****** Page 1 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Basic information Intended use instruments 092019 ULTRA X moisture analysers are solely designed to quickly measure moisture and dry matter in solid, viscous and fluid substances using thermogravimetric analysis. A sample is weighed while being dried by infra-red heat. Common features for all ULTRA X devices: Short measurement time, gentle and even drying of samples, high reproducibility Automatic or time-controlled measurement shut-off UX 3011/ 301 IQ/ 3011HQ Temperature-controlled infra-red heater 250 W/ 375 W Temperature range: 40 – 200 oc/ 360 oc/600 oc Stainless steel drying tray 111 mm diameter UX 3031 Special device for gypsum industry 2 temperature-controlled heaters for separate measurement of free and chemically bound moisture in gypsum products Temperature range: 40 — 360 oc Stainless steel drying tray 111 mm diameter UX 3081/ 3081WQ Special device for extra large samples, e.g. in the chipboard industry 2 temperature-controlled ceramic heaters Temperature range: 40 — 220 oc/ 300 oc Stainless steel drying tray 245 x 120 mm instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 1

****** Page 2 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments Contents Search term Accessories, supplied Assigning a device ID Automatic mode Automatic shut-off Buttons Calibrating the scale Care and maintenance Changing the print ribbon Communication Consumables Data interface Date and time Display Error description and solution Features ID number Installation site Intended use Interface description Loading the printing paper Maintenance and care Measuring moisture Menu Menu lock Menu settings Packaging Parameter setting Ports Printer options Printer self-test Safety instructions Screen Setting the measurement unit Setting the temperature Start-up Technical data Timer Transport Warranty page 4/6/ 1 3 9 7 7 6 11 16 5 17 13 13 10 6 15 1 9 4 1 13 5 16 12 6/ 14 7 4 7 6 9 16 3 6 8 7 4 17 8 4 4 instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 2

****** Page 3 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Safety instructions Always transport the device in an upright position and protect it from impact. Always operate the device on a level surface, using a vibration-free base. Avoid draughts. Never touch the (luminous) heater when it is hot, as there is a risk of injury! Always disconnect the plug from the mains supply before performing any work on the device (e.g. changing the heater)! Repairs should only be carried out by a qualified technician. Never place flammable material next to the device, as there is a risk of fire! 00 instruments O ‘00000 Please take extra care when using flammable samples! You may need to contact the manufacturer for advice. Please be careful not to drop any sample particles next to the tray support plate. 00000 Carefully remove any particles with a vacuum cleaner. Only use accessories specifically designed for use with ULTRA X devices. Do not move the device during measurement, as this will result in measuring errors. Do not leave any substances in the measuring tray, if they are not being measured. Devices should only be shipped in the original packaging using the original packaging material. ULTRA X moisture analysers are stand-alone units and are not designed to be used in continuous production runs. Any such use is at your own risk. instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 3

****** Page 4 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Warranty 1. instruments a&p instruments undertakes to repair faulty devices within the warranty period, free-of-charge, if the defect is a manufacturing fault. This warranty covers the repair of spare parts and working hours required. It does not cover normal wear-and-tear and soiling and staining of the device. a&p instruments will not accept any freight costs. This warranty does not apply to: improper use of the device, use of the device for purposes other than those stated by a&p instruments, mechanical damage or damage caused by fluids or sample matter, incorrect set up or wrong electrical wiring, mechanical damage of the load cell e.g. as a result of being overloaded Transport, packaging, installation site • Please check the packaging and device for any damage upon delivery. • Please retain the original packaging in case you need to return the goods to a&p instruments. Please only use the original packaging when returning the goods. Accessories supplied: 3081 Measuring probe für 3081 2x Tray tongs Tray holder Start-up 2 trays 1. Place the device on a stable, vibration-free surface, as level as possible and protected from draughts. The most suitable locations are the corners of a room, rooms with only one entrance and heavy tables on concrete floors. Mains cable 65 mm 0000 1. 