AEMC INSTRUMENTS J93 (01) Image

Free English AEMC INSTRUMENTS J93 (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS J93 (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

What safety precautions should be followed before using the AEMC INSTRUMENTS J93?

The protection assured by the current probe can be compromised if it is used in a way that is not recommended by the manufacturer. Always adhere to the following precautions:

■ Comply with the rated maximum voltage and current, and the measurement category. Do not use the current probe on networks where the voltage or category exceeds those specified.

■ Comply with the conditions of use (e.g. temperature, humidity, altitude, degree of pollution, location).

■ Do not use the current probe if its housing is open, deteriorated, or incorrectly reassembled. Before each use, check the integrity of the insulation of the unit, jaws, clamps, housing, and leads.

■ Do not subject the current probe to water or other liquids.

■ Keep the jaw contacts of the clamp absolutely clean.

■ Use suitable personal protective equipment when hazardous voltages may be accessible in the installation where the measurement is made.

■ Any repairs must be carried out by accredited skilled personnel.


What do the symbols on the AEMC INSTRUMENTS J93 and in its user manual mean?

A (triangle with exclamation mark)

WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.

A (triangle with lightning bolt)

CAUTION! Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

Lightning bolt symbol

Application or withdrawal authorized on conductors carrying dangerous voltages. Type A current sensor as per IEC 61010-2-032.

Crossed-out lightning bolt symbol

Must not be applied to or removed from conductors at dangerous voltages. Type B current sensor as per IEC 61010-2-032.

□ (Square within a square)

Equipment is protected by double insulation.

— +

Battery

!

Important instructions to read and to fully understand.

i

Useful information or tip to read.

CE

The CE marking guarantees conformity with European directives and with regulations covering EMC.

Crossed-out trash can

The trash can with a line through it means that in the European Union, the product must undergo selective disposal for the recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.


What are the definitions of Measurement Categories (CAT) for the AEMC INSTRUMENTS J93?

CAT IV: Measurement category IV corresponds to measurements taken at the source of low-voltage installations. Example: power feeders, counters and protection devices.

CAT III: Measurement category III corresponds to measurements on building installations. Example: distribution panel, circuit-breakers, machines or fixed industrial devices.

CAT II: Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage installations. Example: power supply to domestic electrical appliances and portable tools.


What are the main control features of the AEMC INSTRUMENTS J93?

The AEMC INSTRUMENTS J93 has the following features:

1. Jaws

2. 4-point CA connector

3. Safety guard

4. Zero adjustment knob

5. Power ON/Low battery indicator

6. Three-position switch: ON, OFF, battery test


How do I use the AEMC INSTRUMENTS J93 to measure current?

The AEMC INSTRUMENTS J93 is used to measure the current flowing in a conductor or bus bar without opening the circuit. Follow these steps for operation:

1. For three-phase measurements, you can use color-coded ID markers to associate a color for each current input to match the phase identifiers on the measured system.

2. Connect the current probe to the current terminals of the measuring instrument. IMPORTANT NOTE: Always connect the probes with the indicating arrows pointing toward the load.

3. Press the jaw opening lever on the probe to open the jaws.

4. Clamp the probe around the conductor to be tested. For best results, center the conductor in the jaws of the clamp.

5. Refer to the user manual of the instrument the current probe is being used with for details on measurement configuration.


How do I clean the AEMC INSTRUMENTS J93?

To clean your AEMC INSTRUMENTS J93, follow these instructions:

■ Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth.

■ Do not use alcohol, solvents, or hydrocarbons.

■ Do not splash water directly on the instrument.

CAUTION: Risk of electric shock. Disconnect the instrument from any source of electricity before cleaning.


How do I replace the battery on the AEMC INSTRUMENTS J93?

Follow these steps to replace the battery in the AEMC INSTRUMENTS J93:

■ Insert a tool, not more than 3mm in diameter, in the hole in the battery compartment cover.

■ Push to unlock the battery compartment cover, then slide it off.

■ Remove it completely by hand.

■ Remove the battery and the shim from the compartment.

■ Place the new battery in the compartment with the polarity as indicated on the label. Then replace the shim.

■ Put the battery compartment cover back in the slide and push it in until you hear a click.

Used batteries must not be treated as ordinary household waste. Recycle them appropriately.


What are the electrical specifications for the AEMC INSTRUMENTS J93?

The electrical specifications for the AEMC INSTRUMENTS J93 are as follows:

Model Measurement Range Compatibility
J93 50 to 3500; 50 to 5000 (DC only) 8333, 8336, 8435 and PEL

Battery: 9V Alkaline NEDA 1604A, 6LR61

Battery Life: 70H typical (NOTE: Battery life is twice the typical value when using Lithium batteries)


What are the mechanical and environmental specifications for the AEMC INSTRUMENTS J93?

Environmental Specifications:

Use: Indoor use.

Operating Temperature: 14° to 131°F (-10° to 55°C); 10% to 85% RH

Storage Temperature: -40° to 158°F (-40° to 70°C); 10% to 90% RH

Degree of pollution: 2

Altitude: < 2000 m

Mechanical Specifications:

Model Lead Length (nominal) Clamping Diameter Dimensions Weight
J93 10 ft (3m) 2.84″ (72mm) 13.23 x 5.00 x 1.65″ (336 x 127 x 42mm) 3.75 lbs (1.7kg)

What are the safety specifications of the AEMC INSTRUMENTS J93?

Protection index: IP 40 for the probes and IP 30 jaws open, according to IEC 60 529

IK Code: IK 04 according to IEC 50102

Drop test: According to IEC 61010-1

Electrical safety: According to IEC 61010-2-032.

Maximum applicable voltage: 600V CAT III; 300V CAT IV

Specifications are subject to change without notice.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD AEMC INSTRUMENTS J93 (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.