COVIDIEN PURITAN BENNETT 980 (01) Image

Free English COVIDIEN PURITAN BENNETT 980 (01) PDF document. Download the PDF, use the online viewer, or browse the FAQs for easy troubleshooting.


CLICK HERE TO DOWNLOAD COVIDIEN PURITAN BENNETT 980 (01) PDF DOCUMENT

If this is not the document you want for this product, click here to see if we have any other documents for this product.

How do I access and use the on-screen help on the COVIDIEN PURITAN BENNETT 980 ventilator?

The COVIDIEN PURITAN BENNETT 980 ventilator is equipped with an on-screen help system called tooltips. If a tooltip is available for an item on the screen, a glowing blue outline will appear around it.

To access tooltips:

1. Touch the item in question for at least 0.5 seconds, or drag the help icon (the question mark icon at the lower right of the screen) to the item. A tooltip with a short description will appear.

2. Touch “more” on the dialog to see an expanded description.

3. Touch “close” to close the dialog, or it will fade away after 5 seconds.

Note:

Dragging the help icon causes the tooltip to display in its unexpanded state.

Dragging the help icon and pausing causes a tooltip to display. Continuing to drag to another item will dismiss the last tooltip and display a new one.


What should I do if the GUI touch screen on my COVIDIEN PURITAN BENNETT 980 is unresponsive or blank?

If the graphical user interface (GUI) display becomes blank, unreadable, or unresponsive, first assess the patient to confirm ventilation is in progress. Examine the status display on top of the breath delivery unit (BDU) to confirm ventilator function. If your clinical judgment determines the patient’s status is satisfactory and ventilation is continuing as set, follow the steps for the specific condition below.

1. Unresponsive screen with display still visible and waveforms active:

If the GUI is unresponsive, inaccurate, or has unintended responses, reset the touch screen.

To reset the touch screen:

1. Touch the display lock key on the GUI bezel to lock the screen. A locked padlock icon will appear.

2. Touch the display lock key again. A progress bar will appear below the padlock icon, after which it will unlock, indicating a successful reset.

Alternatively, ensure no patient is connected and power cycle the ventilator.

Note: Do not touch the screen during the unlock period.

2. Momentary Blank GUI Screen / GUI Transient Reset:

If the GUI resets spontaneously, the status display will show a count-down timer. Assess the patient and wait for the GUI reset to finish (approximately 30 seconds).

3. Blank GUI Screen / GUI Failure:

An error message will appear on the status display. If this occurs, confirm that ventilation continues as set. Provide alternate ventilatory support or transfer the patient to another ventilator according to your institution’s protocols as soon as appropriate. The ventilator should be serviced before being used on patients again.

If, in your clinical judgement, the patient’s status is unsatisfactory, consider patient transfer to another ventilator or provide an alternate source of ventilation in accordance with your institution’s protocols. The ventilator should be replaced as soon as possible and repaired by qualified service personnel.


How do I set up the COVIDIEN PURITAN BENNETT 980 ventilator for a new patient?

To set up the ventilator for a new patient, follow these steps:

1. Connect the ventilator to the electrical and gas supplies.

2. Connect the appropriate patient circuit (adult/pediatric or neonatal) to the ventilator.

3. Turn the ventilator on using the power switch.

4. Before ventilating, run a Short Self Test (SST) to calculate the compliance and resistance with all items included in the patient circuit.

To ventilate a new patient:

1. Touch New Patient on the GUI screen. The New Patient settings screen will appear.

2. Enter the patient’s Predicted Body Weight (PBW) or gender and height.

3. If the default ventilator settings are appropriate, touch Start to confirm. Otherwise, touch a ventilator setting and turn the knob to adjust the parameter. Continue this for all necessary parameters.

4. Touch Accept or Accept ALL to confirm the changes.

5. Connect the circuit to the patient’s airway to start ventilation.


How do I perform an SST (Short Self Test) on the COVIDIEN PURITAN BENNETT 980?

