FREE ENGLISH FLUKE ESA710 (01) PDF USER GUIDE
FREE ENGLISH FLUKE ESA710 (01) PDF USER MANUAL
FREE ENGLISH FLUKE ESA710 (01) PDF OWNER GUIDE
FREE ENGLISH FLUKE ESA710 (01) PDF OWNER MANUAL
FREE ENGLISH FLUKE ESA710 (01) PDF REFERENCE GUIDE
FREE ENGLISH FLUKE ESA710 (01) PDF INSTRUCTION GUIDE
FREE ENGLISH FLUKE ESA710 (01) PDF REFERENCE MANUAL
FREE ENGLISH FLUKE ESA710 (01) PDF INSTRUCTION MANUAL
FREE ENGLISH FLUKE ESA710 (01) PDF OPERATING INSTRUCTIONS


CLICK HERE TO DOWNLOAD FLUKE ESA710 (01) PDF MANUAL

If this is not the document you want for this product, click here to see if we have any other documents for this product.

How should the instrument be unpacked and inspected upon receipt?

Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.


What should be done if physical damage is found upon delivery?

If physical damage is found upon delivery, retain all packing materials in their original condition and contact the carrier immediately to file a claim.


What should be done if the instrument is delivered in good physical condition but does not operate within specifications, or if there are other problems not caused by shipping damage?

If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, contact Fluke Biomedical or your local sales representative.


What is the return procedure for instruments?

All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to the factory location. When you return an instrument to Fluke Biomedical, it is recommended to use United Parcel Service, Federal Express, or Air Parcel Post. It is also recommended that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.

Use the original carton and packaging material for shipment. If they are not available, the following guide for repackaging is recommended:

Use a double-walled carton of sufficient strength for the weight being shipped.

Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.

Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.


What is required for returning a product for partial refund/credit?

Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from the Order Entry Group at 1-440-498-2560 or orders@flukebiomedical.com.


How often should the product be calibrated?

To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once every 12 months.


Who should perform the calibration?

Calibration must be done by qualified personnel.


What are the general safety warnings for using the product to prevent possible electrical shock, fire, or personal injury?

Read all safety information before you use the product.

Carefully read all instructions.

Do not connect the product to a patient or equipment connected to a patient. The product is intended for equipment evaluation only and should never be used in diagnostics, treatment or in any other capacity where the product would come in contact with a patient.

Do not alter the product and use only as specified, or the protection supplied by the product can be compromised.

Do not use the product around explosive gas, vapor, or in damp or wet environments.

Use this product indoors only.

Replace the mains power cord if the insulation is damaged or if the insulation shows signs of wear.

Use only the mains power cord and connector approved for the voltage and plug configuration in your country and rated for the product.

Examine the case before you use the product. Look for cracks or missing plastic. Carefully look at the insulation around the terminals.

Do not put the product where access to the mains power cord is blocked.

Connect an approved three-conductor mains power cord to a grounded power outlet.

Do not use the product if it is altered or damaged.

Do not use the product if it operates incorrectly.

Do not use test leads if they are damaged. Examine the test leads for damaged insulation.

Do not touch metal parts of the device under test (DUT) during analysis. The DUT should be considered an electrical shock hazard when connected to the product as some tests involve high voltages, high currents, and/or the removal of DUT earth bond.

Use the correct terminals, function, and range for measurements.

Use Product-approved Measurement Category (CAT), voltage, and amperage-rated accessories (probes, test leads, and adapters) for all measurements.

Do not apply more than the rated voltage, between the terminals or between each terminal and earth ground.

Do not exceed the Measurement Category (CAT) rating of the lowest-rated individual component of a product, probe, or accessory.

Remove all probes, test leads, and accessories that are not necessary for the measurement.

Keep fingers behind the finger guards on the probes.

Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc.

Turn the product off and remove the mains power cord. Stop for two minutes to let the power assemblies discharge before you open the fuse door.

Do not operate the product with covers removed or the case open. Hazardous voltage exposure is possible.

Use only specified replacement parts.

Use only specified replacement fuses.

Have an authorized service provider repair the product.

