FREE ENGLISH EXTECH EX310 (01) PDF USER GUIDE
FREE ENGLISH EXTECH EX310 (01) PDF USER MANUAL
FREE ENGLISH EXTECH EX310 (01) PDF OWNER GUIDE
FREE ENGLISH EXTECH EX310 (01) PDF OWNER MANUAL
FREE ENGLISH EXTECH EX310 (01) PDF REFERENCE GUIDE
FREE ENGLISH EXTECH EX310 (01) PDF INSTRUCTION GUIDE
FREE ENGLISH EXTECH EX310 (01) PDF REFERENCE MANUAL
FREE ENGLISH EXTECH EX310 (01) PDF INSTRUCTION MANUAL
FREE ENGLISH EXTECH EX310 (01) PDF OPERATING INSTRUCTIONS
CLICK HERE TO DOWNLOAD EXTECH EX310 (01) PDF MANUAL
If this is not the document you want for this product, click here to see if we have any other documents for this product.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
What do the safety symbols on the multimeter mean?
! (Exclamation symbol): This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.
WARNING: This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
CAUTION: This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result in damage to the product.
MAX 600V (with ground symbol): This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds 600V.
(Lightning bolt symbol): This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
(Square within a square symbol): This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
What are the input protection limits for the multimeter?
| Function | Maximum Input |
|---|---|
| V DC or V AC | 600V AC and DC |
| mA AC/DC | 200mA DC/AC |
| A AC/DC | 10A DC/AC (for 30 seconds max. every 15 mins.) |
| Resistance, Diode Test, Continuity | 250V DC/AC |
What are the main safety instructions for using the multimeter?
This meter has been designed for safe use but must be operated with caution. The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum.
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the “COM” input jack exceeds 600V above earth ground.
4. NEVER connect the meter leads across a voltage source while the function switch is in the current, resistance, or diode mode. Doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.
6. ALWAYS turn off power and disconnect test leads before opening the covers to replace the fuse or battery.
7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened securely.
8. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
What are the general cautions for using this multimeter?
• Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter.
• Always remove the test leads before replacing the battery or fuses.
• Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use.
• Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard.
• Warning! This is a class A equipment. This equipment can cause interference to devices in the home; in this case the operator can be required to carry out adequate measures to prevent interference.
• Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or Continuity tests.
• Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not “live”.
• If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
• This device is not a toy and must not reach children’s hands. It contains hazardous objects as well as small parts that children could swallow. In case a child swallows any of the parts, please contact a physician immediately.
• Do not leave batteries and packing material unattended; they can be dangerous for children.
• In case the device is going to be unused for an extended period of time, remove the batteries to prevent them from draining.
• Expired or damaged batteries can cause cauterization on contact with skin. Always use suitable hand protection.
• See that the batteries are not short-circuited. Do not throw batteries into fire.
What are the main parts of the EX310 multimeter?
1. AC Voltage Detector Sensor
2. AC Voltage Detector indicator light
3. Display (LCD)
4. Non-contact AC Voltage Detector test button
5. Rotary function dial
6. 10 ampere test lead jack
7. COM test lead jack
8. Test lead jack for voltage, milliamp, resistance/continuity, and diode functions
9. Protective rubber holster (must be removed to access the rear battery compartment)
What do the symbols on the multimeter display and dial represent?
•))) : Continuity
(diode symbol) : Diode
(battery symbol) : Battery status
~ : AC
— : DC
m : milli (10⁻³) (volts, amps)
k : kilo (10³) (ohms)
V : Volts
A : Amps
Ω : Ohms
AC : Alternating current
DC : Direct current
Why might the display show a random, changing reading on low voltage ranges?
On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random, changing reading. This is normal and is caused by the high-input sensitivity. The reading will stabilize and give a proper measurement when connected to a circuit.
How do I use the Non-Contact AC Voltage Detector (NCV) feature?
The EX310 can detect the presence of AC voltage (from 100 to 600VAC) simply by being held very near to a voltage source. The NCV function works on any rotary switch position.
WARNING: Test the AC voltage detector on a known live circuit before each use.
