What are the general environmental requirements for the KEYSIGHT V3080A?
The KEYSIGHT V3080A Vector Signal Generator Frequency Extender is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for the instrument.
| Environmental Condition | Requirement |
|---|---|
| Temperature | Operating condition, 0 °C to 40 °C Storage condition, -40 °C to 70 °C |
| Maximum Relative Humidity (non-condensing) | Operating condition 95% RH |
| Altitude | Operating condition, <4600 meters (15,000 feet) Storage condition, <4600 meters (15,000 feet) |
What are the physical characteristics and dimensions of the KEYSIGHT V3080A?
The physical characteristics and dimensions for the KEYSIGHT V3080A are:
| Instrument Dimensions and Weight | Weight |
|---|---|
| Instrument Weight | 0.62 kg |
| Base Plate Weight | 0.9 kg |
| Instrument Dimensions (HxWxL) | 8.1 cm H x 4.77 cm W x 11.56 cm L |
| Base Plate Dimensions (HxWxL) | 2.8 cm H x 8.5 cm W x 16.8 cm L |
What are the specifications for the KEYSIGHT V3080A?
The general specifications for the KEYSIGHT V3080A are:
| Description | Specification |
|---|---|
| Input Power Max. | +18 dBm (63 mW), CW or pulsed 0 VDC |
| 2-channel VXG | Requires 220/240 VAC |
| Calibration Cycle | 1 year |
What additional tools are required for setting up the KEYSIGHT V3080A?
Additional tools required for setting up the KEYSIGHT V3080A include:
5/16, 8-in-lb torque wrench for instrument cable connections
2.0 mm hex wrench for attaching the baseplate to the KEYSIGHT V3080A
2.5 mm hex ball driver (or a 3/32 hex ball driver) for the security lock
1/4 inch wrench for 2.4 mm adapter
Small flat head screwdriver to secure the USB-C cable connection
What should I do if the shipping materials for the KEYSIGHT V3080A are damaged or the contents are incomplete?
If the shipping materials are damaged or the contents of the container are incomplete:
Contact the nearest Keysight Technologies office.
Keep the shipping materials for the carrier’s inspection.
If you must return the frequency extender to Keysight Technologies, use the original (or comparable) shipping materials. See “Returning a KEYSIGHT V3080A Frequency Extender for Service”.
How do I attach the Baseplate to the KEYSIGHT V3080A?
Attach the Baseplate to the KEYSIGHT V3080A using the four screws supplied in the case. The screws go through the bottom of the baseplate into the holes in the bottom of the KEYSIGHT V3080A. A 2.0 mm hex wrench (not included in the shipment) is required. Hand tighten the screws. Do not over torque.
How do I connect the V3080A to M9484C?
Connect the cable assembly to the KEYSIGHT V3080A as follows. The cable assembly is marked with connections to each instrument. Torque all these cables to 8-in-lbs.
Use a flat head screwdriver to secure the USB-C connector to the V3080A.
V3080A Cable Connections:
| Cable Assembly | V3080A |
|---|---|
| IF/RF In | IF/RF In |
| USB-C | USB-C |
| LO In | LO In (9.6 GHz In) |
RF Channels Connectivity for V3080A:
You can connect one or two V3080As to a 2-Channel M9484C VXG.
If you want to connect a single V3080A to a 2-Channel M9484C, then you must connect it to Channel 1 of M9484C.
To make cable connections:
Connect the 3.5 mm (f) to 3.5 mm (f) adapter to the M9484C LO 1 or LO 2.
Connect the 1.85 mm (f) to 1.85 mm (f) adapter to the M9484C RF Out channel. Torque the 1.85 mm adapter to 4-in-lbs.
Connect the cable assembly to the M9484C as follows. The cable assembly is marked with connections to each instrument. Torque all cables to 8-in-lbs.
Use a flat head screwdriver to secure the USB-C connector to the M9484C.
M9484C Cable Connections:
| Cable Assembly | M9484C |
|---|---|
| RF OUT | RF Out |
| USB-C | USB-C |
| LO OUT | LO 1 for Channel 1. If using a two-channel M9484C, and are using Channel 2, then connect to LO 2. |
Once you have made all cable connections, verify that the two LEDs on the front of the V3080A are green.
If the V3080A does not power on (LEDs green), the USB-C cable may not be plugged into the M9484C correctly. Turn the USB-C connector 180 degrees, and verify the LEDs are green. For older M9484C models, the USB-C connector can be plugged in either orientation, but it will only work in one direction.
