GE AKTA GO (01) Image

Free English GE AKTA GO (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD GE AKTA GO (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

What are the prerequisites for operating the GE AKTA GO system?

In order to operate the GE AKTA GO system according to its intended purpose, it is important that the following prerequisites are followed:

• you have a general understanding of how the computer and the Microsoft® Windows® operating system work.

• you understand the concepts of liquid chromatography.

• you have read and understood the Safety instructions chapter in ÄKTA go Operating Instructions.

• a user account has been created according to UNICORN Administration and Technical Manual.


What do the safety notices in the GE AKTA GO user documentation mean?

The user documentation for the GE AKTA GO contains the following safety notices:

WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury. It is important not to proceed until all stated conditions are met and clearly understood.

CAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. It is important not to proceed until all stated conditions are met and clearly understood.

NOTICE: Indicates instructions that must be followed to avoid damage to the product or other equipment.


How do I install or replace an optional module on the GE AKTA GO?

Follow the steps below to install or replace a module or module panel on your GE AKTA GO. Module panels must be used in positions not occupied by modules.

Step 1: Disconnect power from the instrument by using the instrument power button.

Step 2: If a module is to be replaced, loosen the tubing connectors and remove the tubing from the existing module.

Step 3: Loosen the screw in the module or module panel with a Torx T20 screwdriver.

Step 4: Remove the module or module panel and disconnect the cable at the back.

Step 5: Connect the cable to the new module or module panel.

Step 6: Insert the module or module panel and fasten it with a Torx T20 screwdriver.

Note: When an optional module is removed or a new module is installed, the system configuration must be updated in UNICORN.


How do I update the system configuration in UNICORN after installing a new module on my GE AKTA GO?

When an optional module is removed or a new module is installed on your GE AKTA GO, the system configuration must be updated in UNICORN. Follow the steps below:

Step 1: On the Tools menu in the Administration module, click System Properties or click the System Properties icon to open the dialog. The System Properties dialog is displayed.

Step 2: Select the system of interest in the System Properties dialog and click Edit. The Edit dialog is displayed.

Step 3: Select the type of module installed/uninstalled from the Component types list. The available modules are shown in the Component selection list. Note: Instrument modules are referred to as Components in UNICORN.

Step 4: Select the appropriate checkbox and click OK to apply the changes. Note: If applicable, choose the appropriate option in the Property column.


How do I connect the tubing to the F9-R fraction collector for my GE AKTA GO?

Follow these steps to connect the tubing to the F9-R fraction collector:

Step 1: Remove the tubing connector from the delivery arm and insert the tubing through the tubing connector.

Step 2: Use the tube adjustment cavity on the delivery arm to expose the correct length of tubing from the tubing connector. Insert the tubing into the tube adjustment cavity and slide the connector down, towards the delivery arm.

Step 3: Re-install the tubing connector on the delivery arm.

Step 4: Adjust the position and height of the delivery arm. Use the adjustment knob to place the tube sensor in the correct position for the used tubes. See Fraction collector F9-C and F9-R Operating instructions for detailed instructions.

Step 5: Connect the tubing from the fraction collector to the Frac port on the outlet valve.

Step 6: Adjust the delay volume setting in UNICORN to the volume of the tubing. Note: The default delay volume is correct and does not need to be set if standard tubing is used.


How do I adjust the delay volume on my GE AKTA GO?

The delay volume is the retention volume that has to be dispensed before collecting the fractions indicated in the chromatogram. It corresponds to the volume in the flow path from the UV monitor to the fraction collector. If a module has been installed after the UV monitor or if the tubing dimensions have been changed, the delay volume has to be adjusted.

You can access the settings through the System Control module in UNICORN by selecting System → Settings.

In the System Settings dialog, select Fraction collection → Delay volume and type in the correct volume for:

• Detector – Frac

• Restrictor volume

• pH cell volume


How can I lock or unlock the instrument control panel on my GE AKTA GO?

The Pause and Continue buttons of the Instrument control panel on the GE AKTA GO can be locked or unlocked. To do this in UNICORN, select Advanced → Instrument control panel in the System Settings dialog, in System control.


How do I enable the power-save mode on my GE AKTA GO?

The GE AKTA GO has a power-save mode that the instrument enters after being in a Ready state for a set period. To enable Power-save in UNICORN, select Advanced → Power-save in the System Settings dialog. Note: To activate Power-save, the system must be connected and in a Ready state.


