Free English BAYER CONTOUR USB (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.
CLICK HERE TO DOWNLOAD BAYER CONTOUR USB (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
What is the intended use of the BAYER CONTOUR USB blood glucose monitoring system?
The BAYER CONTOUR USB blood glucose monitoring system (meter, test strips, and controls) is intended for self-testing by people with diabetes to monitor glucose concentrations in whole blood. The test provides a quantitative measurement of glucose in blood from 20 to 600 mg/dL.
What are the limitations on who can use the BAYER CONTOUR USB system?
The BAYER CONTOUR USB system is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates. In addition, the meter is intended for use with only one person and must not be shared.
What type of test strips does the BAYER CONTOUR USB meter use?
The BAYER CONTOUR USB meter only works with compatible blood glucose test strips designed specifically for this system.
How do you charge the rechargeable battery on the BAYER CONTOUR USB meter?
The meter has a permanent rechargeable battery. Before doing anything else, charge your battery by using the provided wall charger or by plugging your BAYER CONTOUR USB meter into a computer USB port.
Be sure your computer is turned on and not in sleep, hibernate, or power-saving mode. The meter will briefly display “Do Not Test-Charging” and the test strip port light will flash.
Can you perform a blood sugar test while the BAYER CONTOUR USB meter is charging?
No. You cannot do a blood sugar test while the battery is charging. When charging is finished, the test strip port light will turn off. Unplug your BAYER CONTOUR USB meter to test.
How do the buttons next to the screen work on the BAYER CONTOUR USB meter?
Every screen gives you choices. The three buttons located next to the screen allow you to make your selection by pressing the button positioned directly next to your choice on the display.
How do you use the Selection and Scrolling buttons on the BAYER CONTOUR USB meter?
When the up arrow or down arrow appears on the screen, the top and bottom buttons are used for scrolling. When your choice is highlighted, make a selection by pressing the center OK button. Pressing and holding the up or down buttons will scroll through the list faster.
How do you turn on the BAYER CONTOUR USB meter using the Menu button?
To turn on the meter, press and hold the Menu button located on the side of the meter until the device turns on.
What are the three main menu choices on the BAYER CONTOUR USB meter?
The Main Menu has 3 choices: Logbook, Trends, and Setup. You can select any choice by pressing the button next to it on the display.
How do you return to the previous screen on the BAYER CONTOUR USB meter?
To go back to the previous screen, press the Menu button once.
How do you turn off the BAYER CONTOUR USB meter manually?
To turn the meter off manually, press and hold the Menu button for 2 seconds.
How do you turn on the test strip port light on the BAYER CONTOUR USB meter?
With the meter turned off, press the Menu button twice with one second between each press to turn on the test strip port light. Press the Menu button twice again to turn the light off.
What is the AutoLog feature on the BAYER CONTOUR USB meter?
AutoLog is a feature that allows you to mark your test result as Before Meal, After Meal, or No Mark during the 5-second test countdown. The meter comes with AutoLog turned on by default, which helps personalize records and calculate meal-specific averages.
What extra memory partition is included with the BAYER CONTOUR USB meter?
The BAYER CONTOUR USB meter contains a 1 GB portable flash drive with a 500 MB user-accessible extra memory partition for personal use. It appears on your computer as a Removable Disk when plugged into a USB port.
What setup sequence occurs during initial startup of the BAYER CONTOUR USB meter?
When you first turn on the meter after charging, a setup sequence begins:
Step 1. Press and hold the Menu button until the meter turns on. After a welcome screen, the Select Language screen appears.
Step 2. Press the up or down button to highlight your desired language, then press the OK button.
Step 3. A confirmation screen verifies your selected language.
Step 4. On the next screen, choose Quickstart, Customize, or Help.
How do you set up the BAYER CONTOUR USB meter using Quickstart?
Quickstart uses default factory settings so you can begin testing quickly:
Step 1. Press the Quickstart button on the setup screen.
