GW INSTEK GPE-3060 (01) PDF SUMMARY:
What safety instructions should I follow when operating the GPE series?
This chapter contains important safety instructions that you must follow when operating the GPE series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition.
What are the general safety guidelines?
- Do not place any heavy object on the device.
- Avoid severe impacts or rough handling that leads to damaging the device.
- Do not discharge static electricity to the device.
- Do not block or obstruct the cooling fan vent opening.
- Do not perform measurement at circuits directly connected to Mains.
- Do not disassemble the device unless you are qualified as service personnel.
What are the safety guidelines regarding the power supply?
- AC Input voltage: 100V/120V/220V/230VAC ±10%, 50/60Hz
- Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.
What are the safety guidelines for the fuse?
- Fuse type: 100V/120V: T12A/250V 220V/230V: T6.3A/250V
- Make sure the correct type of fuse is installed before power up.
- To ensure fire protection, replace the fuse only with the specified type and rating.
- Disconnect the power cord before fuse replacement.
- Make sure the cause of fuse blowout is fixed before fuse replacement.
How should I clean the device?
- Disconnect the power cord before cleaning.
- Use a soft cloth dampened in a solution of mild detergent and water. Do not spray any liquid.
- Do not use chemicals or cleaners containing harsh products such as benzene, toluene, xylene, and acetone.
What is the recommended operation environment for the device?
- Location: Indoor, no direct sunlight, dust free, almost non-conductive pollution
- Relative Humidity: < 80%
- Altitude: < 2000m
- Temperature: 0°C to 40°C
What is the recommended storage environment for the device?
- Location: Indoor
- Relative Humidity: < 70%
- Temperature: −10°C to 70°C
How should I dispose of the instrument?
Do not dispose of this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
What are the three output modes of the GPE series regulated DC power supply?
The three output modes are independent, series tracking and parallel tracking.
How are the output voltage and current controlled in the independent mode?
In the independent mode, the output voltage and current of each channel are controlled separately.
How are the CH1 and CH2 outputs connected in tracking modes?
In the tracking modes, both the CH1 and CH2 outputs are automatically connected in series or parallel. CH1 is master and CH2 is slave; no need to connect output leads.
How does the output voltage and current change in series and parallel modes?
In the series mode, the output voltage is doubled; in the parallel mode, the output current is doubled.
What is the isolation degree of the GPE series?
The isolation degree, from output terminal to chassis or from output terminal to output terminal, is 500V.
How does each output channel work?
Each output channel works in constant voltage (CV) or constant current (CC) mode.
When is the power supply used as a CV source and when as a CC source?
For a big load, the power supply can be used as a CV source; while for a small load, a CC source.
How can the output current be controlled in CV mode?
When in the CV mode (independent or tracking mode), output current (overload or short circuit) can be controlled via the front panel.
How can the maximum output voltage be controlled in CC mode?
When in the CC mode (independent mode only), the maximum (ceiling) output voltage can be controlled via the front panel.
When does the power supply automatically cross over from CV to CC operation?
The power supply will automatically cross over from CV to CC operation when the output current reaches the target value.
When does the power supply automatically cross over from CC to CV operation?
The power supply will automatically cross over from CC to CV when the output voltage reaches the target value.
What does the front panel display show when operating in tracking mode?
When operating in the tracking mode, the power supply will automatically connect to the auto-tracking mode.
What are the performance features of the GPE series?
- Low noise: Temperature controlled cooling fan
What are the operation features of the GPE series?
- Constant Voltage / Constant Current operation
- Series Tracking / Parallel Tracking operation
- Output On/Off control
- Multi-output: Two channels V/I are adjustable, and one is fixed
- Function for locking the setting
- Output voltage/ current setting view
- Set the displayed digit resolution for the voltage & current output
What are the protection features of the GPE series?
- Overload protection
- Key misoperation protection (Lock)
- Reverse polarity protection
What is the interface feature of the GPE series?
- Remote control (Output ON/OFF)
What are the main components of the power supply?
- AC input circuit
- Transformer
- Bias power supply including rectifier, filter, pre-regulator and reference voltage source.