A Short Self Test (SST) must be run before ventilating a new patient, when changing the patient circuit or its components, or when adding/removing accessories. Always disconnect the patient before running SST.

Required Equipment:

• Proposed patient circuit for ventilation

• Accessories (e.g., water traps)

• Exhalation filter and condensate vial

• Humidifier, if applicable

• A number 1 stopper to block the patient wye

• Two gas sources (air and oxygen) connected to the ventilator at a pressure between 241.3 kPa and 599.8 kPa (35 psi and 87 psi)

To run SST:

1. Ensure a patient is not connected to the ventilator.

2. Ensure the breathing circuit wye is not attached to a test lung or covered.

3. Turn the ventilator on. The ventilator runs a Power On Self Test (POST). Ensure it is operating on full AC power.

4. Wait at least 15 minutes for the ventilator to warm up.

5. At the ventilator startup screen, touch SST or the configure icon (wrench).

6. Connect the patient circuit, filters, condensate vial, and all accessories. Ensure the patient wye is not blocked.

7. Touch Run All SST to perform all tests or Run Leak Test for only the leak test.

8. Touch Accept to continue or Cancel to go back.

9. After accepting, touch the Circuit Type button corresponding to the patient circuit type (adult, pediatric, or neonatal).

10. Touch the Humidification Type button corresponding to the type used. If a humidifier is used, touch Humidification Volume and enter the correct volume.

11. Touch Accept to start SST.

12. Follow the on-screen prompts. Certain tests require operator intervention.

13. After each test, the ventilator displays the results. If a test fails, you can choose to repeat the test or perform the next one.


How do I interpret the SST (Short Self Test) results on the COVIDIEN PURITAN BENNETT 980?

When the SST is complete, the status screen displays the individual test results. The overall outcomes are as follows:

Final Outcome Meaning Response
PASS All SST tests passed. Touch Patient Setup to set up the patient for ventilation.
OVERRIDDEN The ventilator detected one or more faults. Choose to override the Alert status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards. Check the patient circuit to determine the problem or restart SST with a different patient circuit.
FAIL One or more critical faults were detected. The ventilator enters the SVO state and cannot be used for normal ventilation until SST passes. Check the patient circuit to determine the problem or restart SST with a different patient circuit.

WARNING: Overriding an Alert in SST may result in ventilator performance outside of the stated specification for accuracy. Choose to override the Alert status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards.


How do I install or replace the batteries in the COVIDIEN PURITAN BENNETT 980?

WARNING: One primary battery must be installed at all times in the BDU’s primary battery slot for proper ventilator operation. The ventilator will not complete the startup process without the primary battery installed.

The ventilator uses primary and optional extended batteries. The primary battery is in the rearward receptacle on the right side of the BDU, and the extended battery is in the forward receptacle.

To install or replace the primary battery in the BDU or compressor:

1. Check the battery’s charge level by pressing the charge level button on the battery. Five green LEDs indicate ≥90% capacity. A red LED indicates a fault.

2. If the charge is sufficient, orient the battery as shown in the manual. The receptacle towards the rear houses the primary battery.

3. The primary battery is fastened with a thumbscrew. Loosen the thumbscrew approximately four to five turns to allow installation.

4. Insert the battery into its receptacle until it clicks. It will only fit one way.

5. Tighten the thumbscrew to secure the battery.

To remove the primary battery, loosen the thumbscrew and slide the battery ejector to the left to eject it.

To install or remove an extended battery in either the BDU or compressor:

1. Properly orient the battery.

2. Push the battery into the forward receptacle until it clicks, indicating it is latched.

To remove the extended battery, slide the battery ejector to the left. The battery will eject. There is no thumbscrew for extended batteries.


How do I perform a manual purge of the Proximal Flow Sensor on the COVIDIEN PURITAN BENNETT 980?

A manual purge may be performed any time the sensor lines contain excessive condensation, moisture, or secretions.

