What are the operating alert signals for the THERMO SCIENTIFIC FD1530MCN?

Warning alerts you to a possibility of personal injury.

Caution alerts you to a possibility of damage to the equipment.

Note alerts you to pertinent facts and conditions.

Hot surfaces alert you to a possibility of personal injury if you come in contact with a surface during use or for a period of time after use.


What must be done to avoid electrical shock with the THERMO SCIENTIFIC FD1530MCN furnace?

Use a properly grounded electrical outlet of correct voltage and current handling capacity.

Disconnect from the power supply prior to maintenance and servicing.

Have the door switch operating properly.


What must be done to avoid burns with the THERMO SCIENTIFIC FD1530MCN furnace?

Not to be touched on the exterior or interior surfaces during use or for a period of time after use.


What should be done to avoid personal injury with the THERMO SCIENTIFIC FD1530MCN furnace?

Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such materials.

Refer servicing to qualified personnel.


What are the recommended ambient temperatures for the THERMO SCIENTIFIC FD1530MCN?

The recommended ambient temperature is 17°C-27°C. Ambient above this level may result in damage to the controller.


What is the recommended space to allow between the THERMO SCIENTIFIC FD1530MCN furnace and any combustible surface?

Allow at least six inches of space between the furnace and any combustible surface. This permits the heat from the case surface to escape so as not to create a possible fire hazard.


What must be done to avoid electrical shock when installing the THERMO SCIENTIFIC FD1530MCN furnace?

This furnace must be installed by a competent electrician who ensures compatibility among furnace specification, power source and ground code requirements.


How do I turn on the THERMO SCIENTIFIC FD1530MCN furnace?

Both the ON/OFF power switch and the digital display will illuminate when power is switched ON. The furnace will begin to heat to the controller’s current setpoint.


When does the cycle light illuminate on the THERMO SCIENTIFIC FD1530MCN furnace?

The amber cycle light will illuminate whenever the power is being applied to the heating elements. The cycle light will flicker on/off as furnace reaches setpoint.


How can I check if the door safety switches are working properly on the THERMO SCIENTIFIC FD1530MCN furnace?

Open and close the door a few times; note that the amber CYCLE light will be out when door is open. This check must be done when furnace is heating or when cycle light is illuminated. If this condition is not true, consult the Troubleshooting section before proceeding.


What does the digital readout on the THERMO SCIENTIFIC FD1530MCN furnace continuously display?

The digital readout continuously displays chamber temperature.


How do I change the setpoint on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

If you want to change the setpoint, press the SCROLL button until “SP1” is displayed. Press the UP or DOWN button until the desired setpoint value is displayed and then release the button. A few seconds after the button is released, the controller will accept the new value and is indicated by a brief flash of the display. Press PAGE button to return to HOME DISPLAY.


How do I view the display units on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

Press SCROLL until “UNITS” is displayed. The temperature units can be changed by pressing up and down buttons. Choice of Celsius (°C), Fahrenheit (°F), Kelvin (°K), Percentage (%), or None (nonE).


What are the controller parameters listed under the Home display for the Single Setpoint Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

°C, °F, °K, %, or None: Temperature units in Celsius (default), Fahrenheit, Kelvin, Percentage (PErc), or None (nonE).

A1.DHI: Deviation high alarm.

A2.HI: High Limit alarm. Read Only.

A3.LO: Low Limit alarm. Read Only.

A.TUNE (tune): One-shot autotune enable.

WRK.OP: Working Output power. Read Only.

PV.OFS: Process Value Offset. Read Only.

SP.RAT: Ramp Rate Setpoint (default units is minutes).

DWELL: Time for dwell or delay (default units is minutes).

T.STAT: Timer Status. Active only when timer is active.

TM.CFG: Timer configuration.


What are the Pid List parameters for the Single Setpoint Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

PB: Proportional band (in display units).

TI: Integral time in seconds.

TD: Derivative time in seconds.


What are the alarms listed for the Single Setpoint Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

A2.HI: Measured value full scale high alarm.

A1.DHI: Measured value deviation high alarm.

S.br: Sensor break: check that sensor is connected correctly.

LBR: Loop break: check that the heating circuits are working properly.