Do not use the 15 A cable to power devices in excess of 15 A. Doing so may overload the installation.

Do not use in close proximity to strong magnetic fields (for example an MRI machine).

Do not touch the applied part/ECG terminals when adjacent red warning indicators are blinking or continuously illuminated. These terminals source potentially hazardous voltage in these conditions.

Remove the null post adapter from the Ø/Null terminal after a test lead zero is performed. The Ø/Null terminal becomes potentially hazardous during some of the test conditions. Use only cables with correct voltage ratings.


What caution should be observed to ensure the product operates correctly?

Measure a known voltage first to make sure that the product operates correctly.


What do the symbols on the product and in the user manual mean?

Symbol Description
WARNING. RISK OF DANGER.
WARING. HAZARDOUS VOLTAGE. Risk of electric shock.
Consult user documentation.
Power button
FUSE_SYMBOL Fuse
WEEE_SYMBOL This product complies with the WEEE Directive and its marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Do not dispose of this product as unsorted municipal waste. For information about take-back and recycling programs available in your country, consult relevant local authorities or programs.

Note: Actual symbols for Fuse and WEEE may vary. Please refer to the product marking.


What are the general abbreviations used in the user manual?

Abbreviation Description
A Ampere
AC Alternating current, rms
AC + DC True rms sum of ac and dc current
AP Applied part/parts
BPM Beats per minute
BrPM Breaths per minute
DC Direct current
DUT Device under test
ESA Electrical safety analyzer
GFCI Ground fault current interrupter
L Mains, live conductor
mΩ Milliohm
MΩ Megaohm
N Mains, neutral conductor
P2P Point to point
PE Protective earth
SIM Simulation
V Volt
µA Microampere
Ohm

What are the schematic diagram abbreviations used in the user manual?

Abbreviation Meaning
AP Applied part
AP_F/E Applied parts: floating/earthed
AP_SEL Applied part selection
CP Conductive part
DUT Device under test
DUT_L Device under test, hot conductor
DUT_N Device under test, neutral conductor
DUT_PE Device under test, protective earth
EACP Earthed accessible conductive part
FE Functional earth
L Mains, live conductor
MAP_TRANS Mains on applied part transformer
MD Leakage current measuring device
MΩ Insulation resistance measuring device
N Mains, neutral conductor
NEACP Non-earthed accessible conductive part
PE Protective earth
REL_N Neutral: open/closed
REL_PE Earth: open/closed
REL_POL Polarity: normal/reversed
TL Test lead
Resistance measuring device

What are the differences between the ESA710, ESA712, and ESA715 models?

ESA710: The ESA710 is a model with limited functionality when it comes to electrical safety standards selectable.

ESA712: The ESA712 does not synchronize with OneQA. Procedures can only be run remotely from OneQA, with the ESA712 connected to the computer via USB cable.

ESA715: The ESA715 has all functions and features.

Feature ESA710 ESA712 ESA715
Electrical safety standards NFPA 99 / AAMI ES1 only IEC 60601-1, IEC 62353, NFPA 99 / AAMI ES1, AS/NZS 3551, EN 50678 / EN 50699 IEC 60601-1, IEC 62353, NFPA 99 / AAMI ES1, AS/NZS 3551, EN 50678 / EN 50699
Run procedures on analyzer screen
Run procedures remotely

What are the components on the front side of the product?

Item Description
1 Applied parts posts with LEDs
2 Input jacks with LEDs
3 Touch screen display
4 On/off Power button
5 High voltage indicator

What are the components on the back side of the product?

Item Description
6 Carrying handle (removable)
7 Line fuse

What are the components on the left side of the product?

Item Description
8 USB-C for PC communication
9 AC power input

What are the components on the right side of the product?

Item Description
10 2 × USB-A for peripherals
11 Equipment outlet (region dependent)

What are the standard accessories included with the product?

Item Description
12 Power cord (region dependent)
13 Test lead kit (region dependent)
14 USB cable
15 Null post adapter

What optional accessories can be used with the product?

Examples of other accessories, using USB, that can be used with the product:

Barcode scanner, for scanning asset IDs.