2. (not for ULTRA X 3081) Start by adjusting the heater to the height marked on the rod. Quartz heaters are 10 mm higher than radiant heaters. To do this, slightly loosen (never completely unscrew) the knurled screw on the back of the reflector holder. Retighten the knurled screw manually. 65 mm c 00000 Now swing the heater to the right or backwards to its rearmost position. Avoid using force to twist the heater beyond its stop point, as this could damage the device. Place the tray holder on the device by turning it slightly in a clockwise direction. (Also remove it by turning in a clockwise direction). UX 3011 HQ no distance UX 3081 instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 4

****** Page 5 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) 4. Next place one of the two drying trays supplied on 1. the tray holder using one or two sets of tongs. The notch in the reflector tray acts as an access gap for easy positioning of the tray. 1. 1. 1. instruments 000000 5. Check that the voltage indicated on the type plate matches the local mains voltage. 6. The power cable is inserted into the device’s power jack and a suitable earthed mains supply. After the power supply has been connected, the device can be switched on or off using the on/ off button on the back of the device. Once the device is switched on, the current software version is shown, followed by the d me weight display. ULTRA X U. 9.3. value nude 1 7. In order to prevent weighing system measuring errors, the device should be switched on 20 minutes before being used to allow for heat compensation. If weight values vary in the display’s milligram range during this period, this is considered normal within this phase. If the device is switched off using the reset (on/ off) button, it does not require a warm-up period, if switched on again with the same button. 8. ULTRA X moisture measuring devices are supplied, so that in most cases substances can be measured immediately. If you are not satisfied with the result, the measurement parameters can be adjusted: Loading the printing paper Remove the printer cover. To do this, press the back of the cover. Insert the paper into the metal slot below the print ribbon and press the line feed button until approx. 20 mm has been inserted. O Thread the paper through the slot in the printer cover and close the printer cover. Changing the print ribbon Remove the printer cover. To do this, press the back of the cover. Remove the used print ribbon and insert a new one. You may need to turn the wheel on the print ribbon in the direction of the arrow until the print ribbon cartridge is positioned correctly. Thermo_printer The ribbon must be in of the paper. Close the printer cover. The ULTRA X Thermo paper is stable for 10 years. Line feed Change paper line feed instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 5

****** Page 6 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Operation Display and buttons temp. C’ tine ULTRA X 3381 mode value (weight) Tare button for setting the T weight readout back to zero Button for resetting the reset processor programme Button for exiting menu esc levels unit up down set esc reset Button for accessing and scrolling through the menu Button for accessing and scrolling through the menu instruments down P2 purity of gypsum Pl not used Button for confirming entries not used Any weight lower than the one displayed when the device is switched on cannot be reset to zero. ports: Menu Factory settings: Power supply Print Conection (only 3081) for separat Printer Interface RS 232 Power supply switch USB-lnterface Temperature: 105 oc / 360 oc) Automatic shut-off with parameters: Weight loss 5 mg/ Query interval 45 secs / Start of query 3 mins Quartz heater 12mg/ 15 secs / 3 mins Buzzer switched on when measuring process is complete Menu settings The device’s Preferences menu can be accessed using the cup’ or ‘down’ button. In this instruction manual, it is assumed that the ‘down’ button down follows the menu item sequence. The menu is operated using a standard structure: UP or DOWN SET ESC Set parameters Confirmation of menu level and if required, return to previous menu level Return to previous menu level and exit from menu If you confirm a wrong entry with ‘set’, press ‘esc’ and ‘set’ to reselect and change the menu item. instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 6

****** Page 7 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Temperature control temp. co Menu Temperature set Taqperat.ure value temp. C’ instruments Value is stored tem . v IR mcde time unit tinie Sscc unit mode up down up down value {weightl unit value (wei$t) value (wei#tl time m it time Temperature Value is stored a 3sa•c.r value (weight) temp. C4 Manu down made unit Automatic shut-off or timer set set tem . C4 meru gut.0/Timer 1 R value (wei$t) set mcde unit value lwei$tl The automatic mode dries the sample until a constant weight is reached and then switches off the heater. tirne Auto Z Timer IR Rut.omat.ic. temp. C4 time up set down Auto Timer IR Rut.omätic The following settings set can be defined here: mode value {weightl unit mcde value unit 1. Weight loss in mg a value should be defined here which corresponds to approx. 0.1 % of the original sample e.g. original sample 10 g = 10 mg Automatic IR Ideight. loss mg set temp. C’ Weight. loss mode mcde t:irne up down unit set value {weghtl unit l.deight. loss value (wei ght) time mg Init value {wei$tl When the weight loss <= the mg weight specified here within the interval period defined below, the measurement is completed, if at least the start of query time limit has expired. 2. Query interval in secs. an interval should be selected here, so that the sample is not dried for too short a period but also not for far too long. A period of 20 to 30 secs is normally adequate. None / 10-99 secs can be selected. temp. time ten-v. C' Automat. i c. time Automatic IR Ideight. loss mg down up down I nter•l.Æl sz. mcde value (wei#tl temp. C' I nt.ervel unit sec.. unit value (weight) unit temp. set time Z sec.. r value (weight) unit mode value {weightl a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 7 ****** Page 8 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments tme Automatic Stert after min. set mode value (weight) m it 3. Start of query in minutes Specify the drying period required for the samples without automatic mode. The factory-set 3 minutes are usually adequate. This value should be increased for extremely low temperatures, so that the device does not switch itself off prematurely. This affects the UX 3031 in particular, when free moisture is being measured at 450C. You can select 1-99 mins. tænp. C' tin-te temp C' tirrte up Stert after min. Stert after min. 8.3— min. value {weightl down unit set temp. Meru mcde mode 83 min. value (weghtl esc unit mode temp. C4 Timer IR Rid-.omät.ic t:irne unit value (wei#tl down made value (wei$tJ unit The timer setting is the same as described for the quartz heater. Values for the luminous and quartz heaters (only UX 3031 ) are entered separately from each other. Both heaters can be controlled separately from each other via automatic mode or the timer. tem . C4 menu Auto/ Timer a mcde set up down Auto/ Timer a value (wei ght) time unit unit value (wei$t) time up down time Timer Min value fweighO set set esc tem . C • Rule/ Timer a Timer set made value Ti mer tem . v Rulo/Timer Quarz Timer unit esc Min unit value (weighO It is possible to select a drying time of 1 180 mins The automatic settings are the same as described for radiant heaters. mcde Mo i sture value lwei$tJ Remaining æi%t Measurement unit tem . C' I-hit &bist.ure tem . C • Manu Unit made time m it ncrease I l¯ree set Moisture Ors su time unit terce value Iwä$tJ up down unit set m cde value (wei ght) ten-v. C' Moisture value fweighQ 1 øøa time down m it mode value (weight} During measurement, the device optionally displays: % moisture % dry matter (T S) % moisture Atro ( i.e. moisture related to the dry matter Dry matter (TS) in g/1000g Weight loss g (V) Residual weight g (R) (current weight) Weight increase (IW) Example: 50 % moisture = 25 % moisture = 75 % moisture = If measuring has stopped, you can see the purity of gypsum with a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. E-Mail: info@apinstruments.de, Internet: apinstruments.de 100 % Atro 33.3 % Atro 300 % Atro 977820 Seite 8 ****** Page 9 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Menu Printer options Printer value fweghQ time set m it For devices without a printer, the number of prints must be set to '1 '! tem . C' Printer Wmber of instruments set value (weghtl tin-te unit prints unit time temp C' Wmber OF mcde mode set esc set value {weightl tem . v NI-nber• of Printer Wmber ot' time unit time prints mcde ten-v. C' value (wei$t) mode value (weghtl Printer Print. interval value (weight) Print. int.er•l..