Ld.F: Heater Circuit fault: indication of either an open or short solid state relay, a blown fuse, missing supply or open circuit heater.


What happens when Sensor Break Protection is activated on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

If an open thermocouple condition occurs, the digital display will blink “S.br” and the power to the heating element will be shut OFF (Cycle light will extinguish).


How is Over-Temperature Protection (OTP) set for the Single Setpoint Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

The OTP will be in effect during any alarm condition when the temperature of the furnace has deviated beyond the limit. The “Deviation High” alarm is the only alarm value, which can be changed. To change it, press the SCROLL button until “A1.DHi” appears on the display. Press the UP or DOWN button to select the OTP value you desire. We recommend a value of 20° above your working temperature to provide protection for your workload.


How do I start tuning on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

1. Start tuning with the process at ambient temperature. This allows the tuner to calculate the low cutback and high cutback values more accurately.

2. Adjust the setpoint to your desired value.

3. Press the SCROLL button until display reads, “A.TUNE.”

4. Press the UP or DOWN button to select, “on.”

5. Press the PAGE button to return to the HOME DISPLAY. The display will alternately flash between “tunE” and the HOME DISPLAY while tuning is in progress.

6. The controller will then turn the heating on and off to induce an oscillation. When the measured value reaches the required setpoint the first cycle will end.

7. Tuning will be complete after two oscillation cycles and then the tuner will turn itself off.

8. Normal control function will resume after the controller calculates tuning parameters.


What is the purpose of the Single Ramp & Dwell Functions on the THERMO SCIENTIFIC FD1530MCN?

This type of controller has single ramp and dwell programming capabilities. The Ramp and Dwell can be configured to five different modes.

Mode 1 (DWELL): The dwell time begins once the setpoint reaches the set threshold. The END TYPE action is executed when the dwell timer reaches the end.

Mode 2 (DELY): The timer starts immediately upon instrument power-up, or when run is selected. The instrument remains in standby until the time has elapsed. After the time has elapsed, the instrument controls up to setpoint.

Mode 3 (SF.ST): Starts automatically on power up. This is a soft-start function. If the PV is below the Soft Start Threshold, then the power is limited to the Soft Start Limit until the threshold is met.


What is the Program Overview for the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

A program mode can be set by changing the “TM.CFG” variable (in the ‘Timer’ menu) to “DWEL, DELY, or SF.ST. Note: value of “None” deactivates the timer.

A Ramp rate may be set by changing the “SP.RAT” variable (in the ‘SP’ menu) to a value. The Ramp rate units are set with the “RAMPU❞ variable (in the ‘SP’ menu). The selections are Hour / Min / Sec.

The Dwell time can be set by changing the “DWELL” variable (in the ‘Timer’ menu) to the desired value. Dwell time units are set with the “TM.RES” variable (in the ‘Timer’ menu). The selections are Hour / Min.

The program Status can be set by changing the “T.STAT” variable to “run”, “hold”, or “res.” This variable will start, hold, or stop the program.


How do I set the SetPoint Rate Limit for the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

1. Press the SCROLL button until the “SP.RAT” (Ramp Rate) is displayed.

2. Set the desired Ramp rate with the UP or DOWN buttons, if the ramp to setpoint feature is needed. If the Ramp rate is not needed, then set to “OFF” with the UP or DOWN buttons.

3. Press the SCROLL button until “TM.CFG” (Ramp & Dwell mode) will be displayed, select the desired mode with the UP or DOWN buttons. (DWEL, DELY, or SF.ST).

4. Press the SCROLL button until “DWELL” will be displayed set the desired Dwell time with the UP or DOWN buttons.

5. Press the PAGE button and SCROLL button together until the Actual temperature is displayed.


How do I run the program on the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

1. Press the SCROLL button until “T.STAT” is displayed, set to “run” with the UP or DOWN buttons; or from the HOME DISPLAY, press UP and DOWN arrows together.

2. Press the PAGE button to display Actual temperature.


How do I hold the program on the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

1. Press the SCROLL button until “T.STAT” is displayed. Set to “hold” with the UP or DOWN buttons; or from the HOME DISPLAY, press UP and DOWN arrows together.

2. Press the PAGE button to display Actual temperature.


How do I stop the program on the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

Press the SCROLL button until “T.STAT” is displayed, Set to “res” with the UP or DOWN buttons.


