What are the standard components and available accessories for the KANOMAX CLIMOMASTER 6501-C0?
The KANOMAX CLIMOMASTER 6501-C0 comes with several standard components and has a range of available accessories.
Standard Components
| Items | MODEL | Qty | Functions |
|---|---|---|---|
| Main Body | 6501-00 | One of these | – |
| Main Body with analog output | 6501-A0 | With analog output function | |
| Main Body with pressure sensor | 6501-B0 | With pressure measurement function | |
| Main body with pressure sensor and analog output | 6501-C0 | With pressure measurement and analog output functions | |
| Probe | 6531-21 | One of these | Air velocity, air temperature, humidity sensor (directivity) |
| 6541-21 | Air velocity, air temperature sensor (directivity) | ||
| 6542-21 | Air velocity, air temperature sensor (needle-type) | ||
| 6533-21 | Air velocity, air temperature, humidity sensor (spherical) | ||
| 6543-21 | Air velocity, air temperature sensor (spherical) | ||
| 6551-21 | Air velocity sensor (mini-spherical, straight type) | ||
| 6552-21 | Air velocity sensor (mini-spherical, articulating type) | ||
| 6561-21 | Air velocity, air temperature sensor (directivity) | ||
| Probe Cable | 6531-06 | 1 | To connect Probe to Main Body |
| USB Cable | – | 1 | USB Communication Cable |
| Carrying Case | – | 1 | Hard case |
| Operation Manual | – | 1 | – |
| Manganese AA Batteries | – | 6 | – |
Available Accessories
| Items | MODEL | Functions |
|---|---|---|
| Extension Rod (Flexible) | 6531-04 | For measurements in high places |
| Extension Rod (Straight) | 6531-05 | For measurements in high places |
| AC Adaptor | 6113-02 | Power Supply |
| Printer (Recommended) | DPU-S245 | To print out data, calculation results, etc. |
| Printer Cable | 6000-31 | To connect the Main Body to the Printer |
| Mobile ANEMOMASTER Measurement Software | 6000-41 | Data acquisition software |
| Hands-free Case | 6000-61 | For hands-free measurement |
What do the safety symbols used with the KANOMAX CLIMOMASTER 6501-C0 mean?
The symbols indicate different levels of warnings and required actions:
Classifications
Danger: To Prevent Serious Injury or Death
Warnings in this classification indicate a danger that may result in serious injury or death if not observed.
Caution: To Prevent Damage to the Product
Warnings in this classification indicate a risk of damage to the product that may void the product warranty if not observed.
Description of Symbols
Caution Symbol (Triangle with illustration inside): This symbol indicates a condition that requires caution (including danger). The subject of each caution is illustrated inside the triangle (e.g., a high temperature caution symbol).
Prohibition Symbol (Circle with a slash through it): This symbol indicates a prohibition. Do not take the prohibited action shown inside or near this symbol (e.g., the disassembly prohibition symbol).
Mandatory Action Symbol (Blue circle with an exclamation mark): This symbol indicates a mandatory action. A specific action is given near the symbol.
What are the key safety dangers and cautions when using the KANOMAX CLIMOMASTER 6501-C0?
DANGER
• Never bring the probe close to a flammable gas atmosphere. The heated sensor may cause a fire or explosion.
• Never touch the sensor. The sensor is heated during operation. Touching it may cause burns and damage the sensor itself.
• Do not disassemble or heat the batteries, or put them into a fire. This may cause burns and the batteries may burst.
• If abnormal noises, smells, or smoke occur, or if liquid enters the instrument, turn off the instrument immediately, remove the batteries, or pull out the plug. There is a possibility of malfunction, electric shock, and/or fire. Contact your local distributor or service center for repair.
CAUTION
• Always unplug the instrument from the electrical outlet when not in use to avoid electrical shock, fire, or circuit damage.
• Do not use the instrument in a water vapor atmosphere. Condensed steam on the sensor will cause inaccurate measurements and may damage it.
• This instrument is designed for use in a clean air stream without dust or foreign materials, which can cause damage and impede accuracy.
• Do not apply force to the sensor. If the sensor is deformed, its accuracy may be affected.
• When measuring, ensure the direction mark on uni-directional probes is facing into the airflow to prevent inaccurate measurements.
• Do not use or leave the instrument in high temperature, high humidity, or dusty environments, or under direct sunlight for prolonged periods.
• Do not subject the instrument or probe to strong impacts like dropping or placing heavy objects on it.
• Never disassemble, modify, or repair the product.
• Do not pick up or carry the instrument by the probe cable.
• Remove batteries when storing for a long period. Do not leave exhausted batteries in the compartment. Insert batteries with the correct polarity.
