How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

What components are supplied with the WATSON-MARLOW 323E pump?

The WATSON-MARLOW 323E pump is supplied with the following components:

A dedicated 323E pump drive unit fitted with one or more 313 or 314 pumpheads.

The designated mains power lead for your pump.

Compliance documentation.

Safety information documentation.

Note: Some versions of this product may include different components. Check against your purchase order.


What is the procedure for returning a WATSON-MARLOW 323E pump?

Equipment that has been contaminated with or exposed to body fluids, toxic chemicals, or any other substance hazardous to health must be decontaminated before it is returned. A decontamination certificate must be attached to the outside of the shipping carton, even if the WATSON-MARLOW 323E pump is unused. If the pump has been used, the fluids that have been in contact with the pump and the cleaning procedure must be specified along with a statement that the equipment has been decontaminated.


How does the peristaltic pump mechanism of the WATSON-MARLOW 323E work?

A compressible tube is squeezed between a roller and a track on an arc of a circle, creating a seal at the point of contact. As the roller advances along the tube, the seal also advances. After the roller has passed, the tube returns to its original shape, creating a partial vacuum which is filled by fluid drawn from the inlet port.

Before the roller reaches the end of the track, a second roller compresses the tube at the start of the track, isolating a packet of fluid between the compression points. As the first roller leaves the track, the second continues to advance, expelling the packet of fluid through the pump’s discharge port. At the same time, a new partial vacuum is created behind the second roller into which more fluid is drawn from the inlet port.

Backflow and siphoning do not occur, and the WATSON-MARLOW 323E pump effectively seals the tube when it is inactive. No valves are needed.


What are the key safety notes for operating the WATSON-MARLOW 323E pump?

The WATSON-MARLOW 323E pump and tubing should only be used by competent, suitably trained personnel who have read and understood the manual.

A symbol with an exclamation mark inside a triangle means: Caution, refer to accompanying documents.

A symbol of a hand with fingers near gears means: Do not allow fingers to contact moving parts.

A symbol of a crossed-out wheeled bin means: Recycle this product under the terms of the EU Waste Electrical and Electronic Equipment (WEEE) Directive.

Fundamental work regarding lifting, transportation, installation, starting-up, maintenance, and repair should be performed by qualified personnel only. The unit must be isolated from mains power while work is being carried out.

There is a user-replaceable type T1.0AH 250V fuse in the fuse drawer of the IEC mains connector at the back of the pump, which also contains a spare fuse. There are no other user-serviceable fuses or parts inside the pump.


What safety directions must be followed before opening the pumphead on the WATSON-MARLOW 323E?

Before opening the track of the pumphead on your WATSON-MARLOW 323E, ensure the following safety directions are followed:

Ensure that the pump is isolated from the mains power.

Ensure that there is no pressure in the pipeline.

If a tube failure has occurred, ensure that any fluid in the pumphead has been allowed to drain to a suitable vessel, container or drain.

Ensure that protective clothing and eye protection are worn if hazardous fluids are pumped.

Primary operator protection from rotating parts of the pump is provided by the pumphead track.

This product does not comply with the ATEX directive and must not be used in explosive atmospheres.


What are the general operating precautions for the WATSON-MARLOW 323E pump?

The WATSON-MARLOW 323E pump must be used only for its intended purpose and must be accessible at all times. The mains plug is the disconnecting device; do not position the pump so that it is difficult to disconnect. Do not fit any devices to the drive unit other than those tested and approved by Watson-Marlow.

If hazardous fluids are to be pumped, specific safety procedures must be in place.

The exterior surfaces may get hot during operation. Do not take hold of the pump while it is running and let it cool after use before handling.

No attempt should be made to run the drive without a pumphead fitted.


What are the features of the WATSON-MARLOW 323E pump?

The following table lists the features available on the WATSON-MARLOW 323E and comparable models.

Features 323E 323S 323U 323Du
Manual control
Backlight
Beeper
Manual screen: rpm
15-400 rpm 27:1 speed control
3-400 rpm 133:1 speed control
1.5-220 rpm 147:1 speed control
5 rpm scrolling increment
1 rpm scrolling increment
Auto-restart
Keypad lock
MemoDose
Analogue input: 4-20mA, 0-10V
Remote stop/start control
Remote direction control
RS232 control
313D and 314D pumpheads
501RL and 501RL2 pumpheads
314MC and 318MC pumpheads
~100-120V/~220-240V operation
IP31 wipe-down case

What are the pump specifications for the WATSON-MARLOW 323E?