How do I clear the flashing end on the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

Press the PAGE and SCROLL buttons at the same time.


How do I verify a running program on the SINGLE RAMP & DWELL CONTROLLER on the THERMO SCIENTIFIC FD1530MCN?

Press the SCROLL button until “T.STAT” is displayed. The display will show “run” if the program is running, “hold” if it is paused or “res if it is not running. Press the PAGE button to display actual temperature.


How do I change the setpoint on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

If you want to change the setpoint, press the UP or DOWN button until the desired setpoint value is displayed in the lower display and then release the button.


How do I view the display units on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

From the HOME DISPLAY press the SCROLL button. The display will briefly show the temperature units in °C/F/K and then return to the HOME DISPLAY. (If you require a different temperature unit call Customer Service)


How do I view the % Output Power on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

From the HOME DISPLAY press the SCROLL button twice. Press the UP or DOWN button to display the value. This value is a read-only value and cannot be changed.


What are the Button and Indicator functions for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

OP1 (Output 1): illuminates when the heating output of the temperature controller is on.

AUTO/MAN: (Auto/Manual Mode): when the controller is in the automatic mode the output automatically adjusts to keep the temperature or process value at the setpoint. The “AUTO” light will illuminate. The manual mode has been disabled through factory configuration. Call Customer Service for further information.

RUN/HOLD (Run/Hold button):

  • Starts a program when pressed once-RUN light illuminates.
  • Holds a program when pressed again—HOLD light illuminates.
  • Cancels hold and continues running when pressed again-HOLD light is off and RUN light illuminates.
  • Exits a program when the button is held down for two seconds-RUN and HOLD lights are off.
  • At the end of a program the RUN light will flash.
  • During holdback the HOLD light will flash.

PAGE button: allows you to choose a parameter from a list of parameters.

SCROLL button: allows you to choose a parameter within a list of parameters.

UP button: allows you to increase the value in the lower display.

DOWN button: allows you to decrease the value in the lower display.


What are the Controller Parameters listed under Home Display for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

°C: measured temperature in Celsius. Temperature units cannot be changed without entering the configuration. Contact Customer Service if a different temperature unit is required.

OP: % output power demand; displayed in lower display (cannot be changed).

C.id: Controller identification number.

IdHi: Deviation High Alarm.

tunE: One-shot autotune enable.


What are the Controller Parameters listed under run List (Program Run List) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

Stat: Displays the program status [OFF, run (running active program), hoLd (program on hold), HbAc (waiting for process to catch up), End (program completed)] in the lower display. The controller will default to “OFF.”

FAST: Fast run through program (no/YES). The controller will default to “no.”

SEG.d: Flash active segment type in the lower display of the home display (no/YES). The controller will default to “no.”


What are the Controller Parameters listed under ProG LiSt (Program Edit List) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

Hb: Press the UP or DOWN ARROW to select the holdback type [OFF (disables holdback), Lo (deviation low holdback), Hi (deviation high holdback) or bAnd (deviation band holdback)] for the entire program. The controller will default to “OFF”

Hb.U: Press the UP or DOWN ARROW to select the holdback value (in display units).

rmP.U: Press the UP or DOWN ARROW to toggle between ramp units (SEc, min or Hour). Controller will default to “SEC.”

dwL.U: Press the UP or DOWN ARROW to toggle between dwell units (SEc, min or Hour). Controller will default to “SEC.”

Cyc.n: Press the UP or DOWN ARROW to set the number of program cycles (1 to 999 or cont). The controller will default to “cont.”

SEG.n: Press the UP or DOWN ARROW to select the segment number (1-8 in 8 segment models).

tYPE: Press the UP or DOWN ARROW to select the segment type [End (end of program), rmP.r = ramp rate (ramp to a specified setpoint at a set rate), rmp.t = ramp time (ramp to a specified temperature in a set time), dwEll (to maintain a constant temperature for a set time), StEP (climb instantaneously from current to specified tempera- ture). The controller will default to “End.” Other parameters used with tYPE include; tGt target setpoint), Rate (rate of temperature increase) and dur (time to target setpoint or time to dwell)].