æl evens ID sec. mode value (weight} unit It is possible to select no printout or every 10 — 99 secs. time Printer Print interval esc Menu Printer value (wel ght) unit time unit up to 9 down down up down down value (weight) unit Data output for external printer (other manufacturer) tem . C' Menu Oat-a output. value (wei ght) Print. mode time unit value lwei $t) set temp. C' Data out-Fut value (weight) temp. Cleta output time unit unit unit time unit up down Menu set Using line feed sends an additional LF via the interface (factory setting without LF). telnp- C • Menu Data output. time made value (wei$t) Device ID number Mash i re-l O aza i ne-l O value (weight) value (weight) time unit Enter a number up to 9999 to identify the device Imit set up down down set tem . C4 Mash i re-l O 1234 esc made value (wei$t) unit mode temp. C' value (weightl Menu Mash i ne-l O value (weight) Each digit can be selected separately. down a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. E-Mail: info@apinstruments.de, Internet: apinstruments.de 977820 Seite 9 ****** Page 10 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) temp. co Menu Date and time tem . co Date and Time mode tin-.e set Use 'up'/ 'down' to set the date and press 'set'. Use 'up'/ 'down' to set the month and press 'set' Use 'up'/ 'down' to set the year and press 'set' Use 'up'/ 'down' to set the hour and press 'set' Use 'up'/ 'down' to set the minutes and press 'set' value (weight} unit Oate md Time 24.23. 16 15: 31 mcde tem . c- value lwei#tl unit instruments up set down Oate md Time 24.23. IS 15: 31 r The date and time are shown on the printout mcde Buzzer value lwei$tl temp- C • Menu unit time unit temp. C' menü CAL mode set value {weightl 00000 Use 'up'/ 'down' and 'set' to activate or deactivate the buzzer, then press 'set' followed by 'esc'. set and tray must be on the balance! down Buzzer value (weight) Calibrating the scales 00000 unit Tray carrier Follow the menu instructions and use a 200 g Fl weight to calibrate the scales. Cal i brat.e? set or esc value (weight) time unit set temp. C4 If balance empty press set mcde value lwei$tl unit ten-v. C' set please wait ! time unit temp. CAL done Press set place ZEE then wait! value (weight) mode remove 2Zg than wait! value fweighQ time it ten-v. C' unit time unit set Tray being used must be empty. Menü CAL value (weight) 0000 0 time Imit down mode vah•e fweight) mou If the display shows System error press ! value (we@hO value fwei$tl unit switch the power switch off and on, and try it again. a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. E-Mail: info@apinstruments.de, Internet: apinstruments.de 977820 Seite 10 ****** Page 11 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments Language temp. Sprac.he temp. Menu Language value (wei$t) up down set Germ—I angl i sh temp. C4 set mode value (weightl The first two letters on the second line stand for Automatic (A) and Timer (T). They show the mode set for each heater. Combinations: AA= Automatic quartz heater/ Automatic radiant heater TA = Timer quartz heater / Automatic radiant heater, TT = Timer quartz heater / Timer radiant heater, AT = Automatic quartz heater / Timer radiant heater. Two letters you see only by ULTRA X 3031/D all other models have only one letter, because one Heater. You can also lock the menu to prevent accidental changes to drying parameters. temp. C' Merü The 'CAL' menu option mode value (weight) time can still be accessed when the menu is locked. unit The menu can be locked or unlocked as shown below: Press and hold the 'set' button for 5 secs: Lock mt•3e Oder tam . co m de mew H lue cwegntJ vann ('Ingrit' Position set underscored Position set underscored Enter 5-digit pin Enter 5-digit pin etc. etc. up down up down set set trn . co Lack renu IWElgntl tamp. C+ CFen renu ur. h set untt set unlt If you forget the PIN, you can obtain a new one from a&p instruments. a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. E-Mail: info@apinstruments.de, Internet: apinstruments.de 977820 Seite 11 ****** Page 12 ****** 1. 