• Do not wipe the instrument with a volatile solvent. Use a soft dry cloth. For persistent stains, use a cloth soaked in a neutral detergent.
• Discharge any built-up static electricity from your body before touching the instrument.
• Regularly check the probe head for contamination (dust), as it can affect accuracy. To clean, use a blow brush or rinse with water and air-dry completely. Turn power off before cleaning and never use heat to dry the probe.
• Do not move the main unit and probe from a cold to a warm place quickly, as this can cause condensation and lead to inaccurate measurements or malfunction.
• Do not touch the LCD screen with a sharp-pointed object or with excessive pressure.
• When storing, put the instrument in its carrying case and keep it in a place with an ambient temperature of -10 to 50°C and no condensation.
• Do not dispose of the instrument as household waste. Disposal should be in line with local or national legislation.
What are the functions of the keys on the KANOMAX CLIMOMASTER 6501-C0 touch panel?
MENU Key: Enters the Main Menu to select a function. Pressing MENU also cancels the current measurement or setting and returns to the Main Menu.
START/HOLD Key: Starts and stops calculations and/or measurements. It also functions as a hold key during measurements.
▲ ▼ Navigation Keys:
1. In Normal Mode, these keys allow you to select the time constant (TC) of 1, 5, or 10 seconds. TC1 displays the instantaneous value every second, TC5 displays the moving average for 5 seconds, and TC10 displays the moving average for 10 seconds.
2. In the Main Menu, these keys scroll in the desired direction.
MODE Key: Selects the measurement mode. The available parameters, depending on the connected probe and main body model, cycle in the following order: Air Velocity/Air Temp/Humidity → Pressure → Air Flow/Air Temp/Humidity.
SET Key: Executes the selected item.
How do I install or replace the batteries in my KANOMAX CLIMOMASTER 6501-C0?
Follow these steps to install or replace the batteries:
1. On the backside of the body, press down on the battery cover.
2. Slide the cover until it stops.
3. Lift the cover away from the body.
4. Insert six (6) AA size batteries according to the indicated polarity chart. All batteries must be of the same type. Do not mix different types of batteries.
5. Put the cover back on by reversing the above procedure.
Note: Batteries CANNOT be recharged by the AC adapter.
The following types of AA batteries can be used:
• Manganese
• Alkaline
• Ni-Cd
• Ni-MH
How do I connect the probe to the KANOMAX CLIMOMASTER 6501-C0?
Ensure the power is turned OFF before connecting or disconnecting the probe.
1. Align the connector of the probe or probe cable with the probe terminal on the body. When connecting the probe directly, make sure the probe number faces towards the front side of the main unit.
2. Line up the connector and the terminal and gently push until it clicks.
Important: DO NOT FORCE the connector in, and DO NOT ROTATE the probe/probe cable once it is connected, as this may cause serious damage.
How do I disconnect the probe from the KANOMAX CLIMOMASTER 6501-C0?
Ensure the power is turned OFF before connecting or disconnecting the probe.
1. Lightly lift the locking ring of the probe/probe cable.
2. Pull the probe/probe cable out vertically while holding the lock ring up.
Important: DO NOT rotate the connector while it is connected to the main body, as this may cause serious damage.
How do I turn the KANOMAX CLIMOMASTER 6501-C0 on and off, and what do the initial screens indicate?
The power switch is located on the side of the instrument.
• To Turn On: Move the switch to the “|” position.
• To Turn Off: Move the switch to the “O” position.
When you turn the power on with the probe connected, a test screen with the Kanomax logo, model, and firmware version will display for a few seconds before switching to the Normal Mode screen.
“NO PROBE!” Warning: If the probe is not connected properly, this warning will be displayed. Turn the power off and check the probe connection.
Annual Calibration Reminder: A year after the last calibration, a notice to perform the annual calibration will be displayed. The screen will switch to normal measurement mode after 10 seconds or after any key is pressed.
Battery Level Indicator: This icon on the screen shows the remaining battery level. The consumption rate depends on the measured air velocity. When the batteries need to be replaced, the indicator will start blinking. If a high velocity is measured while the indicator is blinking, the screen may freeze. When the locked status icon [X] is displayed, all functions are locked. Turn off the power and replace the batteries.
What precautions should I take when measuring with the KANOMAX CLIMOMASTER 6501-C0?
Air Velocity
• For directional probes (MODEL 6531/6541/6561), make sure the direction mark is facing into the airflow. If unsure of the direction, slowly rotate the probe to find the maximum velocity reading.
• Omnidirectional probes (MODEL 6533/6542/6543/6551/6552) have a horizontal, 360-degree dynamic characteristic.
• Ensure that both the air velocity sensor and temperature compensation sensor are evenly exposed to the airflow under the same conditions.