End.t: End segment type: dwELL (dwell continuous), rSEt (reset) and S OP (End Segment Output power level.


What are the Controller Parameters listed under AL LiSt (Alarm List) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

IdHi: Deviation High Alarm.


What are the Controller Parameters listed under Atun LiSt (Autotune List) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

tunE: One-shot autotune enable.

drA: Adaptive tune enable.

drA.t: Adaptive tune trigger level in display units. Range = 1-9999.


What are the Pid LiSt parameters for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

G.SP (Gain Setpoint): Is the temperature at which the controller switches from the (SEt 1) PID values to the (SEt 2) PID values.

Pb: Proportional band in display units. (SEt 1)

ti: Integral time in seconds. (SEt 1)

td: Derivative time in seconds. (SEt 1)

Pb2: Proportional band. (SEt 2)

ti2: Integral time in seconds. (SEt 2)

td2: Derivative time in seconds. (SEt 2)


What are the Alarms listed for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

IdHi: PV deviation high alarm.

2FSH: PV full scale high alarm.

3FSL: PV full scale low alarm.

LCr: Load current low alarm.

HCr: Load current high alarm.

S.br: Sensor break: check that sensor is connected correctly.

L.br: Loop Break: Check that the heating circuits are working properly.

Ld.F: Heater Circuit Fault: indication of either an open or short solid state relay, a blown fuse, missing supply or open circuit heater.

SSr.F: Solid state relay failure indications in a solid state relay: indicates either an open or short circuit in the SSR.

Htr.F: Heater failure: Indication that there is a fault in the heating circuit: indicates a blown fuse, missing sup- ply or open circuit heater.


What happens when Sensor Break Protection is activated on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

If an open thermocouple condition occurs, the digital display will Blink “S.br” and the power to the heating element will be shut OFF (Cycle light will extinguish).


How is Over-Temperature Protection (OTP) set for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

The OTP will be in effect during any alarm condition when the temperature of the furnace has deviated beyond the limit. The “Deviation High” alarm is the only alarm value which can be changed. To change it, press the SCROLL button until “idHi” appears on the display. Press the UP or DOWN button to select the OTP value you desire. We recommend a value of 20° above your working temperature to provide protection for your workload.


How do I operate the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN as a Single Setpoint Controller?

1. Switch the circuit breaker to the “ON” position. The setpoint temperature presently set in the controller will appear in the lower display. (The upper display indicates the actual chamber temperature.)

2. To change the setpoint, press the UP or DOWN button until the desired setpoint value is displayed; then release the button.

3. The furnace will begin to heat if the new setpoint temperature is higher than the present chamber temperature.


How do I set the holdback type for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “Hb.”

3. Press the UP or DOWN button to toggle between “bAnd, Hi, Lo and OFF.”


How do I set the holdback value (Hb.U) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “Hb.U.”

3. Press the UP or DOWN button to enter a holdback value.


How do I set the Ramp Units (rmP.U) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “rmP.U.”

3. Press the UP or DOWN button to toggle between seconds, minutes and hours.


How do I set the Dwell Units (dwL.U) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “dwL.U.”

3. Press the UP or DOWN button to toggle between seconds, minutes and hours.


How do I set the Number of Cycles (CYC.n) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “CYC.n.”

3. Press the UP or DOWN button to select the number of cycles you want to run or, press the DOWN button to select “cont.” so the program will run continuously.


How do I set the segment type to rmP.r (Ramp) for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

To ramp linearly at a set rate to a specified temperature:

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “tYPE.”

3. Press the UP or DOWN button until display reads, “rmP.r.”

4. Press the SCROLL button until display reads, “tGt.”

5. Press the UP or DOWN button to set a target setpoint.

6. Press the SCROLL button until display reads, “rAtE.”

7. Press the UP or DOWN button to select a value in ramp units (seconds, minutes or hours; set in the “rmP.U” parameter).


How do I set the segment type to rmP.t for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

To ramp to a specified temperature at a set time:

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “tYPE.”

3. Press the UP or DOWN button until display reads, “rmP.t.”

4. Press the SCROLL button until display reads, “tGt.”

5. Press the UP or DOWN button to set a target setpoint.

6. Press the SCROLL button until display reads, “dur.”

7. Press the UP or DOWN button to select a time in ramp units (seconds, minutes or hours; set in the “rmP.U” parameter).


How do I set the segment type to dwEll for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

To maintain a constant temperature for a specified time:

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “tYPE.”