WiFi adapter, enables synchronization of procedures, results, assets and users via WiFi with OneQA.

Information about optional accessories can be found at http://www.flukebiomedical.com.


What are the power safety instructions for the analyzer?

To prevent possible electric shock, fire, or personal injury:

Connect the factory supplied three-conductor line power cord to a properly grounded power outlet.

Do not use a two-conductor adapter or extension cord; this will break the protective ground connection.

Do not touch the applied part/ECG terminals or the device under test when adjacent red warning indicators are illuminated. These terminals source potentially hazardous voltages in these conditions.

Caution: Connect the analyzer to a properly grounded, three-prong outlet. The Analyzer will not properly test a device under test (DUT) when the ground lead is open.

The analyzer is intended for use with single-phase, grounded power. It is not intended for dual, split-phase, or three-phase power configurations. It can, however, be used with any power system that is grounded and supplies the correct voltages for single-phase configurations.


How do I get started with the analyzer?

1. Connect the analyzer with the included power cord to a grounded outlet.

2. Press the power button to turn the analyzer on. The analyzer will show the home screen after successfully completing all self-tests.

3. The analyzer is now ready for making measurements and running procedures.


What happens during the analyzer’s start-up?

During start-up, the analyzer performs a self-test, and checks the ac mains input for proper polarity (when applicable), ground integrity, and voltage level.

The high voltage indicator illuminates briefly during the start-up. If the ground is open, the analyzer shows this fault. If the analyzer model requires a certain polarity and this is not fulfilled, the user will get the option to reverse the polarity in the analyzer.

All LEDs will illuminate during the start-up. If any LED does not illuminate, contact technical support.


What tests can the analyzer run on battery without mains connected?

The analyzer can run the following tests on battery, without mains connected: Point to point, protective earth resistance, ECG and respiration simulations.


What are the main options available from the home screen?

At the home screen, there are two options:

For running procedures, see “Procedures”.

For measurement sessions, see “Measurement types”.


What are the different measurement types the analyzer can perform?

The analyzer can perform several different types of measurements, depending on analyzer model and selected standard:

Mains voltage (Line voltage)

Protective earth resistance (Ground wire resistance)

Insulation resistance

Equipment current

Touch current

Earth leakage current (Ground wire leakage current)

Direct equipment leakage

Alternative equipment leakage

Patient leakage current (Lead to ground)

Mains on applied part leakage (Lead isolation)

Direct applied part leakage

Alternative applied part leakage

Point to point: voltage, leakage and resistance

Simulated ECG waveforms and respiration


What are the key elements of the measurement screen?

1: Tabs with measurement types.

2: Measurement variant selection, when applicable.

3: Applied part configuration, when applicable.

4: Measurement settings, when applicable.

5: Measurement value.

6: Measure button, when applicable.

7: Equipment outlet configuration, when applicable.

8: Store the value in the ongoing measurement session.

9: Access the summary screen.


How is the applied part configuration handled for measurements?

For some measurement types, the applied part posts can be used. The applied part posts can be activated and grouped in many ways. Learn more in “Custom applied parts setup”.


What user-selectable measurement options might be available?

Depending on the selected measurement type, there may be some user-selectable options, such as leakage current mode (ac, dc or ac + dc), applied part posts (floating or earthed), voltage and duration.


How can the equipment outlet be configured for measurements?

Depending on the selected measurement type and variant, the equipment outlet may be configured in the following ways:

Polarity, normal or reversed

Neutral, open or closed

Earth, open or closed


When is it necessary to zero or null test leads?

The protective earth resistance and point to point resistance measurements require test leads to be nulled or zeroed.


How do I zero test leads?

To zero a test lead:

1. Make sure leads are inserted into the proper jack.

For protective earth resistance: Connect the test lead to the V/Ω/A jack. Connect the loose end of the lead to Ø/Null.

For point to point resistance: Connect test leads to both input jacks. Connect the loose ends of the test leads to each other.

2. Tap the zeroing button on the screen.

The screen will also guide you on proper zeroing procedures.