Step 2. A screen confirms that Quickstart is selected.
Step 3. The current date and time screen will appear. If the date and time are correct, press the Accept button. If incorrect, press the Change button to adjust them.
Step 4. Setup is complete and you may begin testing blood sugar.
What default settings are configured during Customize startup on the BAYER CONTOUR USB meter?
When selecting Customize during setup, the default parameters presented are:
AutoLog status: On.
Before Meal Target: 70 to 130 mg/dL.
After Meal Target: 70 to 180 mg/dL.
You can accept each default setting or press Change to modify values.
How do you use the BAYER CONTOUR USB meter for testing in the dark?
To test in dark conditions:
Step 1. With the meter off, press the Menu button twice with one second between each press to activate the lighted test strip port.
Step 2. Insert the test strip into the illuminated port; the display screen will appear.
Step 3. Apply blood to the test strip. Once blood is applied, the port light turns off automatically.
Step 4. Continue with your test. If needed, two quick presses of the Menu button will manually turn the light off.
How should test strips be stored and handled for the BAYER CONTOUR USB meter?
Always keep test strips in their original bottle and tightly close the bottle cap immediately after removing a test strip. The bottle is engineered to keep strips dry; exposure to room humidity or outside storage can cause damage. Test strips are intended for single use only. Never use damaged or already used test strips, and always check expiration dates before use.
What temperature precautions should be observed when using the BAYER CONTOUR USB meter?
The meter and test strips are designed to give accurate results at temperatures between 41 degrees Fahrenheit and 113 degrees Fahrenheit. Always allow 20 minutes for the meter and strips to adjust to the temperature of a new location before performing a test.
Because computer USB ports and self-powered hubs can become much warmer than room temperature, use the provided USB extension cable if you wish to test immediately after disconnecting from your computer.
How do blood glucose measurement units display on the BAYER CONTOUR USB meter?
The meter is preset to display blood glucose results in mg/dL (milligrams of glucose per deciliter) or mmol/L (millimoles of glucose per liter) depending on country standards. Results displayed in mg/dL will never include a decimal point (for example: 93 mg/dL). Results displayed in mmol/L will always include a decimal point (for example: 5.2 mmol/L).
What materials must be prepared before testing blood sugar with the BAYER CONTOUR USB meter?
Before beginning a test, ensure you have the following ready:
Your BAYER CONTOUR USB meter.
Compatible blood glucose test strips.
A lancing device and sterile lancets.
Control solution (if performing a quality control check).
Warm soapy water and a clean towel to wash and dry your puncture site.
How do you prepare the lancing device for testing with the BAYER CONTOUR USB meter?
Step 1. Snap off the endcap from top to bottom while holding the grip indent.
Step 2. Loosen the round protective cap on a lancet by rotating it 1/4 turn, but do not remove it.
Step 3. Insert the lancet firmly into the lancing device until it comes to a full stop.
Step 4. Twist off the round protective lancet cap and save it for safe lancet disposal later.
Step 5. Replace the solid color endcap securely onto the lancing device.
Does the BAYER CONTOUR USB meter require manual coding?
No. The BAYER CONTOUR USB meter features automatic coding and codes itself every time a test strip is inserted, preventing inaccurate results due to miscoding or forgetting to code the meter.
How do you insert a test strip into the BAYER CONTOUR USB meter?
Step 1. Remove a test strip from the bottle and close the bottle lid tightly immediately.
Step 2. Hold the test strip with the gray electrode end facing up.
Step 3. Insert the gray end firmly into the test strip port on the meter.
Step 4. The meter will turn on automatically.
What startup screen sequence appears after inserting a test strip into the BAYER CONTOUR USB meter?
After inserting the test strip, a solid gray screen with a white border will appear for one second to verify that the display screen is functioning properly. Then, the logo screen will briefly display, followed by the “Apply Blood” screen indicating the meter is ready for testing.