To perform a manual purge:

1. Touch the Configure icon in the constant access icons area of the GUI.

2. Touch the Options tab. A screen appears containing the Installed Options and Prox tabs.

3. Touch the Prox tab. The Prox Setup screen appears.

4. Touch START that appears next to the text “Prox Manual Purge: To begin touch the Start button”. During the purge, a message appears in the GUI prompt area stating the purge process is being performed.


What are the recommended steps for cleaning the exterior surfaces of the COVIDIEN PURITAN BENNETT 980?

External surfaces of the GUI, BDU, and compressor base should be cleaned periodically.

To clean the GUI, BDU, or compressor base:

1. Moisten a soft cloth with one of the approved disinfectants or use Sani Cloths (PDI, Inc.).

2. Wipe the GUI, BDU, and compressor base, removing any dirt or foreign substances.

3. Dry all components thoroughly.

4. If necessary, vacuum any cooling vents on the GUI and BDU with an electrostatic discharge (ESD)-safe vacuum to remove any dust.

Approved cleaning agents:

• Mild dish washing detergent solution

• Isopropyl alcohol (70% solution)

• Bleach (10% solution)

• Window cleaning solution (isopropyl alcohol and ammonia)

• Ammonia (15% solution)

• Hydrogen peroxide (3% solution)

• Formula 409™ cleaner (Clorox Company)

• CaviCide™ surface disinfectant (Metrex Research Corporation)

• Control III™ germicide (Maril Products, Inc.)

• Mr. Muscle™ Window & Glass (SC Johnson)

• Sani Cloth™ (PDI, Inc.)

• [Propan-2-ol, Isopropanol, Isopropyl Alcohol]

Cautions:

• Do not allow liquid or sprays to penetrate ventilator openings or cable connections.

• Do not attempt to sterilize the ventilator with ethylene oxide (ETO) gas.

• Do not use pressurized air to clean or dry the ventilator.

• Do not submerge the ventilator or pour cleaning solutions over or into it.


How do I clean and disinfect the Exhalation Flow Sensor Assembly (EVQ) on the COVIDIEN PURITAN BENNETT 980?

WARNING: Damaging the flow sensor’s hot film wire or thermistor in the center port can cause the ventilator’s spirometry system to malfunction.

Removal:

1. Lift up on the exhalation filter latch and open the exhalation filter door.

2. With your thumb in the plastic exhalation port and four fingers under the EVQ, pull it down until it snaps out. To avoid damaging the flow sensor element, do not insert fingers into the center port.

3. Remove and discard the disposable components: exhalation valve diaphragm, exhalation filter seal, and pressure sensor filter.

Disinfection:

WARNING: Do not steam autoclave the EVQ or sterilize with ethylene oxide gas.

1. Presoak the EVQ in an enzymatic solution to break down any bio-film. Follow the manufacturer’s instructions for duration.

2. Rinse in clean, deionized water.

3. Prepare the chemical disinfectant according to the manufacturer’s instructions. Use only approved disinfectants.

4. Immerse the EVQ in the disinfectant solution, oriented as shown in the manual, and rotate to remove trapped air bubbles. Keep immersed for the minimum time specified by the manufacturer.

5. At the end of the immersion period, remove and drain all disinfectant completely.

Rinsing:

1. Rinse the EVQ using clean, deionized water in the same manner as the disinfection step.

2. Drain and repeat rinsing three times with clean, deionized water.

3. After rinsing, immerse in a clean isopropyl alcohol bath for approximately 15 seconds. Slowly agitate and rotate to empty air pockets.

Drying:

1. Dry in a low temperature warm air cabinet designed for this purpose, with temperature not exceeding 60°C (140°F).

Inspection:

After disinfection, inspect the plastic body, diaphragm sealing surface, filter grommet, and seal groove for any damage, degradation, or contamination. Inspect electrical contacts and clean with a soft cloth if necessary. Inspect the hot film wire and thermistor for damage; do not attempt to clean them. If rinsing is unsuccessful or they are damaged, replace the EVQ.