- Main regulator circuit including the main rectifier and filter, series regulator, current comparator, voltage comparator, reference voltage amplifier, remote device and relay control circuit.
What does the front panel of the GPE-3060 include?
- Display
- CH1 parameter display area
- CH2 parameter display area
- CH3 parameter display area
- Status display area
- LCD
- Voltage/OVP Knob
- CH1/CH2 Mode Setting
- Current Knob
- Power Switch
- CH1 Output
- GND Terminal
- CH2 Output
- CH3 Output
- Output key
- OVP/Lock key
- CH2 toggle key
- CH1 toggle key
What does the Voltmeter display?
Displays output voltage of CH1/CH2 channel.
What does the Ammeter display?
Displays output current of CH1/CH2 channel.
What do the CV/CC/OVP indicators show?
You can view the constant current, constant voltage or OVP status for Channel.
What is displayed when the channel output is OFF?
When the channel output is OFF, it is displayed as the set value.
What is displayed when the channel output is ON?
When ON, it is displayed as the output value, when there is a set operation, it is displayed as the set value, and the status bar has a display.
What does the channel indicator show?
Indicates the currently selected channel.
What does the output status of CH3 show?
The output status of CH3 is shown.
What does the overload indicator show?
When the output current is over range, the overloaded indicator will be lit on the LCD display.
How do you set the CH1/CH2 parameters?
When rotating the Voltage/Current knob clockwise, the value increases, and rotating counterclockwise, the value decreases. Click on the Encode to enter the setting. Automatically confirm and exit the set state after stopping the operation for a few seconds.
How do you set the OVP?
Press and hold the Encode to enter the OVP setting. When rotating clockwise, the value increases, and rotating counterclockwise, the value decreases. Click the Encode to enter the setting. Automatically confirm and exit the set state after stopping the operation for a few seconds.
How do you toggle between CH1 and CH2?
Click on the CH1 or CH2 button (there will be a flashing display of the corresponding channel number on the LCD) to set the voltage/current of this channel or view its readback value.
How do you activate parallel/series tracking operation?
Activate the parallel/series tracking operation with the Parallel/Series keys.
How do you enable/disable OVP and Lock key?
Click this button to enable and disable the OVP function; Long press is the Lock function, which can lock or unlock the panel buttons (except for serial and parallel operation and output key functions). The status bar displays.
How do you turn the output on or off?
Turns the output on or off with the Output Key.
How do you turn the main power on or off?
Turns On or Off the main power using the Power Switch.
What does the GND Terminal accept?
Accepts a grounding wire.
What does the CH1 Output output?
Outputs CH1 voltage and current.
What does the CH2 Output output?
Outputs CH2 voltage and current.
What does the CH3 Output output?
Outputs CH3 voltage and current.
What is the Remote Control Terminal used for?
For more information about the remote control terminal, please see page 25.
What does the Power Cord/Fuse Socket accept?
The power cord socket accepts the AC mains.
What does the fuse holder contain?
The fuse holder contains the AC mains fuse.
What does the AC Selector do?
Selects AC input voltage: 100V/ 120V/ 220V/ 230V; 50/60Hz.
How does the GPE switch between constant voltage mode (CV) and constant current mode (CC)?
According to load condition, The GPE can automatically switch between constant voltage mode (CV) and constant current mode (CC).
What happens when the current level is smaller than the output setting in CV mode?
When the current level is smaller than the output setting, the GPE series operates in Constant Voltage mode. The indicator for the corresponding channel appears on the LCD. The Voltage level is kept at the setting and the Current level fluctuates according to the load condition until it reaches the output current setting.
What happens when the current level reaches the output setting in CC mode?
When the current level reaches the output setting, the GPE series starts operating in Constant Current mode. The indicator for the corresponding channel appears on the LCD. The Current level is kept at the setting but the Voltage level becomes lower than the setting, in order to suppress the output power level from overload.
What happens when the current level becomes lower than the setting in CC mode?
When the current level becomes lower than the setting, the GPE series goes back to the Constant Voltage mode.
What is the first step of powering up the power supply?