How do you obtain a blood drop from a fingertip for the BAYER CONTOUR USB meter?
Step 1. Attach the gray endcap to the lancing device and rotate the dial to adjust puncture depth.
Step 2. Press the endcap firmly against the puncture site and press the release button.
Step 3. Stroke your hand and finger toward the puncture site to form a drop of blood. Do not squeeze directly around the puncture site.
Step 4. Test immediately once a small, round blood drop of 0.6 microliters has formed.
When is Alternative Site Testing (AST) recommended with the BAYER CONTOUR USB meter?
Alternative Site Testing (forearm or palm) is recommended only when it is more than two hours after a meal, diabetes medication, or exercise.
Under what conditions should Alternative Site Testing (AST) not be used?
Do not use Alternative Site Testing and use fingertip testing instead under any of these circumstances:
If you think your blood glucose is low (hypoglycemia).
When blood glucose is changing rapidly (after a meal, insulin dose, or exercise).
If you have hypoglycemic unawareness (lack of symptoms).
If alternative site test results do not agree with how you feel.
During illness or times of stress.
If you will be driving a car or operating machinery.
How do you perform forearm testing for the BAYER CONTOUR USB meter?
Step 1. Attach the clear AST endcap to the lancing device.
Step 2. Select a fleshy area on your forearm away from veins, hair, moles, bones, and tendons.
Step 3. Rub the testing area vigorously until it feels warm to increase blood flow.
Step 4. Press the clear endcap firmly against the site and press the release button.
Step 5. Keep the device in constant contact with the skin and pump up and down 2 to 3 times without lifting the device from the skin.
Step 6. Maintain steady pressure until a small round drop forms, then lift the device straight up without smearing.
How do you perform palm testing for the BAYER CONTOUR USB meter?
Step 1. Attach the clear AST endcap to the lancing device.
Step 2. Select a fleshy site on the palm away from veins, bones, and tendons. It is not necessary to rub the skin or pump the device.
Step 3. Press the clear endcap firmly against the puncture site and press the release button.
Step 4. Maintain steady pressure until a small round blood drop forms.
Step 5. Lift the device straight up and away from the skin without smearing the blood.
What blood sample conditions make a sample unsuitable for testing on the BAYER CONTOUR USB meter?
Do not test the blood sample if you observe:
Smeared blood.
Clotted blood.
Runny blood.
Clear fluid mixed with the blood.
How do you apply blood to the test strip on the BAYER CONTOUR USB meter?
Step 1. Immediately touch the tip of the test strip to the drop of blood. Blood is automatically drawn into the test strip tip by capillary action.
Step 2. Hold the tip of the test strip in the blood drop until the meter beeps.
Do not drop blood directly on the flat surface of the test strip, and do not press the tip against your finger, which can block the sample tip.
What occurs if blood is not applied to the test strip within 1 minute on the BAYER CONTOUR USB meter?
If blood is not applied within 1 minute, the meter screen dims and beeps. Pressing any button will reactivate the “Apply Blood” screen. After a total of 3 minutes of inactivity, the meter shuts off automatically.
How does the testing procedure work when AutoLog is turned on in the BAYER CONTOUR USB meter?
Step 1. Touch the test strip tip to the blood drop until the meter beeps.
Step 2. The AutoLog screen appears during the 5-second countdown. Select Before Meal, After Meal, or No Mark by pressing the button next to your choice.
Step 3. If you make your selection in less than 5 seconds, the remainder of the countdown displays.
Step 4. Your result displays in large numbers for 2 seconds before switching to the normal test result screen. If the result is a low or high alert reading, it displays automatically within 5 seconds even if an AutoLog selection is not made.
How does the testing procedure work when AutoLog is turned off in the BAYER CONTOUR USB meter?
When AutoLog is turned off, applying blood to the test strip triggers a standard 5-second countdown screen. After 5 seconds, your test result is displayed directly on the screen without meal prompts, and you may optionally add a Note or Reminder.