2. 3. Operating instructions for moisture analysers instruments UX 3001 (UX 3011, UX 3031, UX 3081) Moisture analysis process The device should be connected to the mains supply and switched on with a drying tray. A weight value is 05232 97780, Fax.: 977820 displayed. Set scales to zero using the 'tare' button, display '0.000 g'. 00000 0000 0 UX 3081 1) Evenly distribute approx. 10 g sample matter in the tray and wait for the readout to stop. 2) glow if red heater is on if measurement ends LED' are off if switch „off" Swing the heater over the sample (for UX 3081 , use heater switch). After a short time, the heater switches itself on and the drying phase begins. Header data is now printed for devices with a built-in printer. The sample is dried until it has reached a constant weight; the heater then switches off automatically. The time and measurement value is printed for devices with a printer. If you do not wish to make any measurements, please always ensure that the heater is swung back to its rearmost position! Measurement values are displayed for the entire length of time the heater is positioned above the sample. If measuring has stopped, you can see the purity of gypsum with 000000 The device should not be moved from its position during measurement, as this causes measurement errors. The built-in printer prints the following at the start of the measurement: Model, device ID, date, time, heater temperature, timer or shut-off parameters, initial weight. And after the device has finished measuring, once the heater has been swung back to its rearmost position: Measurement value, measurement time, substance, initials Please ensure drying trays have cooled down before using for measurement. If the measurement ends before it is complete, the display shows: ttne Drying ww stceped ! U-Taa x 3811Hao mcde weig value (weight' uni For substances which are difficult to remove from the tray after being dried, we recommend using foil trays, which can be made quickly and economically from aluminium foil using the ULTRA X foil press. a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.: E-Mail: info@apinstruments.de, Internet: apinstruments.de samp e. Seite 12 ****** Page 13 ****** Operating instructions for moisture analysers instruments UX 3001 (UX 3011, UX 3031, UX 3081) Data interface Weighing data, temperature and drying data are also shown on the display via a standard USB flash drive and serial interface. Depending on the model, a separate UX 3092 printer or a computer can be connected to the device. ca. 2 m Interface description Model: V24 RS 232 serial data transmission and USB flash drive Data cable Moisture balance – DataChannel 5 Transmission rate: Number of bits Parity check Handshake 9600 baud 8 data bits 1 stop bit none none Data transmission using a USB flash drive If the computer receiving the data has an internet connection, the system automatically installs a driver and creates a COM interface, as soon as the computer is connected to the moisture analyser. If this is not the case, the driver must be manually installed in the device manager. Special software is required for transmitting data to a PC: ULTRA X DataChannel Data can be exported in four different data formats: . xls; .html; .xml; .txt The software is extremely easy to install and use: Connect devices, place CD in PC, the installation is self-explanatory. Accessories: Stainless-steel drying tray AV drying tray, large Tray tongs Mains cable Consumables. Order no.. 10000282 10000230 H1000190 Application: Sample containers for UX 3011/ 3031 (2 trays supplied) Sample containers for UX 3081 (2 trays supplied) for lifting the tray (supplied) Mains connection (supplied) Printing paper 58 mm, durable 10 years Printing paper 58 mm Ribbon 1 10000234 II 000006 11000013 for built-in thermo printer for built-in needle printer for built-in needle printer instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 13 ****** Page 14 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Overview of menu setup via the UP button