Free English PRO’SKIT MT-4011 (01) PDF Document. Free English PRO’SKIT MT-4011 (01) PDF User Guide. Free English PRO’SKIT MT-4011 (01) PDF User Manual. Free English PRO’SKIT MT-4011 (01) PDF Owner Guide. Free English PRO’SKIT MT-4011 (01) PDF Owner Manual. Free English PRO’SKIT MT-4011 (01) PDF Reference Guide. Free English PRO’SKIT MT-4011 (01) PDF Instruction Guide. Free English PRO’SKIT MT-4011 (01) PDF Reference Manual. Free English PRO’SKIT MT-4011 (01) PDF Instruction Manual. Free English PRO’SKIT MT-4011 (01) PDF Operating Instructions.
CLICK HERE TO DOWNLOAD PRO’SKIT MT-4011 (01) PDF DOCUMENT
If this is not the document you want for this product, click here to see if we have any other documents for this product.
How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.
What are the specifications of the PRO’SKIT MT-4011 digital thermometer?
The specifications for the PRO’SKIT MT-4011 are as follows:
Display: 4 digital liquid crystal display.
Resolution:
• 0.1°C / 0.1°F below 1000°C.
• 1°C / 1°F above 1000°C.
Accuracy: (At 23.5°C <80%RH)
• (0.1% reading+0.7°C) -100°C~1372°C
• (0.2% reading+1°C) -250°C~-100°C
• (0.1% reading+1.4°F) -148°F ~2501°F
• (0.2% reading+2°F) -418°F ~-148°F
Measurement Range: -200°C~1372°C (-328°F ~2501°F)
Sampling: 1 time/sec.
Battery life: approx 550 hours.
Auto power: Yes.
Low battery indicator: Yes.
Operating environment: 0°C~50°C (32°F~122°F) 0~80%RH.
Storage environment: -20°C~60°C (-4°F ~140°F) 0~80%RH.
Dimension: 125x80x40mm (L x W x H)
Weight: approx 257g. (Not including batteries)
Accessories:
• (A) K-type thermocouple lead x 2 pieces.
• (B) Users manual x1 piece.
What are the main parts of the PRO’SKIT MT-4011?
The main parts of the PRO’SKIT MT-4011 are:
1. LCD display
2. Power button
3. Function key
4. Input terminal
5. Battery cover
How do I turn the PRO’SKIT MT-4011 on or off, and what is the auto power-off feature?
Press the power button to turn the power on or power off. The meter features an auto power-off function that will turn the device off after 20 minutes if there is no operation or PC communication.
How can I switch the temperature unit between Celsius (°C) and Fahrenheit (°F) on the PRO’SKIT MT-4011?
Press the “°C/°F” button once. The unit will convert from °C to °F, or from °F to °C.
How do I use the HOLD function on the PRO’SKIT MT-4011?
Press the “HOLD” button to enter Hold mode. An “H” will be displayed on the LCD, and the current reading will be frozen. Press “HOLD” again to exit the Hold mode.
How can I disable or enable the auto power-off feature on the PRO’SKIT MT-4011?
To disable the auto power-off mode, press and hold the “HOLD” button for 2 seconds, then press “HOLD” again. The “A” symbol on the LCD will disappear.
To re-enable the auto power-off mode, press and hold the “HOLD” button for two seconds, then press “HOLD” again. The “A” symbol will reappear on the LCD.
How do the MAX/MIN and REL functions work on the PRO’SKIT MT-4011?
MAX/MIN: This function allows you to read the maximum or minimum recorded temperature.
REL (Relative Readout): This function shows the difference between the current temperature and a stored reference temperature.
Step 1: With a stable temperature reading (e.g., T1=25.3, T2=25.2), press the REL button. The display for T1 and T2 will reset to 0.0, and the initial temperatures are saved as the reference.
Step 2: As the temperature changes (e.g., T1=26.1, T2=25.7), the display will show the difference from the saved reference values. In this example, T1 will show 0.8 (26.1 – 25.3) and T2 will show 0.5 (25.7 – 25.2).
How do I use the MEM (memory) function to save data on the PRO’SKIT MT-4011?
The MEM function allows you to save a maximum of 10 data readings.
• Save Data: Press the MEM button to save the current reading.
• Memory Full: If you try to save more than 10 readings, “OU” will appear on the display, indicating the memory is full.
• Clear Memory: To clear all saved data, press and hold the MEM button for 2 seconds. “CLEA” will appear on the screen.
What are the LIMIT and T1-T2 functions on the PRO’SKIT MT-4011?
LIMIT: Press the limit button to turn the Hi/Lo alarm function ON or OFF.
T1-T2: Press this button to display the temperature difference between the T1 and T2 inputs.
How do I set the Hi/Lo alarm temperatures on the PRO’SKIT MT-4011?
The Hi/Lo alarm will sound when the main display temperature goes above the set Hi temperature or below the set Lo temperature. To set these values:
1. Press the SHIFT key, then press the HOLD key to enter the setting mode.
2. Use the HOLD (MAX/MIN) key to move the position of the flashing digit (the adjustable digit).
3. Use the REL (°C/°F) key to set the value of the flashing digit.
4. Use the MEM key to turn a decimal point ON or OFF if needed.
5. Press the SHIFT key to go back to the main display.
What are the temperature setting ranges for the Hi/Lo alarm on the PRO’SKIT MT-4011?
| Hi | Lo |
| 1372°C (2501°F) | -200°C (-328°F) |
What should I do when the low battery symbol appears on the PRO’SKIT MT-4011 display?
When the battery voltage is below the proper operation requirement, a “+ -” symbol will appear on the LCD. This indicates that the battery needs to be replaced with a new one.
How do I replace the battery in the PRO’SKIT MT-4011?
To replace the battery:
1. Use a screwdriver to take the screw off the battery cover.
2. Remove the battery cover.
3. Remove the old battery and then put in a new battery.
4. Put the cover back on and secure it with the screw.
What are the input protection and temperature jack specifications for the PRO’SKIT MT-4011?
Input protection: The temperature jack is protected against a maximum voltage of 24 VDC or VAC.
Temperature jacks: The jacks are designed for the insertion of standard small thermocouple jacks, which have a center spacing of 7.9mm between the two prongs.