3. Press the UP or DOWN button until display reads, “dwell.”

4. Press the SCROLL button until display reads, “dur.”

5. Press the UP or DOWN button to select a time in dwell units (seconds, minutes or hours; set in the “dwL.U” parameter).


How do I set the segment type to StEP for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

To climb instantaneously from the current temperature to a specified temperature.

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “tYPE.”

3. Press the UP or DOWN button until the display reads, “StEP.”

4. Press the SCROLL button until display reads, “tGt.”

5. Press the UP or DOWN button to set a target setpoint.


How do I set the End segment type for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

To end or repeat a program:

1. Press the PAGE button until you reach the program list (ProG LiSt).

2. Press the SCROLL button until display reads, “tYPE.”

3. Press the UP or DOWN button until display reads, “End.”

4. Press the SCROLL button until display reads, “End.t.”

5. Press the UP or DOWN button to toggle between “dwEll” (an indefinite dwell at last target level), “S OP” (End Segment Output power level) and “rSET” (reset will revert back to setpoint in Single Setpoint mode).


How do I run a program on the 8 Segment Programmable Models for the THERMO SCIENTIFIC FD1530MCN?

To run a program, press the RUN/HOLD button. (The RUN light will illuminate).


How do I put a running program on hold on the 8 Segment Programmable Models for the THERMO SCIENTIFIC FD1530MCN?

To put a running program on hold, press the RUN/HOLD button. (The HOLD light will illuminate).


How do I cancel a program on the 8 Segment Programmable Models for the THERMO SCIENTIFIC FD1530MCN?

To cancel a program, hold the RUN/HOLD button down until the RUN and HOLD lights go off.


What is the purpose of tuning for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

The purpose of tuning your furnace is to match the characteristics of your controller to the characteristics of the process being controlled. Good control is evidenced by: stable, straight-line control of the setpoint temperature with no fluctuations; No overshoot or undershoot of the setpoint temperature; rapid restoration of the setpoint temperature when external disturbances cause deviations from the setpoint.

This controller has automatic tuning features which install optimum tuning parameters to give the best temperature accuracy. No manual loading of tuning parameters is needed. We recommend that you tune the furnace to your specific application to obtain the best results. To provide the best temperature accuracy possible, use these features when you install your furnace and whenever you change your application or procedure.


What happens if a Tuning Error occurs on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

The display will flash “tu.ER” if an error occurs during tuning. To clear the error and restart tuning, simultaneously press the PAGE and SCROLL buttons and follow the steps outlined in “Autotuning.”


How do I stop the tuning function on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

To stop the tuning function, simultaneously press the PAGE and SCROLL buttons.


What is G.SP: Gain Scheduling for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

Gain scheduling is the automatic transfer of control between two sets of PID values. The controller does this at a presettable process value. Gain scheduling is used for difficult control processes which show large changes in their response time or sensitivity at high or low temperatures, or when heating or cooling.

The G.SP gain schedule setpoint is factory set at 700° С. The G.SP must be adjusted to 200°C from the desired setpoint temperature when tuning.


How is the Transfer Point set if gain scheduling has been enabled for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

If gain scheduling has been enabled, “G.SP will appear at the top of the PID list. This sets the value at which the transfer will occur. When the process value is below this level, PID1 will be active and when it is above, Pid2 will be active. Set a value between the control regions that show the greatest change to achieve the best point of transfer.


How do I tune the furnace using autotuning on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

1. Load your furnace with a load similar to your normal load and close the door.

2. Set the setpoint temperature.

3. Press the PAGE button until the display reads, “Atun LiSt.”

4. Press the SCROLL button until “tunE OFF” is displayed.

5. Press the UP or DOWN button to select “on.”

6. Simultaneously press the PAGE and SCROLL buttons to return to the HOME DISPLAY. The display will flash “tunE” while tuning is in progress.


When should Adaptive Tuning be used for the 8 Segment Programmable Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

Adaptive tuning continuously evaluates tuning parameters. Adaptive tuning automatically installs new values if better accuracy is possible. Adaptive tuning should be used when the characteristics of a process change due to load or setpoint changes or, in a process that cannot handle the oscillation caused by a one-shot tune.