• In environments with rapid air temperature changes, allow the reading to stabilize for at least 20 seconds before starting the measurement to let the probe acclimate.
Air Temperature
• The response time for temperature is faster with higher wind speeds. The normal response time is approximately 30 seconds at 1 m/s air velocity. Wait for the reading to become stable.
• When there is no airflow, the temperature reading may be higher than actual due to heat from the air velocity sensor. It is recommended to slowly wave the probe to create approx. 0.1 m/s airflow for accurate readings.
Humidity
• In extremely high humidity or with rapid temperature changes, readings may be higher than actual due to condensation on the sensor. If this occurs, allow the probe to dry in an atmosphere with 40%RH or less for 24 hours.
Pressure (for applicable models)
• Do not expose the pressure sensor to more than 75kPa of pressure.
• The operating temperature for measuring pressure is between 5°C and 40°C (41°F and 104°F).
• Perform a zero adjustment before measuring pressure. Leave both pressure ports (+) and (-) open during zero adjustment.
• The zero adjustment range is -0.5 to 0.5 kPa. An error message will display if the reading is outside this range.
How do I perform a pressure zero adjustment and connect the pressure tube on the KANOMAX CLIMOMASTER 6501-C0?
Zero Adjustment Procedure
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “8. PRESSURE ZERO” and press SET.
3. Use the ▲▼ keys to select “1. ZERO ADJUST” and press SET.
4. Use the ▲▼ keys to select “YES” and press SET to display 0.00 kPa.
5. Use the ▲▼ keys to select “2. SAVE INFO” and press SET. This completes the zero adjustment and returns you to the Main Menu.
Connecting the Pressure Tube
• Connect the pressure tube to the (+) or (-) port.
• Insert the other end of the tube into the measuring hole of the duct.
• If the pressure to be measured is positive, connect the tube to the (+) port.
• If the pressure to be measured is negative, connect the tube to the (-) port.
Ensure the tube is properly secured without any leakage or bends for an accurate measurement.
How do I input the duct shape and size on the KANOMAX CLIMOMASTER 6501-C0?
Before measuring volumetric flow rate, you must enter the duct shape and size settings. Up to 25 different duct types can be registered.
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “2. DUCT TYPE” and press SET.
To Set Duct ENTRY No.
• Use the ▲▼ keys to select “1. ENTRY No.” and press SET.
• Use the ▲▼ keys to choose a memory address from 1 to 25 and press SET.
To Set Duct SHAPE (RECTANGLE/CIRCLE)
• Use the ▲▼ keys to select “2. SHAPE” and press SET.
• Use the ▲▼ keys to select “RECTANGLE” or “CIRCLE” and press SET.
To Set Duct Size
• For a Rectangular Duct: Select “3. WIDTH”, press SET, enter the width with ▲▼ keys, and press SET. Then select “4. HEIGHT”, press SET, enter the height, and press SET. The maximum value is 9999mm or 999.9 inch.
• For a Circular Duct: Select “3. DIAMETER”, press SET, enter the diameter with ▲▼ keys, and press SET. The maximum value is 9999mm or 999.9inch.
To Set Units (mm/inch)
• Select “5. UNIT(mm/inch)” (or “4. UNIT (mm/inch)” for a circular duct) and press SET.
• Use the ▲▼ keys to select mm or inch and press SET.
To Save the Setting
• Use the ▲▼ keys to select “6. SAVE INFO” and press SET. This will complete the Duct Shape/Size Input and return you to the Main Menu.
How do I use NORMAL MODE on the KANOMAX CLIMOMASTER 6501-C0?
NORMAL MODE is the default mode when you turn on the instrument. The display is updated every second, and data cannot be saved in this mode.
To Enter NORMAL Mode: From another mode, press MENU, select “1. NORMAL”, and press SET.
Selecting Measuring Parameters:
While in Normal Mode, press the MODE key to change the displayed measurement parameter. The sequence is: Pressure → Air Rate → Air Velocity.
• Air Velocity Measurement Screen: Displays air velocity, air temperature, and humidity (if available).
• Pressure Measurement Screen: Displays pressure. This feature is for models 6501-B0/6501-C0.
• Flow Rate Measurement Screen: Displays the calculated flow rate. The selected duct type is shown at the lower left of the screen. To select a different registered duct type, press and hold SET while using the ▲▼ keys. If a duct type is not selected or available, the flow rate reading will show as “—-“.
How can I hold the current reading or the maximum value in NORMAL MODE with the KANOMAX CLIMOMASTER 6501-C0?
To Hold the Current Reading:
• While measuring, press the START/HOLD key. The “HOLD” indicator will appear on the display, and the reading will be frozen.
