FREE ENGLISH KERN ORF 85BM (01) PDF USER GUIDE
FREE ENGLISH KERN ORF 85BM (01) PDF USER MANUAL
FREE ENGLISH KERN ORF 85BM (01) PDF OWNER GUIDE
FREE ENGLISH KERN ORF 85BM (01) PDF OWNER MANUAL
FREE ENGLISH KERN ORF 85BM (01) PDF REFERENCE GUIDE
FREE ENGLISH KERN ORF 85BM (01) PDF INSTRUCTION GUIDE
FREE ENGLISH KERN ORF 85BM (01) PDF REFERENCE MANUAL
FREE ENGLISH KERN ORF 85BM (01) PDF INSTRUCTION MANUAL
FREE ENGLISH KERN ORF 85BM (01) PDF OPERATING INSTRUCTIONS
CLICK HERE TO DOWNLOAD KERN ORF 85BM (01) PDF MANUAL
If this is not the document you want for this product, click here to see if we have any other documents for this product.
KERN ORF 85BM (01) PDF SUMMARY:
What is the purpose of the refractometer and what should it not be used for?
The refractometer is a measuring instrument for determining the refractive index of transparent substances in liquid or in some cases also in the solid state. It is used to observe the behavior of light as it passes from a prism with known properties to the substance being tested. Use of the refractometer for other purposes is contrary to its intended use and may be hazardous. The manufacturer shall not be liable for any damages caused by improper use.
What are some basic safety precautions to follow when using the refractometer?
- Do not let acids come into contact with skin or eyes. If acid comes into contact with skin, flush with copious amounts of water. Shower if larger areas of skin are affected.
- If acid comes into contact with eyes, keep the eyelid open and flush the eye with running lukewarm water from the outer corner to the inner corner. Flush eyes for at least 15 minutes. Then consult a doctor or ophthalmologist immediately.
- Thoroughly clean the refractometer after each use.
- The refractometer must not be exposed to extreme temperatures, high mechanical stresses, strong direct sunlight or high humidity.
- This refractometer is not a toy. Keep out of reach of children.
- Make sure that you will not be hit by anything else while you are using the refractometer, as this could cause serious eye injuries.
What items are supplied with the refractometer?
- Refractometer
- Storage box
- Pipette
- Screwdriver + batteries
- Cleaning cloth
- Calibration liquid
How does automatic temperature compensation work in the refractometer?
The refractive coefficient is temperature-dependent. Materials expand when they are heated (their density decreases) and contract when they cool (their density increases). The speed of light through a liquid increases with the temperature and the refractive coefficient therefore decreases. In solid substances this thermal effect has only a minor impact. In liquid substances, however, the change in density is considerable. Automatic temperature compensation ensures accurate measurement results for water-based solutions in relation to the temperature of the sample.
How to perform a calibration/zero point calibration?
- Inspect the prism to ensure that the surface is clean and dry.
- Apply a few drops of the calibration liquid onto the prism window.
- Press the “→0” button; the measuring instrument begins the zero-point calibration process.
- The measuring instrument is performing the calibration process when “Scan zero” and a loading bar appear in the display.
- When the calibration process is complete “Scan Result” and “Pass” is shown in the display in case the calibration was successful. When it was not successful the display shows “Fail”. The calibration result will be saved and will be the new zero point even after the device is powered off and on again.
How to perform a measurement?
- Inspect the prism window [4] to ensure it is clean and dry.
- Apply a few drops of the liquid being examined onto the prism window.
- Press the, Meas’ button; the instrument begins to measure and the results is shown in the display for 90 seconds.
- Press the “Meas” button to re-check the previous measurements.
How to perform a battery change?
When the batteries are fully charged the “Battery symbol” symbol appears at the top right of the display in green color. If the batteries are too weak or almost drained the “Battery symbol” symbol appears at the top right in red color. The batteries should then be replaced. Open the battery cover on the rear and replace the batteries. Always replace both batteries!
How to switch temperature unit?
The measuring instrument works in Celsius or Fahrenheit. For changing over just press the “→0” button for 2 seconds.
How do you switch the measuring scale on the device?
- Hold “Meas” for 2 seconds, the scale will change to the next measurement type.
- Repeat Step 1 until desired the scale is shown on the screen.
How to clean and maintain the refractometer?
Clean the refractometer using a soft, lint-free cloth moistened with water, or if necessary alcohol. Do not use any aggressive or abrasive cleaning agents. Never touch the measuring prism [2] with hard tools made from plastic, wood, rubber, metal, glass etc. Hard objects can quickly damage the relatively soft prism glass, resulting in measurement errors. The refractometer is maintenance-free. Cleaning should be carried out immediately before and after each use of the refractometer to maximise its life and optimise measurement results.