How do I exit the M9484C application after connecting the KEYSIGHT V3080A?
After making the connections, ensure that you exit the M9484C. On the VXG, select the Triple-bar icon (top left corner) and then Exit M9484C.
Double-click on the PathWave icon on the Desktop to restart. Note: It takes a moment for the application to start up.
How do I configure the software settings for the KEYSIGHT V3080A?
When you restart the M9484C VXG software application after making hardware connections, the M9484C VXG automatically configures the instrument’s settings for the KEYSIGHT V3080A.
However, the characteristics of the cable used to provide the IF to the V3080A must be entered into the instrument’s software. To do this, you need to select the appropriate IF cable used for V3080A connection in the M9484C software GUI.
On the VXG, select the Triple-bar icon (top left corner) and then Configure Launch.
In the External Upconverter tab, select the IF cable .s2p de-embedding file appropriate for your setup. Your choices should be as follows:
If you are using a cable from the V3080A cable options available from Keysight Technologies, the .s2p characteristics are included in the instrument’s software. You only need to select the appropriate file matching the Keysight cable.
If you are using a cable obtained elsewhere, select Custom and provide the path to the .s2p file in the Custom Cable File setting.
If the selected .s2p file does not match the characteristics of the utilized cable, inaccurate corrections will be applied.
How do I measure a CW Signal using the KEYSIGHT V3080A?
To measure a CW Signal on the VXG:
1. Set the Frequency to 65 GHz and the Power to -10 dBm.
2. Set RF Out to On by selecting the numbered channel indicator switch. This enables the RF Out for the indicated channels. In this case, Channel 1 if using a multi-channel VXG.
On the N9042B:
1. In Spectrum Analyzer mode, select Preset > Mode Preset.
2. From the menu panel, select Frequency menu and set Center Frequency to 65 GHz and Span to 200 MHz.
3. Select the Marker menu and select Peak Search. Verify the peak power is approximately -10 dBm.
Repeat the above procedure at other frequencies of your KEYSIGHT V3080A.
How do I measure a Modulated Waveform using the KEYSIGHT V3080A?
To measure a Modulated Waveform on the VXG:
1. Select the Signal block to open. The screen below is only accessible if Option M9484C-8SG (8 virtual signal generators) is installed. For all other option configurations, continue to the next step.
2. In the Vector Modulation Signal Setup, select the Mode dropdown and set to Waveform File.
3. In the left panel, select the Waveform File tab.
4. Use File Select to navigate to:
D:\Users\Instrument\Documents\Keysight\PathWave\SignalGenerator\Examples
and choose 5G_100MHz_1CC_FR1.wfm, then Select.
5. Select the General tab and Enable to turn on the waveform.
Verify that modulation has been enabled by observing the 100 MHz waveform.
What is the purpose of the security lock for the KEYSIGHT V3080A?
The purpose of the security lock is to protect the expensive ruggedized adapter from theft. The security lock will work with any ruggedized adapter. An example of a ruggedized adapter is the Y1900B 1.0 mm ruggedized (f) to 1.0 mm (f), DC to 120 GHz.
How do I install the security lock on the KEYSIGHT V3080A?
To use the security lock:
1. Torque the V3080A test port and the ruggedized adapter to 10-in-lb using the supplied double-ended torque wrench.
2. Fit the security lock around the connection, and tighten the security lock with a 2.5 mm hex ball driver. A 3/32″ hex ball driver will also work if the 2.5 mm driver is not available.
3. The security lock will fit loosely, because the purpose is to provide additional security from theft. The purpose is not to provide any mechanical means of preventing the ruggedized adapter from loosening from the instrument test port, or not to provide a means to hold the adapter/test port in place when a cable is attached to the ruggedized adapter.
4. Connect the cable or DUT to the ruggedized adapter, by holding the wrench flats on the ruggedized adapter with an open end wrench while torquing the cable or DUT connector. It is very important to use the open end wrench to prevent the ruggedized adapter from spinning.
What is the meaning of the Status and IF Out LEDs on the KEYSIGHT V3080A?
The KEYSIGHT V3080A has two LEDs: Status and IF Out that signal the state of the V3080A Frequency Extender. The table below provides information on how to interpret the LEDs.
| State | Status LED | IF Out LED |
|---|---|---|
| Powered. A successful Power Delivery negotiation took place across a USB-C connection. Ready for measurement. | Green | Green |
| RF Board failed to initialize properly, including a successful Power Delivery negotiation across a USB-C connection. | Red | Red |
What precautions should be taken when working on electronic assemblies to prevent electrostatic discharge (ESD) damage to the KEYSIGHT V3080A?