• Press the START/HOLD key again to release the hold and resume live measurement.
To Hold the Maximum Value:
• While measuring, press and hold the START/HOLD key. The “HOLD” indicator will appear.
• As long as you keep the button pressed, the display will show the maximum value of each parameter (air velocity, air temperature, humidity, pressure) reached during that time.
• When you release the START/HOLD key, the reading shown will be frozen.
• Press the START/HOLD key again to release and return to live measurement.
What is the Time Constant on the KANOMAX CLIMOMASTER 6501-C0 and how do I set it?
The Time Constant (TC) determines the time span of the moving average for the displayed readings. A longer TC results in more stable readings, while a shorter TC makes the readings more responsive to changes. This function is only available in Normal Mode.
To set the Time Constant:
In NORMAL Mode, press the ▲ or ▼ key to cycle through the options:
• TC1: Displays the instantaneous value at every 1 second.
• TC5: Displays the moving average for the past 5 seconds, updated every second.
• TC10: Displays the moving average for the past 10 seconds, updated every second.
By default, the Time Constant is only effective for AIR VELOCITY and AIR FLOW measurements. To apply it to AIR TEMPERATURE, HUMIDITY, and PRESSURE, you must change the application setting.
How do I use the CALCULATION Mode on the KANOMAX CLIMOMASTER 6501-C0?
CALCULATION Mode automatically calculates the maximum, minimum, and mean of the measured data over a set period.
Setup:
1. Press MENU and select “3. CALCULATION”, then press SET.
2. Set the parameters on the CALCULATION setup screen:
• 1. MODE: Select AVERAGE (takes measurements every second and averages them for the sampling time) or INSTANT (takes a single measurement at the end of the sampling time).
• 2. SAMPLING TIME (1 to 999 sec): Set the length of each trial.
• 3. No. TRIAL (N) (1 to 9999 times): Set the number of samples (trials) to be taken.
• 4. DATA STORAGE (YES or NO): Choose whether to save the results.
• 5. SET TO START: Use the ▲▼ keys to select this and press SET to save settings and go to the ready screen.
Measurement:
1. On the “READY” screen, you can press MODE to change the applied parameters (e.g., Air Velocity, Flow Rate).
2. Press START/HOLD to begin the measurement. The screen will show “SMP.” while measuring.
3. Press START/HOLD again to stop. If you selected “YES” for data storage, the data will be stored. Pressing MENU will also stop the measurement, but data will NOT be saved.
4. After all trials are finished, the “RESULT” screen will display the MAX, AVG, and MIN values. Press the ▲▼ keys to cycle through the results for each parameter (e.g., Air Velocity, Air Temp, Humidity, Dew Point Temperature, Discomfort Index).
5. Press START/HOLD to return to the setup menu, or press MENU to return to the Main Menu.
How do I perform a single point air flow measurement using the SINGLE FLOW RATE Mode on the KANOMAX CLIMOMASTER 6501-C0?
This mode is used for performing a measurement of airflow, such as a duct traverse. It calculates the average, maximum, and minimum values from all measured points.
Setup:
1. Press MENU, select “4. FLOW RATE”, and press SET.
2. Select “1. SINGLE FLOW RATE” and press SET.
3. Configure the settings:
• 1. SAMPLING TIME (1 to 999 sec): Set the interval time between each sampling.
• 2. No. TRIAL (N) (1 to 9999 times): Set the number of samples to be taken at each point.
• 3. MEAS. POINT (1 to 2000 points): Set the total number of measuring points in the traverse.
• 4. DATA STORAGE (YES or NO): Choose whether to save the results.
• 5. DUCT ENTRY NO.: Select the registered duct number. If an empty entry is selected, you can input the duct dimensions.
• 6. SET TO START: Select this and press SET to save settings and proceed to the ready screen.
Measurement:
1. On the “Ready” screen, you can change the duct entry number by pressing and holding SET while using the ▲▼ keys.
2. Press START/HOLD to start measuring the first point. You can press START/HOLD to pause and resume measurement of the current point.
3. Once the measurement for the first point is complete, the instrument enters STANDBY (“READY After Point 1” screen) for the next point. The display shows the number of points already measured (e.g., 0001/0015).
4. Move the probe to the next position and repeat the process by pressing START/HOLD.
5. After all points are measured, or if you press SET to complete the measurement early, the results (MAX, AVG, MIN) are displayed. Press the ▲▼ keys to view results for Air Flow, Air Temperature, Humidity, and Air Velocity.
6. Press MENU to return to the Main Menu. If “DATA STORAGE” was set to YES, the results will be stored.
How do I use the MULTI FLOW RATE Mode on the KANOMAX CLIMOMASTER 6501-C0?
