What is included in the package for the BIOLOGIX OXISTAR SENSOR BX1?
The BIOLOGIX OXISTAR SENSOR BX1 comes with the following components in its storage case:
1. 1 BIOLOGIX OXISTAR SENSOR BX1 sensor
2. 1 USB cable
3. 1 hypoallergenic tape for finger attachment
4. 1 storage and transport case
5. 1 quick user guide
What do the symbols on the BIOLOGIX OXISTAR SENSOR BX1 and its packaging mean?
| Symbol | Description | Symbol | Description |
|---|---|---|---|
| Box with number | Pairing limit by number of boxes | Umbrella | Keep dry |
| Sun with rays | Protect from sunlight | Trash can with X | Electronic waste – do not dispose with common waste |
| Blue circle with “i” | Consult instructions | Radiating tower | Non-ionizing electromagnetic radiation |
| SN | Serial number | Factory icon | Manufacturer |
| Person in a box | Type BF applied part | Factory with date | Date of manufacture |
| Hourglass | Expected lifetime | Solid and dashed lines | Continuous current |
| Vertical line | Temperature limit | IPX2 | Protection against water drops (max. incline: 15 degrees) |
| Percent in circle | Moisture limit | %SpO2 | Oxygen saturation |
| Circles with pressure lines | Atmospheric pressure limit | Two triangles over SpO2 | Oxygen saturation |
| Two arrows up | This side up | BPM | Heart rate |
| Wine glass | Fragile, handle with care |
What are the important operational safety guidelines for the BIOLOGIX OXISTAR SENSOR BX1?
Before using the BIOLOGIX OXISTAR SENSOR BX1, please observe the following safety instructions:
Substances containing dyes that alter blood pigmentation may result in inaccurate readings.
Clinical conditions such as hypothermia, vasoconstriction, peripheral vascular disease, and anemia may affect measurements.
The device may not function properly on extremely cold fingers due to low blood perfusion. A Perfusion Index (PI) over 1% is recommended.
Do not apply the hypoallergenic tape too tightly, as it may impair blood circulation.
Avoid excessive finger movement during measurement to prevent inaccurate readings.
Do not use the sensor near intense light sources, as this may interfere with signal processing.
The device does not include an alarm system and is not intended for use as an alert device for dangerous levels.
While taking measurements, remain within Bluetooth range of your smartphone (average range: approx. 4 meters).
Use the sensor exclusively with the authorized mobile application.
The device does not function while charging.
Use the sensor over bare skin only. Do not place it over wounds, rashes, burns, or sensitive skin.
Children may use the device only under adult supervision as it contains small parts that may present choking hazards.
Do not open or disassemble the device. It is fully sealed to prevent access to internal components.
Do not use the device in MRI environments.
Avoid using the sensor in environments with high oxygen concentration, flammable anesthetics, or nitrous oxide.
How does the BIOLOGIX OXISTAR SENSOR BX1 measure oxygen saturation and pulse rate?
The BIOLOGIX OXISTAR SENSOR BX1 is a pulse oximeter that measures blood oxygen saturation (SpO2) non-invasively based on the Beer-Lambert Law. It detects the difference in absorption of red and infrared light by oxygenated and deoxygenated hemoglobin.
The device calculates the pulse rate by analyzing light absorption variations during the cardiac cycle: red light is absorbed more during systole, while infrared light is absorbed more during diastole.
The sensor also includes an actimeter that records movements through an internal accelerometer along three axes (X, Y, and Z). Data is stored internally and transmitted to a connected mobile device every five minutes.
What do the LED light indicators on the BIOLOGIX OXISTAR SENSOR BX1 mean?
The LED indicators on the upper part of the BIOLOGIX OXISTAR SENSOR BX1 verify its status as follows:
| Color of LED | Sensor Status |
|---|---|
| No LEDs on | The sensor is off. |
| Green | Data is being measured properly. |
| Blue | The sensor is on and waiting to be paired with the application. |
| Yellow | The data is NOT being measured properly, likely due to excessive finger movement. |
| Red and blinking | The battery is not charged and is below 10% of its total capacity. |
| Red | The battery is currently charging. |
How do I correctly use the BIOLOGIX OXISTAR SENSOR BX1 to perform a test?