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. An example of a static-safe workstation shows the use of two types of ESD protection:
Conductive table-mat and wrist-strap combination.
Conductive floor-mat and heel-strap combination.
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone. To ensure user safety, the static-safe accessories must provide at least 1 megohm of isolation from ground.
These techniques for a static-safe workstation should not be used when working on circuitry with a voltage potential greater than 500 volts.
How should I handle electronic components and ESD with the KEYSIGHT V3080A?
The possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used. The risk of ESD damage can be greatly reduced by paying close attention to how all components are handled.
Perform work on all components at a static-safe workstation.
Keep static-generating materials at least one meter away from all components.
Store or transport components in static-shielding containers.
Always handle printed circuit board assemblies by the edges. This will reduce the possibility of ESD damage to components and prevent contamination of exposed plating.
What safety precautions should be followed when using the KEYSIGHT V3080A?
Safety precautions for the KEYSIGHT V3080A:
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.
Safety of any system incorporating the equipment is the responsibility of the assembler of the system.
Disconnect the Keysight V3080A from USB power before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
The outer surfaces of the V3080A can become hot during use. Use caution when handling the units. It is recommended to power down the unit to allow the heat to dissipate prior to handling.
The AC Voltage source (outlet) must be in proper working order and provide a secure electrical connection.
Do not use the outlet if the power cord makes a loose connection or if the power cord plug does not match the outlet. Do not use the outlet if it is damaged or if the voltage is outside the required range.
The V3080A frequency extender is not field-repairable, and requires return to Keysight for both repair and calibration. Do not attempt to open the instrument enclosure.
The V3080A is intended to be used in the normal upright position only. The unit includes ventilation openings at the front and top sides. Do not block or impede airflow to these areas. Doing so could cause damage to the device.
The V3080A must be used with the standard base positioner or optional desktop positioner. Secure the V3080A to the base unit with the provided hardware. The unit includes ventilation openings at the front and top sides. Do not block or impede airflow to these areas. The unit is intended to be used in the normal upright position only. Do not stack the units as this could impact cooling to the device.
How do I perform a visual inspection of connectors for the KEYSIGHT V3080A?
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
Magnification is helpful when inspecting connectors, but it is not required and may actually be misleading. Defects and damage that cannot be seen without magnification generally have no effect on electrical or mechanical performance. Magnification is of great use in analyzing the nature and cause of damage and in cleaning connectors, but it is not required for inspection.
Examine the connectors first for obvious defects or damage:
Plating
Bare metal showing
Burrs or blisters
Deformed threads
Center conductors
Bent
Broken
Misaligned
Concentricity
Connector nuts should move smoothly and be free of:
Burrs
Loose metal particles
Rough spots
Any connector that has obvious defects should be replaced.
Inspect the connector contacts for integrity. It is necessary to use good lighting (such as a halogen task light) to see the contacts. Notice the location of the cross hairs in relationship to the center of the figures.
What is the procedure for making connections to the KEYSIGHT V3080A?
Good connections require a skilled operator. Instrument sensitivity and coaxial connector mechanical tolerances are such that slight errors in operator technique can have a significant effect on measurements and measurement uncertainties.
The most common cause of measurement error is poor connections.
1. Ground yourself and all devices (wear a grounded wrist strap and work on an antistatic mat).
2. Visually inspect the connectors (refer to “Visual Inspection”).
3. If necessary, clean the connectors (refer to “Cleaning Connectors”).
4. Carefully align the connectors. The male connector center pin must slip concentrically into the contact fingers of the female connector.
5. Push the connectors straight together. Do not twist or screw these together. As the center conductors mate, there is usually a slight resistance.
Do not twist one connector into the other (like inserting a light bulb). This happens when you turn the device body, rather than the connector nut. Major damage to the center conductor and the outer conductor can occur if the device body is twisted.
How do I make RF Connections to the KEYSIGHT V3080A?
The ruggedized adapter (1 mm to 1.00 mm Y1900B/C/D adapter) can be connected to the V3080A Output port. Due to the physical requirements needed to obtain high performance, the geometry of these, as well as all, 1 mm connectors, demands special care to avoid expensive damage as explained in this document.