MULTI FLOW RATE mode is similar to SINGLE FLOW RATE mode but adds a “locations” parameter. This allows you to take several airflow measurements (e.g., from multiple ducts) and get an overall average.
Setup:
1. Press MENU, select “4. FLOW RATE”, and press SET.
2. Select “2. MULTI FLOW RATE” and press SET.
3. The setup is the same as for SINGLE mode, except for step 3:
• 3. MEAS./LOC.: Select this and press SET. First, use the ▲▼ keys to set the number of MEAS. points per location (1 to 999) and press SET. Then, use the ▲▼ keys to set the number of MEAS. locations (1 to 999) and press SET.
4. Complete the rest of the setup (Data Storage, Duct Entry No.) and select “6. SET TO START”, then press SET.
Measurement:
• The measurement process is the same as for SINGLE FLOW RATE mode. Follow the on-screen prompts for each point and each location.
• After all measurements at all locations are completed, the calculation results will be displayed. Press the ▲▼ keys to cycle through the results for Flow rate, Air Temp, Humidity, and Air velocity.
How do I recall and view stored data on the KANOMAX CLIMOMASTER 6501-C0?
To Recall Stored Data:
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “5. DATA OUTPUT” and press SET.
3. Select “1. DISPLAY” and press SET.
4. The PAGE SET screen appears. Use the ▲▼ keys to select the page of data you wish to recall and press SET. Each page contains information about the measurement, such as mode, date, time, and number of trials.
To View Recalled Data:
• The recalled data for the selected page will be displayed.
• Use the ▲▼ keys to scroll through the data points.
• If you measured in FLOW RATE Mode or CALCULATION Mode (with flow rate), you can press MODE to switch the display between Flow Rate and Air Velocity.
• If you measured pressure, only pressure will be displayed.
To Calculate a Specific Range of Recalled Data:
1. On the data display screen, if the range shown (e.g., START:001 END:050) is correct, press START/HOLD to calculate and display the results (MAX, AVG, MIN).
2. To change the range, press SET. A cursor will appear on “START”. Use the ▲▼ keys to select the first data point for the calculation and press SET.
3. The cursor moves to “END”. Use the ▲▼ keys to select the last data point and press SET.
4. Press START/HOLD to calculate the results for the newly defined range.
How do I print data from the KANOMAX CLIMOMASTER 6501-C0?
Preparation:
1. You need an optional printer (Recommended: Seiko Instruments Model DPU-S245) and printer cable.
2. Ensure the baud rate and data transmission settings are the same on both the CLIMOMASTER and the printer. The factory setting for the CLIMOMASTER is 8 data bits, None parity, and 1 stop bit. The baud rate can be changed in the UTILITY menu.
3. Connect the printer cable to the Serial Communication terminal on the side of the instrument.
4. Turn ON the CLIMOMASTER first, then turn on the printer. Make sure the CLIMOMASTER is in NORMAL Mode.
Print Out in NORMAL Mode:
• While the screen is displaying in NORMAL Mode, press START/HOLD to freeze the display.
• Press MODE to print out the data on the screen.
Print Out in CALCULATION Mode and FLOW RATE Mode:
• After the measurement and calculation have been completed and the results are displayed, press MODE to print out the result.
Print Out Stored Data (Single Page):
1. Go to MENU -> DATA OUTPUT.
2. Select “2. PRINTER(SINGLE)” and press SET.
3. Select the page you want to print and press SET.
4. The stored data for that page will be displayed. Press START/HOLD to bring up the PRINT OUTPUT menu.
5. Select what you want to print: 1. RESULT (Conditions and Calculation Results), 2. DATA (Conditions and Stored Data), or 3. ALL (All information). Press SET to print.
Print Out Stored Data (Multiple Pages):
1. Go to MENU -> DATA OUTPUT.
2. Select “3. PRINTER(MULTI)” and press SET.
3. To print a specific range of pages, select “1. SELECT”. Set the START page and END page. Then select the content to print out (RESULT, DATA, or ALL).
4. To print all stored pages, select “2. ALL DATA”, change the setting to YES, and press SET.
How do I connect the KANOMAX CLIMOMASTER 6501-C0 to a PC and download data?
You can download stored data to your PC using the supplied USB cable and communication software (like Hyper Terminal for Windows).
