What are the technical specifications for the HACH SL1000?
| Specification | Details |
| Dimensions (W x D x H) | 13.08 x 5.89 x 25.83 cm (5.15 x 2.32 x 10.17 in.) |
| Enclosure rating | IP54 with caps installed for the meter probe connector ports or with the probes connected to the instrument |
| Weight | 1.2 kg (2.7 lb) |
| Power requirements (internal) | Lithium-ion polymer rechargeable battery, 7.4 V, 5.0 Ah (5000 mAh) |
| Power requirements (external) | 100V–240 VAC, 50/60 Hz input, 12 V at 3.4 A output |
| Operating temperature (discharging) | 5 to 50 °C (41 to 122 °F), maximum 85% relative humidity (non-condensing) |
| Charging temperature | 5 to 45 °C (41 to 113 °F), maximum 85% relative humidity (non-condensing) |
| Storage temperature | –20 to 60 °C (–4 to 140 °F), maximum 85% relative humidity (non-condensing) |
| Interface | Mini USB port |
| 5-pin input connector | Two M12 connectors for IntelliCal probes |
| External power supply connection | 2.5 mm Jack |
| Chemkey slots | 4 |
| Data memory | 1000 measured values (result, date, time, site ID, user ID) |
| Data storage | Automatic in Press to Read measurement mode. Manual in Continuous measurement mode |
| Temperature correction (for IntelliCal probes) | Off, automatic and manual (parameter dependent) |
| Measurement display lock (for IntelliCal probes) | Press to Read and Continuous measurement mode |
| Protection class | III |
| Certifications | CE, UKCA, FCC, ISED, RCM, KC |
What are the lithium battery safety guidelines for the HACH SL1000?
Fire and explosion hazard. Lithium batteries may get hot, explode or ignite and cause serious injury if exposed to abuse conditions.
Do not use the battery if there is visible damage.
Do not use the battery after strong shock or vibration occurs.
Do not expose the battery to fire.
Keep the battery at temperatures less than 60 Celsius (140 Fahrenheit).
Keep the battery dry and away from water.
Prevent contact between the positive and negative battery terminals.
Do not let unauthorized persons touch the battery.
Discard the battery in accordance with local, regional and national regulations.
Do not use or store the instrument in direct sunlight, near a heat source or in high temperature environments such as a closed vehicle in direct sunlight.
How do I install the battery in the HACH SL1000?
Step 1. Only use the manufacturer-supplied lithium-ion rechargeable battery.
Step 2. To maintain the safety provided by the instrument enclosure, the instrument enclosure covers must be installed and secured with the supplied hardware.
Step 3. Open the battery compartment and connect the battery cable to the internal connector.
Step 4. Place the battery into the compartment and replace the cover, tightening the three screws to secure it.
How do I charge the battery on the HACH SL1000?
Step 1. Connect the external power supply to an AC mains outlet socket.
Step 2. Connect the external power supply to the power access port on the HACH SL1000.
Step 3. The instrument can be operated while the battery charges.
Step 4. Use only the external power supply specified for this instrument. External power supply substitution is not permitted.
Step 5. Do not charge the battery for more than 24 hours.
How do I navigate the HACH SL1000 keypad?
| Key | Description |
| POWER key | Push to set the meter to on or off. |
| MAIN MENU key | Select verification, calibration, routes, settings, data log, information, diagnostics, operator ID and site ID. |
| LEFT selection key (contextual) | Details, select/deselect, cancels or exits the current menu screen to the previous one. |
| RIGHT selection key (contextual) | Read samples, delete, select or confirm options, opens sub-menus. |
| Navigation keys | UP, DOWN, RIGHT, LEFT: Scroll through menus, enter numbers and letters. UP and DOWN scroll through measurements when there are more than four; RIGHT and LEFT scroll through operator and sample IDs. |
| HOME | Go to the main measurement screen. |
How do I change the language, date, and time on the HACH SL1000?
To change the language: Push Settings and then Language. Select the applicable language from the menu.
To change the date and time: Select Settings, then Meter, then Date and Time. Use the arrow keys to select the format for the date and time and then enter the current time and date information. The current data will be shown on the display and on logged measurement data.
How do I use an Operator ID and Site ID on the HACH SL1000?
Operator ID: Push Main Menu and select Operator ID. You can set the mode to on or off, select an ID from a list, create a new ID name, or delete an existing ID.
Site ID: Push Main Menu and select Site ID. Options include:
Mode: Set to manual, auto, off, or routes.
Select: Choose an ID from the list. If Show is set to Yes, you can use the RIGHT or LEFT arrows on the measurement screen to toggle between sites.
Create: Add a new site ID.
Delete: Erase an existing site ID (must be in Manual mode to delete).
Show: Set to on to display the site ID on the measurement screen.
Routes: Supplies an ordered list of site IDs.
Notes: Record special information to an event log.
How do I perform a Chemkey measurement with the HACH SL1000?
Step 1. Set the meter power to on.