WARNING: Before using the meter in the AC Voltage Detector mode, verify that the battery is fresh by confirming characters appear on the LCD when the function dial is turned to any position. Do not attempt to use the meter as an AC Voltage Detector if the battery is weak or bad.
1. Test the detector on a known live circuit before use.
2. Press and hold the NCV button for the duration of the test. The meter will beep once when the button is pushed, and the battery icon will appear briefly before the display goes blank.
3. Hold the top of the meter very close to the voltage source as shown.
4. If voltage is present, the rim of the LCD display will flash a bright orange and an audible warning will sound.
How do I measure AC voltage?
WARNING: Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet has voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present.
CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the 600 VAC position.
2. Insert the black test lead banana plug into the negative COM jack. Insert red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the neutral side of the circuit.
4. Touch the red test probe tip to the “hot” side of circuit.
5. Read the voltage in the display.
How do I measure DC voltage?
CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the highest VDC position.
2. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
4. Read the voltage in the display. Move the function switch to successively lower VDC positions to obtain a higher resolution reading.
How do I test battery voltage?
CAUTION: Do not measure batteries while they are installed in the devices, they are powering. The batteries must be removed from installations before tests can be made.
1. Set the function switch to the 1.5V or 9V BAT switch position. Use the 1.5V position for ‘AAA’, ‘AA’, ‘C’, ‘D’, and other 1.5V batteries. Use the 9V position for square 9V transistor batteries.
2. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the negative side of the battery. Touch the red test probe tip to the positive side of the battery.
4. Read the voltage in the display.
How do I measure AC or DC current?
CAUTION: Do not make current measurements at 10 Amps for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 200mA AC or DC, set the function switch to the 200m AAC or ADC position and insert the red test lead banana plug into the mA jack.
3. For current measurements up to 10A AC or DC, set the function switch to the 10A AAC or 10A ADC position and insert the red test lead banana plug into the 10A jack.
4. Remove power from the circuit under test, then open the circuit at the point where you wish to measure current.
5. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
6. Apply power to the circuit.
7. Read the current in the display.
How do I measure resistance?
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements.
1. Set the function switch to the highest Ω position.
2. Insert the black test lead banana plug into the negative COM jack.
3. Insert the red test lead banana plug into the positive Ω jack.
4. Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
5. Read the resistance in the display. Move the function switch to successively lower Ω positions to obtain a higher resolution reading.
How do I perform a continuity check?
WARNING: To avoid electric shock, never measure continuity on circuits that have a voltage potential.
1. Set the function switch to the continuity position.
2. Insert the black lead banana plug into the negative COM jack.
3. Insert the red test lead banana plug into the positive Ω jack.
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately 100Ω the audible signal will sound. If the circuit is open, the display will indicate “1”.
How do I perform a diode test?
1. Set the function switch to the diode test position.
2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive diode test jack.
3. Touch the test probes to the diode under test.
4. A good diode will indicate approx. 700 ohms for the forward test and “1” for the reverse test.
5. A shorted diode will indicate the same value of resistance in both the reverse and forward test directions. An open diode will indicate “1” in both test directions.
How should I maintain the multimeter?
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover or the battery or fuse covers.
WARNING: To avoid electric shock, do not operate your meter until the battery and fuse covers are in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following care instructions are performed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the battery should be removed to prevent damage to the unit.
How do I know when to replace the battery and how do I do it?
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover. Do not operate meter unless the battery cover is in place.
LOW BATTERY INDICATION
The battery icon will appear in the lower left-hand corner of the display when the battery voltage becomes low. Replace the batteries when this appears.
BATTERY REPLACEMENT
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster. See diagram.
3. Remove the Phillips head screw located on the lower back of the instrument.
4. Remove the fuse/battery compartment cover to access the battery. See diagram.
5. Replace the 9V battery observing polarity.
6. Secure the fuse/battery compartment cover using the Phillips head screw.
7. Place the protective rubber holster on the meter.
What are the safety reminders for handling batteries?
• Please dispose of batteries responsibly; observe local, state, and national regulations.
• Never dispose of batteries in a fire; batteries may explode or leak.