Connecting PC:
1. Connect the PC to the Climomaster using the USB cable.
2. Turn ON the CLIMOMASTER.
3. Make sure the CLIMOMASTER is displaying NORMAL Mode.
PC Commands:
Once connected, you can send commands from your PC’s terminal software to the CLIMOMASTER to retrieve data. All letter inputs need to be capital letters. A space is indicated by `ப` and pressing Enter is indicated by `↵`.
| Command | Function |
|---|---|
| D****↵ | Download On-Time Data (measured at every 1sec). `****` is the number of data points (e.g., D0004 for 4 seconds of data). |
| N↵ | Cancel the current data download. |
| S↵ | Output the Measuring Condition (of On-Time Data). |
| U↵ | Output the Measuring Units. |
| K↵ | Output the Duct Shape/Size settings. |
| P↵ | Output the Page Number of stored data. |
| T****↵ | Output the Stored Data. `****` is the page number. |
| M****↵ | Output the Measuring Condition (of Stored Data). `****` is the page number. |
| B↵ | Output the Measuring Condition of All Pages. |
How do I configure the analog output settings on my KANOMAX CLIMOMASTER 6501-C0?
This feature is only available on models with analog output (6501-A0/-C0). The output is a DC 0 to 1V signal. There is no analog output for Flow Rate.
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “7. UTILITY” and press SET.
3. Select “3. ANALOG OUTPUT” and press SET.
To Select Data Type:
• Select “1. OUTPUT SELECT” and press SET.
• Use the ▲▼ keys to select the parameter you want to output: “VEL.” (Velocity), “PRS.” (Pressure), “HUM.” (Humidity), or “TMP.” (Temperature). Press SET.
To Set the Output Range:
• Select “2. [Selected Parameter]” (e.g., 2. VELOCITY) and press SET.
• Use the ▲▼ keys to select the desired output range from the available options for that parameter and press SET.
To Save the Settings:
• Select “3. SAVE INFO” and press SET to save the changes and return to the UTILITY menu.
How do I set the date and time on the KANOMAX CLIMOMASTER 6501-C0?
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “7. UTILITY” and press SET.
3. Select “1. TIME ADJUST” and press SET.
4. To change the date format, select “1. STYLE” and press SET. Use the ▲▼ keys to choose a style (JP: YYYY/MM/DD, US: MM/DD/YYYY, EU: DD/MM/YYYY) and press SET.
5. To change the date, select “2. DATE” and press SET. Use the ▲▼ keys to change the selected value (year, month, or day) and SET to move to the next value.
6. To change the time, select “3. TIME” and adjust the hour, minute, and second in the same way.
7. After making changes, select “4. SAVE INFO” and press SET to save and return to the Main Menu. Note: When the date is changed, the dates of existing stored measurements will also change.
How do I change the measurement units and baud rate on the KANOMAX CLIMOMASTER 6501-C0?
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “7. UTILITY” and press SET.
3. Select “2. UNIT SELECT” and press SET.
4. On the UNIT SELECT screen, use the ▲▼ keys to highlight the item you want to change (VELOCITY, TEMPERATURE, FLOW RATE, PRESSURE, or BAUD RATE) and press SET.
5. Use the ▲▼ keys to select your desired unit or value from the available options and press SET. For example, for VELOCITY, you can choose m/s or FPM.
6. After setting all desired units, select “6. SAVE INFO” and press SET. The change will be saved, and the screen will return to the Main Menu.
How do I adjust the LCD backlight settings on the KANOMAX CLIMOMASTER 6501-C0?
1. Press MENU to enter the Main Menu.
2. Use the ▲▼ keys to select “7. UTILITY” and press SET.
3. Select “5. BACKLIGHT ON/OFF” and press SET.
4. Select “1. BACKLIGHT” and press SET.
5. Use the ▲▼ keys to choose between BACKLIGHT OFF (always off) and BACKLIGHT ON (turns on when a button is pressed).
6. If you chose BACKLIGHT ON, you can then set the lit-up time. Use the ▲▼ keys to select a duration (e.g., FOR 01 min) or ALWAYS (backlight stays on).
7. To save the changes, select “2. SAVE INFO” and press SET.
How do I delete stored data from the KANOMAX CLIMOMASTER 6501-C0?
To Delete a Specific Page or Range of Pages:
1. Press MENU to enter the Main Menu.
2. Select “6. DATA CLEAR” and press SET.
3. Select “1. CLEAR” and press SET.
4. Use the ▲▼ keys to select the starting page number for deletion (“START”) and press SET.
5. Use the ▲▼ keys to select the ending page number (“END”) and press SET.
6. With “SET” highlighted, use the ▲▼ keys to select YES and press SET. The selected page(s) will be deleted, and the remaining data will shift up.
To Delete All Data:
1. Press MENU to enter the Main Menu.
2. Select “6. DATA CLEAR” and press SET.
3. Use the ▼ key to select “2. ALL CLEAR” and press SET.
4. Use the ▲▼ keys to change the selection from NO to YES and press SET.
5. All data will be deleted, and the remaining memory will be reset to 20000.
How do I adjust the screen contrast on the KANOMAX CLIMOMASTER 6501-C0?