Menu Menu language Language, GERMAN Menu CAL , Calibrate? Set or esc" üa-lbsc.h Engl isch „lf scales are empty, press set!” „Please wait”; „Add 500 g weight, press set” „Please wait”; „Remove 500g weight, then wait” „CAL completed! Press set” Menu buzzer Buzzer activated / deactivated Menu date + time Date. Day . Month . Year Time Hours : Minutes Menu ID Device ID 9999 Device 0000 Menu data output Data output, without LF / with LF Menu printer Printer, Number of print copies Number of print copies, _ Print copies Printer, Interval Printout, every secs Menu Measurement unit Measurement unit, moisture Menu Auto/ Timer instruments Remaining weight igh lass [Jr•w Mo i st.ur•e subst.är.c•.e Moisture ATRO Auto/ Timer quartz, Automatic mode Automatic mode, weight loss mg weight loss mg, _ mg Automatic mode, Interval secs Interval secs, secs Automatic mode, Start mins Start mins, mins Auto/ Timer radiant heater, Timer Auto/ Timer radiant heater, mins Menu Temperature Temperature, radiant heater 1700C (also K heater) Temperature, Quartz heater 360 oc a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. E-Mail: info@apinstruments.de, Internet: apinstruments.de 977820 Seite 14

****** Page 15 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Potential errors instruments Error Display blank: temp. C4 ULTRA X 3E31 mcde value {wei$tl time unit time tem . C4 X 3B.31 Time-out. balarce mcde value (wei#tl unit Scale continuing to run: Drying parameters cannot b entered, only temp C’ CAL mode value {wéghtl tin-te unit Menu lock cannot be unlocked Scale cannot be tared Scale cannot be calibrated: emp- System error press esc ! Ime Imit value (wei ght) Heater will not switch on: Infrared heater will not switch off Measurement not completin Reason Initial load too small or large, Tray support and / or tray missing Location with too much disturbance Draught Area around tray support is dirty Scale is faulty Scale without power Scale is not connected to control card Scale is faulty Location with too much disturbance Draught Area around tray holder is dirty Scale is faulty Menu is locked 5-digit pin for unlocking menu lock is incorrect Weight within minus range Weight difference less than 16 mg or more than 16 mg Initial load too small or large, tray holder and / or tray missing. Location with too much disturbance Draught Area around tray holder is dirty Scale setting is incorrect Scale is faulty System error Calibration process has stopped Heater is faulty Temperature sensor is faulty S202S01 relay is faulty Shut-off settings incorrect Timer settings incorrect Solution Place empty drying dray on centre of tray support, Place on a vibration-free surface Protect from draughts Carefully clean area around tray support Contact a&p instruments Contact a&p instruments Contact a&p instruments Contact a&p instruments Place on a vibration-free surface Protect from draughts Carefully clean area around tray holder Contact a&p instruments Unlock menu lock, contact the person responsible or a&p instruments Contact a&p instruments at a.ukena@apinstruments.de Switch off device, place empty tray on balance and switch on again Less than 16 mg set to zero, press button briefly More than 16 mg tare, press and hold Place empty drying dray on centre of tray holder, Place on a vibration-free surface Protect from draughts Carefully clean area around tray holder Contact a&p instruments Contact a&p instruments Disconnect device from mains and switc on again. Change heater Contact a&p instruments Contact a&p instruments Check shut-off settings Check timer settings instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 15

****** Page 16 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments Error No printout No linefeed Printout distorted Printer not printing a strip in the print graphic Printout too light Care and maintenance Reason Numper of prints = 0 Printer is faulty or no data transmission Linefeed button or printer is faulty Verbindung zum Drucker ist unterbroche No connection to printer Printing unit or printer board is faulty Printing unit is not installed correctly Ribbon is empty Printing unit is dirty Solution Set Number of prints = 1 Contact a&p instruments Contact a&p instruments Contact a&p instruments Contact a&p instruments Contact a&p instruments Contact a&p instruments Contact a&p instruments Change ribbon Contact a&p