APERA INSTRUMENTS DJS-0.1-F (01) Image

FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF USER GUIDE
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF USER MANUAL
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF OWNER GUIDE
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF OWNER MANUAL
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF REFERENCE GUIDE
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF INSTRUCTION GUIDE
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF REFERENCE MANUAL
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF INSTRUCTION MANUAL
FREE ENGLISH APERA INSTRUMENTS DJS-0.1-F (01) PDF OPERATING INSTRUCTIONS


CLICK HERE TO DOWNLOAD APERA INSTRUMENTS DJS-0.1-F (01) PDF MANUAL

If this is not the document you want for this product, click here to see if we have any other documents for this product.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

What are the specifications of the DJS-0.1-F Conductivity/Temperature Electrode?

Model Cell constant Electrode size Dimension size Socket Temp. Sensor Electrode body Cable length Range
DJS-0.1-F 0.1 7×18mm platinum Φ12×155 BNC/RCA 30K Glass 1m (0-200) µS/cm

How do I connect and calibrate the DJS-0.1-F conductivity electrode?

Connect the DJS-0.1-F conductivity electrode to your conductivity meter. Set the cell constant to 0.1. Then calibrate the meter in the lower calibration point (e.g. 84 µS/cm).


How should I maintain the electrode for accurate measurements?

Keep the conductivity electrode clean. Before and after the measurement, rinse the electrode with deionized water/distilled water. In order to maintain the highest measurement accuracy, we recommend rinsing the electrode several times with your test sample solution.


When and how should the included flow cell be used?

A Flow cell (included) is recommended when testing ultra-low conductivity solutions like ultra-pure water. It helps reduce the influence from contact with air and minimizes reading error. Use a rubber tube (7mm in diameter, not included) to connect your water source on the low end of the flow cell. When water flows through the glass cell, conductivity can be measured without air.


How do I clean a conductivity electrode?

(a) Use warm detergent water or alcohol to clean off organic contaminants on the electrode.

(b) Use 10% lemon acid to clean calcium and magnesium sediments.


How should the conductivity electrode be stored?

Store dry when not in use.


CLICK HERE TO DOWNLOAD APERA INSTRUMENTS DJS-0.1-F (01) PDF MANUAL


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.