If the LCD display is too dark or too light, you can adjust the contrast.
1. Open the battery cover on the back of the unit.
2. Locate the contrast adjusting dial at the bottom of the battery compartment.
3. Use a precision screwdriver (-) with a tip size of 0.9 to 1.5mm to turn the dial.
• Turn the dial clockwise to darken the LCD.
• Turn the dial counterclockwise to lighten the LCD.
How should I clean the probe of my KANOMAX CLIMOMASTER 6501-C0?
Dust or particles on the velocity sensor can lead to less precise readings. Regular cleaning is recommended to maintain accuracy.
Procedure:
Clean the sensor by soaking it in water in an ultrasonic cleaner for 10 to 20 seconds. Do not soak it for too long, as this increases the risk of damaging the coating. Use only water for cleaning.
CAUTION:
• When cleaning, make sure the power is turned OFF.
• Make sure the sensor is completely dry before turning the instrument ON.
• For probes with a humidity sensor (MODEL 6531/6533), DO NOT let the humidity sensor come into contact with water. If it gets wet, let it air dry in an environment with less than 40%RH for more than 24 hours.
• NEVER use alcohol or any other organic fluid, as this will cause permanent damage to the humidity sensor.
• After cleaning, DO NOT place the sensor in an upward position until it is completely dry, as residual water could damage the sensor. You can use cool air to help dry it.
What should I do if my KANOMAX CLIMOMASTER 6501-C0 will not turn on?
The battery may be inserted with the wrong polarity. Turn off the power and re-insert the batteries correctly according to the polarity markings in the battery compartment.
What if the display on my KANOMAX CLIMOMASTER 6501-C0 is not visible?
The display contrast may not be set up properly. Locate the contrast adjustment dial inside the battery compartment and adjust it with a small screwdriver.
The battery level indicator is flashing on my KANOMAX CLIMOMASTER 6501-C0. What does this mean?
A flashing battery level indicator means the battery is low. You should turn off the instrument and replace the batteries.
My KANOMAX CLIMOMASTER 6501-C0 displays “[NO PROBE]”. What should I do?
This message indicates that the probe is not connected or not connected properly. Turn off the instrument and check the connection of the probe to the main body.
The screen of my KANOMAX CLIMOMASTER 6501-C0 is displaying “[**.*]”. What does it mean?
This indicates that the measurement is outside the measuring range of the instrument. The instrument must be used within the range stated in its specifications.
Why is my KANOMAX CLIMOMASTER 6501-C0 displaying “[—-]” for air velocity?
This can be caused by a few issues:
• The probe or the probe cable is not connected properly. Check the connection between the probe and the probe cable, and the cable to the main body.
• The probe or the probe cable may be damaged. If the connection is secure, contact your local representative for repair or replacement.
The air speed reading on my KANOMAX CLIMOMASTER 6501-C0 seems incorrect. What could be the cause?
The probe might not be facing into the wind correctly. For directional probes, make sure the wind direction indicator mark is facing directly into the airflow.
Why does my KANOMAX CLIMOMASTER 6501-C0 read a higher temperature with no wind?
The CLIMOMASTER cannot measure the correct temperature with no wind (less than 0.1m/s of wind) because of heat generated by the velocity sensor itself. To get an accurate reading, slowly wave the probe to create a small amount of airflow.
The response time of my KANOMAX CLIMOMASTER 6501-C0 seems slow. What can I do?
The instrument’s Time Constant (TC) setting may be too high. In NORMAL mode, you can set the TC to a shorter value (e.g., TC1 for a 1-second response) to make the readings more responsive.
The volumetric flow measurement on my KANOMAX CLIMOMASTER 6501-C0 is showing “[——-]”. How can I fix this?
This display indicates that the duct information (shape and size) required to calculate the flow rate is not stored or selected. You must go to the DUCT TYPE menu and register the duct information.
Why won’t my KANOMAX CLIMOMASTER 6501-C0 print?
There are several possible causes:
• Printer not connected properly: Confirm the printer connection.
• Baud Rate mismatch: Check that both the instrument and the printer are set to the same Baud Rate.
• Incorrect printer/cable: Make sure to use the optional printer (Model DPU-S245) and printer cable (Model 6000-31).
• Connection procedure: Ensure you connect the printer to the printer port, then turn on the instrument first, and then the printer.
I cannot print out the display from my KANOMAX CLIMOMASTER 6501-C0. What is wrong?
The display is not frozen. First, press the START/HOLD key to hold the display. Then, press the MODE key to print out.
My PC is not loading the measuring data from the KANOMAX CLIMOMASTER 6501-C0. What should I check?
• Connection: Confirm the USB connection is secure.