Note: Use supplied null post adapter when zeroing the test lead with an alligator clip.


How can I create a custom applied part setup?

To create a custom applied part setup:

1. Select a measurement type with applied part posts configuration.

2. Tap the applied part symbol at the top of the screen to open the applied parts setup screen.

3. You can create your own setup by editing group names, types, number of leads and indicate whether adapters will be used or not.

4. Go back to the measurement screen when done.

Up to three groups of applied parts can be tested at the same time. The groups have five, three and two posts, respectively. If a group is selected all posts from that group are active, independent of how many posts are configured. By merging two or three groups together, it is possible to measure leakage current with up to ten posts in one group.

Note: Refer to the test standard when deciding the type of the applied parts and how they should be grouped for testing.


How is the 1-to-10 adapter used?

The 1-to-10 adapter, an optional accessory, increases the number of lead or applied parts connections. The adapter connects a maximum of 10 leads together into one lead that is connected to one of the input jacks of the analyzer. The other input jacks can also be used in conjunction with the adapter. Even more leads can be added with multiple 1-to-10 adapters.


How are measurement sessions handled?

When a measurement value is displayed, you can tap the store button to store that measurement to the current measurement session.

In the upper right corner, the number of measurements stored to the current measurement session is visible. Tap that symbol to go the summary screen for the measurement session. From the summary screen, you can complete the session to store it as a result.


How do I measure mains voltage (Line voltage)?

Mains voltage (Line voltage) measures the voltage on the mains input through three separate measurements of the electrical connection between the mains voltage supply and the analyzer.

To measure mains voltage:

1. Select the V tab.

2. The measured values are continuously updated on the screen.


How do I measure protective earth resistance (Ground wire resistance)?

The Protective earth resistance (Ground wire resistance) measures the impedance between the protective earth terminal of the equipment outlet and the exposed conductive parts of the device under test that are connected to the protective earth of the device under test.

To measure protective earth resistance:

1. Make sure the power cord from the device under test is plugged into the analyzer’s equipment outlet.

2. Select the Ω tab.

3. Connect a test lead to the V/Ω/A jack.

4. Connect the test lead to an earthed accessible conductive part of the device under test.

5. The measured value is continuously updated on the screen.

Notes:

The zero level is important for this test. For more information on proper zero adjustment, see “Zeroing test leads”.

A negative resistance measurement indicates test leads need to be zeroed again.

A low resistance reading is required to confirm a good ground connection through the power cord. Refer to the appropriate electrical safety standard for the specific limit.


How do I measure insulation resistance?

There are several insulation resistance measurement variants. The resistance is measured as:

Mains (L & N) to protective earth

Applied part to protective earth

Mains to applied part

Mains to non-earthed accessible conductive part

Applied part to non-earthed accessible conductive part

All insulation resistance tests can be done using 50 V, 100 V, 250 V, or 500 V dc, at a duration of 1 to 60 seconds.

The insulation resistance measurement is a timed test. The red indicator lights will turn off when the measurement is complete and the applied part posts become safe to handle.

To measure insulation resistance:

1. Select the MΩ tab.

2. Select the desired variant from the drop down menu. LEDs on the analyzer indicate connections for the selected variant.

3. To change the voltage or duration, tap the button that shows voltage and duration in the upper right corner of the screen.

4. Tap the measure button.

5. The resistance will be displayed when the measurement is finished.

Warning: To prevent possible electric shock, fire, or personal injury, do not touch the applied part terminals or device under test when red warning indicators are illuminated.


How do I measure equipment current?

Equipment current measures the current drawn by the device under test.

To measure equipment current:

1. Select the A tab.

2. Connect the device under test to the equipment outlet.

3. The measured value is continuously updated on the screen.


How do I measure touch current?

Touch current measures the current flowing between the device under test’s enclosure and protective earth.

To measure touch current:

1. Select the µA tab.

2. Select Touch current.

3. Connect a test lead between the V/Ω/A jack and the enclosure of the device under test.

4. The measured value is continuously updated on the screen.

The leakage current mode and applied part posts setting is displayed in the top right corner. Tap that button to adjust the settings.