How can I change the wash settings on my GE AKTA GO?

Wash settings, including wash volumes and flow rates, are defined by default and optimized for the system. To change the volume and the flow rate for pump wash and system wash in UNICORN, select Wash settings in the System Settings dialog.


What sample application techniques are available for the GE AKTA GO?

There are three techniques for sample application in the GE AKTA GO: using a sample loop, a Superloop, or the pump. The suitable sample volumes for each technique are:

• Sample loop injection: 25 µL to 10 mL

• Superloop injection: 1 to 150 mL

• Pump injection: > 5 mL

The pros and cons for each technique are as follows:

Sample application technique Pros Cons
Sample loop injection

• Only option for small sample volumes (25 µL to 5 mL)

• No sample in the pump or mixer

• All sample is loaded onto the column

• Manual loading of sample loop from syringe

• Need to empty the loop 3 to 5 times the loop volume to load all sample. Sample is diluted.

Superloop injection

• Suits repeated application of the same sample

• No sample dilution

• Little sample loss for medium volumes

• No sample in the pump or mixer

• Transparent graded sample container, possible to check quality and monitor injected volume

• Manual loading of the Superloop

• Needs to be cleaned after use

Pump injection

• Loading of large amount of sample

• Easy (only requires priming of the sample inlet with correct sample)

• Possibility to use air sensor to load all sample

• No extra equipment needed

• Not possible to load small sample volumes due to the volume in the flow path

• Need to finalize sample application, to load all sample in the system from inlet valve to column

• Not possible to load the sample left in the inlet tubing without an air sensor

• Sample has to travel through the entire flow path up to the column, which has to be cleaned after use


How do I connect, prepare, and fill a Superloop for sample application on the GE AKTA GO?

Prepare the Superloop:

To avoid injecting air into the flow path of the instrument, the Superloop should be prefilled with buffer manually before attaching it to the system. See the corresponding Superloop instructions for information on how to assemble and prefill a Superloop with buffer.

Connect the Superloop:

Step 1: Attach the Superloop to the instrument using a column holder.

Step 2: Connect the top tubing from the Superloop to the LoopE port on the injection valve.

Step 3: Connect the bottom tubing from the Superloop to the LoopF port on the injection valve.

Fill the Superloop:

Step 1: In the Process Picture, make sure that the injection valve is in position Load. (Note: Load is the default position for the valve.)

Step 2: Fill a syringe with sample.

Step 3: Connect the syringe to the Syr port in the injection valve. (Note: Make sure that there is no air in the syringe before connecting it.)

Step 4: Load the sample by emptying the syringe into the injection valve.

Step 5: Disconnect the syringe and plug the Syr port with a stop plug. (Note: Always plug the Syr port on the injection valve with a stop plug after loading a Superloop.)


How do I use pump injection for sample application on my GE AKTA GO?

Use the pump to load large volumes of sample (recommended minimum 5 mL) onto the column using the Sample inlet on inlet valve K9.

To prepare for pump injection, prime the sample manually to the inlet valve. For instructions, refer to ÄKTA go Operating Instructions.

Make sure to load the sample remaining in the flow path between the inlet valve and the injection valve onto the column. This is done by pumping buffer from the A inlet onto the column after loading the sample. When using predefined methods, this is done by clicking the “Finalize Sample injection” tick box.

For multiple sample injections, you can install an optional sample inlet valve and connect it to the Sample inlet of valve K9. Load all sample in the flow path by selecting the “Finalize Sample injection” option in the predefined methods.


How do I use an air sensor for complete sample loading on my GE AKTA GO?

For complete sample loading, place the air sensor on the sample tubing connected to the Sample inlet on inlet valve K9. Install the air sensor at a minimum distance of 150 mm from the inlet valve when working at maximum flow rate (100 mm is sufficient for flow rates of 5 mL/min or less). Disable the air sensor alarm in the method and program it so that when air is detected, sample loading stops and the remaining sample is loaded onto the column with buffer from the A inlet. This is done by selecting “Inject all sample using air sensor” and then “Finalize sample injection” in the predefined method.


How can I use the air sensor to detect when a buffer is finished on my GE AKTA GO?

To detect when a buffer is finished, place the air sensor between inlet valve K9 and the pump. The air sensor alarm needs to be activated. This is done by default in predefined methods. For a manual run, activate the “Air sensor alarm” in System Settings. After replacing the empty buffer bottle, prime the inlet tubing and perform a pump wash to remove the air from the flow path before continuing the run.