• Never mix battery types; install new batteries of the same type.
How do I replace the fuses?
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse cover.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster.
3. Remove the Phillips head screw located on the lower back of the instrument.
4. Remove the fuse/battery compartment cover to access the fuses.
5. Gently remove the fuse(s) and install new fuse(s) into the holder(s).
6. Always use fuses of the proper size and value (200mA/600V fast blow for the mA / µA ranges, 10A/600V fast blow for the A range).
7. Secure the fuse/battery compartment cover.
8. Place the protective rubber holster on the meter.
WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in place and fastened securely.
What are the range specifications for this multimeter?
Notes: Accuracy specifications consist of two elements:
• (% reading) – This is the accuracy of the measurement circuit.
• (+ digits) – This is the accuracy of the analog to digital converter.
Accuracy is stated at 65°F to 83°F (18°C to 28°C) and less than 75% RH.
| Function | Range | Resolution | Accuracy |
|---|---|---|---|
| Non-contact AC Voltage detector | 100 to 600V | Resolution & accuracy do not apply since the meter does not display the voltage in this mode. The lamp at the top of the meter’s display flashes when voltage is sensed and an audible warning will sound | |
| DC Voltage (V DC) | 200mV | 0.1mV | ±(0.5% reading + 2 digits) |
| 2000mV | 1mV | ±(1.0% reading + 2 digits) | |
| 20V | 0.01V | ||
| 200V | 0.1V | ||
| 600V | 1V | ±(1.5% reading + 2 digits) | |
| AC Voltage (VAC) 50/60Hz | 200V | 0.1V | ±(1.5% reading + 3 digits) |
| 600V | 1V | ±(2.0% reading + 4 digits) | |
| DC Current (A DC) | 200mA | 0.1mA | ±(1.5% reading + 2 digits) |
| 10A | 0.01A | ±(2.5% reading + 5 digits) | |
| AC Current (AAC) 50/60Hz | 200mA | 0.1mA | ±(1.8% reading + 5 digits) |
| 10A | 0.01A | ±(3.0% reading + 7 digits) | |
| Resistance | 200Ω | 0.1Ω | ±(1.2% reading + 4 digits) |
| 2000Ω | 1Ω | ±(1.2% reading + 2 digits) | |
| 20kΩ | 0.01kΩ | ||
| 200kΩ | 0.1kΩ | ||
| 2000kΩ | 1kΩ | ||
What are the general specifications of the EX310 multimeter?
| Diode Test | Bias voltage: 2.3VDC |
| Continuity Check | Audible signal will sound if the resistance is less than 100Ω |
| Input Impedance | 50 / 60Hz |
| Display | 2000 count (0 to 1999) LCD |
| Over-range indication | “1” is displayed |
| Polarity | Automatic (no indication for positive); Minus (-) sign for negative |
| Measurement Rate | 2 times per second, nominal |
| Low Battery Indication | Battery symbol is displayed if battery voltage drops below operating voltage |
| Battery | One (1) 9V battery |
| Fuses | mA range; 200mA/600V fast blow. A range; 10A/600V fast blow, ceramic |
| Operating Temperature | 32°F to 122°F (0°C to 50°C) |
| Storage Temperature | -4°F to 140°F (-20°C to 60°C) |
| Operating Humidity | Max 70% up to 87°F (31°C) decreasing linearly to 50% at 122°F (50°C) |
| Storage Humidity | < 80% RH |
| Operating Altitude | 7000 ft (2000 meters) maximum |
| Weight | 9.17 oz (260 g) includes holster |
| Dimensions | 5.8 x 2.9 x 1.6 in (147 x 76 x 42 mm) includes holster |
| Approvals | CE |
| Safety | This meter is intended for indoor use and protected, against the users, by double insulation per EN61010-1 and IEC61010-1 3rd Edition (2010) CAT III 600V; Pollution Degree 2. |
What are the PER IEC1010 Overvoltage Installation Categories?
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level. Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Note – Examples include electricity meters and primary over-current protection equipment.
CLICK HERE TO DOWNLOAD EXTECH EX310 (01) PDF MANUAL