The following equipment outlet configurations can be selected for this measurement:

Polarity, normal or reversed

Neutral, open or closed

Earth, open or closed


How do I measure earth leakage current (Ground wire leakage current)?

Earth leakage current (Ground wire leakage) measures the current flowing into the protective earth circuit of the device under test.

To measure earth leakage current:

1. Select the µA tab.

2. Select Earth leakage current.

3. The measured value is continuously updated on the screen.

The leakage current mode and applied part posts setting is displayed in the top right corner. Tap that button to adjust the settings.

The following equipment outlet configurations can be selected for this measurement:

Polarity, normal or reversed

Neutral, open or closed


How do I measure direct equipment leakage?

Direct equipment leakage measures the leakage current between all applied parts and conductive parts on the enclosure (either non-earthed or both non-earthed and earthed, depending on the class of the device under test), to protective earth.

To measure direct equipment leakage:

1. Select the µA tab.

2. Select Direct equipment leakage.

3. Connect a test lead between the V/Ω/A jack and a non-earthed accessible conductive part of the device under test.

4. The measured value is continuously updated on the screen.

The leakage current mode setting is displayed in the top right corner. Tap that button to adjust the setting.

The following equipment outlet configurations can be selected for this measurement:

Polarity, normal or reversed


How should the product be cleaned?

Clean the Product with a damp cloth and a mild detergent. Do not use abrasives, isopropyl alcohol, or other solvents that can damage the case or the display window.


How should the product be stored?

If the Product is not to be used for an extended period, remove the line power cord and store the Product in an environment as defined in the General Specifications section.


How is the fuse replaced?

The Product has one user-replaceable fuse in the AC power input module. To replace the fuse:

1. Turn off the Product and disconnect the power cord.

2. Let the Product rest for two minutes to allow the internal power supply to discharge.

3. Open the fuse door with a small, flat-blade screwdriver or similar tool.

4. Gently pull out the fuse holder.

5. Remove the fuse and replace it with an identical fuse.

6. Insert the fuse holder.

7. Close the fuse door.

8. Connect the power cord and turn on the Product.


What are common troubleshooting steps for the analyzer?

Problem Solution
The analyzer does not turn on.

Ensure the power cord is securely connected to both the analyzer and a working power outlet.

Check the fuse. Replace if necessary.

The display is blank or difficult to read.

Check the power connection.

Adjust the display contrast if applicable (refer to settings).

Inaccurate readings.

Ensure test leads are properly connected and in good condition.

Zero the test leads if performing resistance measurements.

Verify the correct measurement function and range are selected.

Calibrate the analyzer if due.

Cannot communicate with PC.

Ensure the USB cable is securely connected.

Check that the correct drivers are installed on the PC.

Try a different USB port or cable.

LEDs do not illuminate.

During startup, if any LED does not illuminate, contact technical support.

For specific measurement LEDs, ensure the correct function is selected and connections are made as per the manual for that test.

Ground open fault displayed at start-up.

Verify the analyzer is connected to a properly grounded outlet.

Check the integrity of the power cord.

Polarity reversal option appears at start-up.

This indicates the mains input polarity does not match the expected polarity for the analyzer model or standard. Choose to reverse polarity if required by the test standard or DUT configuration.

Negative resistance reading.

Test leads need to be zeroed again for protective earth or point-to-point resistance measurements.


How do I measure alternative equipment leakage?

1. Select the µA tab.

2. Select Alternative equipment leakage.

3. Connect a test lead between the V/Ω/A jack and a non-earthed accessible conductive part of the device under test.

4. The measured value is continuously updated on the screen.


What equipment outlet configurations can be selected for alternative equipment leakage measurement?

Earth: Open or closed


What does “earth closed” mean during an alternative equipment leakage measurement?

During this measurement, earth closed means that the device under test earth contact is connected to the current measurement device, as depicted in “Figure 38: Alternative equipment leakage, schematic diagram”. The device under test ground is not connected to mains protective earth.


How do I measure patient leakage current (Lead to ground)?