Before powering up the power supply, select the AC input voltage from the rear panel.
What is the second step of powering up the power supply?
Connect the AC power cord to the rear panel socket.
How do you turn on the power?
Press the power switch to turn on the power. The display will first display all the LCD segments before showing settings for each channel.
How do you turn off the power?
Press the power switch again to turn off the power.
How do you connect a load cable using standard accessories?
- Turn the terminal counterclockwise and loosen the screw.
- Insert the cable terminal.
- Turn the terminal clockwise and tighten the screw.
How do you connect a banana plug?
Insert the plug into the socket.
What should you consider when using load cables other than the attached cables?
When using load cables other than the attached, make sure they have enough current capacity for minimizing cable loss and load line impedance. Voltage drop across a wire should not excess 0.5V.
How do you turn on all outputs in each channel?
Press the Output key to turn on all outputs in each channel. The icon will become lit on the LCD display.
How do you turn off all outputs?
Push the Output key again to turn off all outputs. The icon will become lit on the LCD display.
What actions automatically turn the output off?
- Change the operation mode between independent/series tracking/parallel tracking, three channels will be closed together.
- When a single channel OVP is started, the corresponding output will be closed. When CH1/CH2 is simultaneously OVP, three channels will be closed together.
What happens when you connect the GPE in series?
When connecting in series, the output voltage is twice that of a single channel.
What happens when you connect the GPE in parallel?
When connecting in parallel, the output current is twice that of a single channel.
How do you toggle the connection mode of CH1/CH2?
You can toggle the connection mode of CH1/ CH2 by using different combinations of the mode selection key.
What are the different ways to set the connection mode of CH1/CH2?
- For the independent mode, the right key is not pressed.
- Toggle to parallel mode when both keys are pressed.
- Right key is pressed and the left key is not pressed in series mode.
- When CH1 / CH2 is in the series or parallel mode, the corresponding series or parallel icon appears on the LCD display.
What is the lock function used for?
The lock function can be used when you need to keep the output constant to avoid the load from being damaged due to inadvertent operation.
How do you lock the operation in the front panel?
Press the LOCK key (for more than 2 seconds) to lock the operation in the front panel. The icon will become lit.
How do you unlock the front panel?
To unlock, press the LOCK key for more than 2 seconds. The icon will then turn off.
What is not affected by the lock operation?
The OUTPUT key is not affected by the lock operation.
What will unlock the lock function?
Series and parallel operation is not controlled by the locking key, It will unlock the lock function.
How do you set the output state at startup?
- Press and hold the Output key and turn on the power until the or icon flashes on the LCD display.
- Press the “OVP” key to select.
- Press the “ON/OFF” key to confirm.
What is the default output state at startup?
By default the output is set to OFF at startup.
How do you set the displayed digit resolution for voltage and current?
- Press and hold the “OVP” key and turn the power until on the decimal point for the CH1 voltage flashes on the LCD display.
- Press the “OVP” key to select the number of displayed digits.
- Press the “ON/OFF” key to confirm the selection.
What is the default number of displayed digits?
By default the number of displayed digits is set to 3.
How do you use the remote control terminal to turn the power on or off?
- Short pins 7 and 8 (remote control setting). This will put the power state (ON/OFF) under remote control. At this moment, the or icon flashes on the LCD display.
- Output control:
- Pin 9 & 10 Open: ON state.
- Pin 9 & 10 Short: OFF state.
What should you not do with the remote control terminal?
The remote control terminal can only be controlled by shorting (external relay or jumper shunt) /opening the pins. Voltage cannot be applied to the pins. It is strictly prohibited to short pins 5 & 7 or 6 & 8. Pin 1~6 must be set to open.
How do you operate in CH1/CH2 Independent Mode?
- Make sure the Series/Parallel key is not activated (both the and icons are off).
- Connect the load to the front panel terminals, CH1 +/−, CH2 +/−.
- Use the voltage and current knob to set the CH1 output voltage and current.
- Use the voltage and current knob to set the CH2 output voltage and current.
