FREE ENGLISH AGILENT 4107 (01) PDF USER GUIDE
FREE ENGLISH AGILENT 4107 (01) PDF USER MANUAL
FREE ENGLISH AGILENT 4107 (01) PDF OWNER GUIDE
FREE ENGLISH AGILENT 4107 (01) PDF OWNER MANUAL
FREE ENGLISH AGILENT 4107 (01) PDF REFERENCE GUIDE
FREE ENGLISH AGILENT 4107 (01) PDF INSTRUCTION GUIDE
FREE ENGLISH AGILENT 4107 (01) PDF REFERENCE MANUAL
FREE ENGLISH AGILENT 4107 (01) PDF INSTRUCTION MANUAL
FREE ENGLISH AGILENT 4107 (01) PDF OPERATING INSTRUCTIONS
CLICK HERE TO DOWNLOAD AGILENT 4107 (01) PDF MANUAL
If this is not the document you want for this product, click here to see if we have any other documents for this product.
AGILENT 4107 (01) PDF SUMMARY:
What safety practices should I follow when operating the Agilent 4107 Nitrogen Generator?
Careless, improper, or unskilled use of the nitrogen generator can cause death or serious injury, and/or severe damage to equipment and property. The nitrogen generator covers are designed to prevent inadvertent contact with any potential hazards, but if the nitrogen generator is used in any manner not specified, this protection may be impaired. Read each safety topic thoroughly before attempting to operate the nitrogen generator and always operate the nitrogen generator in accordance with the provided safety practices. If there is uncertainty about the effects of any liquid, gas, or chemical on the nitrogen generator, consult an Agilent field service engineer or approved representative before use.
What should I know about the air supply for the nitrogen generator?
To avoid shortening its life, the nitrogen generator requires a supply of clean, dry, oil-free air. To assist in achieving this, the nitrogen generator has integral filters attached to the side of the unit. Depending on the quality of the air supply and the environmental conditions, additional driers and filters may be required. Refer to the Agilent MP-AES Site Preparation Guide for further details and read the instructions regarding filters under the Maintenance chapter of this guide. Under no circumstances should the nitrogen generator be operated without filters or with filters in a saturated condition. Under no circumstances should the nitrogen generator be connected to an air supply containing oil, as permanent damage to the nitrogen generator will occur. Contamination of the sorbent material with oil will invalidate the warranty.
What are the compressed gas hazards associated with the nitrogen generator?
Air compressors must be located and connected strictly in accordance with local safety codes and regulations. All compressed gases (other than air) can create a hazard if they leak into the atmosphere. Even small leaks in gas supply systems can be dangerous. Any leak (except that of air) can result in an oxygen-deficient atmosphere which can cause asphyxiation. The area in which cylinders are stored and the area surrounding the instrument must be adequately ventilated to prevent such gas accumulations. Do not directly inhale the output gas from the nitrogen generator. Periodically inspect the hoses to make sure they are in good condition and properly connected. Keep air hoses off the floor where they can be tripping hazards or become damaged. When disconnecting or connecting the hoses, make sure the hose is depressurized first. Always wear safety glasses when connecting or disconnecting gas lines.
What are the electrical hazards I should be aware of?
Panels or covers which are retained by screws on the nitrogen generator may be opened only by Agilenttrained, Agilent-qualified or Agilentapproved field service engineers (unless specified otherwise). The nitrogen generator is powered by an external AC-DC power supply. If the enclosure of this power supply is damaged it should be replaced immediately. Check the power cord for damage and replace it if it is visibly worn or damaged. If necessary, replace the power cord only with a cord equivalent to the one specified in the site preparation guide. Consult the manuals or product labels supplied with your air compressor for safety information and to determine which parts of it are operator-accessible.
What other precautions should I take?
The nitrogen generator weighs approximately 110 kg (242.5 lb). To avoid injury to personnel or damage to the instrument or property, always use a suitable mechanical lifting device to move the nitrogen generator. Use only Agilent-supplied spares with the nitrogen generator.
What warning symbols are used and what do they mean?
A triangular symbol indicates a warning. The meanings of the symbols that may appear alongside warnings in the documentation or on the instrument itself are as follows:
- Eye hazard
- Heavy weight (danger to feet)
- Heavy weight (danger to hands)
- Noxious gases
The following symbol may be used on warning labels attached to the instrument or in this guide. When you see this symbol, refer to the relevant operation or service manual for the correct procedure referred to by that warning label.