How do I tune the furnace using adaptive tuning on the 8 Segment Programmable Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

1. Load your furnace with a load characteristic of those you intend to heat in it.

2. Press the PAGE button until display reads, “Atun LiSt.”

3. Press the SCROLL button until “drA OFF” is displayed.

4. Press the UP or DOWN button to select “on.”

5. Press the SCROLL button until “drA.t” is displayed.

6. Press the UP or DOWN button until the desired trigger value is achieved.


What is the caution regarding overloading the furnace chamber for the THERMO SCIENTIFIC FD1530MCN?

Do not overload your furnace chamber. If the load is to be heated uniformly it should not occupy more than the center two-thirds of the furnace chamber. Failure to observe this caution could result in damage to furnace components.


What are the best practices for Furnace Loading for the THERMO SCIENTIFIC FD1530MCN?

1. For best results, use only the center 2/3 of the furnace chamber.

2. If you are heating a number of small parts, spread them throughout the center two-thirds of the furnace chamber.

3. Keep objects away from thermocouple.

4. Use insulated tongs and mittens when loading and unloading furnace.

5. Always wear safety glasses.

6. Use the hearth plates supplied to protect bottom of chamber. Part # PHX2.


What is the warning regarding cleaning or decontamination methods for the THERMO SCIENTIFIC FD1530MCN furnace?

Before using any cleaning or decontamination method except those recommended by Thermo Scientific, users should check with Thermo Scientific that the proposed method will not damage the equipment.


What precautions should be taken before cleaning the THERMO SCIENTIFIC FD1530MCN furnace?

Disconnect the furnace from power supply before cleaning.


What is the warning about opening the door for an extended period on the THERMO SCIENTIFIC FD1530MCN furnace?

Opening the door for an extended period of time will cause the painted surfaces above the door to be discolored or burnt.


What are the possible causes and corrective actions if the cycle light does not illuminate on the THERMO SCIENTIFIC FD1530MCN furnace?

POSSIBLE CAUSES CORRECTIVE ACTION
The furnace is not connected to power supply. Reconnect furnace to power supply.
Incorrect power source. Connect to correct power source.
ON and OFF power switch defective. Replace power switch.
Defective cycle light. Replace cycle light.
Door switch malfunction. Realign or replace furnace door safety switch.

What are the possible causes and corrective actions if the THERMO SCIENTIFIC FD1530MCN furnace does not heat?

POSSIBLE CAUSES CORRECTIVE ACTION
No power. Check power source and fuses or breakers.
Defective electrical hookup. Repair electrical hookup.
Thermocouple has oxidized and opened the circuit. Replace thermocouple.
Controller malfunction. Replace controller.
Heating elements burned out. Replace defective elements.
Solid state relay defective. Replace solid state relay.
Door switch malfunction. Re-align or replace door switch.
Defective OTP relay. Replace relay.
Defective solid state relay. Replace relay.

What are the possible causes and corrective actions if the door switch on the THERMO SCIENTIFIC FD1530MCN does not cut power to heating elements?

POSSIBLE CAUSES CORRECTIVE ACTION
Door switch not functioning Realign or replace door safety switch.

What are the possible causes and corrective actions if the controller over-temp. does not cut power to the furnace chamber on the THERMO SCIENTIFIC FD1530MCN?

POSSIBLE CAUSES CORRECTIVE ACTION
Alarm output device malfunction. Replace controller.
OTP relay malfunction. Replace relay.
Element shorted to ground. Replace faulty element.

What are the possible causes and corrective actions if the THERMO SCIENTIFIC FD1530MCN furnace has slow heatup?

POSSIBLE CAUSES CORRECTIVE ACTION
Low line voltage. Install line of sufficient size and proper voltage. (Isolate furnace from other electrical loads.)
Wrong heating element. Install proper element.
Heating element burned out. Replace element.
Heavy load in chamber. Lighten load in chamber to allow heat to circulate.

What are the possible causes and corrective actions for repeated element burnout on the THERMO SCIENTIFIC FD1530MCN furnace?