1. Select the AP tab.

2. Select Patient leakage current.

3. Select the applied part groupings by tapping the applied part symbol at the top of the screen.

4. Tap the side arrows (< and >) to select which applied part groups to use.

5. The measured value is continuously updated on the screen.


What equipment outlet configurations can be selected for patient leakage current (Lead to ground) measurement?

Polarity, normal or reversed

Neutral, open or closed

Earth, open or closed


How do I measure mains on applied part leakage (Lead isolation)?

1. Select the AP tab.

2. Select Mains on applied part leakage.

3. Connect a test lead between the V/Ω/A jack and a non-earthed accessible conductive part of the device under test.

4. Set the applied part groupings by tapping the applied part symbol at the top of the screen.

5. Tap the side arrows (< and >) to select which applied part group to use.

6. Tap the measure button.

7. The value will be displayed when the measurement is finished.


What equipment outlet configurations can be selected for mains on applied part leakage (Lead isolation) measurement?

Polarity, normal or reversed


What warning should be observed during mains on applied part leakage (Lead isolation) measurement?

To prevent possible electric shock, fire, or personal injury, do not touch the applied part terminals or device under test when red warning indicators are illuminated.


How do I measure direct applied part leakage?

1. Select the AP tab.

2. Select Direct applied part leakage.

3. Connect a test lead between the V/Ω/A jack and a non-earthed accessible conductive part of the device under test.

4. Select the applied part groupings by tapping the applied part symbol at the top of the screen.

5. Tap the side arrows (< and >) to move select which applied part groups to use.

6. Tap the measure button.

7. The value will be displayed when the measurement is finished.


What equipment outlet configurations can be selected for direct applied part leakage measurement?

Polarity, normal or reversed


How do I measure alternative applied part leakage?

1. Select the AP tab.

2. Select Alternative applied part leakage.

3. Connect a test lead between the V/Ω/A jack and a non-earthed accessible conductive part of the device under test.

4. Set the applied part groupings by tapping the applied part symbol at the top of the screen.

5. Tap the side arrows (< and >) to move select which applied part groups to use.

6. Tap the measure button.

7. The value will be displayed when the measurement is finished.


What warning should be observed during alternative applied part leakage measurement?

To prevent possible electric shock, fire, or personal injury, do not touch the applied part terminals or device under test when red warning indicators are illuminated.


How do I set up for point to point measurements?

Select the P2P tab and insert test leads in V/Ω/A and Ø/NULL jacks.


How do I measure Point to point leakage?

1. Select Leakage.

2. Select the desired leakage current mode.

3. Connect the leads in series with the circuit to be measured. If one point is earthed, use the Ø/NULL jack for that point.

4. The measured value is continuously updated on the screen.


What should be done before measuring point to point resistance for an accurate result?

For accurate result, zero the test leads before measurement.


How do I measure Point to point resistance?

1. Select Resistance.

2. Place the leads across resistance to be measured.

3. The measured value is continuously updated on the screen.


How do I measure point to point voltage?

1. Select Voltage.

2. Place the leads across the voltage to be measured.

3. The measured value is continuously updated on the screen.


What do negative resistance measurements indicate in point to point resistance mode?

Negative resistance measurements indicate that test leads need to be zeroed again.


How do I set up the ECG waveform simulation?

1. Select the SIM tab.

2. Select the appropriate waveform and rate.


What are the supported waveforms and their selectable rates for ECG waveform simulation?

Waveform type Selectable rate
ECG complex 30, 60, 120, 180, 240 BPM (Beats per Minute)
Pulse (63 ms pulse width) 30, 60 BPM
Ventricular fibrillation
Sine wave 10, 40, 50 60, 100 Hz
Square wave (50% duty cycle) 0.125, 2.0 Hz
Triangle wave 2 Hz

How do I set up the respiration simulation?

1. Select the SIM tab.

2. Select the appropriate mode and/or rate.

3. Select lead to simulate test leads connected to a patient’s arm or leg: LL for left leg lead, LA for left arm lead.


How do I register the product in OneQA?