What symbols appear on the instrument and what do they indicate?
- Direct current
- Compliance with EU directives
- Certification to CSA 61010.1 and UL 61010-1
- Compliance with South Korean Electromagnetic Compatibility (EMC) law
- Compliance with ACMA requirements
- Compliance with WEEE ROHS directive
With what directives does the Agilent 4107 Nitrogen Generator comply?
The Agilent 4107 Nitrogen Generator has been designed to comply with the requirements of the Electromagnetic Compatibility (EMC) Directive and the Low Voltage (electrical safety) Directive (commonly referred to as the LVD) of the European Union. Agilent has confirmed that each product complies with the relevant Directives by testing a prototype against the prescribed EN (European Norm) standards. Proof that a product complies with these directives is indicated by the CE Marking appearing on the rear of the product, and the documentation package that accompanies the product containing a copy of the Declaration of Conformity. The Declaration of Conformity is the legal declaration by Agilent that the product complies with the directives listed above, and shows the EN standards to which the product was tested to demonstrate compliance.
What is the classification of the Agilent 4107 Nitrogen Generator regarding electromagnetic compatibility?
The Agilent 4107 Nitrogen Generator is Group 1 ISM equipment, which contains all ISM equipment in which there is intentionally generated and/or used conductively coupled radio- frequency energy which is necessary for the internal functioning of the equipment itself. It is also Class A equipment, which is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. This device complies with the requirements of CISPR11, Group 1, Class A as radiation professional equipment. Therefore, there may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted as well as radiated disturbances.
What conditions apply to the operation of the device regarding electromagnetic compatibility?
Operation is subject to the following two conditions:
- This device may not cause harmful interference.
- This device must accept any interference received, including interference that may cause undesired operation.
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try one or more of the following measures:
- Relocate the radio or antenna.
- Move the device away from the radio or television.
- Plug the device into a different electrical outlet, so that the device and the radio or television are on separate electrical circuits.
- Make sure that all peripheral devices are also certified.
- Make sure that appropriate cables are used to connect the device to peripheral equipment.
- Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance.
Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
What is the South Korean Class A EMC declaration?
This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
What is the Canadian ICES/NMB-001 compliance?
This ISM device complies with Canadian ICES-001.
What site preparation is required for the Agilent 4107 Nitrogen Generator?
Prior to receiving the Agilent 4107 Nitrogen Generator you will have been provided with an Agilent MP-AES Site Preparation Guide, which describes the environmental and operating requirements of the MP-AES instrument and nitrogen generator. The installation site must be prepared according to these instructions before the nitrogen generator can be installed. The site preparation guide should be kept for future reference.
What documentation is provided with the Agilent 4107 Nitrogen Generator?
You have been provided with the following documentation to help you set up and operate the Agilent 4107 Nitrogen Generator:
- This User’s Guide, with safety practices and hazards information, instructions for finding information about installing and maintaining the components of the nitrogen generator and a brief operation overview.
- The Agilent MP-AES Site Preparation Guide with safety information and site preparation instructions.
What is the purpose of Notes and Tips in the documentation?
A Note is used to give advice or information. A Tip is used to give practical hints to help you achieve the best possible performance from your instrument.
Who should install the Agilent 4107 Nitrogen Generator?
The Agilent 4107 Nitrogen Generator must be installed by an Agilent-trained, Agilent-qualified or Agilent-authorized field service engineer. The form in the site preparation guide should have been completed and returned, stating that you have prepared the laboratory in accordance with the requirements detailed in that guide. Details for unpacking the instrument and what to do in case it has been damaged in transit are also outlined in the site preparation guide.
What are the warnings associated with unpacking and moving the nitrogen generator?
Heavy Weight Danger to feet and hands. The nitrogen generator weighs approximately 110 kg (242.5 lb). To avoid injury to personnel or damage to equipment, always use a suitable lifting device when moving the instrument.
What are the steps for unpacking the nitrogen generator?
- Remove the external packaging.
- Unscrew the four screws attaching the side rails to the pallet on each side of the nitrogen generator.
- While still attached to the rails, use an overhead or portable hand crane and slings, thread the slings between the wheels and the substructure, both front and back and then lift the nitrogen generator.
