BENCHMARK SCIENTIFIC INCU-MIXER MP (01) Image

Free English BENCHMARK SCIENTIFIC INCU-MIXER MP (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD BENCHMARK SCIENTIFIC INCU-MIXER MP (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

What should I do before using the BENCHMARK SCIENTIFIC INCU-MIXER MP for the first time?

Please read and understand this manual completely prior to operation. Upon receipt, please check the instrument and appendix against the packing list. If you find there is something wrong or missing, please contact Benchmark Scientific immediately.


What are the important safety warnings and guidelines for the BENCHMARK SCIENTIFIC INCU-MIXER MP?

The operation, maintenance and repair of the BENCHMARK SCIENTIFIC INCU-MIXER MP should comply with the basic guidelines and warnings listed below. Noncompliance may result in damage to the product and/or bodily harm to the user.

• Do not use the instrument without reading this instruction manual. Read the guidelines and directions below and use the instrument accordingly.

• This product is a 115V electrical instrument suitable for indoor use.

• Only trained electricians should configure internal wiring.

• The operator should not open or repair the instrument. This may void any warranty, written or promised. In the event of product failure, please contact Benchmark Scientific.

• A.C. power’s grounding should be reliable to safeguard against an electric shock. The 3-pin plug supplied with the power cable is a safety device that should be matched with a suitable grounded socket.

• During normal operation, the temperature of the heating block will be very high, and will cause injury. Please avoid personal contact with the heat block while it is on.

• Close all tube lids before placing the tubes into the block. Heated liquids may escape their vessels. Spilled liquids may cause internal damage to the instrument.

• Before powering on, please confirm that the electrical output of the instrument matches the outlet. If the electric line is damaged, you should replace it with an identical line.

• The Instrument should be placed in a cool, dust-free environment, away from direct sunlight. The environment should have good ventilation, lack corrosive gases or a strong disturbing magnetic field, and be located far away from any central heat source. Do not place the instrument where it may be subjected to moisture.

• The vent on the Instrument is designed for ventilation. Do not cover the vent, or the instrument may overheat. Please be sure to leave at least 100cm between the instrument and other instruments or walls.

• The main power switch is on the rear of the device. Push “I” to power on the device, and push “O” to power off the device.

• Power off when you finish your work. Disconnect the power supply and cover the instrument with a cloth when not in use for extended periods of time.


When should I immediately unplug the BENCHMARK SCIENTIFIC INCU-MIXER MP and contact the vendor?

Pull the plug from the jack immediately and contact the vendor if the following occurs:

• There is liquid flowing into the Instrument;

• Subjected to liquid or fire damage.

• Abnormal operation: such as abnormal sound or smell.

• Instrument drop or outer shell damage.

• Any other change to normal operation.


How should I maintain and clean the BENCHMARK SCIENTIFIC INCU-MIXER MP?

The heating plate should be cleaned on a regular basis with a soft cloth and alcohol to prevent variance in heat translation. Contamination on the plate may cause abnormal smell, smoke, and poor heat transfer.

Power off when cleaning the Instrument. When cleaning the heating plate, do not douse or submerge in liquid. Corrosive cleaning solution may damage the instrument.


What are the key features of the BENCHMARK SCIENTIFIC INCU-MIXER MP?

Features of this product are as follows:

1) Microprocessor controlled time, speed and temperature

2) Simultaneous display of set and actual time, temperature and shaking speed

3) Audio signal and cessation of shaking motion after program completion

4) Heating platform for standard micro test plate

5) Soft start, easy to setup and use


What are the specifications for the BENCHMARK SCIENTIFIC INCU-MIXER MP?

Normal operating condition:

Ambient temperature: 5°C ~30°C

Relative humidity: ≤70%

Power supply: AC115V 50-60Hz

The basic parameters and performance:

ParameterIncu-Mixer
Mixing rate100~1500 rpm
Orbit2mm
Temperature rangeRT+5°C~70°C
Timing range1min ~ 99h59min
Accuracy of the temperature≤0.5°C
Heating time≤25min (From room temperature to 70°C)
Standard Block2 or 4 micro test plates
Dimension (mm)350(D)×320(W)×185(H)
Net weight (kg)9.0

What are the main components of the BENCHMARK SCIENTIFIC INCU-MIXER MP?

The main components are:

• Hot lid

• Heating Plate

• Keyboard

• Power connector

• Power Switch

• Universal Serial Bus


How do I read the display panel on the BENCHMARK SCIENTIFIC INCU-MIXER MP?