What are expected non-diabetic fasting blood sugar values according to the BAYER CONTOUR USB manual?
Non-diabetic expected fasting blood sugar values are 75 to 110 mg/dL. Blood sugar values vary based on food intake, medication doses, health, stress, and activity levels. Always consult your healthcare provider to establish personal target values.
What does the BAYER CONTOUR USB meter display if a blood sugar reading is under 20 mg/dL?
If the reading is under 20 mg/dL, the normal screen sequence appears for 5 seconds. The meter then beeps twice and displays the alert screen reading: “Result Under 20 mg/dL Follow Medical Advice Immediately” along with code E23.
What does the BAYER CONTOUR USB meter display if a blood sugar reading is over 600 mg/dL?
If the reading is over 600 mg/dL, the normal screen sequence displays for 5 seconds, the meter beeps twice, and an alert screen displays: “Result Over 600 mg/dL Wash hands or test site. Repeat test with new strip” with code E21. If repeated test results are still over 600 mg/dL, the screen displays: “Over 600 mg/dL, Again Follow Medical Advice Immediately” with code E22.
How does the BAYER CONTOUR USB meter alert you to low or high blood sugar readings?
The meter comes preset with a low blood sugar (hypoglycemia) value of 70 mg/dL and a high blood sugar (hyperglycemia) value of 250 mg/dL:
If reading is under 70 mg/dL: A screen with large orange numbers and an alert symbol warns that your blood sugar is low.
If reading is over 250 mg/dL: A screen with large orange numbers and an alert symbol warns that your blood sugar is high.
If AutoLog was on and not marked prior to the alert, you can press Notes to assign Before Meal, After Meal, or No Mark.
What are the common symptoms of high blood sugar and ketoacidosis?
Symptoms of high blood sugar (hyperglycemia) include:
Frequent urination.
Excessive thirst.
Blurred vision.
Increased fatigue.
Extreme hunger.
Irritability.
Symptoms of ketones (ketoacidosis) include shortness of breath, nausea and vomiting, and very dry mouth.
What are the common symptoms of low blood sugar?
Common symptoms of low blood sugar (hypoglycemia) include:
Shakiness.
Sweating.
Fast heartbeat.
Blurred vision.
Confusion.
Passing out.
Seizures.
When should you perform a control test on the BAYER CONTOUR USB meter?
You should perform a control test:
When using your meter for the first time.
To check if you are testing correctly.
When you open a new vial of test strips.
If you leave the test strip vial open for an extended period of time.
If you think your meter may not be working properly.
If your blood test results do not match how you feel.
How do you perform a control test on the BAYER CONTOUR USB meter?
Step 1. Remove a test strip from the bottle and firmly snap the lid closed.
Step 2. Insert the test strip into the meter to turn it on.
Step 3. Gently rock the control solution bottle before opening to ensure thorough mixing.
Step 4. Squeeze a small drop of control solution onto a clean non-absorbent surface, such as a piece of wax paper. Do not apply solution directly from the bottle to the strip.
Step 5. Replace the cap on the control solution bottle.
Step 6. Immediately touch the tip of the test strip to the drop of control solution and hold it until the meter beeps.
Step 7. In 5 seconds, the meter automatically marks and displays the control test result without affecting your blood sugar averages.
Step 8. Compare the result with the control range printed on the test strip bottle label or box bottom.
Step 9. Remove the test strip to turn off the meter.
How do you set a reminder after testing on the BAYER CONTOUR USB meter?
Step 1. From the test result screen, press the Reminder button.
Step 2. Press the Start button to begin countdown (default is 2 hours or the last time set), or press Change to adjust the time.
Step 3. If changing, press the up or down button to select the hour and press OK.
Step 4. Press the up or down button to select minutes (in 15-minute increments) and press OK.
Step 5. A screen confirms the reminder countdown time, and a clock symbol appears in the blue header.
Step 6. Remove the test strip; the meter retains your set time as the default.