Step 2. If complete traceability is necessary, enter a site ID and operator ID before measurement.
Step 3. Push the applicable Chemkeys fully, in one movement, into the slots. Always use a new Chemkey for each measurement. Do not move or touch the Chemkey once it is in a slot.
Step 4. Prepare the sample cup. If using the multi-sample cup for four samples, fill the large compartment to the fill-line (19 mL) with tap water for the detector to operate, and fill each individual sample compartment to the fill-line (3 to 5 mL) with sample water.
Step 5. Put the HACH SL1000 into the sample cup.
Step 6. Wait for the sound alert and/or the meter removal animation (within 1 to 2 seconds), then immediately remove the meter from the sample cup.
Step 7. Put the meter back into the case and wait until the measurement is completed. A progress bar will show the remaining time.
Step 8. Fully rinse the sample cup with deionized water after use.
How do I perform probe measurements with the HACH SL1000?
Step 1. Calibrate the probes before initial use. For best accuracy, calibrate probes daily.
Step 2. Ensure the date and time in the HACH SL1000 are correct before the probe is attached.
Step 3. Remove the probe connector port caps and attach the probe(s) to the M12 connectors.
Step 4. Rinse the sample flasks with sample water.
Step 5. Fill the sample flask so that the sensor is fully submerged.
Step 6. Put the probe(s) into the sample flasks and push Read if the measurement mode is set to Press to Read. Measurement values will be shown on the display.
How do I calibrate and verify probes on the HACH SL1000?
Calibration: Connect the probe, push Main Menu and select Calibrate. Select the applicable probe and follow the instructions shown on the display. For ISE probes, measure standards from low concentration to high concentration.
Verification: Push Main Menu and select Verification. Select the applicable device and follow the on-screen instructions. To change the verification standard, navigate to Main Menu > Settings > Probe > [Select Probe] > Verification Options > Standard.
How do I import data from the HACH SL1000 to Excel?
Step 1. Set the meter power to on.
Step 2. Connect the HACH SL1000 with a USB cable to a computer. The meter shows as a drive named Hach Portable Parallel Analysis.
Step 3. Open the DataTemplate folder from the drive and open the LogTemplate.xltx file.
Step 4. Right click on cell A2 and select XML and then Import.
Step 5. Select all log files from the Data folder on the HACH SL1000 drive and click Import.
Step 6. Save the file to the computer or a USB flash drive.
Step 7. Safely remove the meter from the computer using the Safely Remove Hardware utility before disconnecting the USB cable.
How do I clean and maintain the HACH SL1000?
General cleaning: Clean the exterior with a moist cloth and a mild soap solution, then wipe dry.
Parts cleaning: Use isopropyl alcohol to clean Chemkey slots, optical windows, barcode windows, and the sample detector. Use water and a mild detergent to clean the light shield. Use a cotton-tipped applicator and move it in one direction only.
Spills: Obey all facility safety protocols for spill control and discard waste according to regulations.
What are the troubleshooting steps for common HACH SL1000 errors?
| Error/Warning | Possible Cause | Solution |
| Barcode read error | Chemkey not installed correctly. | Put the Chemkey into the slot again; use a new Chemkey; or clean the barcode window. |
| Sample aspirate error | Chemkey not fully pushed into the slot. | Use a new Chemkey and ensure it is pushed all the way into the slot. |
| Used Chemkey—Replace | Chemkey used or water in the slot. | Use a new Chemkey. Dry out the slot and try again. |
| Tray leaked | Problem with the measured tray. | Replace the tray. |
| Heater Error | Tray does not operate correctly. | Contact technical support. |
| Remove from sample | Meter put into sample before it was ready. | Wait until the meter is ready before putting it into the sample. |
| Too much time in sample | Chemkeys left in sample too long. | Remove Chemkeys immediately when prompted. |
| Over Range | Concentration higher than method limit. | Dilute the sample and measure again. |
| Tray too cold | Meter too cold to heat tray. | Move to a warmer ambient temperature and measure again. |
| Sample not found | Cup not filled or conductivity too low. | Ensure cup is filled to the line and conductivity is more than 2 microS/cm; clean the detector. |
How do I perform diagnostic checks and format memory on the HACH SL1000?
Push Main Menu and select Diagnostics. You can choose from these options:
Disk Space: Shows disk space for user memory and internal memory.
Config Versions: Shows the configuration versions of meter settings.
Method Versions: Shows the method versions of the meter.
Format Memory: Erases all meter data. Note that all configuration, site IDs, operator IDs, routes, notes, and templates will be lost when this is completed.
How should the HACH SL1000 be prepared for shipping?
Step 1. Drain all water from the unit to prevent damage from freezing temperatures.
Step 2. Clean and decontaminate the instrument before shipping.
Step 3. Ship the instrument with the battery installed in the meter, but ensures the battery is not connected to the internal connector.
Step 4. Disconnect all probes and remove any Chemkeys from the slots.
Step 5. Ship the instrument in its original packaging or an alternative safe packaging.