The display panel is divided into two rows, labeled “P:” and “S:”.

P: line: Shows the current (actual) values. From left to right, it displays Current temperature, Current speed, and Remnant time.

S: line: Shows the set values. From left to right, it displays Setting temperature, Setting speed, and Setting time.


What are the functions of the keys on the BENCHMARK SCIENTIFIC INCU-MIXER MP control panel?

OFF: Stop temperature, speed or time function.

SET: Setting key, used together with the arrow keys to set the temperature, speed and time.

STOP/START: Stop/start key. Press this key to stop or start the procedure. Press momentarily to start, Pressing continuously to stop.

Up Arrow: Increase key for temperature, speed or time.

Down Arrow: Decrease key for temperature, speed or time.


How do I set the temperature, speed, and time on the BENCHMARK SCIENTIFIC INCU-MIXER MP?

a) When the instrument powers on, the LCD will display “System-Testing”. Initialization will be indicated by an audio signal.

b) About 6s later, the “P” line will display the current temperature, speed, and remaining time. The “S” line will display the former set temperature, speed, and time.

c) Press the “SET” key momentarily. A cursor “_” will appear on the “S:” line. Use the ▲ or ▼ key to change the digital indicator above the cursor. Press the “SET” key again to move to the next parameter and save the change.

d) If you want to shut off a temperature, speed, or time function, press “SET” and move the cursor “_” to the desired column. Press the “OFF” key and the column will display “OFF”.


What are some important notes about setting parameters on the BENCHMARK SCIENTIFIC INCU-MIXER MP?

1) When time is set to “OFF”, the instrument functions continuously.

2) Temperature, speed and time can not be set to “OFF” simultaneously.

3) The instrument will start to heat automatically according to the previous set temperature as soon as the instrument is powered on.


How do I run, pause, and stop a program on the BENCHMARK SCIENTIFIC INCU-MIXER MP?

Run:

Press the “STOP/START” key momentarily to run. When the current temperature arrives at the set temperature, the time will start to count backward and the symbol “:” will flicker in 2s intervals. When the run is over, an audio indicator will begin.

Pause:

Press the “STOP/START” key momentarily to pause the program when it is running. Press “STOP/START” again to continue the program.

Stop:

Press “STOP/START” for 1 second continuously to stop. The instrument will maintain temperature, the speed will display 0, and the time will display 00:00.


How do I use the instantaneous mixing function on the BENCHMARK SCIENTIFIC INCU-MIXER MP?

a) Press the “SET” key. After the “S:”, the cursor “_” will appear. Press the “SET” key to move the cursor, and place the cursor under the time column. Press “▼” or “▲” to change the time to 00:00.

b) Press “STOP/START”, and the instrument will mix for 6s and then stop.


How can I troubleshoot common issues with the BENCHMARK SCIENTIFIC INCU-MIXER MP?

No.PhenomenonPossible CausesProcessing Procedure
1No signal on the display when instrument is powered on.No power
Broken Power Line
Broken switch
Others
Check the power
Repair Power Line
Repair the switch
Contact Benchmark Scientific
2The actual and displayed temperatures are quite different.Broken sensorContact Benchmark Scientific
3“OPE.1” in the temperature display with an audio alarmTemperature sensor in the lid is disconnectedContact Benchmark Scientific
4“OPE.2” in the temperature display with an audio alarmTemperature sensor in the heating plate is disconnectedContact Benchmark Scientific
5“SHO.1” in the temperature display with an audio alarmTemperature sensor in the lid is shortedContact Benchmark Scientific
6“SHO.2” in the temperature display with an audio alarmTemperature sensor in the heating plate is shortedContact Benchmark Scientific
7“HHH.1” in the temperature display with an audio alarmOver temperature alarm for lidContact Benchmark Scientific
8“HHH.2” in the temperature display with an audio alarmOver temperature alarm of heating blockContact Benchmark Scientific
9“ERR1” in the speed display with an audio alarmRotor lockedContact Benchmark Scientific
10“ERR2” in the speed display with an audio alarmRotor is running, but display is “0 rpm”Contact Benchmark Scientific
11“ERR3” in the speed display with an audio alarmActual Speed over max. speedContact Benchmark Scientific
12No heating of the heating blockBroken heaterContact Benchmark Scientific
13Press invalidBroken film switchContact Benchmark Scientific

CLICK HERE TO DOWNLOAD BENCHMARK SCIENTIFIC INCU-MIXER MP (01) PDF DOCUMENT