Free English A&D MEDICAL UC-321PBT (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.
CLICK HERE TO DOWNLOAD A&D MEDICAL UC-321PBT (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
What precautions should I take regarding the placement of the A&D MEDICAL UC-321PBT?
Do not place your scale on a soft, wet, or slippery floor. It is recommended that the scale be used on a solid level surface, such as tile, wood, or laminate flooring. If the scale must be used on a carpet, be sure to attach the plastic feet. Avoid extremes in temperature, humidity, direct sunlight, shock, moisture, and dust.
What are the safety usage guidelines for the A&D MEDICAL UC-321PBT?
Please follow these safety guidelines:
1. Do not step onto your scale with wet feet. Make certain that feet are dry to prevent slipping or serious injury.
2. Gently step onto the center portion of your scale.
3. Do not place feet on the display window.
How does the slide switch work on the A&D MEDICAL UC-321PBT?
The measurements can be selected by the slide switch and are classified into two categories: WEIGHT A and WEIGHT B. Within WEIGHT A and WEIGHT B, you have a choice to select and have your weight displayed in kg or lb. The difference between WEIGHT A and WEIGHT B is the speed of communication from the data port to another device (e.g. modem, printer, telephone). Please follow instructions from your telemonitoring company to select the proper position on the slide switch.
What do the display symbols indicate on the A&D MEDICAL UC-321PBT?
Ready/Complete symbol (Circle): This symbol is displayed when the measurement display is stable.
Low battery mark (Battery icon): If this mark is displayed, replace the batteries with new ones.
How do I install or replace the batteries in the A&D MEDICAL UC-321PBT?
Follow these steps to install or replace the batteries:
1. Remove the battery compartment cover on the underside of the scale by gently pushing on the arrow and lifting the cover backward.
2. Place four new batteries with the positive (+) and negative (-) terminals properly aligned with the corresponding symbols in the battery compartment.
3. Replace the cover by sliding it into the compartment and gently pressing into place.
Caution: Replace all four batteries with new ones when the low battery symbol is displayed. Do not mix used and new batteries. Remove the batteries if the scale is not to be used for a long time.
What is the procedure for measuring my weight on the A&D MEDICAL UC-321PBT?
1. Set the slide switch to the correct mode (e.g., “WEIGHT B”).
2. Press the measurement switch gently. All display segments are visible for several seconds.
3. Wait until the Ready/Complete symbol (Circle) is displayed and the display reads 0.0 kg / 0.0 lb.
4. Step on the scale gently and stand during measurement. You will see the display changed to dashes. Please stay still as much as you can.
5. Your weight is displayed after the Ready/Complete symbol (Circle) is displayed.
6. Step off the scale.
Note: If the display 0.0 kg / 0.0 lb is visible for approximately fifteen seconds, the scale will automatically turn off.
How do I measure an infant or small animal on the A&D MEDICAL UC-321PBT?
1. Set the slide switch to the correct mode.
2. Press the measurement switch gently and step on your scale without any baby or animal in your hands. All display segments are indicated for several seconds before displaying 0.0 kg / 0.0 lb.
3. Stay on the scale and have the infant or animal placed securely in your arms within ten seconds after the scale is displaying 0.0 kg / 0.0 lb and the Ready/Complete symbol. Keep still during the measurement until the Ready/Complete symbol is displayed. You will see the display changed to dashes.
4. The weight of the infant or animal can be read after the stable symbol (Circle) is displayed.
5. Step off the scale.
How do I use the A&D MEDICAL UC-321PBT on a carpet?
If the scale must be used on a carpet, be sure to attach the plastic add-on feet following these instructions:
1. There are two pins on each add-on foot side. Match those pins to the holes on the scale leg and insert firmly.
2. Make sure to put add-ons to all 4 legs.
Notes: Please be careful not to fall when using the add-on feet. It is generally recommended to use the scale on firm and level ground.
How should I clean and maintain the A&D MEDICAL UC-321PBT?
Clean the scale with a dry, soft cloth or a cloth moistened with water and a mild detergent. Never use alcohol, benzene, thinner, or other harsh chemicals to clean the scale. Remove the batteries if the scale is not going to be used for a long time.
How does the wireless connection work on the A&D MEDICAL UC-321PBT?
At the end of the measurement, your weight will be sent automatically to your health monitoring device. You do not need to do anything. If the health monitoring device is not working or is not within range of the scale, the weight will be stored in the scale’s memory (up to 40 sets).
How do I troubleshoot problems with the A&D MEDICAL UC-321PBT?
| Cause / Problem | Corrective Action |
|---|---|
| Nothing is displayed by pressing the measurement switch. | Check the battery installation. Replace with four new batteries. |
| The indication is fixed to 88888. H or L is displayed. | Remove and reinstall the batteries. |
| There is large difference between estimated weight and data measurement. | Place the scale on firm surface. Remove any foreign objects that might be touching the bottom of the scale. If the scale is used on a carpet, be sure that the feet are installed or try moving it to a level surface such as tile or wood. |
| E-0 is displayed. | Replace batteries. |