Before using the product with OneQA, you must register the product to a OneQA tenant. Connect the product via USB cable to a computer running OneQA and follow the instructions on the computer screen.


How can I manually initiate a synchronization with OneQA?

To manually initiate a synchronization, check the OneQA section of the settings on the product.


Can the ESA712 be registered to OneQA?

The ESA712 cannot be registered to OneQA.


How are OneQA users added to the product?

To add OneQA users to the product, the product must be registered and synchronized with OneQA. Every OneQA user must be manually added to the product, using the username from OneQA.


What should be considered when adding OneQA users regarding passwords and local users?

Depending on settings in OneQA, the OneQA users may need to enter their password to be added to the product. OneQA can also prohibit local users on a registered product.


How do I create new assets on the analyzer?

To create new assets, tap the plus symbol in the top right corner of the asset list.


How do I remove assets from the analyzer?

To remove assets, tap the check box in the top right corner of the asset list and select which assets to remove. Tap the remove button at the bottom of the screen.


How are assets synchronized with OneQA removed?

Assets synchronized with OneQA must be removed from OneQA running on a computer.


How can I start a procedure on the analyzer?

1. On the analyzer, tap the procedure button on the home screen and follow the setup instructions on the screen. (Functionality not available on ESA712.)

2. Remotely: connect the product via USB cable to a computer running OneQA and start a procedure in OneQA. The analyzer will be remotely controlled by OneQA.


How do I start a procedure from the home screen or procedure list on the analyzer?

To start a procedure from the home screen, tap the procedure button from the home screen and follow the setup instructions on screen. You can also start a procedure from the procedure list, found via the navigation menu.


How can I get an overview of a running procedure or return to a previous step?

To get an overview of the procedure, or return to a previous step, tap the button to the right of the progress bar.


How can the view of a procedure be changed?

The procedure can be viewed with a single step in fullscreen or as a list. You toggle between single step and list in the lower left corner.


Can procedures be paused?

You can at any time pause the procedure and resume later.


What can be done when a procedure is finished?

When the procedure is finished, you have the option to view and edit procedure information before completing the procedure. When completed, the procedure will be stored as a read-only result.


How do I select the electrical safety standard for measurement sessions?

Select the electrical safety standard to be used for measurement sessions. The selected standard defines which leakage measurement variants are available.


What are the selectable electrical safety standard options based on the analyzer model?

ESA710 can be used to analyze equipment against NFPA 99 / AAMI ES1.

ESA712 and ESA715 can be used to analyze equipment against several user-selectable standards: IEC 60601-1, IEC 62353, NFPA 99 / AAMI ES1, AS/NZS 3551, EN 50678 / EN 50699.


What are the measurement names based on IEC 60601-1 and NFPA 99 / AAMI ES1 standards?

IEC 60601-1 NFPA 99 / AAMI ES1
Mains voltage Line voltage
Protective earth resistance Ground wire resistance
Earth leakage current Ground wire leakage
Patient leakage current Lead to ground
Mains on applied part leakage Lead isolation

How is the nominal mains voltage setting used?

Select the nominal mains voltage applicable for the measurement situation. Direct equipment leakage, alternative equipment leakage, direct applied part leakage and alternative applied part leakage values are scaled to the nominal mains voltage value.


What is the function of the GFCI limit setting?

The GFCI (Ground fault current interrupter) protects the device under test from excessive leakage current to earth, for example from a short circuit, when connected to the equipment outlet of the analyzer. When the GFCI trips, the power is removed from the equipment outlet. The analyzer continues to operate, but shows a message to the user. The GFCI limit setting has no effect when the equipment outlet is powered off.


How do I set the resistance unit?

Select unit to display when measuring protective earth resistance and point to point resistance. This setting does not affect procedures.


What is the polarity switching delay setting and when should it be adjusted?

The polarity switching delay is the shortest time the equipment outlet is powered off when switching polarity. Use a longer delay to protect the internal components of the analyzer from transient effects. Transient effects can occur when the device under test has a highly capacitive or inductive power supply. These types of power supplies exist in larger devices, e.g. ultrasound, dialysis, and portable X-ray machines. If you think the device under test has a highly capacitive or inductive power supply, increase the polarity switch delay to at least 5 seconds. The increased time will allow the device under test to self-discharge safely. This setting does not affect procedures.