- Press the Output key to turn on the output. The Output key will be lit and the icon will appear on the LCD display. The or icon appears on the LCD to indicate the output status for each channel.
How do you operate in CH3 Independent Mode?
- Connect the load to the front panel CH3 +/− terminal.
- Press the Output key to turn on the output. The Output key will be lit.
What happens if the output current value is too high in CH3 Independent Mode?
When the output current value is too high, the icon appears on the LCD display and CH3 operation mode switches from constant voltage to constant current.
How do you operate in CH1/CH2 Series Tracking Mode without Common Terminal?
- Press the Series/Parallel key to activate the series tracking mode. The icon will be lit on the LCD display.
- Connect the load to the front panel terminals, CH1+ & CH2− (Single supply).
- Use the current knob to set the CH2 output current to the maximum level.
- Use the voltage and current knob to set the CH1 output voltage and current level.
- Press the Output key to turn on the output. The Output key will be lit.
- Refer to the CH1 (Master) meter and indicators for the output level and CV/CC status.
How do you determine the output voltage level in CH1/CH2 Series Tracking Mode without Common Terminal?
Double the reading on the CH1 voltage meter.
How do you determine the output current level in CH1/CH2 Series Tracking Mode without Common Terminal?
CH1 meter reading shows the output current.
How do you operate in CH1/CH2 Series Tracking Mode with Common Terminal?
- Press the Series/Parallel key to activate the series tracking mode. The icon will be lit on the LCD display.
- Connect the load to the front panel terminals, CH1+ & CH2−. Use the CH1 (−) terminal as the common line connection.
- Use the CH1 voltage knob to set the master & slave output voltage (the same level for both channels).
- Use the CH1 current knob to set the master output current.
- Use the CH2 current knob to set the slave output current.
- Press the Output key to turn on the output. The Output key will be lit.
- Refer to the CH1 (Master) meter and indicators for the output level and CV/CC status.
- Refer to the CH1/CH2 meter and CH2 indicators for the output level and CV/CC status.
How do you determine the CH1 (Master) voltage level in CH1/CH2 Series Tracking Mode with Common Terminal?
CH1 meter reading shows the output voltage.
How do you determine the CH1 (Master) current level in CH1/CH2 Series Tracking Mode with Common Terminal?
CH1 meter reading shows the output current.
How do you determine the CH2 (Slave) voltage level in CH1/CH2 Series Tracking Mode with Common Terminal?
The CH2 meter reading shows the output voltage.
How do you determine the CH2 (Slave) current level in CH1/CH2 Series Tracking Mode with Common Terminal?
The CH2 meter reading shows the output current.
How do you operate in CH1/CH2 Parallel Tracking Mode?
- Press the Series/Parallel key to activate the parallel tracking mode. The icon will be lit on the LCD display.
- Connect the load to the CH1 +/− terminals.
- Use the CH1 voltage and current knobs to set the output voltage and current. CH2 control function is disabled.
- Press the Output key to turn on the output. The Output key will be lit.
- The operating mode of CH2 will appear as the icon on the LCD display.
- Refer to the CH1 meter and indicator for the output level and CV/CC status.
How do you determine the Output voltage level in CH1/CH2 Parallel Tracking Mode?
The CH1 meter reading shows the output voltage.
How do you determine the output current level in CH1/CH2 Parallel Tracking Mode?
Double the amount of CH1 current meter reading.
Q1. I pressed the panel lock key but the output still turns on/off.
A1. For safety reasons the output key is not affected by the panel key lock feature.
Q2. The CH3 overload indicator turned on – is this an error?
A2. No, it simply means that the CH3 output current reached the maximum 5.2A and the operation mode turned from CV (constant voltage) to CC (constant current). You can continue using the power supply, although reducing the output load is recommended.
Q3. The specifications do not match the real accuracies.
A3. Make sure that the power supply is powered on for at least 30 minutes, within +20°C ~ +30°C.
How do you replace the fuse?
- Take off the power cord and remove the fuse socket using a minus driver.
- Replace the fuse in the holder.
What are the fuse ratings?
- 100V/120V: T12A/250V
- 220V/230V: T6.3A/250V.