Supply voltage/frequency 100-120V/200-240V 50/60Hz 1ph
Maximum voltage fluctuation ±10% of nominal voltage. A well regulated electrical mains supply is required along with cable connections conforming to the best practice of noise immunity
Installation category (overvoltage category) II
Power consumption 100VA
Full load current <0.43A at 230V; <0.86A at 115V
Eprom version Accessible through pump software
Enclosure rating IP31
Environment Indoor use only
Operating temperature range 4C to 40C, 40F to 104F
Storage temperature range -40C to 70C, -40F to 158F
Maximum altitude 2,000m, 6,560ft
Humidity (non-condensing) 80% up to 31C, 88F, decreasing linearly to 50% at 40C, 104F
Noise <70dB(A) at 1m

What standards does the WATSON-MARLOW 323E pump comply with?

EC harmonised standards Safety requirements for electrical equipment for measurement, control and laboratory use: BS EN 61010-1 incorporating A2 Category 2, Pollution degree 2
Conducted emissions: BS EN 55011, Class A
Radiated emissions: BS EN 55011, Class A
Electrostatic discharge: BS EN 61000-4-2
Radiated RF immunity: BS EN 61000-4-3
Fast transient burst: BS EN 61000-4-4
Surge immunity: BS EN 61000-4-5
Conducted RF immunity: BS EN 61000-4-6
Voltage dips and interruptions: BS EN 61000-4-11
UL 61010A-1
Other standards CAN/CSA-C22.2 No 61010-1
Conducted emissions FCC 47CFR, Part 15.107
Radiated emissions FCC 47CFR, Part 15

What are the best practices for installing the WATSON-MARLOW 323E pump?

Position: Site the WATSON-MARLOW 323E pump on a flat, horizontal, rigid surface, free from excessive vibration. Allow air flow around the pump for heat dissipation and ensure the ambient temperature does not exceed 40C.

Emergency disconnection: The mains plug is the disconnecting device. Do not position the pump where it is difficult to unplug. The STOP key will always stop the pump, but a local emergency stop device in the mains supply is recommended.

Valves: Peristaltic pumps are self-priming and self-sealing, so no valves are needed. Any valves in the process flow must be open before the pump operates. A pressure relief device is advised on the discharge side if a valve is present.

Tubing materials run-in advice: Sta-Pure and Marprene tubing are hard to compress when new. For the first 30 seconds, run the pump at 10 rpm or greater to avoid an over-current error message.


What are the do’s and do not’s for operating the WATSON-MARLOW 323E pump?

Do observe precautions for reducing electrostatic charge (e.g., ESD protective workwear).

Do operate the WATSON-MARLOW 323E on a flat horizontal surface and do not block air vents.

Do not stack pumps more than three high.

Do use only single phase mains electricity supplies.

Do keep delivery and suction tubes as short and direct as possible (ideally not shorter than 1m). Use bends of a large radius.

Do ensure on longer tube runs that at least 1m of smooth bore flexible tubing is connected to the inlet and discharge ports to minimize pulsation.

Do site the pump at or just below the fluid level for flooded suction.

Do keep the pumphead track and moving parts clean.

Do run at slow speed when pumping viscous fluids.

Do recalibrate after changing pump tubes, fluid, or connecting pipework.

When using Marprene or Bioprene continuous tubing, do re-tension the tube after the first 30 minutes of running.


How do I connect the WATSON-MARLOW 323E pump to a power supply?

A well-regulated electrical mains supply is required. It is not recommended to site the WATSON-MARLOW 323E drive alongside “dirty” electrical mains supplies without special attention to noise.

The voltage selector is at the rear of the pump. Set it to 115V for 100-120V 50/60Hz supplies or 230V for 200-240V 50/60Hz supplies. Always check the selector switch before connecting the mains supply. Connect to an earthed, single-phase mains electricity supply. The mains plug must be a separable plug (not a locking type).

We recommend using commercially available supply voltage surge suppression where there is excessive electrical noise.

The input line fuse is a type T1.0AH 250V 20mm time-delayed cartridge fuse, located in the combined mains IEC inlet socket and fuse drawer at the rear of the pump. A spare fuse is also provided.

Conductor Coding:

European North American
live brown black
neutral blue white
ground green/yellow green

What is the start-up checklist for the WATSON-MARLOW 323E pump?

Before starting your WATSON-MARLOW 323E, please perform the following checks:

Ensure proper connections are achieved between the pump tube and suction and discharge piping.

Ensure proper connection has been made to a suitable power supply.