Follow these steps to operate the BIOLOGIX OXISTAR SENSOR BX1:
Step 1. Fully charge the internal battery before first use. While charging, the red LED remains on. When fully charged, the red LED turns off and the green LED turns on.
Step 2. After charging, place the sensor on the middle finger (thumb or pinky are not recommended) with the LEDs facing the palm. The device turns on automatically upon detecting the finger, and the blue LED will begin blinking once per second. Ensure nails are short and free of polish.
Step 3. Secure the sensor to the finger using a piece of hypoallergenic tape over the protruding part. Avoid excessive pressure.
Step 4. Open the mobile application and start the test when the blue light begins to blink.
Step 5. Once the app connects via Bluetooth, the sensor will start blinking green.
Step 6. The test should last at least 4 hours. It is recommended not to exceed 12 hours.
Step 7. If necessary, you may remove the sensor during the test for up to 30 minutes without the test ending prematurely.
Step 8. Upon waking, finish the test in the application. Remove the sensor and store it properly.
What are the technical specifications for the BIOLOGIX OXISTAR SENSOR BX1?
The technical specifications for the BIOLOGIX OXISTAR SENSOR BX1 include:
| Category | Specification |
|---|---|
| Storage and Operation Temperature | -25 degrees C to +70 degrees C |
| Storage and Operation Humidity | 15% to 93%, non-condensing |
| Atmospheric Pressure | 700 hPa to 1,060 hPa |
| Charger Output Voltage | 4.5 to 5.5 VDC |
| Charger Output Current | Above 0.1 A |
| Battery Type | Lithium-ion Rechargeable |
| Battery Life (Fully Charged) | 30 hours |
| Full Charging Time | Standard Charge: 3 hours; Fast Charge: 1.5 hours |
| SpO2 Measurement Range | 30 to 100% |
| Pulse Rate Measurement Range | 30 to 250 bpm |
| Accuracy (SpO2) | +/- 2% from 70 to 100% |
| Accuracy (Pulse Rate) | +/- 2 bpm from 30 to 250 bpm |
| Dimensions | 50.4 mm Length x 21.9 mm Width x 27.3 mm Height |
| Weight | 20 g |
| Bluetooth Version | 4.0 low energy (BLE) |
| Ingress Protection Code | IPX2 |
How should I clean and maintain the BIOLOGIX OXISTAR SENSOR BX1?
The BIOLOGIX OXISTAR SENSOR BX1 is a reusable device. Sterilization is recommended every three uses:
Clean the exterior area of the sensor with a 70% alcohol wipe.
Do not immerse any part of the sensor in liquid.
Clean the hypoallergenic tape area using an adhesive remover on a damp towel.
Ensure the device is fully dry before reuse.
Do not use abrasive materials, acetone, or acetone-based products.
Do not sterilize the device using an autoclave or gas.
Unplug the device before cleaning; do not clean it while connected to a power source.
To maintain battery health, charge the sensor at least once every three months.
What should I do if the BIOLOGIX OXISTAR SENSOR BX1 is not operating correctly?
Consult the following troubleshooting guide for the BIOLOGIX OXISTAR SENSOR BX1:
The blue LED does not blink: If the battery is depleted, recharge it. If the power button is damaged or the battery is non-functional, contact authorized support.
The battery does not charge (Red LED off): Ensure the charger is connected to a working power source. Verify the charger meets specifications (4.5 to 5.5 VDC). Try a different compatible charger.
The yellow LED blinks: This indicates excessive movement. Minimize movement during the measurement. If the tape is too tight, reposition it. If you have low perfusion or cold fingers, try a different finger and ensure the Perfusion Index exceeds 1%.
The application says “No sensor found”: Press the power button once to activate the sensor. Ensure the sensor is within 10 meters of the smartphone.
The application says “Sensor disconnected”: Move the sensor closer to the smartphone or recharge the battery if it has depleted.
The red LED blinks: The battery charge is below 10%. Recharge the battery immediately.
How should the BIOLOGIX OXISTAR SENSOR BX1 be disposed of?
Dispose of the BIOLOGIX OXISTAR SENSOR BX1 responsibly by keeping it separate from household trash. The internal battery is considered hazardous waste. Take the device to designated collection points for the recycling of electrical and electronic equipment at the end of its useful life. You may also return it to the manufacturer for proper disposal.