What is the difference between system wash and pump wash on the GE AKTA GO?

System wash: This function fills the system with a chosen buffer, for example, to change buffer concentration before an elution step. It consists of two parts: a 15 mL wash at a default flow rate of 25 mL/min, followed by a 6 mL wash at the same flow rate as the prior run. During a system wash, any running method is set to Hold.

Pump wash: This function is used to wash the pump and the flow path between the inlet valve and the injection valve. By default, it uses 65 mL of buffer at a flow rate of 65 mL/min to remove sample from this section of the flow path.


How does pressure control work on the GE AKTA GO and how do I enable it?

Pressure control regulates the flow rate if the pressure approaches a defined limit, which can prevent method stops due to pressure alarms. This is useful during sample application if the sample’s viscosity is higher than the buffer’s.

To enable pressure control when creating a method, select “Control to avoid Overpressure” in the Method Settings phase in the Method Editor.

To enable pressure control during a manual run, select Manual instructions → Pump → Flow in System Control.

The PI (proportional-integral) parameters for pressure control can be optimized in System Settings → Advanced → Pressure control parameters.


What considerations should I take when performing a run in a cold environment with my GE AKTA GO?

When performing a run in a cold environment, pressure can increase by up to 60% compared to room temperature due to increased liquid viscosity. To protect the column, you should decrease the flow rate and set appropriate pressure limits. To automatically decrease the flow rate, select “Reduce for Cold Room” in the Method Settings phase in the Method Editor. This option reduces the flow rate to 50% of the previously set flow rate.


How do I set the conductivity temperature compensation on my GE AKTA GO?

The conductivity temperature compensation factor can be set in System Settings to ensure a stable conductivity signal, especially during runs at different temperatures or in a refrigerator. By default, it is set to 2.0%.

Follow these steps to set the compensation factor:

Step 1: Open the System Settings dialog by selecting System → Settings in the System control module.

Step 2: Select Conductivity → Cond temp compensation.

Step 3: Set the Compensation factor to the desired value.

Note: Do not set the conductivity temperature compensation to zero, as this can generate a fluctuating signal. When running performance tests, a compensation factor of 2.1% should be used.


When should I run performance tests on my GE AKTA GO?

Performance tests on the GE AKTA GO can be run at any time as required. The following table provides recommendations for when to run performance tests:

Occasion Test
After system installation Run the System test and Mixer test. If you have installed additional modules that have performance tests, run the corresponding tests as well.
After module installation Run the performance test applicable to the installed module. (Not all modules have an applicable performance test.)
If a problem is suspected Run the performance test applicable to the module or modules causing the problem.
After a prolonged stop Run the System test and Mixer test. If you have installed additional modules that have performance tests, run the corresponding tests as well.
If the System test fails due to the pump. Run the System test or the Pump test after troubleshooting.

What should I do before running a performance test on my GE AKTA GO?

Before running a performance test on your GE AKTA GO, the following aspects should be considered:

• Disconnect all columns from the flow path. Columns should not be used when running a performance test.

• Make sure that the correct path length is set for the UV flow cell. The correct path length is set by default, but has to be set if the flow cell is changed to a 5 mm flow cell.

• Set the temperature compensation factor to 2.1%, which is suitable for a solution of 1 M NaCl 1% acetone.


What should I do if a performance test fails on my GE AKTA GO?

If a performance test fails on your GE AKTA GO, review the test report to check which part of the test failed and consult the troubleshooting sections for possible causes and corrective actions. If a failed performance test cannot be corrected by the suggested actions or if it indicates a malfunction not listed, the module may need repair or replacement. After performing corrective actions, verify the fix by rerunning the test.


What is the maintenance schedule for the GE AKTA GO?

The user should perform the following periodic maintenance on the GE AKTA GO:

Interval Maintenance action
Daily or before each run Calibrate the pH monitor
Weekly Change pump rinsing solution (daily if aqueous buffers are used)
Calibrate the pressure sensors
Clean the fraction collector sensor
Monthly Check the flow restrictor
Every 6 months Clean the UV flow cell
Replace the pH electrode

Additionally, the following maintenance should be performed when required:

• Clean the instrument surfaces

• Run System CIP (System cleaning-in-place)

• Clean and store the pH electrode

• Clean the conductivity flow cell

• Calibrate the conductivity monitor

• Calibrate the temperature sensor

• Reset the pressure sensors

• Replace tubing and connectors

• Replace the UV flow cell

• Replace pump rinsing system tubing

• Replace main fuses


How do I exchange the online filter on my GE AKTA GO?