Making RF Connections:
1. Work at a static-safe workstation.
2. Visually inspect the connectors. If necessary, clean the connectors. Carefully align the connectors. The male connector center pin must slip concentrically into the contact fingers of the female connector.
3. Push the connectors straight together.
4. Tighten lightly using only your fingers as, at this point, all you want is a connection in which the outer conductors make gentle contact at all points on both mating surfaces. Very light finger pressure (no more than 2 pound-inches of torque) is enough.
Do not twist one connector into the other (like inserting a light bulb). This happens if you turn the device body rather than the connector nut. Major damage to the center conductor can occur if the device body is twisted.
5. Use a torque wrench to make the final connection. This guarantees perfectly tight, consistent connection that prevents connector damage.
Rotate only the connector nut when you make the connection. Do not rotate the cable or adapter.
Hold the torque wrench lightly at the groove located at the end of the handle.
Apply force perpendicular to the wrench handle. This applies torque to the connection through the wrench. Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unlimited amount of torque.
Tighten the connection just to the torque wrench “break” point. Do not tighten the connection further.
The maximum torque setting is 4 in-lb (0.45 Nm) for non-ruggedized 1.0 mm connectors.
What torque should be applied for the 1 mm Output Port connector on the KEYSIGHT V3080A?
For the 1 mm Output Port connector, use the dual torque wrench. For matching non-ruggedized connectors using the inner threads, to the KEYSIGHT V3080A, use the 4 lb-in torque wrench.
Two wrenches must be used to avoid stress on semi-rigid cables.
The silver end of the dual torque wrench is 4 lb-in. The red end of the torque wrench is 10 lb-in.
The torque specification for the Output Port threads is:
10 lb-in torque for ruggedized adapters (outer threads)
4 lb-in torque (inner threads)
How should wrenches be positioned when connecting a ruggedized adapter to the 1 mm Output Port of the KEYSIGHT V3080A?
The correct placement involves using the 10 lb-in torque wrench (the red end) and the 8 mm open end wrench. On the ruggedized adapter, connect the 8 mm wrench on the flats that are adjacent to the knurled ring.
To connect a cable, or device, to the ruggedized test port adapter, use a 7 mm open end wrench to hold the test port adapter stationary. For a 2.4 mm cable or device, use a 5/16 inch torque wrench to torque the cable or device to 8 inch-pounds (0.90 N-m). For a 1 mm cable or device, use a 6 mm torque wrench to torque the cable or device to 4 inch pounds (0.45 N-m).
What is the torque limit when connecting an adapter directly to the 1 mm Output Port of the KEYSIGHT V3080A without a ruggedized adapter?
It is recommended that you always connect a ruggedized adapter to the KEYSIGHT V3080A Output port. However, if you want to connect an adapter that uses the inner threads, or cable directly to a test port connector, use the 4 lb-in end of the dual torque wrench (silver end of the dual torque wrench). Do not use the red end of the dual torque wrench which is only used when connecting the stronger outside test port connector threads to the ruggedized adapter.
Never apply in excess of 4 lb-in of torque to the KEYSIGHT V3080A Output port inner threads.
To connect a cable or device to the 1 mm adapter, use a 6 mm open end wrench to hold the test port adapter stationary, and torque the cable or device to 4 inch pounds (0.45 N-m). A common 1mm cable connection nut is 6 mm.
How should I clean the connectors of the KEYSIGHT V3080A?
1. Inspect the connectors for dust, dirt, metal fragment, oils or film, and debris.
2. Blow off any dust with a filtered, clean supply of compressed air.
3. Add a few drops of high-purity isopropyl alcohol to a small cleaning swab (do not apply alcohol directly to the parts).
Cleaning connectors with alcohol shall only be done with the instruments power cord removed, and in a well-ventilated area. Allow all residual alcohol moisture to evaporate and the fumes to dissipate prior to energizing the instrument.
4. Gently wipe connecting surfaces with the end of the cleaning swab.
5. Blow dry with compressed air.
6. Inspect and repeat cleaning procedure if necessary.
What is the procedure for shipping the KEYSIGHT V3080A for service or repair?
If you wish to send your KEYSIGHT V3080A product for service or repair:
Include a complete description of the service requested or of the failure and a description of any failed test and any error message.
Remove and retain the baseplate that is held in place with four screws.
Return the 4-wire interconnect cable assembly with the instrument. The cable assembly may contribute to the fault seen.
Ship the product using the original or comparable anti-static packaging materials.
Contact Keysight for instructions on where to ship your product.