Ensure that the recommendations in the good pump installation practice section are followed.

Check the position of the voltage selector switch.

Check the mains power switch at the rear of the pump.

Check the fuse in the mains inlet socket at the rear of the pump.

Ensure that the mains IEC plug is correctly fitted in the mains IEC inlet socket.


How do I switch on the WATSON-MARLOW 323E pump for the first time?

Switch on the power supply at the rear of the WATSON-MARLOW 323E pump. The pump will run a power-on test. If a fault is found, an error message is displayed. If the pump starts running, look for the ‘!’ symbol on the display, which indicates auto-restart is set. Press the STOP key if you need to stop the pump.

The pump is now ready to operate according to the defaults listed below. All operating parameters may be changed by means of key-presses.

First-time start-up defaults:

323E 323S 323U 323Du
Direction Clockwise Clockwise Clockwise Clockwise
Pumphead, 400 313 313 313 313
Pumphead, 220 501RL 501RL 501RL
Keypad lock Off Off Off
Auto-restart Off Off Off
Pump status Stopped Stopped Stopped Stopped
Remote stop Open=run Open=run

How do I perform manual operation of the WATSON-MARLOW 323E pump?

You can adjust the speed on the display while the pump is stopped or running.

Use the UP key to increase the set speed. Use the DOWN key to reduce the set speed. We recommend reducing the speed to a minimum before starting the pump.

The WATSON-MARLOW 323E increments in steps of 5rpm.

Press the DIRECTION key to reverse the direction of rotation. The direction is shown by the rotation symbol and may be changed while the pump is stopped or running.

Start the pump with the START key. The rotation symbol will move to confirm operation.

Stop the pump with the STOP key. The pump will stop immediately, and the display will show the previous speed and direction.

You can reduce the pump speed to 0 rpm with the DOWN key. The pump is still in the running state and the rotation symbol will continue to move. Press the UP key to return the pump to the minimum speed.


How do I perform care and maintenance on the WATSON-MARLOW 323E pump?

The WATSON-MARLOW 323E pump is sealed to IP31 and is suitable for wipe-down cleaning. Do not use solvents, mechanical scourers, strong organic acids, or alkali-based cleaning solutions.

Remove any tubing, detach the pumphead, and wash the pumphead thoroughly with a mild solution of detergent in water.

Check moving parts of the rotor from time to time for freedom of movement. Lubricate pivot points and rollers occasionally with Teflon lubricating oil.

The pump case has good chemical resistance but may be damaged by contact with strong acids or strong solvents.

There are no user-serviceable parts inside the pump. The unit should be returned for service.


How do I troubleshoot the WATSON-MARLOW 323E pump?

If the WATSON-MARLOW 323E display remains blank when the pump is on, make the following checks:

Check the position of the voltage selector switch at the rear of the pump.

Check the mains power switch at the rear of the pump.

Check that mains power is available to the pump.

Check the fuse in the fuse drawer of the IEC mains power connector at the rear of the pump.

Check the fuse in the mains power plug if one is present.

If the WATSON-MARLOW 323E pump runs but there is little or no flow, make the following checks:

Check that the tube and rotor are in the pumphead.

Check that fluid is supplied to the pump.

Check that the tube is not split or burst.

Check for any kinks or blockages in the lines.

Check that any valves in the lines are open.

Check that the correct wall-thickness tube is being used.

Check direction of rotation.

Check that the rotor is not slipping on the drive shaft.


What do the error messages on the WATSON-MARLOW 323E mean?

If a fault condition is detected on your WATSON-MARLOW 323E, the pump will stop and display an error number. All keys will be disabled.