How can screen brightness and sound options be adjusted?

Adjust the screen brightness and sound options as desired.


How can the interface language and keyboard language be changed?

The interface and the user manual will be available in several languages. You can also select what keyboard language to use.


How are the date and time settings managed?

Select time zone and date and time format. If the product is connected to OneQA via USB cable, or has an internet connection, date and time will automatically be adjusted.


How can the product be connected to a wireless network?

The product may be connected to a wireless network using a WiFi adapter. Connect the adapter to one of the USB-A ports on the side of the product and edit the network settings to connect.


Where can I find information about the OneQA connection and registration?

Status and functions related to the OneQA connection. For registration, see “Register the product in OneQA”.


How can a factory reset be initiated and what are its effects?

Factory reset and system update can be initiated from this screen. A factory reset will remove everything added by the user, such as assets, users, results, OneQA registration and reset all settings to their default values.


How do I perform a system update?

Before updating the system, complete all procedures and measurements. System update files are published by the manufacturer.

1. Save the system update file on a USB drive.

2. Insert the USB drive to one of the USB host ports on the right side of the product.

3. Tap the system update button and follow the instructions displayed.


How should the product be cleaned?

Turn off the product and disconnect the power cord before cleaning with a cloth dampened with a mild detergent solution.


What should be done before storing or transporting the product?

Before storing or transporting the product, turn off the product.


How do I replace a fuse?

First, find the correct replacement fuse, see specification on the product label located on the bottom side of the analyzer.

To replace a fuse, follow these steps:

1. Make sure that the product is turned off and disconnect any power cords and test leads.

2. Remove the fuse cover from the back side of the analyzer, using a narrow slotted screwdriver.

3. Unscrew the fuse holder and take out the fuse.

4. Make sure the new fuse is identical in type, voltage rating, and current rating as specified by the product label.

5. Insert the fuse holder with the new fuse and turn to lock.

6. Insert the fuse cover.


What are the available spare fuse parts?

#6017274, Fuse glass T 10A 250V ac, 5×20 mm

#6044658, Fuse glass T 15A 250V ac, 5×20 mm

#6017290, Fuse glass T 16A 250V ac, 5×20 mm

#6017288, Fuse glass T 20A 250V ac, 5×20 mm


What should I do if the mains power cable is connected, but the battery is not charging?

Possible cause: Fuse blown or mains power cable damaged.

Solution: Check the fuse, see “Replace a fuse”. Try with another mains power cable with the same rating.


What should I do if the product is not responding?

Possible cause: Unknown.

Solution: Press the power button for 10 seconds to force a shut down. Wait 30 seconds before turning the power on again. If the problem persists, contact support.


What if my WiFi adapter does not work?

Possible cause: The accessory is not compatible with the product.

Solution: Check with the manufacturer for compatible accessories to use with the product.


What if some procedures are not synchronized to the product?

Possible cause: All steps in the procedure are not compatible with the product.

Solution: Run the procedure from OneQA on a computer, with the product connected with USB cable.


Why can’t I register my analyzer to OneQA?

Possible cause: Limited model (ESA712).

Solution: The ESA712 cannot be registered in OneQA. To run procedures remotely, connect the analyzer to a computer running OneQA and start a procedure in OneQA.


What are the CSA approval details for different models and versions?

Model ESA710 ESA712 ESA715
US version, 90-132 V ac, 20 A MAX, 47-63 Hz CSA approved CSA approved CSA approved
NEMA 6-15 version, 90-132 V ac & 180-264 V ac, 15 A MAX, 47-63 Hz CSA approved CSA approved CSA approved

What happens if a deviation from the specified supply mains voltage and frequency is detected?

The supply mains voltage and frequency are continuously monitored. If a deviation from specification is detected, measurements are stopped, and a message is displayed until the user confirms.


CLICK HERE TO DOWNLOAD FLUKE ESA710 (01) PDF MANUAL


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.