- Remove the pallet from underneath and then lower the nitrogen generator to the floor.
- Once the instrument is completely off of the pallet, unscrew the three screws attaching the rail to the nitrogen generator.
- The nitrogen generator can then be wheeled into position.
Position the nitrogen generator to allow easy access to the power button located on the front of the instrument.
How do I connect the nitrogen generator to the air supply?
The air supply hose is attached at the factory to the nitrogen generator.
- Ensure that the air supply meets the requirements specified in the Agilent MP-AES Site Preparation Guide.
- Using the supplied fitting, connect the free end of the hose to the air supply connection. PTFE tape may be used to ensure a leak-free connection.
The supplied fitting can be removed from the hose using the push-release feature.
How do I connect the nitrogen generator output to the MP-AES instrument?
- Ensure a 1/4 inch Swagelok fitting is attached to the end of the gas supply hose.
- Connect one end of the nitrogen gas supply hose to the nitrogen output on the rear of the nitrogen generator and the other end to the nitrogen inlet on the MP-AES instrument or EGCM.
The External Gas Control Module (EGCM) will need a separate nitrogen generator to the one supplying nitrogen to the MP-AES instrument.
- Ensure that the fittings are tightened firmly.
How do I connect the electrical power to the nitrogen generator?
- Plug the small DC output connector from the power supply to the receptacle on the rear of the nitrogen generator.
- Plug the AC supply cable into the power supply.
Plug the AC supply cable in to an outlet complying with the nitrogen generator requirements specified in the Agilent MP-AES Site Preparation Guide.
What should I do before operating the nitrogen generator?
Before starting the nitrogen generator, carefully read the ‘Safety Practices and Hazards’ section at the front of this guide and ensure that the laboratory is set up according to the details specified in the Agilent MP-AES Site Preparation Guide. In operation, the nitrogen generator cycles approximately every 40 seconds. At the end of each cycle the oxygen enriched air is vented to the room from a vent in the bottom of the nitrogen generator. A noise may be heard as this gas is released.
How do I turn on the nitrogen generator?
- Ensure that the gas filters are installed correctly and do not need replacing.
- Ensure the air and nitrogen hoses are connected to the nitrogen generator and that the air supply is turned on.
- Check the pressure indicated on the pressure gauge of the filter regulator unit on the right side of the nitrogen generator meets the specification.
- Switch on the nitrogen generator. The green LED will illuminate to indicate that it is powered and operating.
If the pressure on the pressure gauge does not meet the specification, check the air supply system. The regulator is preset and is not user-adjustable.
What is the caution and warning related to turning on the nitrogen generator?
Do not use the gas produced by the nitrogen generator to run the plasma until the nitrogen generator has been purged.
Eye hazard Do not disconnect the air supply or nitrogen outlet hoses without first depressurizing the nitrogen generator. The rapid escape of gas could cause injury. Always wear safety goggles and other appropriate safety clothing when disconnecting the gas supply.
How do I select the nitrogen generator for use in the MP Expert software?
- Turn on the instrument exhaust system.
- Turn on the MP-AES instrument.
- Turn on the nitrogen generator.
- From within the MP Expert software, click Instrument > Configuration.
- Select Use nitrogen generator.
- Close the Instrument window.
Why and how do I purge the nitrogen generator?
The nitrogen generator does not produce nitrogen of the required purity immediately after connection and turning on. A period of purging is necessary to produce the required purity of nitrogen gas.
- Turn on the instrument exhaust system.
- Turn on the MP-AES instrument.
- Turn on the nitrogen generator.
- Start the MP Expert software.
- Ensure that the nitrogen generator is enabled for use by selecting Use nitrogen generator in Instrument > Configuration.
- Click Status.
- Click Start Purge.
- The system will purge for up to 20 minutes.
The nitrogen generator is now ready for use. The purge will stop once the O2 sensor indicates a safe level for ignition. If the 20 minute time out is reached, an error will be shown and automatic purge will be disabled until reconnection of the instrument. From a powered down state the nitrogen generator may require up to 20 minutes of purge before it delivers nitrogen of the required purity.
When and how do I turn off the nitrogen generator?