What are the error messages that may show?
| Error Message | Description |
|---|---|
| Fail | Calibration was not successful. Make sure you are using the correct calibration solution and there is sufficient amount and perform the calibration again. |
| High | Measurement outside of measuring range at top of scale. |
| Low | Measurement outside of measuring range at bottom of scale. |
| Temp. 45.0°C | Temperature too high for measuring instrument. |
| Temp. -10.0°C | Temperature too low for measuring instrument. |
| Error | No liquid on the prism or not sufficient amount. |
What are the troubleshooting steps to take if the refractometer is not working correctly?
- If you suspect that the measuring instrument is not working correctly or displays incorrect results, perform a zero point calibration.
- Then perform a measurement using water to check the zero point.
- If this does not lead to any improvement, please perform a factory reset by pressing “On/Off” and “Meas” simultaneously and confirming with “→0”.
- If the measured data still appear untrustworthy, please replace the batteries and perform the above calibration once again.
- If the measuring instrument jams or no longer reacts, separate the measuring instrument from the power supply by removing the batteries and replacing them.
- If no data appears in the display when it is switched on, examine the position and polarity of the batteries.
- If the instrument still does not work, please examine the battery charge status and/or replace the batteries.
- If the batteries are good and the measuring instrument still does not display, please call your dealer’s technical customer service.
What should be the lighting conditions be considered when taking measurements?
It is important to take care of the lighting conditions in the user environment when calibrating the device and subsequent measuring a sample. The measuring prism of the refractometer does not have any coverage and therefore a measurement in addition to the light of the internal light source is also influenced by ambient light.
What is the effect of the ambient light with the measurement?
Too bright lighting (e.g., by a lamp directly above the device) may trigger an error message already during the calibration. Ordinary (diffuse) light is not a problem.
Can the lighting condition be changed during measurement?
The lighting conditions (or the location of the device) may not be changed after the calibration or between different measurements, otherwise the refractometer performs each of the processes on a different basis and the results would not be comparable.
How to avoid influence of ambient light on measurements?
We recommend that both for the calibrations and for measurements, after the sample was placed on the measuring prism, to create an improvised opaque cover over the prism (incl. sample).
What is the optimal solution temperature and ambient temperature when doing measurements?
The refractometer is temperature-compensated for water-based, sugary solutions. Temperature differences in a range of 5 °C-40 °C (41 °F-104 °F) can be compensated. However, the temperature of the sample has only a minor influence on the accuracy of the measurement. The volume of the sample compared to the mass of the refractometer is so small that in most cases the sample is almost immediately at the temperature of the refractometer and assumes the ambient temperature, and is therefore in equilibrium with the temperature range of the measuring instrument. The temperature coefficient of the majority of water-based solutions is generally very close to the coefficient of the cane sugar / water solution. However, the temperature coefficient of oils, hydrocarbons and other, non-water-based solutions is generally greater. In order to achieve as accurate a result as possible, non-water based samples should be measured at an ambient temperature as close to 20 °C (68 °F) as possible.
What is the ideal ambient and calibration liquid temperature during calibration?
Ideally, the ambient and calibration liquid temperature should be exactly 20.0 °C (68.0 °F) during calibration.
What do you need for calibration/zero point calibration?
For the majority of models all you need for calibration/zero point calibration is distilled and deionised water. Calibration liquid with 60% Brix is required for a variety of models.
When should the calibration function be used?
The calibration function should only be considered when the original condition of the measuring prism or other optical components changes due to wear and tear causing measuring inaccuracies while using the refractometer.
How often should you do a calibration/zero point calibration?
The models of the ORF series have been calibrated at the factory. In general, this condition does not change even over a long period of usage, when carefully handled. Therefore it is not necessarily needed to carry out any additional calibration before use.
What are specifications of the refractometer?
- Temperature range: 0 °C-40 °C (32 F-104 F)
- Temperature measuring accuracy: +/- 0.5 °C (1 F)
- Automatic temperature compensation: 5°C – 40°C
- Minimum volume of sample: Approx. 3 drops
- Measurement period: ≤ 3 seconds
- Power supply: 2 x AAA batteries
- Battery life: ≥ 3,750 measurements
- Dimensions: 145 x 67 x 38 mm
- Weight: 185 g
- Language: English
- Auto-Off function: after 90 seconds
- IP protection class: IP 65
What are the models and specifications?
