What does the Ethernet Interface Kit include?
The kit includes:
- Ethernet PC Board
- Screws (4)
- Instruction Manual
- Cable
Where and how do I install the Ethernet Interface?
Install the Ethernet Interface on the underside of the balance. Follow these instructions:
- Remove the red interface options cover.
- Install the Interface PCB. Push it through the pins and secure it with the 4 screws.
- Connect an Ethernet Cable.
- Place the cord in the groove (see step 2) and put the red interface option cover back on.
Upon installation, will the scale recognize the Ethernet Interface Kit?
Yes, the scale will recognize the Ethernet Interface Kit upon installation.
What is the preset IP Address of the interface?
The interface has a preset IP Address which can be displayed by choosing Ip Adr in the menu. Use the “No” and “Back” keys to show the entire number on the display. The Port Number is 9761.
How will the scale operate with the interface installed?
With the interface installed, the scale will operate in one of the following two ways:
- Auto Print = off, Stable = on or off: Pressing PRINT will send the display data. If Stable is set to On, the balance will wait for a stable reading before sending data.
- Auto Print = not off, Stable = on or off: The scale will automatically send data based on the settings in the Menu. If Stable is set to On, only stable will be sent.
What format is the data sent in?
Data sent is in standard ASCII format terminated with a [CRLF] (carriage return-line feed).
What is the breakdown of the data sent?
The data sent is broken down as follows:
- [weight] 9 characters
- [space] 1 character
- [unit] 5 characters max (left justified)
- [space] 1 character
- [stability indicator] 1 character “?” when unstable, blank when stable
- [G/N] 1 character
- [LF] 1 character
What are the input commands that the scale will respond to via the interface?
The scale will respond to the following commands sent via the interface:
- IP: Immediate Print of displayed weight (stable or unstable).
- P: Print displayed weight (stable or unstable).
- CP: Continuous Print.
- SP: Print on Stability.
- xS: 0S: Turn off “Stable Only” menu item and allow unstable print. 1S: Turn on “Stable Only” menu item and only print stable print.
- xP: Interval Print x = Print Interval (1-3600 sec), 0P turns auto print OFF.
- Z: Same as pressing Zero Key.
- T: Same as pressing Tare Key.
- xT: Download Tare value in grams (positive values only). Sending 0T clears tare (if allowed).
- PU: Print current unit: g, kg, lb, oz, lb:oz, t, c (custom).
- xU: Set scale to unit x: 1=g, 2=kg, 3=lb, 4=oz, 5=lb:oz, 6=t.
- xM: Set scale to mode x. M will scroll to next enabled mode.
- PV: Version: print name, software revision and LFT ON (if LFT is set ON).
- H x “text”: Enter Header line, where x = line number 1 to 5, “text” = header text up to 24 alphanumeric characters.
- F x “text”: Enter Footer line, where x = line number 1 to 2, “text” = rear text up to 24 alphanumeric characters.
- \EscR: Global reset to reset all menu settings to the original factory defaults (Escape key + ‘R’ Key).
How does Auto-Print operation work?
Once Auto-Print is activated in the menu, the balance will send data as required. To temporarily stop Auto-Printing, press the PRINT key. If there is data in the print buffer the printer will finish printing this data. A second press will resume Auto-Printing. (Note: In some modes, stopping or resuming Auto-Print must be done when the balance is not reading “0”.)