• Baud Rate: The Baud Rate might not be set properly. Check that both the instrument and the PC’s terminal software are set to the same Baud Rate.
• Command: Confirm you are using the correct communication command (e.g., in Hyper Terminal).
What are the technical specifications for the KANOMAX CLIMOMASTER 6501-C0?
| Product | CLIMOMASTER Air Velocity Meter | |||||||
| Model No. of the Main Body | 6501 | |||||||
| Model No. of the Probe | 6531-21 | 6541 -21 | 6561 -21 | 6542 -21 | 6551 -21 | 6552 -21 | 6533-21 / 6543 -21 | |
| Measuring Object | Clean air flow | |||||||
| Air Velocity [m/s] | Measuring Range | 0.01 to 30.0, 0.01 to 50.0 (6561-21 only) | – | 0.01 to 5.00 | ||||
| Resolution | 0 to 9.99: 0.01 10 to 30: 0.1 30 to 50: 0.1 (6561-21 only) |
0.01 | ||||||
| Accuracy | ±2% of reading or ±0.015 (±3FPM), whichever is greater | – | 0.01 to 0.99:±0.02 0.99 to 5.00: ±2% of reading or ±0.015 (±3FPM), whichever is greater |
|||||
| Response Time [sec.] | Approx. 1 | Approx. 4 | Approx. 7 (at 1m/s, 90% response time) |
Approx. 7 | ||||
| Air Temperature [°C] | Measuring Range | -20.0 to 70.0 | – | -20.0 to 70.0 | ||||
| Resolution | 0.1 | – | 0.1 | |||||
| Accuracy | ±0.5 | – | ±0.5 | |||||
| Response Time [sec.] | Approx. 30 (at 1m/s, 90% response time) |
– | Approx. 30 (at 1m/s, 90% response time) |
|||||
| Humidity *1 [%RH] | Measuring Range | 2.0 to 98.0 | – | 2.0 to 98.0 | ||||
| Resolution | 0.1 | 0.1 | ||||||
| Accuracy | 2 to 80: ±2.0 80 to 98: ±3.0 |
2 to 80: ±2.0 80 to 98: ±3.0 |
||||||
| Response Time [sec.] | Approx. 15 | – | Approx. 15 | |||||
| Pressure *2 [kPa] | Measuring Range | -5.00~+5.00 | ||||||
| Resolution | 0.01 | |||||||
| Accuracy | ±3% of reading or ±0.01 | |||||||
| Response Time [sec.] | Approx. 1 | |||||||
| Temperature Compensation [m/s] | ±5% of reading or ±0.015 (±3FPM), whichever is greater (in the temperature range of 5 to 60.0°C) | |||||||
| Functions | Readings Hold, Max Hold, Time Constant (1, 5, 10 sec.), Battery Level Indicator (5 Levels), Unit selection (Air Velocity: m/s or FPM, Flow Rate: m³/min, m³/h, ft³/min, or ft³/h, Temperature*1: °C or °F, Humidity*1: %RH, Static Pressure*2: kPa or Pa), Max., Min., and Average Calculation, Sampling Time: 1 to 999 sec. No. of trails: 1 to 9999, Store up to 25 Duct Sizes (Square/Round, Size Range: 1 to 9999mm or 0.1 to 999.9inch (diameter, height, width)), Max. Data Storage: 20000 | |||||||
| Output | Digital Output: USB (when connected to a printer, automatically switch to RS-232C, Baud Rate 4800, 9600, 19200 and 38400 bps) Analog Output*3: DC 0 to 1V (select one from Air Velocity, Temperature*1, Humidity*1 and Pressure*2) |
|||||||
| Power | 6 × AA Batteries, AC Adaptor*4: AC 100 to 240V (50/60Hz) | |||||||
| Battery Life | Approximately 10 continuous hours (at air velocity 5m/s, 20°C, with alkaline batteries) | |||||||
| Operating Environment | Main Body: 5 to 40°C (41 to 104°F) with no visible condensation Probe: -20 to 70°C (-4 to 158 °F) with no visible condensation |
|||||||
| Storage Environment | -10 to 50°C (14 to 122°F) with no visible condensation | |||||||
| Weight | Approximately 400g (batteries included) | |||||||
| Accessories | Carrying case, Operation manual, 6×AA Batteries, Probe cable (2m), USB cable | |||||||
| Options | Spare probe, Analog output, Pressure sensor, Extension rod, Printer, Printer cable, Software (for Windows), AC adaptor, Hands-free Case | |||||||
*1: Humidity is only available on MODEL6531/6533.
*2: Pressure is only available on MODEL6501-B0/-C0.
*3: Analog Output is only available on MODEL6501-A0/-C0.
*4: Optional.