Exchange the online filter every week, or when required (e.g., if clogged). Use forceps and gloves to avoid contamination.

Step 1: Unscrew the nut from the online filter.

Step 2: Remove the old filter (1) using forceps. Replace the support net (2) if damaged.

Step 3: Mount the new filter. Carefully push the filter onto the holder (3).

Step 4: Refit the nut and tighten the online filter by hand.


How do I clean or replace the mixer membrane and O-ring on my GE AKTA GO?

Clean the mixer membrane if the mixer performance test fails or if system pressure is high. Replace the membrane if the problem persists, and replace the O-ring if damaged. Use gloves to avoid contamination.

Step 1: Disconnect all tubing to the mixer.

Step 2: Unscrew the top section of the mixer.

Step 3: Remove the O-ring from the top section.

Step 4: Remove the inner components of the mixer (membrane and O-ring).

Step 5: Clean the membrane (1) based on usage:
• For biological agents: Immerse in 20% ethanol in an ultrasonic bath for a few minutes, then repeat with distilled water.
• For salt buffers: Rinse with distilled water.

Step 6: Remove the O-ring (2) from the bottom section.

Step 7: Wet the new O-ring with 20% ethanol and fit it into position.

Step 8: To reassemble, hold the top section upside-down and place the membrane on top, fitting the indentations to the protrusions.

Step 9: Place the bottom section on top of the membrane, ensuring alignment.

Step 10: Screw the metal section onto the inner components and connect the tubing.

Note: Incorrect assembly can cause leaks.


How do I clean the UV flow cell on my GE AKTA GO?

Clean the UV flow cell every six months or when required. Do not allow solutions to dry out in the flow cell.

Step 1: Disconnect all tubing from the UV monitor.

Step 2: Replace the top fingertight connector with a Luer connector and connect a piece of waste tubing to the bottom of the UV monitor, directing it to a waste container.

Step 3: Fill a syringe with distilled water and connect it to the Luer connector.

Step 4: Inject the distilled water through the UV flow cell carefully. Disconnect the syringe.

Step 5: Fill a syringe with a 10% surfactant detergent solution. (Tip: Heat the solution to 40°C to improve cleaning.)

Step 6: Connect the syringe and inject the detergent solution. Repeat this step about five times.

Step 7: Disconnect the syringe and leave the detergent solution in the flow cell for at least 20 minutes.

Step 8: Fill a syringe with distilled water and connect it to the Luer connector.

Step 9: Inject the distilled water to rinse the flow cell. Disconnect the syringe.

Step 10: Disconnect the Luer connector and waste tubing. Reconnect the original tubing to the UV monitor.


How do I replace the UV flow cell on my GE AKTA GO?

Follow these steps to replace the UV flow cell. Ensure the used cell is thoroughly flushed with a bacteriostatic solution (e.g., NaOH) and distilled water before service.

Step 1: Switch off the instrument.

Step 2: Disconnect the tubing from the UV flow cell.

Step 3: Unscrew the wheel at the bottom of the UV monitor and press it upwards to release the flow cell.

Step 4: Pull the flow cell upwards out of the monitor. Hold it by the top part with the O-ring and do not touch the optical surfaces.

Step 5: Insert a new flow cell into the monitor.

Step 6: Tighten the wheel firmly.

Step 7: Connect the tubing to the new flow cell.

Step 8: Switch on the instrument.

Step 9: If a cell with a different path length is used, update the UV flow cell path length in the Calibrate dialog in System Control. If a real cell path length is needed, calibrate the new cell instead of using the nominal value.

After replacement, run a system performance test to verify proper function.


How do I replace tubing and connectors on my GE AKTA GO?

Replace tubing and connectors when required, for example if a tubing is clogged or kinked.

Step 1: Unscrew the connectors and disconnect the tubing.

Step 2: If the old tubing has labels, remove them to use with the new tubing.

Step 3: Cut the new tubing to the same length as the old tubing using a tubing cutter for a straight angle cut. (CAUTION: The tubing cutter is very sharp.)

Step 4: Put the old labels on the new tubing.