How do you set a reminder from the Setup menu on the BAYER CONTOUR USB meter?
Step 1. Open the Main Menu and press the Setup button.
Step 2. If Reminder is set to Off, press the OK button to modify Reminder settings.
Step 3. Press Start to initiate countdown, or press Change to adjust the time, using the up and down buttons to set hours and minutes.
Step 4. A screen confirms the reminder time set, and the clock symbol appears in the header.
How do you turn off an active reminder on the BAYER CONTOUR USB meter?
To turn off a reminder after testing:
Step 1. Press the Reminder button on the test result screen (the screen displays remaining countdown time).
Step 2. Press the Stop button. A confirmation screen verifies that the reminder is stopped, and the clock icon disappears.
To turn off a reminder from the Setup menu:
Step 1. In the Setup menu, highlight Reminder and press OK.
Step 2. Press the Stop button. A confirmation screen verifies that the reminder is stopped.
What happens when a programmed reminder alarm sounds on the BAYER CONTOUR USB meter?
When the reminder time is reached, 20 beeps will sound, the meter turns on, and a reminder screen displays: “Time for your next test.” You can stop the beeps by pressing the OK button or by inserting a test strip. If you perform a test within 15 minutes prior to the reminder time, the reminder cancels automatically.
How do you add notes to a test result on the BAYER CONTOUR USB meter?
Step 1. From the test result screen, press the Notes button.
Step 2. Use the up or down button to scroll through available choices (such as Sick, Stress, Activity, Don’t Feel Right).
Step 3. Press the OK button when your desired note is highlighted.
Step 4. A confirmation screen confirms the note was added. You can add additional notes by repeating the steps; added notes will scroll across the bottom of the result screen.
How do you view stored results in the Logbook of the BAYER CONTOUR USB meter?
Step 1. Turn on the meter and press the Logbook button on the Main Menu.
Step 2. Scroll through stored results using the up and down arrow buttons. The most recent result appears first, and the oldest appears last.
Step 3. When you reach the oldest record, the “End of Logbook” screen displays. The meter stores up to 2,000 test results in memory.
How do you view Trends and averages on the BAYER CONTOUR USB meter?
Step 1. From the Main Menu, press the Trends button.
Step 2. If AutoLog is enabled, the default Before Meal and After Meal target ranges will appear.
Step 3. Press the down button to view the 14-day Before Meal Average screen.
Step 4. Press the down button again to display the 14-day After Meal Average screen.
Step 5. Press the down button a third time to display the 14-day Results total average screen. If AutoLog is off, only the 14-day Results total average screen is displayed.
What information is shown on the 14-Day Average trend screens of the BAYER CONTOUR USB meter?
Each trend average screen displays:
The calculated average glucose value in mg/dL.
The total number of test results included in that average.
The number of test results that fell above the target range (indicated by an up arrow).
The number of test results that fell within the target range (indicated by a check mark).
The number of test results that fell below the target range (indicated by a down arrow).
How do you change the date on the BAYER CONTOUR USB meter?
Step 1. Press the Menu button and select Setup.
Step 2. Press the up or down button to highlight Date, then press OK.
Step 3. Press the Change button.
Step 4. Select your preferred date format (Month/Day/Year or Day.Month.Year).
Step 5. Use the up or down button to select the year and press OK.
Step 6. Use the up or down button to select the month and press OK.
Step 7. Use the up or down button to select the day and press OK.
Step 8. Verify the date on the confirmation screen and press Done.
How do you change the time and time format on the BAYER CONTOUR USB meter?
Step 1. In the Setup menu, select Time and press OK.
Step 2. Press the Change button.
Step 3. Choose either the 12 hour clock or 24 hour clock format.
Step 4. Press the up or down button to set the hour and press OK.
Step 5. Press the up or down button to set the minute and press OK.
Step 6. If using 12-hour format, select AM or PM and press OK.
Step 7. Confirm the time entered and press Done.