POSSIBLE CAUSES CORRECTIVE ACTION
Overheating furnace. Keep furnace under maximum temperature. Closer supervision of control setting.
Heating harmful materials. Enclose material in container. Clean up spills in and on chamber. Ventilate chamber by leaving top vent slightly open when heating known harmful reagents.
Contamination present from previous burnout. Replace insulation material.

What are the possible causes and corrective actions for inaccurate temperature readout on the THERMO SCIENTIFIC FD1530MCN furnace?

POSSIBLE CAUSES CORRECTIVE ACTION
Oxidized or contaminated thermocouple. Replace thermocouple.
Poor thermocouple connection. Tighten connections.
Improper loading procedures. Use proper loading procedures.
Poor ventilation of control. Clear area around furnace control.
Thermocouple connections reversed. Reconnect thermocouple correctly.

How do I replace a Heating Element on the THERMO SCIENTIFIC FD1530MCN furnace?

1) Disconnect furnace from power supply.

2) Remove back terminal cover.

3) Loosen the nuts on the terminals of the element to be replaced.

4) Straighten the leads of the old element.

5) Open the door and pull the defective element out. (It may be easiest to turn the furnace so that the element to be removed is on top.)

6) Slide the new element into place, threading the leads through the insulating porcelain bushing in the back of the furnace.

7) Reverse the disassembly procedure, making sure you thread all element lead wires through the insulating porcelain bushings on the back of the furnace and cut off any excess element lead wire after securing the leads to the terminal points.

8) Replace the back terminal cover.

9) Reconnect furnace to power supply.

10) Test the operation of the furnace.


How do I replace the Platinel II Thermocouple on the THERMO SCIENTIFIC FD1530MCN furnace?

1) Disconnect the furnace from power supply.

2) Remove the back terminal cover. (Note placement and connection of wires.)

3) Remove the screw and washer from clip holding the thermocouple in place and remove the nuts and washers on the thermocouple terminals.

4) Remove the thermocouple.

5) Insert the new thermocouple straight through hole in the chamber until the tip extends approximately 1″ into the chamber. Make sure the thermocouple lead wire with the colored bead is connected to the terminal marked positive.

6) Secure the thermocouple extension wires to the terminals (yellow to +).

7) Secure the thermocouple with clip, screw and washer.

8) Replace the back cover.

9) Reconnect furnace to power supply.

10) Test the operation of the furnace.


How do I replace the Door Switch (Micro-Switch) on the THERMO SCIENTIFIC FD1530MCN furnace?

1) Disconnect the furnace from power supply.

2) Remove the screws from the front dial and the three bottom screws and lock washers on the back cover of the control section.

3) Remove the back terminal cover.

4) Disconnect the two wires on terminals T1 and T2 going to the control section.

5) Slide the control section forward.

6) Remove two screws and nuts from the door switch and slide the switch from the wire rod.

7) Disconnect the wires from the door switch. Identify or mark the wires disconnected to insure proper placement and connection when reinstalling.

8) Reconnect the wires identified or marked in step 7 to the new door switch.

9) Install the new door switch while sliding it over the wire rod and secure with the screws and nuts removed in step 6.

10) Slide the control section back and replace the screws and lock washers described in step 2.

11) Connect wire marked T1 to terminal T1 and wire marked T2 to terminal T2 with nuts removed in step 4.

12) Replace the back terminal cover with screws and lock washers removed in step 3.

13) Test operation of door switch. (See next section “To Realign Door Switches” step 5.)

14) Reconnect to power supply.


How do I realign the Door Switches on the THERMO SCIENTIFIC FD1530MCN furnace?

1) Disconnect the furnace from power supply.

2) Loosen the setscrew on the cam. Open the door approximately 1″ and rotate the cam counterclockwise until you hear a click then rotate the cam clock- wise until you hear a click, tighten the set screw on adjusting cam.

3) Open and close the door; the switch should click when the door is opened approximately 2″ and 1″ to 2″ before the door is closed.

4) Reconnect to power supply.

5) To test the operation of the door switch: turn the power switch on, set the control to a setting high enough to keep the control from cycling, open and close the door; the cycle light should turn OFF when the door is opened approximately 2” and turn back ON 1″ to 2″ before the door is closed.