Step 5: Mount the new connectors on the tubing.
• For fingertight connectors, slide the connector onto the tubing.
• For 1/8″ tubing connectors, slide the connector onto the tubing, then slide the ferrule onto the tubing with the thick end towards the tubing end.

Step 6: Insert the tubing with the connector into the port, ensuring it is fully inserted to the bottom.

Step 7: Tighten the connector fully. Use the fingertight wrench for areas that are difficult to access.


How do I troubleshoot my GE AKTA GO if it does not turn on?

If your GE AKTA GO instrument does not turn on, check the following possible causes and perform the corresponding actions:

Possible cause Action
Power cord not connected Connect the power cord to the wall outlet and to the electrical inlet on the instrument. Make sure that the cord is attached using the clip to prevent detachment.
No electric current in the wall outlet Make sure that there is electric current in the wall outlet.
Broken fuse in the instrument Replace the fuse as described in the ÄKTA go Operating Instructions.
Overheated instrument Switch off the instrument and wait until the temperature has decreased before restarting it. If the problem persists, generate a System error report and contact Service.

What should I do if the internal temperature of my GE AKTA GO is high?

If the GE AKTA GO has a high internal temperature, consider these causes and actions:

Possible cause Action
Insufficient instrument ventilation Make sure that the air ventilation on the back of the instrument is not covered. There should be at least 10 cm clearance at the back to allow adequate air circulation.
Hot surroundings Decrease the room temperature. The maximum operating temperature is 35°C.
Faulty fan Contact Service.

How can I troubleshoot leakage issues on my GE AKTA GO?

If you observe leakage around a connector on your GE AKTA GO, follow these troubleshooting steps:

Possible cause Action
Connector not tightened Tighten the connector. If necessary, replace the connector.
Worn connector Unscrew the connector to check it. If necessary, replace the connector.
Crystallized material around the connector Unscrew to clean the connector. If necessary, replace the connector.

What should I do if I experience high pressure or a pressure alarm on my GE AKTA GO?

If you encounter higher than expected pressure or a pressure alarm on your GE AKTA GO, consult the following table for causes and solutions:

Possible cause Action
Pressure increase due to viscosity (e.g. samples or run in a cold environment) Lower the flow rate or activate Pressure Control.
Pressure alarm incorrectly set Check the pressure alarm setting in System Settings.
Clogged online filter Replace the online filter.
Tubing i.d. changed Be aware that the back pressure for a 0.25 mm i.d. tubing is 16 times higher than for a 0.50 mm i.d. tubing under the same conditions.
Blocked flow path Remove obstructions from the flow path. For example, remove stop plugs and replace constricted tubing.
Inappropriate pressure control parameter selected (when pressure control is active) Refer to the section on Pressure control for correct parameter selection.

How do I troubleshoot the pump on my GE AKTA GO if it shows erratic flow or pressure?

If you experience erratic flow, pressure, or other pump-related issues with your GE AKTA GO, use the following guide to troubleshoot:

Problem Possible cause Action
Erratic flow or pressure Air trapped in the pump heads Purge the pump according to the Operating Instructions.
Buffer finishing Check that there is sufficient buffer volume in the flask. Prime the inlets.
Bad piston seal Replace the piston seal and rinse membrane.
Bad piston Replace the piston.
Bad check valve Replace the check valves.
Small pressure pulsations that remain after a proper purge Air bubble stuck in the pump seal Run 100% methanol at >5 mL/min and >0.5 MPa for a few minutes. Alternatively, run water at 25 mL/min and >2 MPa for 1 to 2 hours.
Liquid leaking between the pump head and the side panel Piston seal or rinsing membrane incorrectly fitted or worn Replace or reinstall the piston seal or membrane.
Strange noise from the pump Piston spring damaged or incorrectly mounted Disassemble the pump head and examine the piston spring.

What should I do if the UV monitor on my GE AKTA GO is not working correctly?