How do you turn sound on or off on the BAYER CONTOUR USB meter?
Step 1. In the Setup menu, highlight Sound and press OK.
Step 2. Press Turn Off to disable operational sounds, or press Go Back to keep sound enabled.
Step 3. A confirmation screen verifies that sound is off.
Note: Certain error messages and the reminder alarm will override the sound off setting. A single long beep indicates confirmation, while two beeps indicate an error or alert situation.
How do you turn AutoLog on or off in the Setup menu of the BAYER CONTOUR USB meter?
Step 1. From the Setup menu, press the up or down button to select AutoLog, then press OK.
Step 2. Press Turn Off to deactivate AutoLog, or press Accept to keep it enabled.
Step 3. A confirmation screen confirms that AutoLog is off.
How do you customize blood sugar target ranges on the BAYER CONTOUR USB meter?
Step 1. In the Setup menu, select Target and press OK.
Step 2. Press Change to modify values.
Step 3. Use the up or down button to select your desired Before Meal low target and press OK.
Step 4. Repeat the process to set your Before Meal high target, After Meal low target, and After Meal high target.
Step 5. Review the confirmation screen showing all set targets and press Done.
How do you change the system language on the BAYER CONTOUR USB meter?
Step 1. In the Setup menu, select Language and press OK.
Step 2. Press the up or down button to highlight your desired language.
Step 3. Press OK. A confirmation screen confirms your selection, and menus will display in the chosen language.
How are error messages identified on the BAYER CONTOUR USB meter?
When an error occurs (such as hardware, software, or testing errors), the meter beeps twice and displays an error code consisting of the letter “E” followed by a number in the lower left corner of the screen. The first line explains the specific error, and subsequent lines tell you what corrective action to take. Press the OK button next to the screen prompt to continue.
How does the battery indicator work on the BAYER CONTOUR USB meter?
The battery symbol appears on the Apply Blood screen and Main Menu screen. As battery charge is depleted, the fill color gradually decreases. The fill color turns yellow when the battery is low, and red when the battery is almost out of charge. If uncharged, the meter displays “Shutting Down, Battery is dead” and requires immediate charging.
What is Rapid Charge on the BAYER CONTOUR USB meter?
If the battery is depleted when you plug the meter in, it initiates Rapid Charge for approximately 1 minute. As soon as Rapid Charge completes, you can unplug the meter and perform a blood glucose test immediately.
How long does a normal battery charge take on the BAYER CONTOUR USB meter?
After Rapid Charge completes, a normal charging cycle takes up to 2 hours. While charging, the strip port light flashes slowly. When the battery is completely charged, the port light turns off.
What does the message “Low Power Charging” mean on the BAYER CONTOUR USB meter?
If the charging status screen displays “Low Power Charging,” the meter is connected to a low-power USB port that cannot deliver optimal power. Try connecting the meter to a different, high-power USB port on your computer to allow full access to onboard features and minimize charging time.
How do you safely remove the battery for disposal at the end of life of the BAYER CONTOUR USB meter?
Note: Carry out this process only when disposing of the meter permanently; the meter will no longer function once opened.
Step 1. Beginning near the test strip port, insert the tip of a screwdriver between the upper and lower casing and twist to pry them apart along the seam.
Step 2. Pry off the black plastic cover located over the battery.
Step 3. Disconnect the rechargeable lithium polymer battery from its connector.
Step 4. Dispose of the meter housing and lithium polymer battery in accordance with applicable local regulations.
What is the electrochemical principle of the BAYER CONTOUR USB blood glucose test?
The test is based on measuring electrical current produced by the reaction of glucose with reagents on the strip electrode. Blood is drawn into the test strip tip by capillary action. Glucose reacts with FAD glucose dehydrogenase (FAD-GDH) and potassium ferricyanide to generate electrons. This produces an electrical current proportional to the glucose concentration in the blood sample, which the meter converts into a quantitative display without requiring user calculation.
