menu, select “Remove” to delete the last added measurement, or “Remove All” to clear all measurements from the display.
How do I use Cursor Measurements on the HANTEK HRDO4104C?
Cursors are used for manual measurements of time (X-axis) and voltage (Y-axis). Press the [Cursors] button to enter the cursor setting menu.
1. Select Mode:
• None: Cursors are off.
• Manual: You can freely move two cursors (A and B) to measure points on the screen. The measurement window shows the absolute position of each cursor and the difference between them.
• Track: The cursors “snap” to the waveform of the selected source. As you move one cursor, the other tracks along the waveform.
2. Using Manual Mode:
• Select Cursor Type: Choose X (vertical lines for time/frequency) or Y (horizontal lines for voltage).
• Select Source: Choose the channel (CH1-CH4, Math) you want to measure.
• Adjust Cursors: Select AX/AY to adjust cursor A and BX/BY to adjust cursor B using the multifunction knob (V0). Select AXBX/AYBY to move both cursors together.
• Read Results: The measurement window displays values like AX, BX, AY, BY, the difference (BX-AX or BY-AY), and for X cursors, the reciprocal (1/dX) which represents frequency.
How do I use the Pass/Fail test function on the HANTEK HRDO4104C?
The Pass/Fail test function monitors a signal and compares it to a predefined mask, counting pass and fail events. Access it via [Utility] -> Pass/Fail.
1. Enable the Function: In the Pass/Fail menu, select “Enable” and turn it on.
2. Select the Source: Choose the channel (CH1-CH4) to test.
3. Create a Mask (Rule):
• Select “Mask” -> “Create Rule”.
• Set the horizontal (X Mask) and vertical (Y Mask) tolerance range around the source waveform. This creates a template.
• Click “Create Rule” again to apply the created mask.
4. Set Options:
• Go to the “Option” menu.
• Stop Output: Choose whether the test should stop upon a failure.
• Error Action: Choose what happens on failure (e.g., sound a buzzer).
• Aux Output: Configure the rear panel “Aux” port to output a pulse on a pass or fail event.
5. Start the Test:
• Select “Operate” and press the play icon to start the test. The oscilloscope will now compare the incoming waveform against the mask.
6. View Information:
• Select “Information” and turn it on to display a counter on the screen showing the total number of waveforms tested, as well as the number of passes and fails.
How can I save and load settings, waveforms, and images on my HANTEK HRDO4104C?
You can save and load data to the oscilloscope’s internal memory or an external USB flash drive. Press the [Save/Recall] button to open the menu.
To Save Data (Storage Type):
• Save Image: Captures the current screen display.
• Format: Choose from *.png, *.bmp, *.jpg, or *.tif.
• File Name: Enter a name using the on-screen keyboard.
• Save: Select the save location (internal or USB) and save the file.
• Save Wave: Saves the waveform data points from the screen or the entire acquisition memory.
• Data Source: Select “Screen” or “Memory”.
• Format: Choose from *.bin or *.csv.
• Save: Select the location and save the file.
• Save Setup: Saves the current configuration of the oscilloscope.
• File Type: Saved as *.stp or *.ptr.
• Save: Select the location and save the file.
To Load Data (Load Type):
• Load Wave: Loads a previously saved waveform file (*.wfm) for display as a reference.
• Load Setup: Loads a previously saved setup file (*.stp or *.ptr) to restore the oscilloscope to a specific configuration.
Disk Management:
In the [Save/Recall] menu, select “Disk” to manage files. Here you can create new folders, copy, paste, rename, and delete files on both internal storage and the connected USB drive.
What should I do if the waveform display on my HANTEK HRDO4104C is ladder-like?
If the waveform appears as a series of steps or “ladder-like,” try the following:
1. The horizontal time base might be too slow for the signal being displayed, resulting in too few sample points per cycle. Try increasing the horizontal time base (i.e., decreasing the time/division value) to improve the display resolution.
2. If the display type is set to “Vector” in the [Display] menu, the oscilloscope draws straight lines between sample points. This can cause a ladder-like appearance on some waveforms. Changing the display mode might alter the appearance, but the primary solution is to adjust the time base.
What should I do if my HANTEK HRDO4104C shows no waveform on the screen?
If no signal waveform is displayed, follow these troubleshooting steps:
1. Check that the probe is correctly and securely connected to the BNC input connector of the signal channel.
2. Check if the channel you are using is enabled. Press the corresponding channel button (e.g., [CH1]) to ensure it is turned on.
3. Verify that the circuit or device you are testing is actually outputting a signal.
4. If it is a DC signal with a large amplitude, you may need to increase the vertical scale (V/div) to bring the signal into view.
5. Press the [Auto Scale] button. This function will automatically detect the signal and adjust the oscilloscope settings to display it.
If you still see no waveform, please contact technical support.
Why can’t I use the touch screen functions on my HANTEK HRDO4104C?
If the touch screen functions cannot be used normally, please check the following:
1. Check if the touch screen function has been disabled. If the [Touch Lock] button in the front panel button area is illuminated, the touch function is locked. Press it to turn off the lock and enable the touch screen.
2. Check for any dirt or debris on the screen or on your fingers. If there is, please clean it first.
3. Check if the oscilloscope is close to a strong magnetic field. Strong magnetic interference can affect touch screen performance. If it is, please move the oscilloscope away from the source of interference.
If you still cannot use the touch screen, please contact Hantek technical support.
What should I do if a USB storage device is not recognized by my HANTEK HRDO4104C?
If the oscilloscope cannot recognize your USB storage device, try these steps:
1. Check if the USB drive is functioning properly by testing it on another device.
2. Confirm the capacity of the USB flash drive. It is recommended to use a USB flash drive that does not exceed 16GB for this oscilloscope.
3. Confirm that the format of the USB drive is FAT32.
4. Restart the oscilloscope and reinsert the USB drive for inspection.
If you still cannot use the USB drive, please contact Hantek Technical Support.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
CLICK HERE TO DOWNLOAD HANTEK HRDO4104C (01) PDF DOCUMENT