For issues with the UV monitor on your GE AKTA GO, refer to this troubleshooting table:

Problem Possible cause Action
No UV signal The lamp is turned off Turn the lamp on in the Manual instructions dialog, in Monitors → UV lamp.
Sharp dips or peaks in the UV signal. Ghost peaks Air in the UV flow cell due to missing flow restrictor Add the flow restrictor. If a pH valve is installed, set the flow restrictor in-line via Manual instructions.
Noisy UV signal Dirt in the UV flow cell Clean the UV flow cell.
Dirt in the flow path Perform a System CIP and Column CIP as per the Operating Instructions.
The UV lamp is broken or worn out. Contact Service.
Broken UV flow cell Replace the flow cell.
Auto zero out of accepted range Wrong UV flow cell for current buffer Change to a shorter UV flow cell or change buffer.
Incorrectly installed UV flow cell Check that the UV flow cell is fitted correctly.
Broken UV flow cell Replace the flow cell.
Distorted protein peaks in IEX gradients The refractive index of the buffer changes rapidly and disturbs the shape of protein peaks in the 2 mm flow cell Run with reversed flow direction: connect the inlet tubing at the bottom and the outlet tubing at the top of the flow cell.

How do I troubleshoot issues with the conductivity monitor on my GE AKTA GO?

If you are experiencing problems with the conductivity monitor on your GE AKTA GO, use the following guide:

Problem Possible cause Action
Unstable conductivity Air bubbles in the conductivity flow cell, due to missing flow restrictor Add the flow restrictor to the flow path. If a pH valve is installed, set the flow restrictor in-line in the Manual instructions dialog.
Air or dirt in the conductivity flow cell Clean the conductivity flow cell.
Increasing/decreasing conductivity measurement with the same buffer over time Dirt in the conductivity flow cell Clean the conductivity flow cell.
Changes in ambient temperature Set the temperature compensation factor in System Settings → Conductivity → Cond temp compensation.
Uncalibrated conductivity monitor Check the calibration with a solution with known conductivity. Calibrate the conductivity monitor.
A “knee” in the beginning of a linear gradient B inlet not properly primed Prime the B inlet.
Fluctuating conductivity during a gradient Missing mixer or missing membrane in the mixer Check or replace the membrane in the mixer.
Buffers used are too difficult to mix Change buffers or use an external large mixer.

How do I troubleshoot the pH monitor and pH valve on my GE AKTA GO?

For issues related to the pH monitor and pH valve on your GE AKTA GO, refer to the table below:

Problem Possible cause and action
Leakage pH or dummy electrode not properly installed in the valve. Remove the dummy electrode and wet it properly with distilled water. Insert the dummy electrode into the pH valve and rotate it before securing it with the nut.
No or strange pH signal Dummy electrode installed. Replace the dummy electrode with a pH electrode.
Electrode cable not connected properly. Connect the electrode cable.
Faulty or old electrode. Replace the electrode.
Incorrect pH measurement pH not properly calibrated. Calibrate the pH monitor.
Bad pH electrode. Replace the electrode.
Dirty pH electrode. Regenerate the pH electrode. Place the electrode in water for 30 minutes followed by 30 minutes in a buffer with pH 4. If the problem persists, replace it.
Not possible to inject calibration solution Blocked waste tubing or W3 outlet. Check connector and tubing.
Alarm in UNICORN: (Alarm) The pH cell can only be run at pressure below 0.8 MPa. Post-column pressure is too high. Check the tubing and/or lower the flow rate.

How do I execute manual instructions on my GE AKTA GO?

To manually interact with the GE AKTA GO system using Manual instructions, follow the steps below:

Step 1: In the System Control module, select Manual → Execute Manual Instructions or use the shortcut Ctrl + m.

Step 2: In the Manual instructions dialog that opens, do the following:
a. Click the + symbol to expand the instruction group you want to modify.
b. Select the specific instruction.
c. Enter the new values for the instruction.

Step 3: To execute several instructions at once, select and edit an instruction, then click Insert. Repeat for additional instructions.

Step 4: To update parameter fields during a method run, check the “Auto update of parameters during run” box.

Step 5: To perform the instructions, click Execute.


How do I change the Node ID for a module on my GE AKTA GO?

To check or change the Node ID of a module on your GE AKTA GO, follow these steps:

Step 1: Remove the module according to its respective instruction.

Step 2: The Node ID is set by the position of an arrow on one or two rotating switches at the back of the module.
a. The first switch, labeled A, sets the tens digit.
b. The second switch, labeled B, sets the units digit.
For example, to set a Node ID of 6, switch A should be set to 0 and switch B to 6.

Step 3: Check the current Node ID and compare it with the listed Node IDs for core and optional modules.

Step 4: To change the Node ID, use a screwdriver to turn the arrows of the switches to the desired numbers.

Step 5: If applicable, re-install the module in the instrument.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD GE AKTA GO (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.