What substance limitations and interferences affect the BAYER CONTOUR USB test results?
Blood preservatives: Only heparin-treated blood collection tubes may be used; do not use other anticoagulants or preservatives.
Altitude: Testing up to 10,000 feet (3,048 meters) does not significantly affect results.
Lipemic specimens: Cholesterol levels over 500 mg/dL or triglyceride levels over 3,000 mg/dL may produce elevated readings.
Peritoneal dialysis solutions: Icodextrin does not interfere with the test strips.
Xylose: Do not test during or shortly after a xylose absorption test, as xylose causes interference.
Peripheral blood flow: Capillary testing may not be appropriate in persons experiencing shock, severe hypotension, hyperosmolar hyperglycemia, or severe dehydration.
What are the computer system requirements for the BAYER CONTOUR USB meter?
Operating System: Windows XP SP3 or Windows Vista SP2.
USB Port: High-Power USB port.
Java: Version 1.6.0_07 or higher.
What are the regression accuracy statistics for the BAYER CONTOUR USB system?
In correlation studies with 100 capillary blood specimens across three test strip lots compared against reference plasma glucose (range 58 to 411 mg/dL), the following regression statistics were obtained:
| Lot | Linear Equation (y =) | Correlation Coefficient (r) |
|---|---|---|
| 1 | 0.97x + 4.57 mg/dL | 0.975 |
| 2 | 1.04x – 4.07 mg/dL | 0.971 |
| 3 | 1.01x + 2.68 mg/dL | 0.979 |
What are the precision test results for the BAYER CONTOUR USB system?
Precision testing using whole blood specimens across 100 replicates yielded the following statistics:
| Mean Glucose | Standard Deviation (SD) | Coefficient of Variation (CV) |
|---|---|---|
| 44 mg/dL | 1.0 mg/dL | 2.4% |
| 84 mg/dL | 1.4 mg/dL | 1.7% |
| 132 mg/dL | 1.8 mg/dL | 1.3% |
| 204 mg/dL | 3.5 mg/dL | 1.7% |
| 332 mg/dL | 3.8 mg/dL | 1.2% |
What are the technical specifications of the BAYER CONTOUR USB meter?
| Feature | Specification |
|---|---|
| Test Sample | Whole blood |
| Calibration / Result | Referenced to plasma/serum glucose |
| Sample Volume | 0.6 microliters |
| Measuring Range | 20 to 600 mg/dL |
| Measuring Time | 5 seconds |
| Memory Capacity | Stores most recent 2,000 test results |
| Battery Type | Non-serviceable 280mAh rechargeable lithium polymer battery (3.4V – 4.2V) |
| Meter / Battery Life | 5 years |
| Operating Temperature | 41 to 113 degrees Fahrenheit |
| Operating Humidity | 10% to 93% Relative Humidity |
| Dimensions | 3.82 in (H) x 1.18 in (W) x 0.63 in (T) |
| Weight | 1.2 oz. (34 grams) |
| Onboard Storage | 1 GB portable flash drive with 500 MB partition |
How do you clean and maintain the BAYER CONTOUR USB meter?
Only connect the device to a computer or charger certified to UL 60950-1, and store it in the provided carrying case whenever possible.
The exterior can be cleaned using a moist (not wet) lint-free tissue dampened with a mild detergent or disinfectant solution, such as 1 part bleach mixed with 9 parts water. Wipe dry with a lint-free tissue after cleaning.
Do not allow cleaning solution to run in or around the buttons, USB cap, or test strip port to prevent damage to internal electronics.
What replacement parts are available for the BAYER CONTOUR USB system?
| Part Number | Item Description |
|---|---|
| 82286110 | BAYER CONTOUR USB User Guide |
| 81363014 | USB Extension Cable |
| 81675369 | USB Cap |
| 81675377 | Wall Charger |
CLICK HERE TO DOWNLOAD BAYER CONTOUR USB (01) PDF DOCUMENT