instruments Please ensure that no pollutants enter the device. Always keep the area around the tray holder clean. Do not use compressed air to clean the device. Do not overload the scales, max. weight 750 g. Protect device from moisture and impact. For ideal care and maintenance, we recommend taking out a service contract. Printer self-test If your printout is incorrect, you will need to run a printer self-test: Disconnect the device from the mains supply using the main switch or plug. Press and hold the linefeed button and connect the device to the mains, then release the linefeed button. The printer will print a self-test: u-TRR X Cryxkeselbsttest U 1.85 JKl..MOCRSIU.U41’ZAäöift& If this printout is correct, the printing unit and printer board are working but the data transmission is not functioning. If the printout is incorrect, there may be several reasons for this error. a&p instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax. E-Mail: info@apinstruments.de, Internet: apinstruments.de 977820 Seite 16

****** Page 17 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments Technical data Scale: Weight range Resolution Sample volume Weighing tray up to 220 g/ 400g (models 3081) 0.1 g max. 95 cm3 / 506 cm3 (models 3081) 110 mm round/ 230 x 110 mm (models 3081) For moisture analysis: Initial weight any weight from approx. 5 g, recommended from 10 g Heater capacity Infrared heater 250 watts Quartz heater 375 watts Ceramic heater 2x 250 watts (3081 )/ 2x 400 watts (3081 WQ) Automatic shut-off Automatic drying to a constant weight, using 3 factors for I heaters and Q heaters, Temperature Measuring units which can be set separately: Query start, query interval, weight loss/ query interval Infrared heater adjustable from 400C to 170 oc Quartz heater adjustable from 105 oc to 360 oc (301 IQ)/ 1050C to (3011 HQ) Ceramic heater adjustable from 400C to 180 oc (3081 1050C to 360 oc (3081 QW) % moisture % dry matter % ATRO moisture g solid matter/ kg Reading accuracy 0.1% readable Communication: Data terminal Data output V24 RS 232 all weight and measuring data meet GLP requirements with a start time, optional variable interval printout, date, time, total measuring time etc. Using ULTRA X DataChannel software can be transmitted to a PC as a .xls file in Microsoft Excel, .html file, .xml file and .txt file. For operating the device: Mains connection 230 V 2%/ + 15 % Mains frequency 48-63 Hz Power input Dimensions Weight 250 watts 850 watts, depending on the model approx. B 385 x D 275 x H 420 mm approx. 9 kg/ 15 kg (3081 model) Technical data subject to change without notice ULTRA X moisture analysers are stand-alone units and not suitable for use in continuous production runs. We shall not accept any liability for incorrect readings! instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 17

****** Page 18 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) Accessories: Stainless steel drying tray AV drying tray, large Tray tongs Mains cable Order no.. 10000282 10000288 10000230 H1000019 Use: Sample carriers UX 3011/ 3031 (2 trays supplied) Sample carriers UX 3081 (2 trays supplied) for lifting the tray (supplied) Mains connection (supplied) for calibrating the Calibration weight Fl 200g 10000403 Aluminium foil 130 mm x 0.1 mm Aluminium foil 130 mm x 0.3 mm Aluminium foil 130 mm x 260 mm scales H1000016 H1000017 H1000013 for placing on the drying tray for placing on the drying tray for placing on the drying tray Foil press Software Consumables. Printing paper 58 mm Printing paper 58 mm Ribbon Infrared heater 250 W Quartz heater 375 W Ceramic heater 250 W Ceramic heater 400 W Ceramic heater 600 W 10000062 for shaping aluminium foil DataChannel for transmitting data to a PC instruments durable 10 years 110000234 11000006 11000013 H1000013 10000139 10000405 10000318 for built-in thermo printers for built-in needle printers for built-in printers Spare part Spare part Spare part Spare part Spare part instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 18

****** Page 19 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 19

****** Page 20 ****** Operating instructions for moisture analysers UX 3001 (UX 3011, UX 3031, UX 3081) instruments instruments, Albert-Schweitzer-Straße 16, 32758 Detmold, Tel.:05232 97780, Fax.: 977820 E-Mail: info@apinstruments.de, Internet: apinstruments.de Seite 20