How do I reassemble and install the Exhalation Flow Sensor Assembly (EVQ) on the COVIDIEN PURITAN BENNETT 980?

You will need an EVQ Reprocessing Kit which contains a new diaphragm, pressure sensor filter, and exhalation filter seal.

EVQ Reassembly:

1. After drying the EVQ, take the new pressure sensor filter from the kit. Install its large diameter into the filter grommet with a twisting motion until it is flush with the plastic valve body. The narrow end should face out.

2. Take the new exhalation filter seal from the kit. Turn the EVQ so its bottom is facing up.

3. Install the seal into the exhalation flow sensor, ensuring it fits completely within the recess and sits flat.

4. Remove the new diaphragm from the kit and install it. Ensure the diaphragm bead is located in the EVQ’s groove.

5. Carefully inspect component placement and the complete assembly.

To install the EVQ into the ventilator:

1. With the exhalation filter door open, insert the assembly directly under the exhalation valve and push straight up until it snaps into place.

2. Install the exhalation filter by sliding it onto the tracks in the door, orienting the filter’s from patient port through the hole in the door.

3. Close the exhalation filter door and lower the exhalation filter latch.

4. Calibrate the flow sensor.


What do the non-technical alarms on the COVIDIEN PURITAN BENNETT 980 indicate and what are the suggested responses?

The following table details non-technical alarms, their meanings, and suggested actions for the COVIDIEN PURITAN BENNETT 980 ventilator.

Alarm message Meaning What to do
AC POWER LOSS The ventilator or compressor is running on battery power. Monitor the battery charge level to ensure there is enough power remaining to operate the ventilator/compressor.
APNEA (patient data alarm) The time between patient breaths exceeds the set apnea interval. Check patient and settings.
CIRCUIT DISCONNECT The patient circuit has become disconnected or there is a large leak in the patient circuit. Reconnect the patient circuit, or eliminate the leak.
Compliance limitedVT (patient data alarm) Compliance volume required to compensate delivery of a VC, VC+, or VS breath exceeds the maximum allowed for three consecutive breaths. Check patient and circuit type. Inspired volume may be less than set.
COMPRESSOR INOPERATIVE Air pressure not detected in the compressor’s accumulator. Status display indicates the compressor is inoperative. Service or replace compressor.
SEVERE OCCLUSION The patient circuit is severely occluded. The ventilator enters occlusion status cycling. The elapsed time without ventilatory support appears. Check the patient. Obtain alternative ventilation, if necessary. Check the patient circuit for bulk liquid, crimps, blocked filter. If problem persists, remove ventilator from use and obtain service.
PROCEDURE ERROR The patient is attached before ventilator startup is complete. Safety ventilation is active. Provide alternate ventilation, if necessary. Complete ventilator startup procedure.
NO AIR SUPPLY The air supply pressure is less than the minimum pressure required for correct ventilator operation. Check the patient. Check the air and oxygen sources. Obtain alternative ventilation, if necessary.
NO O₂ SUPPLY The oxygen supply pressure is less than the minimum pressure required for correct ventilator operation. Check the patient. Check the air and oxygen sources. Obtain alternative ventilation, if necessary.

How to use the PDF below:
* Touchscreen: Swipe up/down with-in the PDF to scroll and pinch or spread with two fingers to zoom.
* Mouse: While your mouse is hovering over the PDF, use the mouse wheel to scroll and click on the – / + buttons at the bottom of the PDF to zoom.

NO IMAGE CONTENT AT THIS TIME. PLEASE CHECK BACK LATER

CLICK HERE TO DOWNLOAD COVIDIEN PURITAN BENNETT 980 (01) PDF DOCUMENT


Leave comments, questions, reviews, tips, tricks, hacks or page errors below. Account not required.