| Model | Scales | Measuring range | Accuracy | Division | Calibration solution |
|---|---|---|---|---|---|
| ORF 45BM | Brix | 0-45% | ± 0.2% | 0.1% | 0% (distilled water) |
| ORF 92BM | Brix | 58-92% | ± 0.2% | 0.1% | 60% solution |
| ORF 85BM | Brix | 0-85% | ± 0.2% | 0.1% | 0% (distilled water) |
| ORF 92HM | Brix | 58-92% | ± 0.2% | 0.1% | 60% solution |
| Baume | 38-43 °Be | ± 0.1 °Be | 0.1 °Be | ||
| Water content | 13-25% | ± 0.2% | 0.1% | ||
| Refractive index | 1.4370-1.5233 nD | ±0.0003 nD | 0.0001 nD | ||
| ORF 3SM | Brix | 0-35% | ± 0.2% | 0.1% | 0% (distilled water) |
| Salt (NaCl) | 0-28% | ± 0.2% | 0.1% | ||
| Refractive index | 1.3330-1.3900 nD | ± 0.0003 nD | 0.0001 nD | ||
| ORF 2WM | Mass SW | 0-35% | ± 0.2% | 0.1% | 0% (distilled water) |
| KMW (Babo) | 0-25 °KMW | ± 0.1 °KMW | 0.1 °KMW | ||
| Oechsle | 30-150 °Oe | ± 1 °0e | 1 °Oe | ||
| Vol. AP | 0-22% | ±0.1% | 0.1% | ||
| ORF 1PM | Serum protein | 0-12 g/dl | ± 0.1 g/dl | 0.1 g/dl | 0% (distilled water) |
| Urine (spec. weight) | 1.000-1.050 sgU | ±0.001 sgU | 0.001 sgU | ||
| Refractive index | 1.3330-1.4098 nD | ±0.0003 nD | 0.0001 nD | ||
| ORF 2UM | EG | -50-0°C | ±0.5°C | 0.1 °C | 0% (distilled water) |
| PG | -50-0 °C | ±0.5°C | 0.1 °C | ||
| BF | 1.00-1.50 kg/l | ± 0.01 kg/l | 0.01 kg/l | ||
| CW | -40-0 °C | ±0.5°C | 0.1 °C | ||
| ORF 5UM | EG | -50-0 °C | ±0.5°C | 0.1 °C | 0% (distilled water) |
| PG | -50-0 °C | ±0.5°C | 0.1 °C | ||
| Urea | 0-40% | ± 0.2% | 0.1% | ||
| CW | -40-0 °C | ±0.5°C | 0.1 °C | ||
| ORF 6US | Urea | 0-40% | ± 0.2% | 0.1% | 0% (distilled water) |
| ORF 1RS | Refractive index | 1.3330-1.5400 nD | ± 0.0003 nD | 0.0001 nD | 0% (distilled water) |
What is the Brix to Refractive index (nD) conversion table?
| BRIX | RI nD | BRIX | RI nD | BRIX | RI nD |
|---|---|---|---|---|---|
| 0 | 1.3330 | 30 | 1.3812 | 60 | 1.4419 |
| 1 | 1.3344 | 31 | 1.3830 | 61 | 1.4442 |
| 2 | 1.3359 | 32 | 1.3848 | 62 | 1.4465 |
| 3 | 1.3373 | 33 | 1.3866 | 63 | 1.4488 |
| 4 | 1.3388 | 34 | 1.3885 | 64 | 1.4511 |
| 5 | 1.3403 | 35 | 1.3903 | 65 | 1.4535 |
| 6 | 1.3418 | 36 | 1.3922 | 66 | 1.4558 |
| 7 | 1.3433 | 37 | 1.3941 | 67 | 1.4582 |
| 8 | 1.3448 | 38 | 1.3960 | 68 | 1.4606 |
| 9 | 1.3463 | 39 | 1.3979 | 69 | 1.4630 |
| 10 | 1.3478 | 40 | 1.3999 | 70 | 1.4655 |
| 11 | 1.3494 | 41 | 1.4018 | 71 | 1.4679 |
| 12 | 1.3509 | 42 | 1.4038 | 72 | 1.4704 |
| 13 | 1.3525 | 43 | 1.4058 | 73 | 1.4729 |
| 14 | 1.3541 | 44 | 1.4078 | 74 | 1.4754 |
| 15 | 1.3557 | 45 | 1.4098 | 75 | 1.4779 |
| 16 | 1.3573 | 46 | 1.4118 | 76 | 1.4804 |
| 17 | 1.3589 | 47 | 1.4139 | 77 | 1.4830 |
| 18 | 1.3605 | 48 | 1.4159 | 78 | 1.4855 |
| 19 | 1.3622 | 49 | 1.4180 | 79 | 1.4881 |
| 20 | 1.3638 | 50 | 1.4201 | 80 | 1.4907 |
| 21 | 1.3655 | 51 | 1.4222 | 81 | 1.4933 |
| 22 | 1.3672 | 52 | 1.4243 | 82 | 1.4960 |
| 23 | 1.3689 | 53 | 1.4265 | 83 | 1.4986 |
| 24 | 1.3706 | 54 | 1.4286 | 84 | 1.5013 |
| 25 | 1.3723 | 55 | 1.4308 | 85 | 1.5040 |
| 26 | 1.3741 | 56 | 1.4330 | ||
| 27 | 1.3758 | 57 | 1.4352 | ||
| 28 | 1.3776 | 58 | 1.4374 | ||
| 29 | 1.3794 | 59 | 1.4397 |
CLICK HERE TO DOWNLOAD KERN ORF 85BM (01) PDF MANUAL