Error Error condition Suggested action
0 RAM write error Attempt to reset by switching power OFF / ON. Or seek support
1 RAM corruption Attempt to reset by switching power OFF / ON. Or seek support
2 OTP ROM error / corruption Attempt to reset by switching power OFF / ON. Or seek support
3 OTP ROM read error Attempt to reset by switching power OFF / ON. Or seek support
5 Unknown pump type Check the interface card and cables. Attempt to reset by switching power OFF / ON. Or seek support
7 Display failure Seek support
8 Wrong key-press Attempt key-press again. Attempt to reset by switching OFF / ON
9 Motor stalled Stop pump immediately. Check pumphead and tube. Power OFF/ON may reset. Or seek support
10 Tacho fault Stop pump immediately. Power OFF/ON may reset. Or seek support
14 Over speed Stop pump immediately. Power OFF/ON may reset. Or seek support
15 Over current Stop pump immediately. Check system. Power OFF/ON may reset. Or seek support
16 Over voltage Stop pump immediately. Check mains voltage selector switch. Check supply. Power OFF/ON may reset. Or seek support
17 Under voltage Stop pump immediately. Check mains voltage selector switch. Check supply. ON/OFF may reset. Or seek support
18 Watchdog error Attempt to reset by switching power OFF / ON. Or seek support
19 Over temperature Stop pump immediately. Turn OFF. Seek support
20 Signal out of range Check analogue control signal range. Trim signal as required. Or seek support
21 Over signal Reduce the analogue control signal
22 No signal Connect analogue control signal or return to manual control
25 Network not detected Turn OFF. Check network and connections. Or seek support
26 RS232 fault Turn OFF. Check network and connections. Or seek support
27 RS232 lost Turn OFF. Check network and connections. Or seek support
33 Unrecognised keypress Attempt keypress again. Attempt to reset by switching power OFF/ON. Otherwise seek support.
35 Work overload Turn OFF. Check power supply. Check pumphead and tubing. Wait 30 minutes. Power ON may reset. Otherwise seek support.
ERR General error condition Turn OFF. Seek support

What are the safety procedures for the pumpheads on the WATSON-MARLOW 323E?

Before opening the pumphead track on your WATSON-MARLOW 323E, please ensure that the following safety directions are followed:

Ensure that the pump is isolated from mains voltage.

Ensure that there is no pressure in the pipeline.

If a tube failure has occurred, ensure that any product in the pumphead has been allowed to drain to a suitable drain.

Ensure that protective clothing and eye protection are worn if hazardous products are being pumped.


How do I install or remove a 313 or 314 pumphead on the WATSON-MARLOW 323E?

The WATSON-MARLOW 323E drives have an integral mounting plate to attach a 313 or 314 pumphead.

Installation:

Engage the pumphead drive slot with the end of the pump drive shaft. Continue to align the pumphead until the bayonet engages with the mounting plate. Turn the pumphead clockwise until it locks into an upright position.

Removal:

Push the locking lever back and turn the pumphead anti-clockwise until it is free from the mounting plate.


How do I load a tube into a 313 or 314 pumphead on the WATSON-MARLOW 323E?

1. Switch off the WATSON-MARLOW 323E pump before tube loading. Lift the “flip top” track until fully open.

2. Set the tube clamps to the correct tube size. The track must be fully open. Align the scale on both sides of the pumphead. If the tube is dirty or there is a high suction lift, the tube clamps may need a smaller setting.

3. Select enough tube length for the curve of the pump track. Slide the tube into the open pumphead. The tube must not be twisted or stretched against the rollers.

4. Ensure the tubing locates in the centre of the tube clamps. Carefully lower the track. Check that the tube is not crushed in the clamps or overstretched.

When using Marprene tubing: Re-tension new tube after the first 30 minutes of running. Stop the pump, release the flip top, and allow the tube to resettle naturally. Re-clamp the tube and restart the pump. This corrects normal stretching and is essential for good tube life.


What are the flow rates for the WATSON-MARLOW 323E with 313D and 314D pumpheads?

Flow rates were obtained using silicone tubing with the pumphead rotating clockwise, pumping water at 20°C with zero suction and delivery pressures. Note: 314D pumpheads should run no faster than 300 rpm in continuous use. Speeds up to 400 rpm are permissible for intermittent use.

Flow rates, 313D, 1.6mm wall (ml/min)

bore 0.5 0.8 1.6 3.2 4.8 6.4 8.0
mm 0.5 0.8 1.6 3.2 4.8 6.4 8.0
in 1/50 1/32 1/16 1/8 3/16 1/14 5/16
# 112 13 14 16 25 17 18
15-400 rpm 0.45-12 1.1-28 4.1-110 15-400 33-880 54-1400 75-2000

Flow rates, 314D, 1.6mm wall (ml/min)

bore 0.5 0.8 1.6 3.2 4.8 6.4 8.0
mm 0.5 0.8 1.6 3.2 4.8 6.4 8.0
in 1/50 1/32 1/16 1/8 3/16 1/14 5/16
# 112 13 14 16 25 17 18
15-400 rpm 0.45-12 0.90-24 3.8-100 13-340 29-760 45-1200 60-1600

Is the WATSON-MARLOW 323E pump designed for use in patient-connected applications?

Warning: The WATSON-MARLOW 323E pump is not designed for use in, and should not be used for, patient-connected applications.


CLICK HERE TO DOWNLOAD WATSON-MARLOW 323E (01) PDF DOCUMENT