The nitrogen generator does not need to be turned off for short idle periods. It uses a small amount of electric power and air when the connected instrument or accessory is off but the nitrogen generator is left running. This gives the fastest possible start-up when the idle period is over. For longer idle periods the nitrogen generator may be turned off to save air and electric power.
- Press the power button on the front panel to turn off the power to the nitrogen generator.
- If desired, close the valve between the air supply and the nitrogen generator. The nitrogen generator does not use any air when turned off.
The nitrogen generator will require purging when restarted.
When and how do I depressurize the nitrogen generator?
Eye hazard Do not disconnect the air supply or nitrogen outlet hoses without first depressurizing the nitrogen generator. The rapid escape of gas could cause injury. Always wear safety goggles and other appropriate safety clothing when disconnecting the gas supply. To replace the filters you must depressurize the nitrogen generator before disconnecting the air supply or nitrogen outlet hoses.
- Ensure that the nitrogen generator is switched on.
- Shut off the valve between the air supply and the nitrogen generator.
- Allow 10 minutes for the reservoir in the nitrogen generator to empty.
The nitrogen generator may now be safely disconnected or the filters replaced.
What maintenance is required for the Agilent 4107 Nitrogen Generator?
This chapter includes the Agilent 4107 Nitrogen Generator maintenance requirements that may be carried out by an operator. Any maintenance procedures not specifically mentioned in this chapter should be carried out only by Agilent-trained, Agilent-qualified or Agilent-authorized field service engineers. This section refers to maintenance procedures for the nitrogen generator. Refer to your instrument guide for the Agilent MP-AES instrument maintenance procedures, and to the MP Expert Help for the maintenance procedures for any accessories you ordered.
What routine maintenance is required?
The following parts of the nitrogen generator require routine maintenance:
Daily
- Check the filter sight glasses for collected liquid and drain if required
Monthly
- Check power supply and cords for damage
- Check gas tubing for leaks
Yearly
- Confirm the air supply system still complies with all applicable local regulations
- Replace the gas filters
How do I drain liquid from the filter housing?
- Ensure you have a container to catch the liquid as it begins to drain to avoid spills.
- Loosen the screw at the bottom of the filter housing. Liquid will begin to drain immediately.
Disposal of all chemical waste must comply with any rules and/or regulations that may be imposed by the local authorities.
How do I replace the gas filters?
The compressed air filter assembly for the nitrogen generator consists of three separate different filters. The first (closest to the pressure gauge) is a large particle filter, the middle filter is a small particle filter and the final filter is for oil vapor removal. The filter locations are not interchangeable. The filter must be replaced in the correct location; otherwise damage to the nitrogen generator may occur.
- Depressurize the nitrogen generator.
- Unscrew and remove the filter housing, this should be hand tight but may require the use of a spanner (24 mm).
- Unscrew the filter and replace it.
- Replace the filter housing. Repeat for each filter.
Before using the system, purge the nitrogen generator.
How do I check the gas supply tubing for leaks?
When any gas filter is changed, inspect all gas hoses. Replace any hoses showing signs of damage or deterioration.
Eye Hazard Pressurized gases pose a potential risk to sight. It is essential that appropriate protective eyewear be worn at all times when maintaining gas lines that may still be under pressure.
Pressurize the system and test all connections for leaks using a commercial leak testing solution, a mild detergent, or an electronic leak detector.
When changing air supply:
- check all gas supply hoses for leaks
- using the air pressure gauge on the filter regulator unit, check the air pressure meets the specifications listed in the Agilent MP-AES Site Preparation Guide
- ensure that shut-off valves are functioning correctly
What spare parts are recommended?
For details on additional spare parts and consumables ordering information not listed here, refer to the Agilent Technologies website: http://www.agilent.com
The following is a list of recommended spares to keep on hand to minimize down time during maintenance and repairs:
| Part number | Part Description |
|---|---|
| G8001-64004 | Filter kit |
| G8001-64003 | Hose kit |
To replace the items listed above, you must use Agilent-manufactured parts, which can be ordered online from the Agilent website or through your local sales representative.
How can I get technical support or troubleshooting information?
For technical support contact information, refer to the Agilent Technologies website for details: http://www.agilent.com
For troubleshooting information, refer to the MP Expert Help. http://www.agilent.com
CLICK HERE TO DOWNLOAD AGILENT 4107 (01) PDF MANUAL