What are the specifications of the DJS-0.1-F Conductivity/Temperature Electrode?
| Model | Cell constant | Electrode size | Dimension size | Socket | Temp. Sensor | Electrode body | Cable length | Range |
|---|---|---|---|---|---|---|---|---|
| DJS-0.1-F | 0.1 | 7×18mm platinum | Φ12×155 | BNC/RCA | 30K | Glass | 1m | (0-200) µS/cm |
How do I connect and calibrate the DJS-0.1-F conductivity electrode?
Connect the DJS-0.1-F conductivity electrode to your conductivity meter. Set the cell constant to 0.1. Then calibrate the meter in the lower calibration point (e.g. 84 µS/cm).
How should I maintain the electrode for accurate measurements?
Keep the conductivity electrode clean. Before and after the measurement, rinse the electrode with deionized water/distilled water. In order to maintain the highest measurement accuracy, we recommend rinsing the electrode several times with your test sample solution.
When and how should the included flow cell be used?
A Flow cell (included) is recommended when testing ultra-low conductivity solutions like ultra-pure water. It helps reduce the influence from contact with air and minimizes reading error. Use a rubber tube (7mm in diameter, not included) to connect your water source on the low end of the flow cell. When water flows through the glass cell, conductivity can be measured without air.
How do I clean a conductivity electrode?
(a) Use warm detergent water or alcohol to clean off organic contaminants on the electrode.
(b) Use 10% lemon acid to clean calcium and magnesium sediments.
How should the conductivity electrode be stored?
Store dry when not in use.