What are the general specifications for the THERMO SCIENTIFIC FD1530M-33?

The THERMO SCIENTIFIC FD1530M-33 has the following general specifications:

MODEL NUMBER FD1540M-33 FD1530M-33
CHAMBER WIDTH 4 (10 CM)
CHAMBER HEIGHT 3-3/4 (9.5 CM)
CHAMBER DEPTH 9 (22.9 CM)
OVERALL WIDTH 11 (28 CM)
OVERALL HEIGHT 16-1/2 (42 CM)
OVERALL DEPTH 18 (46 CM)
WEIGHT LBS. (KG) 42 LBS (19 KG)
ELECTRICAL RATINGS VOLTS 220-240
ELECTRICAL RATINGS AMPS 6.5
ELECTRICAL RATINGS WATTS 1560
ELECTRICAL RATINGS FREQ. 50/60
ELECTRICAL RATINGS PHASE 1
OPERATING TEMP RANGE CONT. 100°C-1066°C
OPERATING TEMP RANGE INTERMITTENT 1066°C-1200°C

What are the environmental conditions for operating the THERMO SCIENTIFIC FD1530M-33?

The operating environmental conditions are: 17°C – 27°C; 20% to 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit: 2,000 meters. Storage conditions are: -25°C to 65°C; 20% to 80% relative humidity.


What should be done when unpacking the THERMO SCIENTIFIC FD1530M-33?

Visually check for any physical damage to the shipping container. Inspect the equipment surfaces that are adjacent to any damaged area. Unpack furnace from box. The owner’s manual and door handle are included in the box. After unpacking the furnace, attach the door handle and remove packing material from inside furnace chamber. The furnace is supplied with one hearth plate. The 120V models are not supplied with a power cord because current requirements are too great to be handled by ordinary power cords and standard wall supply.


What precautions should be taken regarding ambient temperature during installation of the THERMO SCIENTIFIC FD1530M-33?

Be sure ambient temperature does not exceed 104°F (40°C). The recommended ambient temperature is 17°C-27°C. Ambient above this level may result in damage to the controller.


What clearance space is required around the THERMO SCIENTIFIC FD1530M-33 during installation?

Allow at least six inches of space between the furnace and any combustible surface. This permits the heat from the case surface to escape so as not to create a possible fire hazard.


How should the power switch on the THERMO SCIENTIFIC FD1530M-33 be set before connecting to electrical supply?

Prior to connecting your Type 1500 furnace to your electrical supply, be sure the two position power switch is in the OFF position.


How can the door safety switches on the THERMO SCIENTIFIC FD1530M-33 be checked?

Open and close the door a few times; note that the amber CYCLE light will be out when door is open. This check must be done when furnace is heating or when cycle light is illuminated. If this condition is not true, consult the Troubleshooting section before proceeding.