How do I replace the Controller on the THERMO SCIENTIFIC FD1530MCN furnace?

The controller plugs into a panel mounting sleeve, which should be left permanently installed in the furnace housing. To remove the controller, release the side clips and slide the controller out. Do not attempt to dismantle this unit further; replace it with a Thermolyne loaner or a new unit.


What is the specification for the THERMO SCIENTIFIC FD1530MCN furnace regarding the operating temperature range?

The operating temperature range for continuous use is 100°C-1066°C and for intermittent use is 1066°C-1200°C.


What are the electrical ratings for the THERMO SCIENTIFIC FD1530MCN furnace?

The electrical specifications for the THERMO SCIENTIFIC FD1530MCN are:

RATING VALUE
VOLTS 220-240
AMPS 6.5
WATTS 1560
FREQ. 50/60
PHASE 1

What are the dimensions of the Chamber and Overall for the THERMO SCIENTIFIC FD1530MCN furnace?

Chamber Dimensions (Width x Height x Depth): 4 in (10 CM) x 3-3/4 in (9.5 CM) x 9 in (22.9 CM).

Overall Dimensions (Width x Height x Depth): 11 in (28 CM) x 16-1/2 in (42 CM) x 18 in (46 CM).


What is the weight of the THERMO SCIENTIFIC FD1530MCN furnace?

The weight is 42 LBS (19 KG).


What are the environmental conditions for operating the THERMO SCIENTIFIC FD1530MCN furnace?

Operating: 17°C – 27°C; 20% to 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit: 2,000 meters.

Storage: -25°C to 65°C; 20% to 80% relative humidity.


What are the electrical standards the THERMO SCIENTIFIC FD1530MCN (-33 model only) conforms to for EMC and Safety?

EMC: EN 61326-1 Electrical equipment for measurement, control, and laboratory use; Part I: General Requirements

Safety:

  • ΕΝ 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use; Part I: General Requirements
  • ΕΝ 61010-2-010 Part II: Particular requirements for laboratory equipment for the heating of materials

These are per the provisions of the Electromagnetic Compatibility Directive 2006/95/EC, and per the provisions of the Low Voltage Directive 2004/108/EC.


What should be done after unpacking the THERMO SCIENTIFIC FD1530MCN furnace?

Attach the door handle and remove packing material from inside furnace chamber. The furnace is supplied with one hearth plate. The 120V models are not supplied with a power cord because current requirements are too great to be handled by ordinary power cords and standard wall supply.


What are the safety requirements when loading or unloading the THERMO SCIENTIFIC FD1530MCN furnace?

Always wear safety glasses or a safety shield and high temperature gloves when loading or unloading the furnace. Long sleeved, fire retardant clothing and a fire retardant apron is also recommended.


What should be done before connecting the THERMO SCIENTIFIC FD1530MCN furnace to the electrical supply?

Be sure the two position power switch is in the OFF position.


What should be done if the electrical service for the THERMO SCIENTIFIC FD1530MCN furnace is different than those listed on the specification plate?

Consult Thermo Scientific if your electrical service is different than those listed on the specification plate.


What wire gauge is required for 120V supply connections to the THERMO SCIENTIFIC FD1530MCN furnace?

For 120V supply connections, use 14 AWG or larger wires suitable for at least 105°C (221°F).


What note is provided regarding the HOME DISPLAY return function for the Eurotherm 3216 Controller Operation on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

If at any time you want to return to the HOME DISPLAY, press PAGE button.


What is the factory default setting for the temperature units in the Home display for the Single Setpoint Model w/OTP on the THERMO SCIENTIFIC FD1530MCN?

Temperature units in the Home display are Celsius (°C) by default.


What should be done before starting a tune on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

Furnace must be at ambient temperature before starting a tune. “Stat” and “Sp.rr” must be set to OFF or “tunE” will not initiate.


What indicates that the tune has completed on the Single Setpoint Model w/OTP for the THERMO SCIENTIFIC FD1530MCN?

Tune has completed when “tunE” stops flashing on display.


What is the note regarding the program status for the Single Ramp & Dwell Controller on the THERMO SCIENTIFIC FD1530MCN?

The program must be stopped and the controller must be displaying the actual temperature before beginning the Setup.